K223414 · Guilin Refine Medical Instrument Co., Ltd. · EBZ · Nov 15, 2023 · Dental
Device Facts
Record ID
K223414
Device Name
Curing Light, Model: MaxCure 9
Applicant
Guilin Refine Medical Instrument Co., Ltd.
Product Code
EBZ · Dental
Decision Date
Nov 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515mm waveband of visible light.
Device Story
MaxCure 9 is a dental curing light used by trained professionals to polymerize light-cured resins, composites, and orthodontic materials. The device consists of a handpiece with an integrated LED light source and a light guide. The operator selects a mode (Turbo or Normal) and irradiation time; the device emits high-intensity visible light (385-515nm) to initiate polymerization. The device is powered by a rechargeable lithium-ion battery, charged via contact. It is used in clinical settings to restore teeth or bond orthodontic brackets. The output is controlled by the operator to ensure proper curing of dental materials, directly affecting the structural integrity of the dental restoration or orthodontic appliance. Patient benefit is achieved through efficient, reliable hardening of dental materials.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), dental equipment safety (IEC 80601-2-60), and performance testing per ANSI ADA Standard No. 48-2004. Biocompatibility testing (cytotoxicity, skin sensitization, oral mucosa irritation) was performed.
Technological Characteristics
Handheld LED curing light; wavelength 385-515nm; dual peak 400nm/452nm. Powered by rechargeable lithium-ion battery. Contact-based charging. Patient-contacting material: PE (polyethylene). Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, ANSI ADA No. 48, ISO 10993-5/10. Non-sterile.
Indications for Use
Indicated for polymerization of resinous dental materials, restorative composites, and orthodontic bonding/sealing materials in the 385-515nm range. For use by trained dental professionals.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2023
Guilin Refine Medical Instrument Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China
Re: K223414
Trade/Device Name: Curing Light, Model: MaxCure 9 Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: August 14, 2023 Received: August 15, 2023
Dear Kevin Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmoham madi -S
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and
{2}------------------------------------------------
ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K223414
Device Name Curing Light, Model: MaxCure 9
### Indications for Use (Describe)
The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515mm waveband of visible light.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 10px;"> <span style="border: 1px solid black; padding: 0px; margin-right: 2px;"><b>X</b></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="border: 1px solid black; padding: 0px; margin-right: 2px;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
EF
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
## 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
### Prepared Date: 2023/11/9
#### 1. Submission sponsor
Name: Guilin Refine Medical Instrument Co., Ltd.
Address: No.8-3, Information Industrial Park, High-Tech Zone, Qixing District, Guilin, GuangXi, China 541004
Contact person: Chen Judong
Title: Management Representative
E-mail: chenjudong(@refine-med.com
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P. R. China 518067
Contact person: Kevin Wang E-mail: kevin(@chonconn.com
Tel: +86-755 33941160
| Trade/Device Name | Curing Light, Model: MaxCure 9 |
|-------------------|--------------------------------------------------------------------|
| Model | MaxCure9 |
| Common Name | Curing Light |
| Regulatory Class | Class II |
| Classification | 21 CFR 872.6070 / Activator, Ultraviolet, For Polymerization / EBZ |
| Submission type | Traditional 510(K) |
#### 3 Subiect Device Information
#### 4. Predicate Device
- 1. Diadent Group International, D-Lux+ under K200809 (Primary predicate)
#### Device Description ട്.
This Curing Light is a device which generates high intensity light for polymerization of light curing materials used for dental curing purpose. This curing light is intended for use by trained dental professionals for the purpose of light curing dental resins and composites and used to restore teeth.
#### Intended use & Indication for use 6.
The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515nm waveband of visible light.
Section 5 510(k) Summary
{5}------------------------------------------------
| Features | Predicate device<br>D-Lux+ | Subject device<br>Curing Light, Model:<br>MaxCure 9 (K223414) | Note |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K200809 | \ | Same |
| Product code | EBZ | EBZ | Same |
| Intended Use &<br>Indications for<br>use | The D-Lux+ is intended to<br>polymerize resinous dental<br>materials, restorative composite<br>materials, and orthodontic<br>brackets, bonding and sealing<br>materials that are photo-<br>polymerized in the 385~515nm<br>waveband of visible light. | The MaxCure9 is intended<br>to polymerize resinous<br>dental materials, restorative<br>composite materials, and<br>orthodontic brackets,<br>bonding and sealing<br>materials that are photo-<br>polymerized in the<br>385~515nm waveband of<br>visible light. | Same |
| Principles of<br>operation | 1. Insert the Light Probe into the<br>handpiece. Attach the Light<br>Protector to the handpiece.<br>2. Cover the Light Probe with the<br>Disposable Sheath.<br>3. Turn on the power and select<br>the mode and set-up time.<br>4. Press the Operation button to<br>perform light irradiation for the<br>set time and press the button<br>once again to stop the light<br>irradiation. The light irradiation<br>progress time is displayed<br>numerically in the display<br>window. | a) Insert the light guide into<br>the mounting hole of the<br>curing light. Clamp and<br>install the light hood at the<br>root of the light guide.<br>b) Cover the Light Guide<br>with the disposable isolation<br>sleeve.<br>c)Turn on the power and<br>select the mode and set-up<br>time.<br>d) Press the mode key to<br>perform light irradiation for<br>the set time and press the<br>button once again to stop<br>the light irradiation. | Same |
| Delivery form | -D-Lux+ Handpiece<br>-D-Lux+ Charger<br>-Light Probe<br>-Light Protector<br>-C-Battery (Included in the<br>handpiece)<br>-Adapter<br>-Power Cord<br>-Disposable Sheaths(200ea/Box) | Main unit,<br>Charger,<br>Light hood,<br>Power adapter,<br>disposable PE isolation<br>sleeve (100ea/Box),<br>Instruction Manual. | Same |
| Features | Predicate device<br>D-Lux+ | Subject device<br>Curing Light, Model:<br>MaxCure 9 | Note |
| | -User Manual | | |
| Technical<br>Specifications –<br>Wavelength<br>range | 385 – 515 nm | 385 – 515 nm | Same |
| Peak<br>wavelength | Dual peak:<br>405nm, 460nm | Dual peak:<br>400nm, 452 nm | SE, The<br>values are<br>similar. This<br>difference<br>does not<br>raise issues<br>of<br>substantial<br>equivalence. |
| Depth of cure | ≥2mm | ≥2mm | Same |
| Light<br>source/Probe | a light probe with an embedded<br>LED | a light probe with an<br>embedded LED | Same |
| Operation<br>Modes and<br>Light Intensity | 5 modes:<br>STD (Standard) mode :900 mW/cm²,<br>5, 10, 15 and 20 seconds<br>SFT (Soft Start) mode: 1,300<br>mW/cm²,<br>10 and 15 seconds<br>HIG (High Power) mode: 1,300<br>mW/cm²,<br>5, 10 and 15 seconds<br>ORT (Orthodontic) mode: 1,800<br>mW/cm²,<br>3, 4, 5 seconds and P<br>MAX (Max Power) mode:2,400<br>mW/cm²,<br>1, 2 and 3 seconds | 2 modes:<br>Turbo mode: 2000 mW/cm2,<br>1S, 3S<br>Normal mode: 1000<br>mW/cm2,<br>5S, 10S, 15S, 20S | Different 1 |
| Cross Infection<br>Prevent | disposable sheath | disposable PE isolation<br>sleeve | Same |
| Use | Prescription / Hospital | Prescription / Hospital | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Power Source | rechargeable lithium ion battery | rechargeable lithium ion | Same |
| Features | Predicate device<br>D-Lux+ | Subject device<br>Curing Light, Model:<br>MaxCure 9 | Note |
| Battery<br>Charging | wirelessly charged | charging method by the<br>contact | Different 2 |
| Power supply | 100-240VAC, 50-60Hz | 100V-240V AC 50Hz/60Hz | Same |
| outputs | 6VDC/2A | DC 5V/1A | Different 2 |
| Electrical Safety | IEC60601-1 and IEC 60601-1-1-<br>2 | IEC60601-1 and IEC<br>60601-1-1-2 | Same |
| Chemical<br>composition of<br>patient<br>contacting<br>portions of the<br>device | N/A | PE | Different 3 |
| Biocompatibility | Direct contact with issue in not<br>intended. Therefore ISO10993-1<br>is not applicable | ISO 10993-5<br>ISO 10993-10 | Different 3 |
# 7. Comparison to the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
Different (1): Operation Modes and Light Intensity
The subject device has several modes corresponding to the light output intensity and available times. The light output safety and performance test was conducted according to IEC 60601-1, IEC60601-1-2 and FDA guidance performance testing requirements with the difference. The testing results show that these differences do not affect safety and effectiveness .
Different (2): Battery Charging and Power Supply
The subject device is charged by the charging method of contact. However, the predicate device is wirelessly charged without a contact using a wireless charger. As a power source for the charger, the AC/DC adapter provided by the manufacturer is used. The input specifications of the adapter(100-240VAC, 50-60Hz) are the same, but the outputs differ from each other at 5VDC/1A and 6VDC/2A respectively. These differences do not affect safety and effectiveness . Different (3): Biocompatibility
Biocompatibility testing of the subject device was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur. However, the predicate device is not evaluated. The testing results show that these differences do not affect safety and effectiveness.
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
Biocompatibility testing was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur.
The biocompatibility evaluation for the Curing light was conducted in accordance with the FDA Guidance Section 5 510(k) Summary 5- 5 / 6
{8}------------------------------------------------
for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:
- Cytoxicity
- Skin Sensitization
- Oral Mucosa Irritation
## Software Validation
The Curing light software has been fully validated and the device software meets its Design Specifications. FDA Guidance Document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices has been followed in connection with the design of the device.
## Non-clinical data
The Curing light has been tested according to the following standards:
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 80601-2-60 Edition 2.0 2019-06 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- ANSI ADA Standard No. 48-2004 (R2015) Visible Light Curing Units
## Clinical data
The subject of this premarket submission, Curing light, did not require clinical studies to support substantial equivalence.
#### 9. Conclusion
Based on the similarities in intended use, principles of operation, design rationale, test results, and performance, the subject Curing light is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.