K223381 · Welch Allyn, Inc. · HKI · Mar 15, 2023 · Ophthalmic
Device Facts
Record ID
K223381
Device Name
iExaminer System with Panoptic Plus
Applicant
Welch Allyn, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Mar 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The iExaminer system with PanOptic Plus, consisting of PanOptic Plus, SmartClip, iExaminer application, and one of the following: iPhone 11 Pro, iPhone 11 Pro Max, iPhone 12 Pro, is intended to be used to capture images as an aid to clinicians in the evaluation, and documentation of ocular health. The images from the iExaminer System with PanOptic Plus are not intended to be used as a sole means of diagnosis.
Device Story
System captures ocular images using PanOptic Plus ophthalmoscope, smart device attachment (SmartBracket/SmartClip), and compatible iPhone (X, 11 Pro, 11 Pro Max, 12 Pro) running iExaminer Pro app. Clinician mounts smartphone to ophthalmoscope; captures images of eye/surrounding area; stores/retrieves images via app. Used in clinical environments by trained personnel. Output provides visual documentation of ocular health to aid clinical evaluation; not intended as sole diagnostic tool. Benefits include improved documentation and potential for easier viewing via larger smartphone displays compared to predicate.
Clinical Evidence
Clinical evidence consists of an image comparison study comparing images from the iExaminer System with PanOptic Plus against the predicate (K121405). Study objective was to demonstrate equivalence in usefulness for documentation and clinical referrals. Results confirmed the device met all acceptance criteria and performed comparably to the predicate.
Technological Characteristics
System comprises PanOptic Plus ophthalmoscope (LED light source), SmartBracket/SmartClip attachment, and compatible iPhone. Complies with ISO 10940 (fundus cameras), ANSI Z80.36 (light hazard), ISO 10993-1 (biocompatibility), and IEC 60601-1-2 (EMC). Features 3mm minimum pupil diameter, 24.4mm working distance, 0.90 magnification, and -20D to +25D diopter compensation. Software-based image capture and storage.
Indications for Use
Indicated for clinicians to capture images of the eye and surrounding area for evaluation and documentation of ocular health in adult and pediatric patients. Not for use as a sole means of diagnosis.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
March 15, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Welch Allyn, Inc. Jeffrey Thompson Manager, Regulatory Affairs (a subsidiary of Baxter Healthcare Corporation) 4341 State Street Road Skaneateles Falls, New York 13153
Re: K223381
Trade/Device Name: iExaminer System with Panoptic Plus Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: February 17, 2023 Received: February 17, 2023
Dear Jeffrey Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223381
Device Name iExaminer System with PanOptic Plus
#### Indications for Use (Describe)
The iExaminer system with PanOptic Plus, consisting of PanOptic Plus, SmartClip, iExaminer application, and one of the following: iPhone 11 Pro, iPhone 11 Pro Max, iPhone 12 Pro, is intended to be used to capture images as an aid to clinicians in the evaluation, and documentation of ocular health. The images from the iExaminer System with PanOptic Plus are not intended to be used as a sole means of diagnosis.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right, giving it a dynamic appearance.
#### 510(k) Summary
#### DATE PREPARED: 09 MAR 2023
#### OWNER:
Welch Allyn Inc. (a subsidiary of Baxter Healthcare Corporation) 4341 State Street Road Skaneateles Falls, NY 13153-0220
### CONTACT PERSON:
Jeffrey E. Thompson Manager, Regulatory Affairs Baxter Healthcare Corporation 32650 N. Wilson Road Round Lake, IL 60073 T 224.270.3806
### IDENTIFICATION OF THE DEVICE:
Trade/Device Name: iExaminer System with PanOptic Plus Classification Panel: Ophthalmic Regulation Number: 21 CFR 886.1120 Regulation Name: Camera, Ophthalmic, Ac-Powered Regulatory Class: Class II Product Code: HKI
| Table 1. Proposed Model number for iExaminer System with PanOptic Plus | |
|------------------------------------------------------------------------|--|
|------------------------------------------------------------------------|--|
| Model Number | Name |
|--------------|-------------------------------------|
| 901161 | iExaminer System with PanOptic Plus |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be slightly italicized. The word is the logo for the healthcare company Baxter International.
#### Figure 1. Proposed Product Diagram for iExaminer System with PanOptic Plus
Image /page/4/Picture/2 description: The image shows a medical device with a smartphone attached to it. The device has a camera lens at the top, labeled as 1 and 2. The smartphone is mounted on the device and displays a medical image, labeled as 3 and 4. The device appears to be designed for capturing and displaying medical images, possibly for diagnostic purposes.
- 1. PanOptic Plus ophthalmoscope
- 2. Smart device attachment mechanism (made of SmartBracket and SmartClip)
- 3. Compatible Smart device
- 4. iExaminer Pro Software Application
#### Predicate Device
#### Table 2. Predicate Device
| Device | Company | Predicate 510(k) | Clearance Date |
|----------------------------|-------------------|------------------|-------------------|
| iExaminer with<br>PanOptic | Welch Allyn, Inc. | K121405 | December 20, 2012 |
### Description of the Device
The iExaminer System with PanOptic Plus is a medical device that allows the user to capture images through the use of a PanOptic Plus ophthalmoscope and a smart device. The iExaminer System with PanOptic Plus consists of (also see Figure 1):
- 1. PanOptic Plus Ophthalmoscope:
- a. Ophthalmoscope Head
- b. Compatible energy sources (i.e. battery handles or wall units)
- c. Optional Patient Eyecup
- 2. Smart device attachment instrument (made of SmartBracket and SmartClip);
- 3. Compatible smart device (iPhone X, iPhone 11 Pro, iPhone 11 Pro Max, iPhone 12 Pro).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif, and the letters are closely spaced together.
- 4. iExaminer Pro Software Application.
The iExaminer system with PanOptic Plus is intended to take photographs of the eye and surrounding area.
The device is for prescription use only.
#### Indications For Use
The iExaminer system with PanOptic Plus, consisting of PanOptic Plus, SmartBracket, SmartClip, iExaminer application, and one of the following: iPhone X, iPhone 11 Pro, iPhone 11 Pro Max, iPhone 12 Pro, is intended to be used to capture images as an aid to clinicians in the evaluation, and documentation of ocular health. The images from the iExaminer System with PanOptic Plus are not intended to be used as a sole means of diagnosis.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly italicized, giving it a sense of movement. The background is plain white, which makes the word stand out.
## Device Comparison and Substantial Equivalence
| Characteristic | Predicate - iExaminer with PanOptic<br>(K121405) | Proposed Device - iExaminer System with<br>PanOptic Plus | Discussion of Differences |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Product Code | HKI; Camera Ophthalmic, AC-Powered | Same | N/A |
| FDA Regulation # | 21 CFR 886.1120 | Same | N/A |
| Indications for<br>Use/Intended Use | The iExaminer is an attachment and software<br>used only with the iPhone 4 and iPhone 4S in<br>conjunction with the Welch Allyn PanOptic<br>Ophthalmoscope to allow users to capture,<br>send, store, and retrieve images of the eye. The<br>device is intended to be used by trained<br>personnel within a medical or school<br>environment. | The iExaminer system with PanOptic Plus,<br>consisting of PanOptic Plus, SmartBracket,<br>SmartClip, iExaminer application, and one of<br>the following: iPhone X, iPhone 11 Pro,<br>iPhone 11 Pro Max, iPhone 12 Pro, is<br>intended to be used to capture images as an<br>aid to clinicians in the evaluation, and<br>documentation of ocular health. The images<br>from the iExaminer System with PanOptic<br>Plus are not intended to be used as a sole<br>means of diagnosis. | The proposed device uses similar<br>hardware components including<br>a different list of smart devices<br>as clearly listed in the Welch<br>Allyn® iExaminer® Pro<br>Ophthalmoscopes Instructions<br>for Use (IFU).<br>The proposed device IFU does<br>not constitute a new Intended<br>Use from the predicate device as<br>both the predicate and the<br>proposed devices are intended to<br>aid the clinician in the evaluation<br>and documentation of ocular<br>health. The use environment and<br>use conditions are the same as<br>the predicate device for which it<br>is intended to be used. |
| Components | Ophthalmoscope, adapter for aligning smart<br>device optics and ophthalmoscope optics,<br>smart device and the iExaminer software<br>application. | Same | N/A |
| Compatible<br>Smartphones | iPhone 4, iPhone 4S | iPhone X, iPhone 11 Pro, iPhone 11 Pro Max,<br>and iPhone 12 Pro | The proposed device uses<br>different list of smart devices as<br>clearly listed in the Welch<br>Allyn® iExaminer® Pro |
| Characteristic | Predicate - iExaminer with PanOptic<br>(K121405) | Proposed Device - iExaminer System with<br>PanOptic Plus | Discussion of Differences |
| | | | Ophthalmoscopes Instructions<br>for Use (IFU). The proposed<br>device IFU does not constitute a<br>new Intended Use from the<br>predicate device as both the<br>predicate device and the<br>proposed device are intended to<br>allow users to capture images of<br>the eye.<br>Design control activities have<br>been conducted and confirmed<br>that there is no impact to safety<br>or effectiveness for this<br>application. |
| Intended Population | Adult, Pediatric | Same | N/A |
| Characteristic | Predicate - iExaminer with PanOptic<br>(K121405) | Proposed Device - iExaminer System with<br>PanOptic Plus | Discussion of Differences |
| Prescription Use<br>Only | Yes | Same | N/A |
| Minimum Pupil<br>Diameter Size | 2mm (per PanOptic) | 3mm (per PanOptic Plus) | Both ophthalmoscopes provide<br>the ability to function on patients<br>with small pupil sizes. |
| Working Distance | 25.4mm | 24.4mm | Both ophthalmoscopes provide a<br>similar working distance. |
| Pixel Pitch | Measured<br>5.37 microns for iPhone 4 High Res<br>4.25 microns for iPhone 4S High Res | Measured<br>Approximately 2 microns for each of the<br>compatible smartphones | Both opthalmascopes provide the<br>appropriate pixel pitch based on<br>the models of iphones used. The<br>proposed device has a lower<br>pixel pitch due to advancements<br>in the resolution of the camera<br>and the telephoto lens. |
| Optical<br>Magnification | 1.183 (per PanOptic) | 0.90 (per PanOptic Plus) | Both ophthalmoscopes provide<br>similar magnification (~1x). |
| Data Output/ Output<br>Terminals | iExaminer App iOS (version 1.0) | iExaminer Pro App iOS (version 4.0.0.13) | Minor differences have been<br>made to the application since the<br>predicate device, but do not<br>impact the safety or effectiveness<br>of the product.<br>Design control activities have<br>been conducted and confirmed<br>that there is no impact to safety<br>or effectiveness for this<br>application. |
| Use Conditions | With or without mydriatic | Same | N/A |
| Characteristic | Predicate - iExaminer with PanOptic<br>(K121405) | Proposed Device - iExaminer System with<br>PanOptic Plus | Discussion of Differences |
| Photographic Light<br>Source | As per PanOptic observation light source:<br>Halogen lamp visible light | As per PanOptic Plus observation light<br>source: LED visible light | See "Observation Light Source"<br>in this table. |
| Observation Light<br>Source | As per PanOptic observation light source:<br>Halogen lamp visible light | As per PanOptic Plus observation light<br>source: LED visible light | Both provide a light output that<br>allows for the clinician to view<br>anatomy. Also see "Exposure<br>Parameters" in this table below |
| Observation and<br>display system | As per iPhone 4 or 4S: 3.5" screen size | iPhone X and 11 Pro - 5.8"inch screen size<br>iPhone 11 Pro Max – 6.5" inch screen size<br>iPhone 12 Pro - 6.1" inch screen size | The proposed system includes a<br>larger display, which should<br>permit easier viewing. Other<br>screen parameters of the user's<br>system are verified via the "Test<br>the Resolution" section of the<br>IFU. |
| Camera Spec | 5 megapixel / 8 megapixel | 12 megapixels | Both opthalmascopes provide the<br>appropriate camera spec based<br>on the models of iphones used.<br>The proposed device has a higher<br>camera spec due to<br>advancements in the resolution<br>of the camera. |
| Diopter<br>Compensation | As per PanOptic: -20D to +20D | As per PanOptic Plus: -20D to +25D | PanOptic Plus provides a wider<br>diopter range, which is a benefit<br>to the user. |
| Apertures | As per PanOptic: Multiple | Same | N/A |
| Picture Angle | 25 degrees | 20 degrees | Both provide a relatively large<br>field of view for performing eye<br>exams. |
| Characteristic | Predicate - iExaminer with PanOptic<br>(K121405) | Proposed Device - iExaminer System with<br>PanOptic Plus | Discussion of Differences |
| Storage Media | Internal storage capacity | Same | N/A |
| Exposure Parameters | As per PanOptic: Compliance with ISO 15004-<br>1 as Group 1 (ophthalmic instrument for which<br>no potential light hazard exists) | As per PanOptic Plus: Compliance with ANSI<br>Z80.36 as Group 2 (ophthalmic instrument for<br>which a potential light hazard exists.) | Both devices comply with<br>applicable eye safety standards<br>(e.g. ISO 15004-2 and ANSI<br>Z80.36: 2021) at the time of the<br>respective submission dates. |
| Image Data Format | JPEG | Same | N/A |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly italicized, giving it a sense of movement. The background is plain white, which makes the word stand out.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, giving it a sense of depth.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, giving it a sense of depth.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly blurred, giving it a sense of depth. The background is plain white, which makes the blue text stand out.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly italicized. The word is centered and takes up most of the image.
#### DISCUSSION OF NONCLINICAL TESTS:
Non-Clinical testing of the iExaminer System with PanOptic Plus has been performed against requirements for performance, physical attributes, environmental conditions and safety, and to provide objective evidence that the device's intended use is met. A summary of testing performed is identified in the summary below.
- . Software: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005.
- . Software: Guidance for Off-the-Shelf Software Use in Medical Devices. September 2019.
- Software: General Principles of Software Validation; January 2002.
- FDA Guidance on Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices.
- . Biocompatibility: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices--Part 1: Evaluation and testing within a risk management process", Issued June 2016.
The iExaminer System with PanOptic Plus was tested to evaluate its safety and effectiveness based on the standards and associated methods below:
- . ISO 10940 (Second edition 2009-08-01) Ophthalmic Instruments—Fundus Cameras
- . ANSI Z80.36 (2021) American National Standard for Ophthalmics—Light Hazard Protection for Ophthalmic Instruments
- . EN/IEC 62304 (2006/A1:2016) Medical Device Software – Software Life Cycle Processes
- . EN/IEC 60601-1-60 (ED 3.1 2013-10) Medical Electrical Equipment – Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard Usability
- . IEC 62366-1 (2015+AMD1: 2020) Medical devices – Part 1: Application of usability engineering to medical devices
- . EN/ISO 10993-1 (Fifth edition 2018-08) Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement.
- . EN ISO/ISO 15004-1 (Second edition 2020-5) Ophthalmic Instruments-Fundamental Requirements and Test Methods - Part 1: General Requirements Applicable to all Ophthalmic Instruments
- EN IEC 60601-1-2 (2014) Medical Electrical Equipment Part 1-2: General . Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility – Requirements and Tests
In addition to the above, and in consideration of IEC 62366-1: 2015+AMD1: 2020, Medical devices - Part 1: Application of usability engineering to medical devices as well as FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices - Guidance For Industry And Food And Drug Administration Staff (document Issued On: February 3, 2016), results demonstrated that the iExaminer System with PanOptic Plus is suitable for its intended use.
The results of verification and validation demonstrate that the iExaminer System with PanOptic Plus has passed all established acceptance criteria and is as safe and effective as the predicate device for its intended use. Based on the comparison analysis and test results, Welch Allyn has concluded that the iExaminer System with PanOptic Plus to be substantially equivalent to the predicate device.
### DISCUSSION OF CLINICAL TESTS:
Image Comparison Study: An image comparison study was performed on the iExaminer System with PanOptic Plus. The primary objective of this study was to demonstrate that the iExaminer System with PanOptic Plus images are substantially equivalent to the predicate device (i.e. PanOptic with iPhone 4/4S and iExaminer (version 1.0) FDA 510(k) K121405) images in their usefulness for documentation and clinical referrals.
The results of the image comparison study demonstrate that the iExaminer System with PanOptic Plus has passed all established acceptance criteria and is as safe and effective as the predicate device for its intended use. Based on the image comparison study results, Welch Allyn has concluded the iExaminer System with PanOptic Plus to be substantially equivalent to the predicate device.
#### CONCLUSION:
The iExaminer System with PanOptic Plus has been verified and validated against design input requirements, user needs and intended uses. The non-clinical and clinical testing (image comparison study) demonstrate that the subject device raises no new questions
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement.
concerning safety and effectiveness, is substantially equivalent, and performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.