Antimicrobial Silicone PHMB Foam Wound Dressing
K223310 · Advanced Medical Solutions Limited · FRO · Apr 5, 2023 · SU
Device Facts
| Record ID | K223310 |
| Device Name | Antimicrobial Silicone PHMB Foam Wound Dressing |
| Applicant | Advanced Medical Solutions Limited |
| Product Code | FRO · SU |
| Decision Date | Apr 5, 2023 |
| Decision | SESE |
| Submission Type | Special |
| Device Class | Class U |
| Attributes | Therapeutic |
Intended Use
Antimicrobial Silicone PHMB Foam Wound Dressings are indicated for use in the management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, lacerations, superficial and partial thickness burns, dermatologic disorders, other wounds inflicted by trauma and, as a secondary dressing or cover dressing for packed wounds.
Device Story
Trilaminate polyurethane foam dressing impregnated with 0.8-1.1% w/w PHMB; microporous hydrophilic structure absorbs exudate; maintains moist healing environment; perforated silicone adhesive contact layer ensures non-traumatic removal. Used in clinical settings or home care for various wound types; applied by healthcare professionals or patients. PHMB provides antimicrobial barrier against bacterial penetration, colonization, and proliferation of bacteria, yeast, and mold within dressing for up to 7 days. Semi-occlusive film allows gas/moisture exchange while blocking bacteria. Benefits include infection protection, exudate management, and atraumatic dressing changes.
Clinical Evidence
Bench testing only. In-vitro microbial efficacy testing performed using modified AATCC TM 100 method against mold challenge organisms. Biocompatibility (ISO 10993-1, USP <151>), absorbency (BS EN 13726-1), moisture vapor transmission (BS EN 13726-2), waterproofness (BS EN 13726-3), peel resistance (ASTM D6282-11), and distribution testing (ASTM D4169) leveraged from predicate device.
Technological Characteristics
Trilaminate polyurethane foam; 0.8-1.1% w/w PHMB antimicrobial agent; silicone adhesive contact layer. Semi-occlusive film. Dimensions: 10.24 in² to 64 in². Terminal sterilization via ethylene oxide (SAL 10⁻⁶). Standards: ISO 10993-1, USP 41-NF36, BS EN 13726-1/2/3, ASTM D6282-11, ASTM D4169.
Indications for Use
Indicated for management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, lacerations, superficial/partial thickness burns, dermatologic disorders, trauma wounds, and as secondary/cover dressing for packed wounds.
Predicate Devices
- Silicone PHMB Foam Wound Dressing (K190819)
Related Devices
- K190819 — Silicone PHMB Foam Wound Dressing · Advanced Medical Solutions, Ltd. · Jul 25, 2019
- K181197 — PHMB Foam Wound Dressing · Advanced Medical Solutions, Ltd. · Aug 3, 2018
- K192463 — PHMB Foam Dressing Non-Adhesive, PHMB Foam Dressing Adhesive, Silicone PHMB Foam Dressing, Silicone PHMB Foam Dressing with Border · Winner Medical Co., Ltd. · May 14, 2020
- K163062 — PHMB Antimicrobial Foam Wound Dressing · Advanced Medical Solutions, Ltd. · Jul 25, 2017
- K071371 — COPA AMD ANTIMICROBIAL WOUND DRESSING · Tyco Healthcare Group LP · Nov 19, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2023
Advanced Medical Solutions Limited Kay McGrath Regulatory Affairs Specialist Premier Park, 33 Road One Winsford Industrial Estate Winsford, Cheshire CW7 3RT United Kingdom
Re: K223310
Trade/Device Name: Antimicrobial Silicone PHMB Foam Wound Dressing Regulatory Class: Unclassified Product Code: FRO Dated: March 6, 2023 Received: March 6, 2023
Dear Kay Mcgrath:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Julie A. Morabito -S
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K223310
### Device Name
Antimicrobial Silicone PHMB Foam Wound Dressing
Indications for Use (Describe)
Antimicrobial Silicone PHMB Foam Wound Dressings are indicated for use in the management of
- · post-surgical incisions,
- · pressure sores,
- · venous stasis ulcers,
- · diabetic ulcers,
- · donor sites.
- · abrasions,
- · lacerations,
- · superficial and partial thickness burns,
- · dermatologic disorders,
- · other wounds inflicted by trauma and,
- · as a secondary dressing or cover dressing for packed wounds.
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue "A" with a swoosh above the text. The text "Advanced Medical Solutions Ltd" is written in blue font below the "A" symbol.
Advanced Medical Solutions Limited Premier Park, 33 Road One, Winsford Industrial Estate, Winsford, Cheshire, CW7 3RT, UK Tel: +44 (0) 1606 863500 Fax: +44 (0) 1606 863600 Web: www.admedsol.com Registered in England 2666957 VAT No. GB 636 5551 27
# K223310
# 510(k) Summary
| Submitted by: | Advanced Medical Solutions Ltd<br>Premier Park<br>33 Road One<br>Winsford Industrial Estate<br>Winsford<br>Cheshire<br>CW7 3RT<br>Tel: +44 1606 863500 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kay McGrath |
| Date of Summary: | April 04, 2023 |
| Trade Name: | Antimicrobial Silicone PHMB Foam Wound Dressing |
| Common Name: | Wound Dressing |
| Classification Name: | Dressing, Wound, Drug |
| Classification: | Unclassified (Pre-amendment) |
| Classification Code: | Product code: FRO |
| Predicate Device(s): | Silicone PHMB Foam Wound Dressing (K190819) |
Image /page/3/Picture/5 description: The image contains two logos. The first logo is for the UK Authorised Economic Operator, and it features the letters AEO in bold black font, with a map of the UK inside of a circle. The second logo is for BSI, and it features the letters BSI in a circle above the words UKAS MANAGEMENT SYSTEMS 0003.
Certificate No. MD 78010
Page 1 of 5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Advanced Medical Solutions Ltd. The logo features a blue stylized letter "A" with a curved line extending from the top right. Below the logo, the text "Advanced Medical Solutions Ltd" is written in blue.
# K223310
#### Device Antimicrobial Silicone PHMB Foam Wound Dressing, is a polyurethane foam trilaminate dressing Description: impregnated with Polyhexamethylene Biguanide (PHMB), an agent that protects the dressing from bacterial penetration and colonization. The foam in the dressing has a microporous hydrophilic foam structure that absorbs wound exudate and maintains a moist wound healing environment.
Based on in vitro performance data, the Antimicrobial Silicone PHMB Foam Wound Dressing provides a barrier to bacterial penetration through the dressing and the PHMB prevents colonization and proliferation of bacteria, yeast and mold within the dressing for up to 7 days. Antimicrobial Silicone PHMB Foam Wound Dressing, when tested in-vitro has demonstrated to be effective against gram positive bacteria, gram negative bacteria, yeast and mold challenge organisms within the dressing.
The perforated wound contact layer contains a gentle silicone adhesive that provides secure, nonirritating adhesion and supports non-traumatic removal during dressing changes.
The device is presented in a border (adhesive) version. The dressing is supplied sterile in a range of sizes between 10.24 in2 (64cm) to 64 in2 (400cm).
Indications for Antimicrobial Silicone PHMB Foam Wound Dressing is indicated for use in the management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, Use: lacerations, superficial and partial thickness burns, dermatologic disorders, other wounds inflicted by trauma and as a secondary dressing or cover dressing for packed wounds.
Image /page/4/Picture/8 description: The image contains two logos. The first logo is for the UK Authorized Economic Operator, and it features the letters "AEO" in bold, black font, followed by the words "UK Authorized Economic Operator" in a smaller font. To the right of the letters is a black circle containing a white outline of the United Kingdom. The second logo is for BSI (British Standards Institution) and UKAS (United Kingdom Accreditation Service). The BSI logo features the letters "bsi" above a heart-shaped symbol, while the UKAS logo features a crown above a checkmark, with the words "UKAS MANAGEMENT SYSTEMS" below. The number "0003" is at the bottom of the UKAS logo.
Page 2 of 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Advanced Medical Solutions Ltd. The logo features a stylized letter A in blue, with a curved line extending from the top right of the A. The text "Advanced Medical Solutions Ltd" is written in blue below the logo.
# K223310
Comparison of Technological Characteristics:
Antimicrobial Silicone PHMB Foam Wound Dressing is a modification of the predicate device to include updated product labelling in support of an antimicrobial product offering.
Antimicrobial Silicone PHMB Foam Wound Dressing has substantially equivalent intended use and performance characteristics, identical design, materials and manufacture process to the predicate device Silicone PHMB Foam Wound Dressing (K190819). Antimicrobial Silicone PHMB Foam Wound Dressing is a multi-layer one piece dressing design incorporating an absorbent polyurethane foam pad containing 0.8-1.1 %w/w PHMB. Based on in vitro testing, the PHMB within the dressing is efficacious against gram positive bacteria, gram negative bacteria, yeast and mold. The dressing is semi-occlusive allowing the exchange of gases within the dressing such as oxygen and moisture, and has a film that provides a barrier to bacterial penetration through the dressing. The silicone wound contact layer contains a gentle silicone adhesive that provides secure, non-irritating adhesion and supports non-traumatic removal during dressing changes. The following table shows the comparison of technological characteristics between the subject and predicate devices.
| | Subject<br>(modified device) | Predicate | Similarities and<br>Differences |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Product Name | Antimicrobial Silicone<br>PHMB Foam Wound<br>Dressing | Silicone PHMB Foam<br>Wound Dressing | |
| Manufacturer | Advanced Medical<br>Solutions Ltd | Advanced Medical<br>Solutions Ltd | Identical |
| 510(k) | K223310 | K190819 | - |
| Classification | Unclassified<br>(pre-amendment) | Unclassified<br>(pre-amendment) | Identical |
| Product code | FRO (Dressing,<br>Wound, Drug) | FRO (Dressing,<br>Wound, Drug) | Identical |
| Indications for use | Antimicrobial Silicone<br>PHMB Foam Wound<br>Dressings are<br>indicated for use in<br>the management of<br>post-surgical incisions,<br>pressure sores,<br>venous stasis ulcers,<br>diabetic ulcers, donor<br>sites, abrasions,<br>lacerations, superficial<br>and partial thickness<br>burns, dermatologic<br>disorders, other<br>wounds inflicted by<br>trauma and,<br>as a secondary<br>dressing or cover | Silicone PHMB Foam<br>Wound Dressings are<br>indicated for use in<br>the management of<br>post-surgical<br>incisions, pressure<br>sores, venous stasis<br>ulcers, diabetic<br>ulcers, donor sites,<br>abrasions,<br>lacerations, 1st and<br>2nd degree burns,<br>dermatologic<br>disorders, other<br>wounds inflicted by<br>trauma and,<br>as a secondary<br>dressing or cover | Different - Subject<br>device includes an<br>antimicrobial claim |
Image /page/5/Picture/7 description: The image shows the text "Page 3 of 5". This indicates that the image is a page from a document, specifically page 3 of a total of 5 pages. The text is written in a clear, bold font, making it easily readable. The purpose of this text is to provide context and indicate the position of the page within the larger document.
Image /page/5/Picture/8 description: The image shows three logos. The first logo is for the UK Authorised Economic Operator, and it features the letters AEO in bold black font, with a graphic of the UK in a circle. The second logo is for BSI, and it features the letters BSI in a circle with a heart-shaped design. The third logo is for UK Management Systems, and it features a checkmark inside of a circle with a crown on top.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue "A" with a swoosh extending from the top right. Below the logo, the company name "Advanced Medical Solutions Ltd" is written in blue font.
Advanced Medical Solutions Limited Premier Park, 33 Road One, Winsford Industrial Estate, Winsford, Cheshire, CW7 3RT, UK Tel: +44 (0) 1606 863500 Fax: +44 (0) 1606 863600 Web: www.admedsol.com Registered in England 2666957 VAT No. GB 636 5551 27
# K223310
| | dressing for packed<br>wounds. | dressing for packed<br>wounds. | |
|------------------------------------|-----------------------------------------|-----------------------------------------|-----------|
| Primary Material | Polyurethane (PU)<br>foam | Polyurethane (PU)<br>foam | Identical |
| Antimicrobial agent | PHMB<br>(0.8-1.1%w/w) | PHMB<br>(0.8-1.1%w/w) | Identical |
| Sterilization method<br>(terminal) | Ethylene oxide<br>SAL 10⁻⁶ | Ethylene oxide<br>SAL 10⁻⁶ | Identical |
| Biocompatibility | Biocompatible | Biocompatible | Identical |
| Storage | Store below 25°C<br>(77°F) | Store below 25°C<br>(77°F) | Identical |
| Range of available<br>sizes | 10.24 in² (64 cm) to 64<br>in² (400 cm) | 10.24 in² (64 cm) to<br>64 in² (400 cm) | Identical |
Image /page/6/Picture/4 description: The image shows the text "Page 4 of 5". This indicates that the image is a page from a document. The document has at least 5 pages. This is page number 4.
Image /page/6/Picture/5 description: The image contains two logos. The first logo is for the UK Authorised Economic Operator, and it features the letters "AEO" in bold black font, with a graphic of the United Kingdom inside of a circle. The second logo is for BSI (British Standards Institution), and it features the BSI logo with the UKAS Management Systems logo. The certificate number MD 78010 is printed below the BSI logo.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue "A" with a curved line extending from the top right. The text "Advanced Medical Solutions Ltd" is written in blue below the logo.
# K223310
# Performance Testing Summary:
Page 5 of 5
Microbial efficacy performance data submitted in support of this 510(k) includes in-vitro testing against a mold challenge organism. Testing was performed on real time aged predicate device in accordance with the well-established modified AATCC TM 100 method previously used for the predicate device.
No other performance tests were conducted for this submission. All performance data leveraged from the predicate device was submitted as part of the original 510(k) submission of the predicate, Silicone PHMB Foam Wound Dressing (K190819) and includes:
## Biocompatibility
ISO 10993-1; Biological evaluation of medical devices USP 41-NF36; <151> Pyrogenic Test
## Performance testing
BS EN 13726-1; Test methods for primary wound dressings – aspects of absorbency. BS EN 13726-2; Test methods for primary wound dressings – moisture vapour transmission rate of permeable film dressings. BS EN 13726-3; Test methods for primary wound dressings - waterproofness. ASTM D6282-11; Standard Test Method for 90 Degree Peel Resistance of Adhesives. Bacterial barrier.
## Distribution
ASTM D4169 – Standard Practice for Performance Testing of Shipping Containers and Systems
The subject device, Antimicrobial Silicone PHMB Foam Wound Dressing, is manufactured with the exact same materials and processes as the predicate.
- Rationale for The modified device, Antimicrobial Silicone PHMB Foam Wound Dressing, is identical to the predicate, Silicone PHMB Foam Wound Dressing (K190819), with regard to technology, materials, Substantial Equivalence: manufacture process, intended use, and target population. The only difference between the predicate and subject device is that the subject device has an antimicrobial claim, this minor modification does not raise any new questions of safety or effectiveness. Therefore, the Antimicrobial Silicone PHMB Foam Wound Dressing, is substantially equivalent to the predicate, Silicone PHMB Foam Wound Dressing (K190819).
- Conclusion: Antimicrobial Silicone PHMB Foam Wound Dressing is substantially equivalent to the predicate device listed when compared to the technological characteristics such as design, materials, chemical composition, and manufacture and are supplied sterile for single use. Based on the information provided within this 510(k), Advanced Medical Solutions Ltd. concludes that the proposed Antimicrobial Silicone PHMB Foam Wound Dressing is substantially equivalent to the predicate device listed, Silicone PHMB Foam Wound Dressing (K190819).
Image /page/7/Picture/15 description: The image shows two logos side by side. The first logo has the letters "AEO" in bold, followed by a graphic of the UK map inside a circle, and the words "UK Authorised Economic Operator" underneath. The second logo is a square containing two circular logos, one with the letters "bsi" and a heart-shaped design, and the other with a crown, a checkmark, and the words "UKAS MANAGEMENT SYSTEMS" and the number "0003".
Certificate No. MD 78010