The Quantra® QPlus® System is composed of the Quantra Hemostasis Analyzer, QPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra QPlus System is intended for in vitro diagnostic use. The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The QPlus Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer. The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Time with Heparinase (CTH), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS) and Clot Time Ratio (CTR). The Quantra QPlus System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in cardiovascular or major orthopedic surgeries before, during, and following the procedure. Results obtained with the Quantra QPlus System should not be the sole basis for patient diagnosis.
Device Story
Quantra QPlus System measures whole blood viscoelastic properties using SEER Sonorheometry (ultrasound-based). Input: 3.2% citrated venous or arterial whole blood in single-use multi-channel cartridge. Analyzer performs four parallel measurements without reagent prep. Output: Clot Time (CT), Heparinase Clot Time (CTH), Clot Stiffness (CS), Fibrinogen Contribution (FCS), Platelet Contribution (PCS), Clot Time Ratio (CTR). Used at point-of-care or clinical labs by trained professionals. Results assist clinicians in assessing coagulation status during surgery; not sole basis for diagnosis.
Clinical Evidence
No clinical data provided; substantial equivalence based on analytical performance demonstrating equivalency between arterial and venous whole blood samples.
Indicated for evaluation of blood coagulation in perioperative patients age 18+ to assess hypocoagulable and hypercoagulable conditions in cardiovascular or major orthopedic surgeries before, during, and following procedures.
Regulatory Classification
Identification
A coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients is an in vitro diagnostic device used to evaluate blood coagulation, fibrinolysis, or both, in perioperative patients, as an aid in the assessment of coagulopathies when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.
Special Controls
The special controls for this device are:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include detailed documentation of, and results from, the following:
(i) A study assessing precision using protocols determined to be acceptable by FDA, to cover the measurement range for each reported parameter (test output). Testing must include native specimens with coagulation profiles representative of the intended use population. In order to cover the measuring range, testing may include a limited number of contrived specimens, not to exceed 10 to 20 percent, or as otherwise deemed appropriate by FDA. The contrived specimens must be prepared to resemble clinical specimens. This testing must evaluate repeatability and reproducibility and provide assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system;
(ii) Studies that demonstrate the performance of each parameter (test output) throughout the claimed measurement range, to include linearity studies or dose-response studies, as applicable to the parameter (test output);
(iii) Potential interferent study that includes evaluation of hemolyzed and lipemic samples as potential interferents; exogenous and endogenous interferents associated with each patient population intended for use with the device, and which might be expected to affect assay performance, must be evaluated; and potential interferents that are specific for, or related to, the technology or methodology of the device. Evaluation of all potential interferents must be performed using a protocol determined to be acceptable to the FDA (
*e.g.,* an FDA-recognized standard) and include both normal and abnormal specimens covering coagulation profiles representative of the intended use population;(iv) A study that evaluates specimen stability under the intended conditions for specimen collection, handling, and storage, using samples that cover the coagulation profiles representative of the intended use population, and using protocols determined to be acceptable by FDA;
(v) A multisite clinical study, determined to be acceptable by FDA, demonstrating performance, relative to clinically relevant and clinically validated laboratory test(s) for each parameter (test output). Further, the study must meet all of the following criteria:
(A) The study must be performed in the intended use population and include representation from all patient populations for whom the device is intended to be used. Potential endogenous and exogenous interferents for each target patient population must be evaluated or known prior to the study;
(B) The study must be conducted at a minimum of three external sites representative of the intended use setting by the intended operators;
(C) Test samples must be collected at time intervals relevant to the device's use in the intended use population;
(D) Clinical specimens, which cover coagulation profiles representative of the intended use population, must be evaluated at each of the three clinical sites in the study;
(E) Analysis of the concordance of clinical interpretation of patient coagulation status made from individual test parameter (test output) results as compared to clinical interpretation of coagulation status from a clinically relevant laboratory test or tests (
*e.g.,* a comparative viscoelastic device or standard laboratory tests) must be conducted; and(F) Expected (reference) values for each parameter (test output) must be demonstrated by testing a statistically appropriate number of samples from apparently healthy normal individuals;
(vi) For a device with a user interface that has information that needs to be interpreted by the user in correctly using the device to achieve the intended test results or a device that does not provide a final output that is a comprehensive interpretation of all parameter (test output) results, a study evaluating the ability of device users to correctly interpret results;
(vii) For any device indicated to guide blood product use, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding blood product use; and
(viii) For any device indicated to guide use of medication, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding use of medication.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A summary of results from the study required by paragraph (b)(1)(i) of this section, including repeatability, reproducibility, and assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system.
(ii) The claimed measurement range of each parameter (test output), as supported by demonstrated performance of the parameter (test output) throughout the claimed measurement range, including studies required by paragraphs (b)(1)(i) through (iii) and (v) of this section, and, if applicable, paragraphs (b)(1)(vii) and (viii) of this section.
(iii) Identification of known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to each parameter (test output). The information must include the concentration(s) or level(s) at which interference was found to occur and the concentration range or levels at which interference was not found to occur.
(iv) Information regarding the multisite clinical study required by paragraph (b)(1)(v) of this section, including:
(A) Each patient population evaluated;
(B) Each intended use setting and the operators;
(C) A summary of the results, including the concordance analysis to clinically relevant laboratory test(s); and
(D) Demonstrated expected (reference) values for each parameter (test output).
(3) The labeling required under § 809.10 of this chapter must include the following:
(i) A limiting statement that the result(s) from the device is(are) not intended to be used as the sole basis for a patient diagnosis.
(ii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(vii) of this section that specifically validate an indication for the device's use in guiding blood product use, a limiting statement that the device has not been evaluated to guide blood product use.
(iii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(viii) of this section that specifically validate an indication for the device's use in guiding use of medication, a limiting statement that the device has not been evaluated to guide use of medication.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2022
HemoSonics, LLC Deborah Winegar Vice-President, Clinical Affairs 4020 Stirrup Creek Drive. Suite 105 Durham, North Carolina 27703
Re: K223433
Trade/Device Name: Quantra QPlus System Regulation Number: 21 CFR 864.5430 Regulation Name: Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients Regulatory Class: Class II Product Code: OFR Dated: November 14, 2022 Received: November 14, 2022
Dear Deborah Winegar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the text "Min Wu-S" in a large, sans-serif font. The text is black and is set against a white background. The letters are evenly spaced and the text is easy to read. The image is simple and straightforward, with no other elements present.
Min Wu Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name QPlus® Cartridge
#### Indications for Use (Describe)
The Quantra® QPlus® System is composed of the Quantra Hemostasis Analyzer, QPlus Cartridge, and Quality Controls Level 1 and 2. The Quantra OPlus System is intended for in vitro diagnostic use.
The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The QPlus Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer.
The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time with Heparinase (CTH), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS) and Clot Time Ratio (CTR).
The Quantra QPlus System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in cardiovascular or major orthopedic surgeries before, during, and following the procedure. Results obtained with the Quantra QPlus System should not be the sole basis for patient diagnosis.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from it on the left side. To the right of the blood drop is the word "HEMOSONICS" in black, sans-serif font.
# SECTION 5. SPECIAL 510(K) SUMMARY
# A. APPLICANT INFORMATION
| Submission Date: | November 14, 2022 |
|------------------------|---------------------------------------------------------------------------------------------------|
| Submitter Information: | HemoSonics, LLC<br>4020 Stirrup Creek Drive, Suite 105<br>Durham, NC 27703<br>Phone: 919-244-6990 |
| Contact Person: | Deborah Winegar, PhD<br>Email: dwinegar@hemosonics.com<br>Phone: 919-244-6990 |
#### B. PROPRIETARY AND ESTABLISHED NAMES
Quantra® QPlus® System
## C. REGULATORY INFORMATION
| Trade/Device Name: | Quantra QPlus System |
|----------------------------|---------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 864.5430 |
| Regulation Name: | Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients |
| Regulatory Classification: | Class II |
| Product Code: | QFR |
#### D. PURPOSE OF SUBMISSION
To add a sample matrix to the intended use and indications for use statement for the QPlus Cartridge. This change does not alter the device's fundamental scientific technology.
## E. MEASURAND
The combination of clot time and clot stiffness parameters measured from the four channels of the cartridge provides information about the functional role of coagulation factors, fibrinogen, and platelets in the sample.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from it on the left. To the right of the blood drop is the company name, "HEMOSONICS", in black sans-serif font.
# F. TYPE OF TEST
The Quantra QPlus System is an in vitro diagnostic device designed to assess a patient's coagulation system by measuring the viscoelastic properties of a blood sample during clot formation in surgical and intensive care settings. The system consists of the Quantra Hemostasis Analyzer (instrument), a single-use disposable cartridge, (OPlus Cartridge) and Quantra Quality Controls (external Quality Control materials).
## G. INTENDED USE AND INDICATIONS FOR USE STATEMENT
The Quantra® OPlus® System is composed of the Quantra Hemostasis Analyzer, OPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra QPlus System is intended for in vitro diagnostic use.
The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The QPlus Cartridge is a multi-channel cartridge that provides semiquantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer.
The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Time with Heparinase (CTH), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS) and Clot Time Ratio (CTR).
The Quantra QPlus System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in cardiovascular or major orthopedic surgeries before, during, and following the procedure. Results obtained with the Quantra OPlus System should not be the sole basis for patient diagnosis.
# H. DEVICE DESCRIPTION
The Quantra QPlus System consists of an instrument (the Quantra Hemostasis Analyzer), a single-use disposable cartridge (QPlus Cartridge), and external quality control materials (QPlus Controls). The OPlus System is intended for use with patients 18 years or older by professionals in a hospital setting (point of care or laboratory). The measurements are performed in four test channels of the disposable cartridge.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from it on the left, followed by the word "HEMOSONICS" in black, sans-serif font. The logo is simple and modern, and the use of red and black gives it a bold and professional look.
The QPlus Cartridge is a multichannel disposable cartridge which enables four independent measurements to be performed in parallel with different sets of reagents without the need for any reagent preparation or controlled pipetting. The cartridge utilizes a citrated evacuated blood collection tube filled with a patient whole blood sample The proprietary technology SEER Sonorheometry measures the evolution of shear modulus (i.e., clot stiffness) in all four channels as a function of time.
Clot times and clot stiffness values obtained from the measurements performed by the QPlus Cartridge are combined to form parameters that depict the functional status of the patient's coagulation system. Four (4) of the parameters are measured and two (2) are calculated. The parameters are summarized in Table 5-1.
| Channel | Reagents | Measurement | Description | Units | Reportable<br>Range |
|------------|----------------------------------------------------------------------------|----------------------------------|---------------------------------------------------------------------|-----------------------|---------------------|
| 1 | Kaolin, calcium buffers<br>&stabilizers | Clot Time (CT) | Clot time in citrated<br>whole blood | sec | 60 to 480 |
| 2 | Kaolin, heparinase I, calcium<br>buffers & stabilizers | Heparinase Clot<br>Time (CTH) | Clot time in citrated<br>whole blood with heparin<br>neutralization | sec | 60 to 480 |
| 3 | Thromboplastin, polybrene,<br>calcium, buffers & stabilizers | Clot Stiffness (CS) | Stiffness of the whole<br>blood clot | hPa<br>(hectopascals) | 2 to 65 |
| 4 | Thromboplastin, polybrene,<br>abciximab, calcium, buffers<br>& stabilizers | Fibrinogen<br>Contribution (FCS) | Contribution of<br>functional fibrinogen to<br>clot stiffness | hPa<br>(hectopascals) | 0.2 to 30 |
| Calculated | Calculated by subtracting<br>FCS from CS | Platelet<br>Contribution (PCS) | Contribution of platelet<br>activity to clot stiffness | hPa<br>(hectopascals) | 2 to 50 |
| Calculated | Calculated ratio of CT over<br>CTH | Clot Time Ratio<br>(CTR) | Assessment of residual<br>heparin anticoagulation | Unitless | 0.8 to 4 |
Table 5-1. QPlus Cartridge Output Parameters
# I. DEVICE MODIFICATION DESCRIPTION
The Quantra System was previously cleared under DEN180017 with a sample matrix claim of venous whole blood. HemoSonics is submitting this Special 510(k) to demonstrate the equivalency of arterial and venous whole blood samples analyzed on the Quantra Hemostasis Analyzer with the QPlus Cartridge in order to expand the sample matrices accepted on this system.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for HemoSonics. The logo features a red blood drop with sound waves emanating from it on the left. To the right of the blood drop is the company name, "HEMOSONICS", in a sans-serif font.
## J. SUBSTANTIAL EQUIVALENCE INFORMATION
Predicate Device Name: Quantra Plus System
Predicate 510(k) Number: DEN180017
Comparison with the Predicate:
Table 5-2 provides an overall comparison of the modified Quantra QPlus System with the previously cleared Quantra QPlus System.
| | Modified Device | Predicate Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Quantra System | Quantra System |
| | (Subject of Special 510(k)) | (DEN180017) |
| | Similarities | |
| Manufacturer | HemoSonics, LLC | Same |
| Trade Name | Quantra QPlus System | Same |
| Common Name | Whole Blood Hemostasis System | Same |
| Classification Name | Coagulation system for the measurement<br>of whole blood viscoelastic properties in<br>perioperative patients | Same |
| Regulation Number | 21 CFR 864.5430 | Same |
| Product Code | QFR | Same |
| Device Class | II | Same |
| Location of Use | Point of care and laboratory settings | Same |
| Disposables | QPlus Cartridge (multichannel cartridge)<br>QStat Cartridge (multichannel cartridge)<br>Quantra Quality Controls (Level 1 and<br>Level 2) | Same |
| Analyzer Hardware | Quantra Hemostasis Analyzer HS-002 | Same |
| Analyzer Software | Ver 2.1.37 | Same |
| | Differences | |
| Indications for use | The Quantra® QPlus® System is<br>composed of the Quantra Hemostasis<br>Analyzer, QPlus Cartridge, QStat<br>Cartridge, and Quantra Quality Controls<br>Level 1 and 2. The Quantra System is<br>intended for in vitro diagnostic use.<br>The Quantra Hemostasis Analyzer uses<br>Sonic Estimation of Elasticity via<br>Resonance (SEER) Sonorheometry, an<br>ultrasound-based technology, to measure<br>the shear modulus of whole blood during<br>coagulation. The cartridges are multi-<br>channel and provides semi-quantitative<br>indications of the coagulation state of a | The Quantra® QPlus® System is composed<br>of the Quantra Hemostasis Analyzer,<br>QPlus Cartridge, QStat Cartridge, and<br>Quantra Quality Controls Level 1 and 2.<br>The Quantra System is intended for in<br>vitro diagnostic use.<br>The Quantra Hemostasis Analyzer uses<br>Sonic Estimation of Elasticity via<br>Resonance (SEER) Sonorheometry, an<br>ultrasound-based technology, to measure<br>the shear modulus of whole blood during<br>coagulation. The cartridges are multi-<br>channel and provides semi-quantitative<br>indications of the coagulation state of a |
#### Table 5-2: Comparison between DEN180017 and Modified Quantra System
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from the bottom left, followed by the company name "HEMOSONICS" in black, sans-serif font. The logo is clean and modern, with a focus on the company name.
| | Modified Device | Predicate Device |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Quantra System | Quantra System |
| | (Subject of Special 510(k)) | (DEN180017) |
| | 3.2% citrated venous or arterial whole<br>blood sample. Cartridges include tests to<br>assess coagulation characteristics via the<br>intrinsic pathway, via the extrinsic<br>pathway, and includes tests with a<br>heparin neutralizer.<br>The system is intended to be used by<br>trained professionals at the point-of-care<br>and in clinical laboratories to evaluate<br>the viscoelastic properties of whole<br>blood.<br>The Quantra System is indicated for the<br>evaluation of blood coagulation in<br>perioperative patients age 18 years and<br>older to assess possible hypocoagulable<br>and hypercoagulable conditions in<br>cardiovascular or major orthopedic<br>surgeries before, during, and following<br>the procedure.<br>Results obtained with the Quantra<br>System should not be the sole basis for<br>patient diagnosis. | 3.2% citrated venous whole blood sample.<br>Cartridges include tests to assess<br>coagulation characteristics via the intrinsic<br>pathway, via the extrinsic pathway, and<br>includes tests with a heparin neutralizer.<br>The system is intended to be used by<br>trained professionals at the point-of-care<br>and in clinical laboratories to evaluate the<br>viscoelastic properties of whole blood.<br>The Quantra System is indicated for the<br>evaluation of blood coagulation in<br>perioperative patients age 18 years and<br>older to assess possible hypocoagulable<br>and hypercoagulable conditions in<br>cardiovascular or major orthopedic<br>surgeries before, during, and following the<br>procedure.<br>Results obtained with the Quantra System<br>should not be the sole basis for patient<br>diagnosis. |
| Sample Type(s) | Arterial and venous whole blood | Venous whole blood |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.