K223227 · Paragon 28, Inc. · HSN · Nov 17, 2022 · Orthopedic
Device Facts
Record ID
K223227
Device Name
MAVEN™ Patient-Specific Instrumentation
Applicant
Paragon 28, Inc.
Product Code
HSN · Orthopedic
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3110
Device Class
Class 2
Indications for Use
The MAVEN™ Patient-Specific Instrumentation System is intended to be used as patient specific surgical instrumentation to assist in the positioning of total ankle replacement components intraoperatively and in guiding the marking of bone before cutting. The MAVEN™ Patient- Specific Instrumentation System is intended for use with the Paragon 28® APEX 3D™ Total Ankle Replacement System and its cleared indications for use, provided that anatomic landment and positioning of the implant are identifiable on patient imaging CT scans. The MAVEN™ Patient-Specific Instrumentation System is intended for single use only. The Paragon 28® TAR Patient- Specific Case Reports are intended for use with the Paragon 28® APEX 3D™ Total Ankle Replacement System and its cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging CT scans.
Device Story
System assists surgeons in pre-operative planning and intraoperative execution of total ankle replacement procedures. Inputs: patient CT scan data. Process: design engineers use OTS software to create patient-specific anatomical models and surgical guides based on patient anatomy. Outputs: physical patient-specific guides and digital/printed patient-specific case reports. Used in clinical settings by surgeons to guide bone cutting and implant positioning. Benefits: facilitates accurate transfer of pre-operative plans to the surgical site, ensuring alignment of the APEX 3D Total Ankle Replacement System.
Clinical Evidence
Bench testing only. Design and manufacturing process validation demonstrated that modifications did not alter the output (patient-specific instruments) and that the device remains substantially equivalent to predicates.
Technological Characteristics
Patient-specific surgical guides and anatomical models. Material: Nylon. Manufacturing: Additive manufacturing/3D printing. Connectivity: N/A (physical instrumentation). Software: OTS design software used by engineers for pre-operative planning.
Indications for Use
Indicated for use as patient-specific surgical instrumentation to assist in positioning total ankle replacement components and guiding bone marking during surgery. Intended for use with the Paragon 28 APEX 3D Total Ankle Replacement System, provided anatomic landmarks are identifiable on patient CT scans.
Regulatory Classification
Identification
An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 17, 2022
Paragon 28, Inc. % Jan Triani Regulatory Affairs Consultant Triani Consulting, LLC 14445 Grasslands Drive Englewood, Colorado 80112
Re: K223227
Trade/Device Name: MAVENTM Patient-Specific Instrumentation Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSN. OYK Dated: October 17, 2022 Received: October 18, 2022
Dear Jan Triani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223227
Device Name MAVEN™ Patient Specific Instrumentation
#### Indications for Use (Describe)
The MAVEN™ Patient-Specific Instrumentation System is intended to be used as patient specific surgical instrumentation to assist in the positioning of total ankle replacement components intraoperatively and in guiding the marking of bone before cutting. The MAVEN™ Patient- Specific Instrumentation System is intended for use with the Paragon 28® APEX 3D™ Total Ankle Replacement System and its cleared indications for use, provided that anatomic landment and positioning of the implant are identifiable on patient imaging CT scans. The MAVEN™ Patient-Specific Instrumentation System is intended for single use only. The Paragon 28® TAR Patient- Specific Case Reports are intended for use with the Paragon 28® APEX 3D™ Total Ankle Replacement System and its cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging CT scans.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# Paragon 28, Inc. MAVEN Patient Specific Instrumentation
| Manufacturer: | Paragon 28, Inc<br>14445 Grasslands, Dr.<br>Englewood, CO 80112<br>Phone: 855-786-2525 |
|---------------------|----------------------------------------------------------------------------------------|
| Official Contact: | Jan Triani<br>Regulatory Consultant<br>Patient Specific, Paragon 28, Inc. |
| Date Prepared: | October 11 ,2022 |
| Device Trade Name: | Paragon 28 MAVEN™ Patient-Specific Instrumentation |
| Device Common Name: | Patient Specific Guide |
| Classification: | 21 CFR 888.3110 |
| Product Code: | HSN, OYK |
# Purpose of the Special 510(k) notice:
The modification to the subject device is to allow an additional material and supplier to make the subject device.
#### Intended Use
The intended use and indication are the same as the main predicate device. K202019. Indications are:
The MAVEN™ Patient-Specific Instrumentation System is intended to be used as patient specific surgical instrumentation to assist in the positioning of total ankle replacement components intraoperatively and in guiding the MAVEN™ marking of bone before cutting. The Patient-Specific Instrumentation System is intended for use with the Paragon 28® APEX 3D™ Replacement System indications Total Ankle and its cleared for use, provided that anatomic landmarks necessary for alignment and positioning of the MAVEN™ implant are identifiable on patient imaging CT scans. The Patient-Specific Instrumentation System is intended for single use only. The 286 TAR Patient-Specific Case Reports are intended Paragon for use with the Paragon 28® APEX 3D™ Total Ankle Replacement System and its that anatomic landmarks necessary cleared indications for use, provided for alignment and positioning of the implant are identifiable on patient imaging CT scans.
#### Device Description
The Paragon 28® MAVEN™ Patient Specific Instrumentation (PSI) system is intended to assist a surgeon with pre-operative planning and transfer of the pre-operative plan to the surgery in total
{4}------------------------------------------------
ankle replacement procedures. The system contains several physical and digital outputs including patient-specific anatomical models and guides (physical outputs); and a patient-specific case report (digital or documentation output).
The MAVEN™ PSI System is for use with the APEX 3D™ Total Ankle Replacement System.
# Performance Data
The company conducted a design and manufacturing process validation, to demonstrate that the changes presented within this premarket submission, did not change the outputs, which is the creation of a patient specific instrument. The validation successfully demonstrated the minor changes made to the subject device are substantially equivalent to the predicate.
## Substantial Equivalence
The modification made to the MAVEN Patient Specific Instrumentation are minor. The subject device has the same intended use and same indications, and principles of operation, as the primary predicate (K202019). The minor differences in the grade of nylon and OTS design software do not raise any new questions of safety or effectiveness. The subject device is made by the same supplier, from the same nylon material and manufacturing methods, and uses the same software as the secondary predicate (K211076). Performance data demonstrates that the subject device is as safe and effective as the predicate devices. Thus, the subject device is substantially equivalent to its predicate devices.
## Conclusions
In summary, the company's modified MAVEN Patient Specific Instrumentation has the same intended use as all of the previously cleared MAVEN Patient Specific Instrumentation (K202019). In addition, the modified device has the same indications, technological characteristics, and principles of operation as its primary predicate. Although there are minor differences between the subject and its primary predicate device, namely the grade of nylon material and the OTS software used by the design engineers, those differences do not raise new questions of safety or efficacy, and data provided in this and previous 510(k)s establish equivalence. Thus, the subject device is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.