← Product Code [NHA](/productcode/NHA) · K223395

# Luna Dental Implant System (K223395)

_Shinhung Mst Co., Ltd. · NHA · Feb 3, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223395

## Device Facts

- **Applicant:** Shinhung Mst Co., Ltd.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Feb 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The Luna Dental Implant System is intended for delayed loading and immediate loading is possible when good primary stability is achieved and with appropriate occlusal loading.

## Device Story

The Luna Dental Implant System consists of endosseous dental implants (fixtures) and various abutments (Healing, Duo, Duo Plus, Temporary). These components are surgically placed into the jawbone by a dental professional to support prosthetic restorations like crowns, bridges, or overdentures. The system restores chewing function for edentulous or partially edentulous patients. The device acts as a mechanical interface between the jawbone and the prosthetic tooth. Clinicians select specific abutment sizes based on patient anatomy and clinical needs. The system supports both delayed and immediate loading protocols depending on primary stability. The device is intended for use in a clinical dental setting.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and performance were supported by risk analysis (ISO 14971) and non-clinical MRI safety evaluations (ASTM F2052, F2213, F2182, F2119) using scientific literature. Previous testing for sterilization, shelf life, and biocompatibility was leveraged from K123155.

## Technological Characteristics

Materials: Pure Titanium Grade 4 (ASTM F67) and Titanium Alloy Ti-6Al-4V (ASTM F136). Components: Fixture, abutment, mount screw, cover screw. Sterilization: Radiation or user-performed sterilization. Form factor: Various diameters (4.0-6.1mm) and post/gingiva heights. Connectivity: None (mechanical device).

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- LUNA DENTAL IMPLANT SYSTEM ([K123155](/device/K123155.md))
- LUNA DENTAL IMPLANT SYSTEM ([K160106](/device/K160106.md))
- ET/SS IMPLANT SYSTEM ([K120847](/device/K120847.md))
- Temporary Snap Abutment ([K161435](/device/K161435.md))

## Reference Devices

- ET/SS IMPLANT SYSTEM ([K120847](/device/K120847.md))
- Temporary Snap Abutment ([K161435](/device/K161435.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Shinhung MST Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. , STE 160 Irvine. California 92612

Re: K223395

Trade/Device Name: Luna Dental Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 1, 2022 Received: November 8, 2022

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K223395

Device Name Luna Dental Implant System

Indications for Use (Describe)

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The Luna Dental Implant System is intended for delayed loading is possible when good primary stability is achieved and with appropriate occlusal loading.

Type of Use (Select one or both, as applicable)

V Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K223395)

This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: 2/2/2023

# 1. Submitter

SHINHUNG MST CO., LTD. 110-2, Donghwagondan-ro, Munmak-eup, Wonju-si, Gangwon-do, Republic of Korea, 26365

Tel: +82-33-730-1901

# 2. U.S Agent/Contact Person

Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman STE 160, Irvine CA 92612 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com

## 3. Device

- Trade Name: Luna Dental Implant System ●
- Common Name: Dental Implant System ●
- Classification Name: Endosseous Dental Implant ●
- Product Code: NHA
- . Classification regulation: 21CFR 872.3640

## 4. Predicate Device:

- · Primary predicate device: LUNA DENTAL IMPLANT SYSTEM by SHINHUNG MST CO., LTD (K123155)
- · Reference predicate device: LUNA DENTAL IMPLANT SYSTEM by SHINHUNG MST CO., LTD (K160106) ET/SS IMPLANT SYSTEM by OSSTEM IMPLANT CO.,LTD (K120847) Temporary Snap Abutment by Nobel Biocare AB (K161435)

{4}------------------------------------------------

# 5. Description:

The Luna Dental Implant System is a device of pure titanium (ASTM F67) and titanium alloy (ASTM F1136) intended to be surgically placed in the bone of the upper or lower arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It consists of fixture, abutment, mount screw, cover screw. This 510k is intended to add the new models of the following abutments to the Luna Dental Implant System.

- Healing Abutment I
- י Duo Abutment
- I Duo Plus Abutment
- I Temporary Abutment

## 6. Indication for use:

The Luna Dental Implant system is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The Luna Dental Implant System is intended for delayed loading and immediate loading is possible when good primary stability is achieved and with appropriate occlusal loading.

# 7. Basis for Substantial Equivalence

#### Comparison Chart

| Device Name            | Subject Device                                                                                                                                                                                                                            | Primary Predicate Device                                                                                                                                                                                                                                     | Reference Device                                                                                                                                                                                                                               |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                 | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                                    | Osstem Implant Co., Ltd.                                                                                                                                                                                                                       |
| Device Name            | Healing Abutment                                                                                                                                                                                                                          | Healing Abutment                                                                                                                                                                                                                                             | Healing Abutment                                                                                                                                                                                                                               |
| Design                 | Image: Subject Device Design                                                                                                                                                                                                              | Image: Primary Predicate Device Design                                                                                                                                                                                                                       | Image: Reference Device Design                                                                                                                                                                                                                 |
| 510(k)<br>Number       | K223395                                                                                                                                                                                                                                   | K123155                                                                                                                                                                                                                                                      | K120847                                                                                                                                                                                                                                        |
| Indications for<br>Use | The Luna Dental Implant<br>system is intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as artificial teeth, and to<br>restore the patient's | The Luna Dental Implant<br>System is intended to be<br>surgically placed in the bone<br>of the upper or lower jaw<br>arches to provide support<br>for prosthetic devices, such<br>as artificial teeth, and to<br>restore the patient' s<br>chewing function. | The ET/SS Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture |

- · Healing Abutment

{5}------------------------------------------------

|                     | chewing function. The<br>Luna Dental Implant<br>System is intended for<br>delayed loading and<br>immediate loading is<br>possible when good<br>primary stability is<br>achieved and with<br>appropriate occlusal<br>loading.                                                                                                                                                                                                                                                                                                                          |                                  | restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading. The<br>abutment is intended for<br>use with a dental implant<br>fixture to provide support<br>for prosthetic restorations<br>such as crowns, bridges, |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                  | or overdenture.                                                                                                                                                                                                                                                                   |
| Diameter<br>(mm)    | 4.0 ~ 5.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 4.5 ~ 7.0                        | 4.0 ~ 8.0                                                                                                                                                                                                                                                                         |
| Post Height<br>(mm) | 2.0 ~ 10.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 2.0~7.0                          | 3.0 ~ 9.0                                                                                                                                                                                                                                                                         |
| Material            | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Pure Titanium Gr 4<br>(ASTM F67) | Titanium Alloy Ti-6Al-<br>4V<br>(ASTM F136)                                                                                                                                                                                                                                       |
| Sterilization       | Radiation Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Radiation Sterilization          | Radiation Sterilization                                                                                                                                                                                                                                                           |
| S.E                 | The subject device has the same intended use, design, and technological<br>characteristics as the primary predicate device (k123155, k160106). The diameter of<br>the new model is smaller than the primary predicate device. We identified a reference<br>device which has the diameter size (4mm). The post height is slightly higher than the<br>reference device, however, this minor difference does not raise a question in<br>substantial equivalence since the user can select an appropriate height model<br>according to each patient need. |                                  |                                                                                                                                                                                                                                                                                   |

| Device Name            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device                                                                                                                                                                                                                           |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | SHINHUNG MST CO., LTD.                                                                                                                                                                                                                                                                                                                                                                                                                            | SHINHUNG MST CO., LTD.                                                                                                                                                                                                                             |
| Device Name            | Duo Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                      | Duo Abutment                                                                                                                                                                                                                                       |
| Design                 | Image: Duo Abutment                                                                                                                                                                                                                                                                                                                                                                                                                               | Image: Duo Abutment                                                                                                                                                                                                                                |
| 510(k) Number          | K223395                                                                                                                                                                                                                                                                                                                                                                                                                                           | K123155                                                                                                                                                                                                                                            |
| Indications for<br>Use | The Luna Dental Implant system is<br>intended to be surgically placed in the<br>bone of the upper or lower jaw arches to<br>provide support for prosthetic devices,<br>such as artificial teeth, and to restore the<br>patient's chewing function. The Luna<br>Dental Implant System is intended for<br>delayed loading and immediate loading<br>is possible when good primary stability<br>is achieved and with appropriate occlusal<br>loading. | The Luna Dental Implant System is<br>intended to be surgically placed in the<br>bone of the upper or lower jaw arches to<br>provide support for prosthetic devices,<br>such as artificial teeth, and to restore the<br>patient's chewing function. |
| Diameter (mm)          | 4.6 ~ 5.1                                                                                                                                                                                                                                                                                                                                                                                                                                         | 4.6 ~ 6.1                                                                                                                                                                                                                                          |
| Gingiva Height<br>(mm) | 1.0 ~ 5.0                                                                                                                                                                                                                                                                                                                                                                                                                                         | 1.0 ~ 5.0                                                                                                                                                                                                                                          |
| Post Height<br>(mm)    | 7.0                                                                                                                                                                                                                                                                                                                                                                                                                                               | 4.0 ~ 7.0                                                                                                                                                                                                                                          |

{6}------------------------------------------------

| Material      | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                            | Pure Titanium Gr 4<br>(ASTM F67) |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Sterilization | User Sterilization                                                                                                                                                                                                          | User Sterilization               |
| S.E.          | The subject device has the same intended use, design, and technological<br>characteristics as the primary predicate device (k123155). The modification is only for<br>adding a few more size options within the same range. |                                  |

| Device Name            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Osstem Implant Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Name            | Duo Plus Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Transfer Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Duo Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Design                 | Image: Duo Plus Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Image: Transfer Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Image: Duo Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for<br>Use | The Luna Dental Implant<br>system is intended to be<br>surgically placed in the<br>bone of the upper or lower<br>jaw arches to provide<br>support for prosthetic<br>devices, such as artificial<br>teeth, and to restore the<br>patient's chewing<br>function. The Luna Dental<br>Implant System is<br>intended for delayed<br>loading and immediate<br>loading is possible when<br>good primary stability is<br>achieved and with<br>appropriate occlusal<br>loading.    | The ET/SS Implant<br>System is indicated for use<br>in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading. The<br>abutment is intended for<br>use with a dental implant<br>fixture to provide support<br>for prosthetic restorations<br>such as crowns, bridges, or<br>overdenture. | The Luna Dental Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or lower<br>jaw arches to provide<br>support for prosthetic<br>devices, such as artificial<br>teeth, and to restore the<br>patient' s chewing<br>function.                                                                                                                                                                                                                                                                                                     |
| 510(k) Number          | K223395                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K120847                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K123155                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Diameter (mm)          | 4.1 ~ 6.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 4.0 ~ 7.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 4.6 ~ 6.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Gingiva Height<br>(mm) | 1.0 ~ 5.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 1.0 ~ 5.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1.0 ~ 5.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Material               | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                          | Titanium Alloy Ti-6Al-4V<br>(ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterilization          | User Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                        | User Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | User Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| S.E.                   | The subject device has the same intended use, design, and technological<br>characteristics as the primary predicate device (k120847). The only difference would<br>be area of TiN coating, but this difference does not raise a question in substantial<br>equivalence.                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Name            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer           | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | SHINHUNG MST CO.,<br>LTD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Nobel Biocare AB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Name            | Temporary Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Temporary Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Temporary Snap Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Design                 | Image: Temporary Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Image: Temporary Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Image: Temporary Snap Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 510(k)<br>Number       | K223395                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K123155                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K161435                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Indications for<br>Use | The Luna Dental Implant<br>system is intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as artificial teeth, and to<br>restore the patient's<br>chewing function. The<br>Luna Dental Implant<br>System is intended for<br>delayed loading and<br>immediate loading is<br>possible when good<br>primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. | The Luna Dental Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or lower<br>jaw arches to provide<br>support for prosthetic<br>devices, such as artificial<br>teeth, and to restore the<br>patient's chewing function.                                                                                                                                                                                                                                                                                            | The Temporary Snap<br>Abutment is intended to be<br>used to fabricate and<br>support provisional<br>restorations that aid in<br>creating<br>an esthetic emergence<br>through the gingiva during<br>the healing period and<br>prior to final restoration.<br>The Temporary Snap<br>Abutment can be used for<br>cement retained or screw-<br>retained provisional<br>restorations. The<br>abutments can be used for<br>single-unit and multi-unit<br>restorations. Use of the<br>Temporary Snap Abutment<br>is not to exceed one<br>hundred and eighty<br>(180) days. |
| Diameter<br>(mm)       | 4.6~5.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 4.1 ~ 4.6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 4.0 ~ 6.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Gingiva<br>Height (mm) | 2.0 ~ 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 1.0 ~ 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1.5 ~ 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Material               | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                          | Pure Titanium Gr 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Titanium vanadium alloy<br>(ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sterilization          | User sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                              | User sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | User sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| S.E                    | The subject device has the same intended use, design, and technological characteristics<br>as the primary predicate device (k123155). The only difference is that the diameter of<br>the new models is outside of the range. We identified a reference device which cover<br>the diameter range of the subject device.                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{7}------------------------------------------------

## Substantial Equivalence Discussion

The Luna Dental Implant System (Healing Abutment, Duo Abutment, Duo Plus Abutment, and Temporary Abutment) are substantially equivalent to its predicate device (k123155) in terms of indications for use, raw material, and design. The wording of the indications for use statement is slightly different but that does not affect the intended use.

{8}------------------------------------------------

The difference is in size range. We have identified a reference device that encompass the size range of the subiect device. Based on the information submitted here in we conclude that the subject device is substantially equivalent to the predicate devices.

# 8. Non-Clinical Testing

Risk analysis and validation activities were performed according to the ISO 14971 and the result support that the predicate device does not raise a new question in safety and performance.

A non-clinical worst-case MRI review was conducted to evaluate the Luna Implant System device in an MRI environment using scientific evidence and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795). Titanium Grade 4 and Titanium alloy (Ti-6Al-4V, ELI) were assessed according to magnetic induction displacement force (ASTM F2052), magnetic induction torque (ASTM F2213), RF induction heating (ASTM F2182), and image artifact (ASTM F2119) by T. O. Woods et al. Based on that rationale, we have addressed parameters per FDA guidance " Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetic induced displacement force and torque.

The following test items were previously submitted and reviewed under K123155.

- . Sterilization - Radiation and Moist Heat
- . Shelf life for sterile devices
- Biocompatibility ●

## 9. Conclusion

The subject devices and the predicate device have the same intended use and have the same technological characteristics.

Based on the similarities, we conclude that the Luna Dental Implant System is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K223395](https://fda.innolitics.com/device/K223395)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
