Identity Imprint Porous Total Knee Replacement System, Identity Imprint Porous Cruciate Retaining Total Knee Replacement System
Applicant
Conformis, Inc.
Product Code
MBH · Orthopedic
Decision Date
Jul 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - Painful joint disease due to osteoarthritis, traumatoid arthritis or osteonecrosis of the knee. - Post traumatic loss of joint function. - Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. - Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. - Revision procedures, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for uncemented use, but porous implants may be used with cement if desired by the surgeon.
Device Story
Total knee replacement system; includes femoral, tibial, and patellar components; utilizes patient-specific CT imaging for implant selection and ancillary surgical instrumentation. Femoral component: CoCrMo alloy with CP Ti porous scaffold; Tibial tray: Ti6Al4V-ELI alloy with CP Ti porous scaffold; Tibial insert: Vitamin-E enriched highly cross-linked UHMWPE; Patellar component: UHMWPE with Ti6Al4V-EL backing and CP Ti porous scaffold. Porous coating facilitates biological fixation (cementless), though cement may be used. Used in sterile OR by orthopedic surgeons. Patient-specific nylon guides assist in component positioning and bone cutting. System provides standardized implant sizes combined with patient-specific instrumentation to address severe knee pain/disability.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical performance evaluation including femoral and tibial fatigue testing, patella shear/tensile testing, porous bond strength testing, tibial micromotion, patella durability/wear testing, MRI compatibility, and cadaveric testing.
Technological Characteristics
Implants: CoCrMo femoral, Ti6Al4V-ELI tibial tray, Vitamin-E enriched UHMWPE insert, UHMWPE patella with Ti6Al4V-EL backing. All feature CP Ti porous scaffold for biological fixation. Manual surgical instruments: biocompatible nylon (single-use) and reusable metal. Semi-constrained design. Sterilization: Gamma irradiation (implants/single-use instruments).
Indications for Use
Indicated for patients with knee joint pain/disability due to osteoarthritis, traumatic/rheumatoid arthritis, or osteonecrosis; post-traumatic loss of function; moderate varus, valgus, or flexion deformity; or failed prior implants (osteotomies, hemiarthroplasties, unicondylar/patellofemoral/bicompartmental).
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Identity Imprint Knee Replacement System / Identity Imprint CR Knee Replacement System (K221404)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
July 21, 2023
Conformis Inc. Marv Kruitwagen Sr. Regulatory Affairs Specialist 600 Technology Park Drive, 4th Floor Billerica, Massachusetts 01821
Re: K223316
Trade/Device Name: Identity Imprint Porous Total Knee Replacement System, Identity Imprint Porous Cruciate Retaining Total Knee Replacement System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH, JWH, OIY, OOG Dated: October 19, 2022 Received: October 28, 2022
Dear Mary Kruitwagen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Lixin Liu-S
Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K223316
#### Device Name
Identity Imprint Porous Total Knee Replacement System, Identity Imprint Porous Cruciate Retaining Total Knee Replacement System
#### Indications for Use (Describe)
The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
- · Painful joint disease due to osteoarthritis, traumatoid arthritis or osteonecrosis of the knee.
- · Post traumatic loss of joint function.
- · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
- · Revision procedures, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for uncemented use, but porous implants may be used with cement if desired by the surgeon.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written below it. The first three letters of the word are in black, and the last five letters are in orange.
# Traditional 510(k) Summary
| 510(k) Number | K223316 |
|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Conformis, Inc.<br>600 Technology Park Drive, Fourth Floor Billerica,<br>Massachusetts 01821<br>USA |
| Main Telephone Number Main<br>Fax Number Establishment<br>Registration Numbers(s): | 781-345-9164<br>781-345-0147<br>3009844603 and 3004153240 |
| Date of Summary Preparation | October 19, 2022; updated November 30, 2022 |
| Contact Name:<br>Title:<br>Contact Telephone:<br>Contact email: | Mary Kruitwagen<br>Sr. Regulatory Affairs Specialist<br>781-345-9038<br>Mary.Kruitwagen@conformis.com |
| Alternate Contact:<br>Title:<br>Alternate Contact Telephone:<br>Alternate Contact email: | Liz Haines<br>Vice President of Regulatory Affairs<br>(978) 569-6862<br>Elizabeth.Haines@conformis.com |
| Subject Device: (Proprietary/<br>Trade Name) | Identity Imprint Porous Total Knee Replacement System,<br>Identity Imprint Porous Cruciate Retaining Total Knee<br>Replacement System |
| Common Usual Name | Knee Replacement System |
| Type of Submission | Traditional 510(k) |
| Device Class | II |
| Regulation Number(s): | 21 CFR 888.3565 - Knee joint patellofemorotibial metal/<br>polymer porous-coated uncemented prosthesis |
| | 21 CFR 888.3560 - Knee joint patellofemorotibial polymer/<br>metal/polymer semiconstrained cemented prosthesis |
| Product Code(s) | MBH: prosthesis, knee, patello/femorotibial, semi-<br>constrained, uncemented, porous, coated, polymer/metal/<br>polymer<br>JWH: prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer<br>OIY: prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer + additive/metal/polymer + additive<br>OOG: knee arthroplasty implantation system |
| Classification Panel | Orthopedics |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter C in orange and black, with the word "CONFORMIS" written in black, except for the "ORM" which is in orange. The logo is simple and modern, and the colors are eye-catching.
| Manufacturing Site | Conformis Inc.<br>600 Research Drive<br>Wilmington, Ma. 01887<br>USA |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration | 3004153240 |
| Sterilization Site | Isomedix Operations Inc.<br>Steris Isomedix Services<br>3459 South Clinton Ave.<br>S. Plainfield, NJ 07080<br>USA |
| Establishment Registration | 2246552 |
| Primary Predicate Device: | Arthrex iBalance® TKA System |
| Primary Predicate Device 510(k): | K141635, September 3, 2014 |
| Primary Predicate Device Product<br>Code(s): | MBH: prosthesis, knee, patello/femorotibial, semi-<br>constrained, uncemented, porous, coated, polymer/metal/<br>polymer<br>JWH: prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer |
| Secondary Predicate Device: | Identity Imprint Knee Replacement System<br>Identity Imprint CR Knee Replacement System |
| Secondary Predicate Device 510(k): | K221404 |
| Secondary Predicate Device Product<br>Codes: | JWH: prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer<br>OIY: prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer + additive/metal/polymer + additive<br>OOG: knee arthroplasty implantation system |
### Device Description:
The subject device, Identity Imprint Porous Total Knee Replacement System (including Identity Imprint Porous Cruciate Retaining Total Knee Replacement System) is a knee replacement system including standardized implant sizes combined with standard and patient-specific ancillary surgical instrumentation. As with other Conformis Knee Replacement Systems (KRS), the Identity Imprint Porous Knee Replacement System (KRS) is semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis or trauma. The device is intended for cementless fixation however the surgeon may also use cement.
Using patient imaging (CT scans), an Identity Imprint set of implants is selected. The femoral component is manufactured from a cobalt chromium molybdenum (CoCrMo) alloy with a porous Commercially Pure
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the ConforMIS logo. The logo features the word "CONFORMIS" in a sans-serif font, with the "FORMIS" portion of the word in orange. Above the word is a stylized "C" shape, with the top portion in orange and the bottom portion in black.
titanium (CP Ti) scaffold on the interior surface. The tibial tray is manufactured from alloy (Ti6Al4V-ELI), with a CP Ti porous scaffold on the interior surface. The tibial insert component is manufactured from highly cross-linked ultra-high molecular weight Vitamin-E enriched polyethylene (iPoly XE). The patellar component is manufactured and offered in ultra-high molecular weight polyethylene (iPoly) with a solid titanium alloy (Ti6Al4V-EL) and CPTi porous metal backing. The layer of CPTi scaffolding bonded to the femoral, tibial, and patellar implants provides a surface for porous ingrowth, promoting biological fixation and obviating the need for bone cement to achieve fixation. Porous tibial, femoral, and patellar implants are designed for use without cement, but may be used with a cemented technique if necessary.
For user convenience, single-use, standard and patient-specific ancillary orthopedic manual surgical instruments designed for use with the selected Identity Imprint implants are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding of bone. These guides are pre-navigated to fit the contours of the patient's femoral and tibial anatomies and to facilitate a simpler surgical technique. The ilig instrument set is designed for single-use, and manufactured from biocompatible nylon material and supplied sterile along with the implants. In addition, reusable orthopedic manual surgical instruments are provided separately.
The device is intended to be used in a sterile field by trained orthopedic surgeons (Use Environment).
The Imprint Porous Knee Replacement System is compatible with cemented Identity Imprint CR KRS implants. Cemented and uncemented implants may be used together for a hybrid technique.
### The Indications for Use:
The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
- . Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee.
- Post traumatic loss of joint function.
- Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- . Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
- . Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans
The Identity Imprint Porous Cruciate Retaining Total Knee Replacement System is intended for uncemented use, but porous implants may be used with cement if desired by the surgeon.
### Technological Characteristics:
The subject device single-use, disposable patient-specific instrumentation, implants, and reusable instruments are similar to the predicate device in operating principle, fundamental technology, design, and materials. The subject device implants use the same materials as the predicate Identity Imprint CR
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written in black and orange letters below it. The "ORM" in Conformis is in orange.
KRS device, with a porous coating on the bone cut sides of the femoral and tibial implants. The patellar component also has a porous coating on metal backing. The same as used in the predicate Arthrex iBalance® TKA device. The fixation method of the subject device is for cementless application, although the surgeon may use cement if desired. The subject device uses the same packaging materials and sterilization methods as those of the predicate device. The manual and automated manufacturing methods remain the same as or similar to the predicate Identity Imprint CR KRS.
The focus of this submission is to seek clearance for a porous version of the Identity Imprint Porous Total Knee Replacement System along with hybrid combinations of the cemented implants in K221404.
### Non-Clinical Performance Evaluation:
There is no change to the articulating surfaces, UHMWPe inserts or interlocks compared to the predicate Identity Imprint CR KRS device. The testing performed and included in this submission is relative to the areas of change from the - predicate Identity Imprint CR KRS device. The following confirmatory testing was performed:
- . Femoral and Tibial Fatigue testing
- Patella Shear Testing
- Patella Tensile Testing
- Porous Bond Shear and Tensile Strength
- Tibial Micromotion
- Patella Durability/Wear Thru Testing
- MRI Compatibility Testing
- Cadaveric Testing
The subject device verification and validation data provided in this submission supports that the subject device is as safe, effective, and performs as well as or better than the predicate device. No different issues of safety or effectiveness were raised.
### Conclusion:
Based on a comparison of the intended use and technological characteristics to the predicate devices and on the results of confirmatory testing, it is concluded that the proposed Identity Imprint Porous Total Knee Replacement System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.