V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System
Applicant
Samsung Medison Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223387 · Feb 13, 2023
V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System
Samsung Medison Co., Ltd.
Retrospective and prospective clinical practice ultrasound images
Retrospective and prospective clinical images were used to validate the performance (segmentation, view recognition, and measurement accuracy) of AI-based features including BiometryAssist, ViewAssist, and HeartAssist.
AI algorithm validation; Retrospective clinical data; Ultrasound imaging; Performance testing
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
BiometryAssist Validation; Retrospective and prospective clinical data collection
Fetal ultrasound patients; Sample Size: 320 fetal biometry images; Number of Sites: 2
The V8 / V7 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Sketal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via transducers; displays images in 2D, M, Color/Power/Pulsed/Continuous Wave/Tissue Doppler, 3D/4D, and ElastoScan modes. Used in hospitals/clinics by trained healthcare professionals for clinical diagnosis. Integrates AI-based analysis tools (NerveTrack, BiometryAssist, ViewAssist, HeartAssist) to assist in anatomical measurement, view recognition, and segmentation. Output displayed on-screen for clinician review; aids in anatomical measurement and clinical decision-making. Benefits include improved workflow, standardized measurements, and diagnostic support.
Clinical Evidence
No clinical studies required. Bench testing only. AI algorithm performance validated using retrospective and prospective datasets. NerveTrack: 90.3% accuracy (95% CI 88.6-92.0), 3.54 FPS. BiometryAssist: 0.928 Dice-score, <8% circumference error. ViewAssist: 94.56% recognition accuracy, 0.898 Dice-score. HeartAssist: Fetal/Adult view recognition (93.21%/98.31%), segmentation (0.88/0.93 Dice-score), and size measurement error rates/PCC values met acceptance criteria.
Technological Characteristics
Mobile diagnostic ultrasound system; 2D, Doppler, 3D/4D imaging modes. Materials evaluated for biocompatibility per ISO 10993-1. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and NEMA UD 2-2004. Software-controlled with AI-based analysis features. Connectivity via standard ultrasound interfaces. Sterilization/disinfection per manufacturer instructions.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal/Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal, Urology, Cardiac (Adult/Pediatric/Trans-esophageal), Thoracic, and Peripheral vessel examinations.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
RS85 / RS80 EVO Diagnostic Ultrasound System (K221117)
HS50 / HS60 Diagnostic Ultrasound System (K210068)
HERA W9 / HERA W10 Diagnostic Ultrasound System (K220043)
Submission Summary (Full Text)
{0}------------------------------------------------
February 13, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Samsung Medison Co., Ltd. % Ju Jee Young Regulatory Affairs Specialist Samsung Medison Co. Ltd. 3366, Hanseo-ro, Nam-myeon Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA
Re: K223387
Trade/Device Name: V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 16, 2023 Received: January 18, 2023
Dear Ju Jee Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K223387
#### Device Name
V8 Diagnostic Ultrasound System
V7 Diagnostic Ultrasound System
#### Indications for Use (Describe)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|---------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K223387
### 510(K) Summary:
In accordance with 21 CFR 807.92, the following summary of information is provided:
- Date Prepared November 04, 2022 1.
- 2. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do, Republic of Korea
- 3. Primary Contact Person Ju, Jee-Young Regulatory Affairs Specialist Phone: +82.2.2194.0861 Fax: +82. 2.2194.0273 Email: jee.ju(@samsungmedison.com
- Secondary Contact Person 4. Ninad Gujar Vice President Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com
- Proposed Device 5.
-
-
- Common/Usual Name : Diagnostic Ultrasound System and Accessories -
- Proprietary Name : V8 Diagnostic Ultrasound System
- V7 Diagnostic Ultrasound System
- Regulation Name : Ultrasonic pulsed doppler imaging system
- Regulatory Class : Class II -
- Product Code : IYN, IYO, ITX -
- -Regulation Number : 21 CFR 892.1550, 892.1560, 892.1570
- Predicate Devices 6.
- V8 / V7 Diagnostic Ultrasound System (K220975) Primary -
- RS85 / RS80 EVO Diagnostic Ultrasound System (K221117) Reference -
- HS50 / HS60 Diagnostic Ultrasound System (K210068) Reference -
- HERA W9 / HERA W10 Diagnostic Ultrasound System (K220043) Reference -
- Device Description 7.
- The V8 / V7 are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8 / V7 also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SAMSUNG" in a bold, sans-serif font. The letters are all capitalized and are a dark blue color. The background is white, which makes the word stand out.
diagnosis by competent health care professionals. The V8 / V7 have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
- 8. Intended Use
The V8 / V7 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Sketal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
- Technological Comparison to Predicate Devices 9.
The V8 / V7 employ the same fundamental scientific technology as its predicate devices V8 / V7 (K220975), RS85/ RS80 EVO(K221117), HS50 / 60 (K210068) and HERA W9 / W10(K220043) System.
- 10. Determination of Substantial Equivalence
Comparison to Predicate: The V8 / V7 are substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
- The systems are all intended for diagnostic ultrasound imaging and fluid flow . analysis.
- . The proposed V8 / V7 and the primary predicate V8 / V7 (K220975) have the same clinical intended use.
- . The proposed V8 / V7 and the primary predicate V8 / V7 (K220975) have the same imaging modes and modes of operation.
- . The proposed V8 / V7 have expanded the detection nerve type (Head: Greater Occipital / Leg : Sciatic, Common Peroneal, Tibial) of NerveTrack based on AI in the previously cleared V8/V7 (K220975)
- . The proposed V8 / V7 have included the BiometryAssist and ViewAssist which are based on AI version from the previously cleared HERA W9/ HERA W 10 (K220043).
{5}------------------------------------------------
- The proposed V8 / V7 have included the HeartAssist which is based on the AI ● version from the previously cleared HERA W9/ HERA W10 (K220043) and expanded the clinical application to adult.
- . The proposed V8 / V7 have included the QUS(TSI/TAI) and EzHRI from the previously cleared RS85/ RS80 EVO (K221117).
- . The proposed V8 / V7 have included the additional matching methods of S-Fusion in the previously cleared V8/V7 (K220975)
- The proposed V8 / V7 have included the EzPrep for work-flow improvement. .
- The proposed V8/V7 have included one new transducer miniER7. PA4-12B and . LA2-9S have been migrated from the predicate RS85/ RS80 EVO (K221117). Only V7 has included four new transducers LA4-18AD, EV2-11ARE, EV2-11AVE and CV1-8AE, and LA3-14AD has been migrated from the predicate HS50 / 60 (K210068). Biocompatibility test has been conducted for the new transducers, and image performance tests have been conducted for the new and migrated transducers.
- . The proposed V8 / V7 and primary predicate V8 / V7 (K220975) have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- . The proposed V8 / V7 and primary predicate V8 / V7 (K220975) have been designed in compliance with approved electrical and physical safety standards.
- The system is manufactured with materials which have been evaluated and found ● to be safe for the intended use of the device.
- . The system has acoustic power levels which are below the applicable FDA limits.
- 11. Summary of Non-Clinical Testing
The device has been evaluated for acoustic output, biocompatibility, software function. cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable FDA guidances and medical device safety standards. The V8 / V7 and its applications comply with the following FDA-recognized standards.
| Reference No. | Title |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012<br>and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -<br>Part 1: General requirements for basic safety and essential performance<br>(IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 | IEC60601-1-2: 2020-09(4.1 Edition) , Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential performance -<br>EMC |
| IEC 60601-2-37 | IEC 60601-2-37 Edition 2.0 2007, Medical electrical equipment – Part 2-<br>37: Particular requirements for the basic safety and essential performance |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "SAMSUNG" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
| | of ultrasonic medical diagnostic and monitoring equipment |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO10993-1 | AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
| ISO14971 | ISO 14971:2019, Medical devices - Application of risk management to medical devices |
| NEMA UD 2-2004 | NEMA UD 2-2004 (R2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3 |
[ The validation for expansion of the detection nerve of NerveTrack based on AI ]
These are the details on validation of the AI algorithm used for all the new and improved AI-based features that includes expansion of the use of NerveTrack feature and the information about the dataset the device was tested on.
Acceptance Criteria:
| Validation Type | Definition | Acceptance Criteria |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Accuracy (%) | $ \frac{\text{Number of correctly detected frames}}{\text{Total number of frames with nerve}} \times 100 $ | ≥ 80% |
| Speed (FPS) | $ \frac{1000}{\text{Average latency time of each frame (msec)}} $ | ≥ 2 FPS |
Summary Performance data, Standard Deviations & Confidence Intervals:
| Validation Type | Average | Standard Deviation | 95% CI |
|-----------------|---------|--------------------|--------------|
| Accuracy (%) | 90.3 | 4.8 | 88.6 to 92.0 |
| Speed (FPS) | 3.54 | 0.13 | 3.47 to 3.61 |
Testing Data Information:
| | Females | Males | Total |
|--------------------|-------------|-------|---------|
| Number of Subjects | 13 | 5 | 18 |
| Number of Images | 1,168 | 978 | 2,146 |
| Age range | 32~68 | 22~50 | 22~68 |
| Average age | 45.7 | 35.0 | 42.7 |
| BMI range | 16~27.1 | 31.5 | 16~31.5 |
| Average BMI | 20.5 | 31.5 | 21.5 |
| Ethnicity | All Koreans | | |
The standalone performance of NerveTrack was evaluated for BMI (Body Mass Index) subgroups, which are potential subject and image confounder. For subgroup analysis, we divided BMI values into four groups according to the CDC (Centers for Disease Control and Prevention) definition for adult overweight & obesity. They are underweight (BMI < 18.5), healthy weight (18.5 ≤ BMI < 25), overweight (25 ≤ BMI < 30), and obesity (BMI ≥
30).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "SAMSUNG" in all capital letters. The letters are a dark blue color. The background is white, which makes the word stand out.
We evaluated the performance for the four groups and observed very good generalizability for BMI, because the average accuracy of all subgroups is included within the confidence interval of the accuracy for the full dataset. The depth range was set 2.5cm to 4.5cm, and the maximum depth of the detected nerve was 3.31 cm from skin. When we calculate the relationship between BMI and the maximum depth of nerves, it illustrates a low correlation and no significant difference of NerveTrack performance according to the patient's BMI information. Therefore, we believe that our model is robust even if the cases with various BMI information exist.
When collecting scan data, the speed of the probe varies approximately from 1cm to 5cm per second. The orientation of the probe was short-axis imaging and perpendicular to the skin, obtained by sliding the probe in both the proximal and distal directions. All the nerve data were acquired on Samsung ultrasound devices including V8.
Ten anesthesiologists and five sonographers with more than 10 years of experience participated to establish the ground truth (GT) for the location of 12 different kinds of nerves. One anesthesiologist who scanned the ultrasound directly drew the GT of nerve location, and two or more other anesthesiologists and sonographers reviewed and confirmed that it was correct. If there was any mistake during the review, it was revised again.
The training data used for the training of the NerveTrack algorithm are independent of the data used to test the NerveTrack algorithm.
[The Summary of Testing for BiometryAssist]
- Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
We tested on two areas: Segmentation and Size measurement.
- □ Segmentation test
- ் A deep learning based segmentation algorithm was validated using 320 fetal biometry images collected at two hospitals.
- · The average dice-score is 0.928 (threshold 0.8)
- □ Size measurement test
- · We use the same datasets of segmentation test.
- · The error rate of circumference measured value is 8% or less.
- · The error rate of distance measured value is 4% or less.
- ਼ The error rate of NT(Nuchal Translucency) measured value is 1mm or less.
- Demographic distribution:
{8}------------------------------------------------
- □ Gender: Female
- [ Age: Reproductive age, specific age not collected
- [ Ethnicity/Country: Americans and Koreans
- Information about clinical subgroups and confounders present in the dataset:
- [ We divided the fetal ultrasound images, depending on the ISUOG and AIUM guidelines, into 8 views.
- Information about equipment and protocols used to collect images
- [ ] We acquired the data set with the three of SAMSUNG MEDISON's ultrasound systems (W10, W9 and V8) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice (Only V8 images have been used for the Performance Test).
- Information about how the reference standard was derived from the dataset (i.e. the "Truthing" process):
- [ All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images.
- □ The participating experts were composed of obstetricians with more than 20 years of experience and two sonographers with more than 10 years of experience; all in fetal cardiology. The entire process was supervised by another obstetrician with more than 25 years of experience.
- Description of how the independence of test data from training data was ensured:
- [ ] Data used for training, tuning and validation purpose are completely separated from the ones during training process, and there is no overlap among the three.
[The Summary of Testing for ViewAssist]
- Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
We tested on two areas: view recognition and anatomy annotation(segmentation).
- □ View recognition test
- · A deep learning based view recognition algorithm was validated using 680 fetal heart and fetal biometry images collected at two hospitals.
- · The average recognition accuracy is 94.56% (threshold 89%)
{9}------------------------------------------------
- □ Anatomy annotation(segmentation) test
- · We use the same datasets of view recognition test.
- · The average dice-score is 0.898 (threshold 0.8)
- Demographic distribution:
- □ Gender: Female
- □ Age: Reproductive age, specific age not collected
- □ Ethnicity/Country: Americans and Koreans
- Information about clinical subgroups and confounders present in the dataset:
- □ We divided the fetal ultrasound images, according to the ISUOG and AIUM guidelines, into 17 views.
- Information about equipment and protocols used to collect images
- | | We acquired the data set with the three of SAMSUNG MEDISON's ultrasound systems (W10. W9 and V8) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice (Only V8 images have been used for the Performance Test).
- Information about how the reference standard was derived from the dataset (i.e. the "Truthing" process):
- □ All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images.
- □ The participating experts were composed of obstetricians with more than 20 years of experience and two sonographers with more than 10 years of experience; all in fetal cardiology. The entire process was supervised by another obstetrician with more than 25 years of experience.
- Description of how the independence of test data from training data was ensured:
- [ ] Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three.
[The Summary of Testing for HeartAssist]
- Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
{10}------------------------------------------------
We tested adult and fetal hearts individually in the three areas: view recognition, segmentation and size measurement.
- □ View recognition test
- ் A deep learning based view recognition algorithm was validated using 888 fetal heart and adult heart images collected at five hospitals.
- (Fetus) The average recognition accuracy is 93.21% (threshold 89%)
- ं (Adult) The average recognition accuracy is 98.31% (threshold 84%)
- □ Segmentation test
- · We use the same datasets of view recognition test.
- (Fetus) The average dice-score is 0.88 (threshold 0.8).
- ் (Adult) The average dice-score is 0.93 (threshold 0.9).
- □ Size measurement test
- · We use the same datasets of segmentation test.
- (Fetus) The error rate of area measured value is 8% or less.
- (Fetus) The error rate of angle measured value is 4% or less.
- (Fetus) The error rate of circumference measured value is 11% or less.
- · (Fetus) The error rate of diameter measured value is 11% or less.
- ं (Adult, B-mode) Pass if the *PCC value is 0.8 or greater than the specification.
- (Adult, M-mode) Pass if the *PCC value is 0.8 or greater than the specification.
- ं (Adult, Doppler-mode) Pass if the *PCC value is 0.8 or greater than the specification.
* The Pearson correlation coefficient(PCC) that is a representative way to measure similarity is a measure of linear correlation between two sets of data. The HeartAssist's PCC was calculated to evaluate the auto measurement using the ground truth, defined as the cardiologist's measurements.
- Demographic distribution:
- □ Gender: Male and Female
- □ Age: Reproductive age, specific age not collected
- [ Ethnicity/Country: Americans and Koreans
- Information about clinical subgroups and confounders present in the dataset:
- [ ] We divided the fetal ultrasound images, according to the ASE and AIUM guidelines, into 7 fetal heart views and 16 adult heart views.
{11}------------------------------------------------
- I Information about equipment and protocols used to collect images
- | | We acquired the data set with the three of SAMSUNG MEDISON's ultrasound systems (W10, W9 and V8) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice (Only V8 images have been used for the Performance Test).
- Information about how the reference standard was derived from the dataset (i.e. the "Truthing" process):
- | (Fetus) All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image. The participating experts were composed of an obstetrician with more than 20 years of experience and two sonographers with more than 10 years of experience, all in fetal cardiology. The entire process was supervised by another obstetrician with more than 25 years of experience.
- [ (Adult) We employed the help of four professionals: two cardiologists with at least 10 years of experience and two sonographers with at least 10 years of experience. These experts manually traced the contours of the heart and the signal outline on the images.
- Description of how the independence of test data from training data was ensured:
- [ ] Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three.
- 12. Summary of Clinical Tests
The proposed device V8 / V7 Ultrasound System did not require clinical studies to demonstrate substantial equivalence.
#### 13. Conclusion
Since the predicate devices and the subject device have a similar intended use and key technological features, the non-clinical data support the safety of the device and demonstrate that the V8 / V7 Ultrasound System should perform as intended in the specified use conditions. Therefore, SAMSUNG MEDISON CO., LTD. considers the subject device to be as safe, as effective, and performance is substantially equivalent to the primary predicate device that is currently marketed for the same intended use.
- END of 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.