Monoject Magellan Insulin Safety Syringe 1mL, 31G x 6mm (8881893110); Monoject Magellan Insulin Safety Syringe 0.5mL, 31G x 6mm (8881893150); Monoject Magellan Insulin Safety Syringe 0.3mL, 31G x 6mm (8881893130)
K223376 · Cardinal Health200, LLC · MEG · Jun 21, 2023 · General Hospital
Device Facts
Record ID
K223376
Device Name
Monoject Magellan Insulin Safety Syringe 1mL, 31G x 6mm (8881893110); Monoject Magellan Insulin Safety Syringe 0.5mL, 31G x 6mm (8881893150); Monoject Magellan Insulin Safety Syringe 0.3mL, 31G x 6mm (8881893130)
Applicant
Cardinal Health200, LLC
Product Code
MEG · General Hospital
Decision Date
Jun 21, 2023
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Primarily intended for intra-muscular, subcutaneous, and/or intra-dermal medication delivery. Secondary use is a safety feature that helps prevent sharps injuries.
Device Story
Piston syringe with permanently attached needle and integrated safety shield; designed for manual, one-handed activation to cover and lock over needle post-injection. Used in clinical or home settings for U-100 insulin delivery. Operator (clinician or patient) fills syringe, administers dose, and activates shield to prevent sharps injury. Single-use device; sterile; orange cap indicates insulin use. Benefits include reduced risk of accidental needle-sticks during disposal.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 8537 (syringe delivery accuracy, hub/needle bond strength), ISO 10993-1 (biocompatibility), ISO 11607 (packaging), and USP <788> (particulate testing).
Technological Characteristics
Piston syringe; polypropylene barrel; polyethylene safety shield; Type 304 stainless steel needle; silicone lubricants. Sterilized via gamma radiation (SAL 10^-6). Dimensions: 0.3mL, 0.5mL, 1.0mL volumes; 31G x 6mm needle. Conforms to ISO 8537:2016, ISO 10993-1:2018, ISO 7864, and USP <85>.
Indications for Use
Indicated for delivery of U-100 insulin. Includes needle stick prevention feature to guard against accidental needle-sticks.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 21, 2023
Cardinal Health 200, LLC Emily Hunter Senior Specialist, Regulatory Affairs 3651 Birchwood Drive Waukegan, Illinois 60085
## Re: K223376
Trade/Device Name: Monoject Magellan Insulin Safety Syringe 1mL, 31G x 6mm (8881893110); Monoject Magellan Insulin Safety Syringe 0.5mL. 31G x 6mm (8881893150); Monoject Magellan Insulin Safety Syringe 0.3mL, 31G x 6mm (8881893130) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG, FMF, FMF, FMI Dated: Mav 24, 2023 Received: May 25, 2023
Dear Emily Hunter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan St
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223376
#### Device Name
Monoject Magellan Insulin Safety Syringe 1 ml, 31 G x 6mm (888189311 0); Monoject Magellan Insulin Safety Syringe 0.5ml, 31G x 6mm (8881893150); Monoject Magellan Insulin Safety Syringe 0.3ml, 31G x 6mm (8881893130)
#### Indications for Use (Describe)
Monoject Magellan Insulin Safety Syringes are intended for delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223376 510(k) Summary
### Submitter Information:
Cardinal Health 200, LLC 3651 Birchwood Drive Waukegan, IL 60085, USA
Contact Person: Emily Hunter Phone: 847-887-6367 Email: emily.hunter()1(@cardinalhealth.com
Date of Preparation: June 20, 2023
| Device | |
|----------------------|--------------------------------------------------------|
| TRADE NAME | Monoject™ Magellan™ Insulin Safety Syringe |
| COMMON NAME | Insulin Syringe, with sharps injury prevention feature |
| CLASSIFICATION NAME | Piston Syringe |
| REGULATION NUMBER | 21 CFR 880.5860 |
| REGULATORY CLASS | II |
| PRODUCT CODES | MEG, FMF, FMI |
| CLASSIFICATION PANEL | General Hospital |
### Predicate Device
510(k) Number: K061492 Device Name: Monoject™ Magellan™ Insulin Safety Syringe
### Device Description
The Monoject™ Magellan™ Insulin Safety Syringe consists of a piston syringe with a permanently attached needle, and safety shield designed to extend, fully cover, and permanently lock over the needle when activated by a fingertip or thumb to reduce the occurrence of accidental needle sticks. The device is provided sterile and is intended for single use only. An orange needle cap indicates for use with insulin. The barrel of the syringe is graduated for measuring a prescribed dose. The syringe is intended for use immediately after filling and is not intended to contain or store insulin for extended periods of time.
A complete list of the subject devices proposed model numbers is included in Table 1
| Product Code | Product Description |
|--------------|-------------------------------------------------------------------|
| 8881893130 | Monoject Magellan Insulin Safety Syringe, 0.3mL, 31G x 6mm needle |
| 8881893150 | Monoject Magellan Insulin Safety Syringe, 0.5mL, 31G x 6mm needle |
| 8881893110 | Monoject Magellan Insulin Safety Syringe, 1.0mL, 31G x 6mm needle |
# Table 1: Proposed Subiect device model numbers
### Indications for Use
Monoject Magellan Insulin Safety Syringes are intended for delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.
{4}------------------------------------------------
## Comparison of Technological Characteristics with the Predicate Device
Subject Monoject Magellan Insulin Safety Syringe has similar materials, and identical design, components, technological characteristics, intended use, and indications for use as the predicate device. Biocompatibility evaluation and testing in conformance with ISO 10993 series standards was completed to verify similar materials. Differences exist in needle gauge, length, and wall thickness; needle of subject device is smaller, shorter, and thinner than needle of predicate. Differences have been verified in conformance with FDA recognized standard ISO 8537:2016 Sterile single-use syringes, with or without needle, for insulin.
| Description | Subject Device | Predicate Device | Comparison |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Monoject Magellan | Kendall Monoject | |
| | Insulin Safety Syringe | Magellan Insulin Safety | |
| | K223376 | Syringe | |
| | | K061492 | |
| Indications for use | Monoject Magellan Insulin<br>Safety Syringes are intended<br>for delivery of U-100<br>insulin. The needle stick<br>prevention feature of the<br>device, once activated,<br>guards against accidental<br>needle-sticks. | The device is intended for<br>the delivery of U-100<br>insulin. The needle stick<br>prevention feature of the<br>device, once activated,<br>guards against accidental<br>needle-sticks. | Same |
| Intended Use | Primarily intended for intra-<br>muscular, subcutaneous,<br>and/or intra-dermal<br>medication delivery.<br>Secondary use is a safety<br>feature that helps prevent<br>sharps injuries. | Primarily intended for intra-<br>muscular, subcutaneous,<br>and/or intra-dermal<br>medication delivery.<br>Secondary use is a safety<br>feature that helps prevent<br>sharps injuries. | Same |
| Syringe Type | Piston Syringe | Piston Syringe | Same |
| Principle<br>of<br>Operation | Manually extendable safety<br>shield, one handed<br>activation | Manually extendable safety<br>shield, one handed<br>activation | Same |
| Specific Drug Use | Insulin | Insulin | Same |
| Needle size | 31G | 29G, 30G | Different |
| Needle length | 6mm (15/64") | 8mm (5/16")<br>13mm (1/2") | Different |
| Wall thickness | Extra thin wall | Regular Wall | Different |
| Needle bevel | Conforms to ISO 7864 | Conforms to ISO 7864 | Same |
| Needle<br>Sheath<br>Color | Orange | Orange | Same |
| Needle Lubricant | Silicone | Silicone | Same |
| Needle Material | Type 304 Stainless Steel | Type 304 Stainless Steel | Same |
| Syringe volumes | 0.3mL<br>0.5mL<br>1.0mL | 0.3mL<br>0.5mL<br>1.0mL | Same |
| Barrel Lubricant | Silicone | Silicone | Same |
| Description | Subject Device<br>Monoject Magellan<br>Insulin Safety Syringe<br>K223376 | Predicate Device<br>Kendall Monoject<br>Magellan Insulin Safety<br>Syringe<br>K061492 | Comparison |
| Barrel Materials | Polypropylene | Polypropylene | Similar |
| Barrel Nozzle | Fixed Needle | Fixed Needle | Same |
| Barrel<br>Transparency | Clear | Clear | Same |
| Safety Shield<br>Material | Polyethylene | Polyethylene | Same |
| Sterilization<br>Method | Gamma Radiation | Gamma Radiation | Same |
| SAL | 10-6 | 10-6 | Same |
| Reuse Durability | Single use only | Single use only | Same |
| Delivery Accuracy | Conforms to ISO 8537 | Conforms to ISO 8537 | Same |
| Hub/Needle Bond<br>Strength | Conforms to ISO 8537 | Conforms to ISO 8537 | Same |
| Biocompatibility | Conforms to ISO 10993-1 | Conforms to ISO 10993-1 | Same |
### Table 2: Comparison of Technological Characteristics of Subject Device with Predicate Device
{5}------------------------------------------------
### Biocompatibility
Biocompatibility evaluation was conducted considering worst-case, clinical use, and potential repeat contact in conformance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Applicable testing was completed utilizing the ISO 10993 series standards following Good Laboratory Practices. Monoject™ Magellan™ Insulin Safety Syringes with permanently attached 31G x 6mm needle are concluded to be toxicologically acceptable for the intended use.
### Sterility
Monoject™ Magellan™ Insulin Safety Syringes with permanently attached 31G x 6mm needle is sterilized using gamma irradiation to a sterility assurance level of 10 %. Devices are nonpyrogenic as determined by the Limulus Amebocyte Lysate (LAL) Test. Validation and testing was completed following FDA Guidance Submission and Review of Sterility Information in Premarket Notification (510(k) Submissions for Devices Labeled as Sterile, ISO 11137-2:2013 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, ANSI AAMI ST72:2019 Bacterial Endotoxins - Test methods, routine monitoring, and alternatives to batch testing, and USP Chapter <85>, Bacterial Endotoxins Test.
## Performance Data
### Nonclinical tests
Monoject™ Magellan™ Insulin Safety Syringes with permanently attached 31G x 6mm needle underwent evaluation following FDA Guidance documents and FDA recognized standards to support substantial equivalence determination:
- Guidance on the Content of Premarket Notification [510(K)] Submissions for Piston Syringes ●
{6}------------------------------------------------
- ISO 8537:2016 Sterile single-use syringes with or without needle for insulin
- . ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- . ISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ISO 11607-2:2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- USP <788> Particulate Testing .
### Conclusion
Results of non-clinical testing support the conclusion that Monoject™ Magellan™ Insulin Safety Syringes with permanently attached 31G x 6mm needle are as safe, as effective, and performs as well as the predicate device, therefore subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.