AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223240 · Apr 3, 2023
Annalise Enterprise CTB Triage Trauma
Annalise-Ai Pty , Ltd.
Retrospective, anonymized clinical CT brain imaging studies from five US hospital network sites
The device's AI algorithm performance was validated using a large, retrospective, multi-site clinical dataset of non-contrast brain CT scans to demonstrate sensitivity and specificity for triage findings.
Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings: - acute subdural/epidural hematoma* - acute subarachnoid hemorrhage * - intra-axial hemorrhage* - intraventricular hemorrhage* *These findings are intended to be used together as one device. The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage. The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings. Its results are not intended: - to be used on a standalone basis for clinical decision making - to rule out specific findings, or otherwise preclude clinical assessment of CTB studies Intended modality: Annalise Enterprise identifies suspected findings in non-contrast brain CT studies. Intended user: The device is intended to be used by trained clinicians who, as part of their scope of practice, are qualified to interpret brain CT studies. Intended patient population: The intended population is patients who are 22 years or older.
Device Story
Annalise Enterprise CTB Triage Trauma is a software-only workflow tool; operates in medical care environments; processes non-contrast brain CT studies. Device ingests DICOM images; utilizes deep-learning convolutional neural network (CNN) to identify four specific intracranial hemorrhage findings. Output consists of study-level notifications sent to PACS/RIS worklists to prioritize studies containing suspected findings. Device does not alter image data; does not downgrade existing worklist priorities; operates in parallel to standard clinical workflows. Intended for use by clinicians qualified to interpret brain CTs. Benefits include reduced turnaround time for critical findings; improved triage efficiency. Results serve as notification only; clinicians must perform final diagnostic assessment.
Clinical Evidence
Bench testing only. Retrospective study of 1,485 cases (slice thickness ≤1.5mm) and 1,878 cases (slice thickness >1.5mm) from five US hospital networks. Sensitivity ranged from 82.4% to 98.0%; specificity ranged from 85.1% to 100% across findings. Triage effectiveness study (n=277) demonstrated a mean turnaround time of 81.6 seconds (95% CI: 80.3-82.9).
Technological Characteristics
Software-only; DICOM-compliant; operates on non-contrast brain CT images. Uses deep-learning convolutional neural network (CNN). No hardware components. Connectivity via integration with PACS/RIS. Software-based triage and notification.
Indications for Use
Indicated for triage and prioritization of non-contrast brain CT studies in patients 22 years or older to identify suspected acute subdural/epidural hematoma, acute subarachnoid hemorrhage, intra-axial hemorrhage, and intraventricular hemorrhage. Not for standalone clinical decision-making or ruling out findings.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
April 3, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION".
Annalise-AI Pty Ltd. % Haylee Bosshard Regulatory Affairs Manager Level P. 24 Campbell Street Sydney. New South Wales 2000 AUSTRALIA
### Re: K223240
Trade/Device Name: Annalise Enterprise CTB Triage Trauma Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: March 1, 2023 Received: March 1, 2023
### Dear Haylee Bosshard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
for
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223240
Device Name
Annalise Enterprise CTB Triage Trauma
### Indications for Use (Describe)
Intended context:
Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings:
- · acute subdural/epidural hematoma*
· acute subarachnoid hemorrhage *
· intra-axial hemorrhage*
· intraventricular hemorrhage*
*These findings are intended to be used together as one device.
The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage.
The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings.
Its results are not intended:
· to be used on a standalone basis for clinical decision making
· to rule out specific findings, or otherwise preclude clinical assessment of CTB studies
Intended modality:
Annalise Enterprise identifies suspected findings in non-contrast brain CT studies.
Intended user:
The device is intended to be used by trained clinicians who, as part of their scope of practice, are qualified to interpret brain CT studies.
Intended patient population:
The intended population is patients who are 22 years or older.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K223240
### 510(k) Summary
#### I. SUBMITTER
| Company Name | Annalise-AI Pty Ltd |
|----------------|--------------------------------------------------------------|
| Address | Level P, 24 Campbell Street<br>Sydney, NSW 2000<br>Australia |
| Phone Number | +61 1800-958487 |
| Contact Person | Haylee Bosshard |
| Date Prepared | March 31, 2023 |
#### II. SUBJECT DEVICE
| Manufacturer Name | Annalise-AI Pty Ltd |
|---------------------|---------------------------------------------------------------------------------|
| Device Name | Annalise Enterprise CTB Triage Trauma |
| Classification Name | Radiological computer aided triage and notification software<br>(21CFR892.2080) |
| Regulatory Class | II |
| Product Code | QAS |
#### PREDICATE DEVICE III.
| Manufacturer Name | Infervision Medical Technology Co., Ltd. |
|---------------------|---------------------------------------------------------------------------------|
| Device Name | InferRead CT Stroke.AI |
| 510(k) reference | K211179 |
| Classification Name | Radiological computer aided triage and notification software<br>(21CFR892.2080) |
| Regulatory Class | II |
| Product Code | QAS |
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{5}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Annalise Enterprise CTB Triage Trauma is a software workflow tool which uses an artificial intelligence (Al) algorithm to identify suspected findings on non-contrast brain CT studies in the medical care environment. The findings identified by the device include acute subdural/ epidural hematoma, acute subarachnoid hemorrhage, intra-axial hemorrhage, and intraventricular hemorrhage.
Radiological findings are identified by the device using an AI algorithm - a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets that included a range of equipment manufacturers including Toshiba, GE Medical Systems, Siemens, Philips, and Canon Medical Systems. This dataset, which contained over 200,000 CT brain imaging studies, was labelled by trained radiologists regarding the presence of the four findings of interest.
The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified neuroradiologists.
The device interfaces with image and order management systems (such as PACS/RIS) to obtain noncontrast brain CT studies for processing by the AI algorithm. Following processing, if any of the clinical findings of interest are identified in a non-contrast brain CT study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these clinical findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results.
The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of non-contrast brain CTs. The device is intended to aid in prioritization and triage of radiological medical images only.
{6}------------------------------------------------
#### INDICATIONS FOR USE V.
The Indications for Use statement is as follows:
### Intended context Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings:
- acute subdural/ epidural hematoma* ●
- acute subarachnoid hemorrhage * ●
- intra-axial hemorrhage* ●
- intraventricular hemorrhage*
*These findings are intended to be used together as one device.
The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage.
The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings. Its results are not intended:
- to be used on a standalone basis for clinical decision making ●
- to rule out specific findings, or otherwise preclude clinical assessment . of CTB studies
Intended Annalise Enterprise identifies suspected findings in non-contrast brain CT modalitv studies. The device is intended to be used by trained clinicians who, as part of their Intended user scope of practice, are qualified to interpret brain CT studies. Intended patient The intended population is patients who are 22 years or older.
population
The Indications for Use statement of the subject device differs to the predicate device only in the clinical conditions of interest, however a standalone performance evaluation was conducted and demonstrated that the device is as safe and effective for its intended use. Both the subject and predicate device are intended for use to assist with worklist triage by providing notifications of suspected findings and their associated priority.
{7}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device was evaluated and compared to the predicate device with respect to the following characteristics:
- 1. Indications for Use
- 2. Anatomical site and modality
- 3. Intended user and clinical use environment
- 4. Technical method for notification and prioritization
- 5. Device input and radiological image protocol
- 6. System components
- 7. Location where results are received
- 8. Prioritization relationship to standard of care workflow
- 9. Ability to support effective triage
- 10. Device output and means of notification to user
The following characteristics showed a difference between the subject and predicate devices. The different characteristics include:
- 1. Set of findings and algorithm
The difference between the subject and predicate device is the set of findings that the subject device identifies and the underlying artificial intelligence algorithm. The performance of the subject device algorithm for each of the findings was addressed in a standalone performance evaluation and showed that the subject device is as safe and effective for its intended use as the predicate device.
{8}------------------------------------------------
### VII. PERFORMANCE DATA
The following performance data have been provided to support evaluation of substantial equivalence.
#### Software Verification and Validation Testing A.
Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005.
#### Performance Testing B.
Performance of the subject device was assessed in four performance studies to satisfy requirements set forth in the special controls per 21CFR892.2080. These included standalone performance and triage effectiveness evaluations.
Standalone performance was assessed via a retrospective, anonymized study of adult patient, DICOMcompliant non-contrast brain CT cases. The test dataset used during the standalone performance evaluation was newly acquired and independent from the training dataset used in model development. The standalone performance study was conducted on four independently assessed cohorts which equated to a total dataset of 1,485 cases for slice thickness ≤1.5mm (positive n=1,003 and negative n=482) and 1,878 cases for slice thickness >1.5mm (positive n=1,257 and negative n=621), collected consecutively from five US hospital network sites.
The performance testing datasets included representation across subgroups for patient demographics (gender [female: 44.9-52.2%, male: 47.8-55.1%], age [mean: 66.5-68.0 years, min: 22, max: 99-105], ethnicity [Hispanic: 5.9-11.3%], race [White/Caucasian: 76.6-82.1%, Other: 13.6-19.3%, Unknown: 2.7-6.9%]), co-existing findings or abnormalities and technical parameters (imaging equipment make, model). The datasets included GE Healthcare, Siemens and Toshiba CT scanners for the pivotal study. Additional analyses were conducted with GE, Philips, Siemens and Toshiba scanners to demonstrate the generalizability of the device.
To determine the ground truth, each deidentified case was annotated in a blinded fashion by at least two ABR-certified and protocol-trained neuroradiologists (ground truthers), with consensus determined by two ground truthers and a third ground truther in the event of disagreement. The key results of the study are summarized in the table below.
{9}------------------------------------------------
| Finding | Slice Thickness Range | Operating Point | Sensitivity % (Se)<br>(95% CI) | Specificity % (Sp)<br>(95% CI) |
|--------------------------------------|-----------------------|-----------------|--------------------------------|--------------------------------|
| Acute subdural/ Epidural<br>hematoma | <1.5mm | 0.060177 | 91.4 (88.1,94.4) | 86.7 (79.6,92.9) |
| | | 0.101143 | 89.1 (85.5,92.4) | 94.9 (89.8,99.0) |
| | | 0.135700 | 86.5 (82.5,90.1) | 96.9 (92.9,100.0) |
| | >1.5mm & ≤5.0mm | 0.060177 | 82.4 (78.6,86.1) | 89.6 (83.7,94.8) |
| Acute subarachnoid<br>hemorrhage | <1.5mm | 0.014372 | 98.0 (95.2,100.0) | 89.4 (82.4,95.3) |
| | | 0.060162 | 93.9 (89.8,97.3) | 96.5 (91.8,100.0) |
| | | 0.082652 | 89.8 (85.0,94.6) | 100 (100.0,100.0) |
| | >1.5mm & ≤5.0mm | 0.020255 | 90.7 (86.3,95.1) | 92.4 (86.7,97.1) |
| | | 0.030010 | 87.4 (82.4,91.8) | 96.2 (92.4,99.0) |
| Intra-axial hemorrhage | <1.5mm | 0.322700 | 93.1 (90.8,95.2) | 85.6 (81.1,89.6) |
| | >1.5mm & ≤5.0mm | 0.203600 | 93.4 (91.3,95.1) | 85.1 (80.9,88.9) |
| | | 0.322700 | 90.3 (87.9,92.5) | 90.3 (86.8,93.8) |
| Intraventricular hemorrhage | <1.5mm | 0.015487 | 95.9 (90.4,100.0) | 90.9 (84.4,97.4) |
| | | 0.051859 | 90.4 (83.6,97.3) | 97.4 (93.5,100.0) |
| | >1.5mm & ≤5.0mm | 0.008430 | 95.6 (91.2,98.9) | 86.0 (78.5,92.5) |
| | | 0.015487 | 92.3 (86.8,96.7) | 89.2 (82.8,94.6) |
| | | 0.051859 | 87.9 (80.2,94.5) | 97.8 (94.6,100.0) |
The results demonstrate the subject device establishes effective triage within a clinician's queue based on high sensitivity and specificity. Further, these results are substantially equivalent to those of the predicate device.
Triage effectiveness (turn-around time) was assessed by an internal bench study using a dataset of n=277 cases positive for any of the findings eligible for prioritization. These cases were collected from multiple data sources spanning a variety of geographical locations, patient demographics and technical characteristics. The results demonstrated a triage turn-around time of 81.6 (95% CI: 80.3 - 82.9) seconds, which is substantially equivalent to the total performance time published for the predicate device.
Therefore, the subject device has been shown to satisfy the performance requirements per 21CFR892.2080, for 'Radiological computer aided triage and notification software', by providing clinically effective triage for non-contrast brain CT studies containing features suggestive of clinical findings of interest. This data demonstrates the subject device is safe and effective for its intended use, and thereby supports substantial equivalence.
{10}------------------------------------------------
### VIII.CONCLUSIONS
The subject device and the predicate device are both software only packages, devices intended to assist with worklist triage by providing notification of findings. The subject and predicate devices utilize the same principles of operation and work in parallel to the current standard of care workflow.
Both the subject and predicate devices use an artificial intelligence algorithm to identify findings in images and require the same inputs (DICOM image data) and provide the same outputs (prioritization for a medical worklist).
The technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness.
Standalone performance testing and the comparison of technological characteristics with the predicate devices shows that the subject device:
- performs as intended, ●
- is safe and effective for its intended use, and
- is therefore substantially equivalent to the predicate device. ●
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.