PathLoc-TA Expandable Lumbar Cage System
K223474 · L&K BIOMED Co., Ltd. · MAX · Jan 13, 2023 · Orthopedic
Device Facts
| Record ID | K223474 |
| Device Name | PathLoc-TA Expandable Lumbar Cage System |
| Applicant | L&K BIOMED Co., Ltd. |
| Product Code | MAX · Orthopedic |
| Decision Date | Jan 13, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
PathLoc-TA Expandable Lumbar Cage System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device system is designed for use with bone graft (autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft) to facilitate fusion. PathLoc-TA Expandable Lumbar Cage System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. Patients should have at least six months of non-operative treatment with a lumbar intervertebral fusion device. The device is intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
Device Story
Expandable lumbar interbody fusion cage; manufactured from Titanium 6AL-4V Alloy (ASTM F136). Inserted in unexpanded state; expanded in situ via system-specific instruments to restore disc height. Features serrated superior/inferior surfaces for vertebral endplate contact. Used in lumbar spine (L2-S1) by orthopedic or neurosurgeons in hospital settings. Requires concurrent use of supplemental fixation (e.g., pedicle screws/rods). Facilitates spinal fusion when packed with autograft or allogenic bone graft. Single-use only; provided sterile. Benefits patients by stabilizing spinal segments and promoting fusion in DDD cases.
Clinical Evidence
Bench testing only. Mechanical performance validated via static/dynamic axial compression, static/dynamic compression-shear, static expulsion, and static subsidence testing per ASTM F2077-18, ASTM F2267-04 (2018), and ASTM draft F-04.25.02.02. Results demonstrate mechanical properties equal to or exceeding predicate devices.
Technological Characteristics
Material: Ti-6Al-4V ELI titanium alloy (ASTM F136). Form factor: Expandable lumbar cage with serrated endplates. Mechanism: Mechanical expansion via dedicated instrumentation. Connectivity: None. Sterilization: Provided sterile. Software: None.
Indications for Use
Indicated for skeletally mature patients requiring intervertebral body fusion at one or two contiguous lumbar levels (L2-S1) for degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. Requires six months of failed non-operative treatment. Must be used with bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- ABTross ALIF Expandable Lumbar Cage System (K221719)
Related Devices
- K182349 — CONCORDE LIFT · Medos International SARL · Apr 3, 2019
- K190708 — AccelFix Lumbar Interbody Fusion Cage System · L&K BIOMED Co., Ltd. · Sep 16, 2019
- K221719 — ABTross ALIF Expandable Cage System · Aegis Spine, Inc. · Nov 3, 2022
- K241413 — sagAlign Lumbar Cage System (Various PNs) · Acuity Surgical Devices, LLC · Jul 17, 2024
- K253377 — Expandable Titanium PLIF/TLIF System · Spectrum Spine, Inc. · Oct 24, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
January 13, 2023
L&K Biomed Co., Ltd. Katherine Kim RA 101, 201, 202 16-25, Dongbaekjungang-ro16 beon-gil Giheung-gu Yongin-si, Gyeonggi-do 17015 Korea
Re: K223474
Trade/Device Name: PathLoc-TA Expandable Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: November 15, 2022 Received: November 18, 2022
Dear Katherine Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Katherine D. Kavlock -S
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K223474
### Device Name
PathLoc-TA Expandable Lumbar Cage System
### Indications for Use (Describe)
PathLoc-TA Expandable Lumbar Cage System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device system is designed for use with bone graft (autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft) to facilitate fusion. PathLoc-TA Expandable Lumbar Cage System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. Patients should have at least six months of non-operative treatment with a lumbar intervertebral fusion device. The device is intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems).
Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# 510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
# 1. SUBMITTER
| Submitter's Name: | L&K BIOMED Co., Ltd. |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Submitter's Address: | #201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea |
| Submitter's Telephone: | +82-2-6717-1983 |
| Contact Person: | Katherine Kim<br>khkim@lnkbiomed.com / ra@lnkbiomed.com |
#### 2. DEVICE NAME
| Trade or Proprietary Name | PathLoc-TA Expandable Lumbar Cage System |
|---------------------------|----------------------------------------------------------------|
| Common or Usual Name | Intervertebral Body Fusion Device, Intervertebral cage, Spacer |
| Regulation class / Number | ClassII, CFR 888.3080 |
| Regulation Name | Intervertebral Body Fusion Device |
| Product Code | MAX |
| Classification Panel | Spinal Devices (DHT6B) |
#### 3. PREDICATE DEVICE
The subject PathLoc-TA Expandable Lumbar Cage System is substantially equivalent to the following devices:
Primary Predicate Device: ABTross ALIF Expandable Lumbar Cage System(K221719)
#### 4. DESCRIPTION OF THE DEVICE
The PathLoc-TA Expandable Lumbar Cage System is interbody fusion devices. This cage system is made of Titanium 6AL-4V Alloy (ASTM F136). And cages are offered in a variety of widths, lengths, heights and lordotic angles designed to adapt to a variety of patient anatomies.
The cage can be expanded in height using the system instrument after being inserted in the unexpanded state. The cages have serrations on the superior endplate and inferior endplate surfaces area to contact vertebrae bone endplate. The device system is designed for use with bone graft (autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft) to facilitate fusion. The device is intended to be used with supplemental fixation systems that have been cleared for use in the lumbrosacral spine (e.g., posterior pedicle screw and rod systems, anterior or lateral plate systems, and anterior screw and rod systems). All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances.
#### INDICATION FOR USE న్.
PathLoc-TA Expandable Lumbar Cage System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device system is designed for use with bone graft (autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft) to facilitate fusion. PathLoc-TA Expandable Lumbar Cage System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K223474 Page 2 of 3" at the top. Below that is the logo for L&K BIOMED. The logo consists of the text "LEARNING & KNOWLEDGE BASED L&K BIOMED" in a sans-serif font, with the "L&K BIOMED" text being larger and bolder. To the right of the text is a circular graphic made up of smaller blue circles.
(DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should have at least six months of non-operative treatment prior to treatment with a lumbar intervertebral fusion device. The device is intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (>20° lordosis) must be used with at least anterior supplemental fixation.
#### 6. PERFORMANCE DATA
Performance testing was performed to demonstrate that the subject PathLoc-TA Expandable Lumbar Cage System is substantially equivalent to other predicate devices.
Static compression, dynamic compression, static and dynamic shear testing according to ASTM F2077, was presented to demonstrate the substantial equivalency of the PathLoc-TA Expandable Lumbar Cage System to the predicate devices.
- Static Axial Compression Test ASTM F 2077 -18 .
- . Static Compression-Shear Test - ASTM F 2077 -18
- . Static Expulsion Test - ASTM draft F-04.25.02.02
- . Static Subsidence Test – ASTM F 2267 – 04 (Reapproved 2018)
- . Dynamic Axial Compression Test- ASTM F 2077 -18
- . Dynamic Compression-Shear Test - ASTM F 2077 -18
Bench testing to evaluate the mechanical properties of the PathLoc-TA Expandable Lumbar Cage System showed a higher or similar mechanical value than predicate marketed devices.
#### 7. MATERIAL
The PathLoc-TA Expandable Lumbar Cage System is manufactured from Ti-6A1-4V ELI titanium alloy (ASTM F136). This this is the same material used in the predicate devices.
#### COMPARISON OF TECHNICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 8.
The subject and predicate devices have identical technological characteristics and there is not any new issues of safety and effectiveness. the following characteristics are same between the subject and predicates:
- י Instruction for use
- י Design and sizes
- -Expanding Mechanism
- . Material
- י Approach
- . Sterilization & Method
- . Manufacturing process
#### SUBSTANTIAL EQUIVALENCE AND CONCLUSION 9.
The subject PathLoc-TA Expandable Lumbar Cage System have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text 'K223474' and 'Page 3 of 3' at the top. Below that is the logo for L&K BIOMED. The logo has the text 'LEARNING & KNOWLEDGE JOINED' above 'L&K BIOMED'. To the right of the text is a circular graphic made of smaller blue circles.
equivalence of the subject device to the predicate devices.
The overall data lead to the conclusion that the PathLoc-TA Expandable Lumbar Cage System is substantially equivalent to the predicate devices.