K223183 · Olympus Winter & Ibe GmbH · HBI · Jan 13, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223183
Device Name
Light-Guide Cables
Applicant
Olympus Winter & Ibe GmbH
Product Code
HBI · General, Plastic Surgery
Decision Date
Jan 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
Transmission of light energy from the light source to an optical instrument. The light-guide cable is used to transmit light during endoscopic procedures or for other medical illumination application when the properties and operating instructions are complied with. The light-guide cable is designed for use with halogen, xenon or LED based light which are utilized in medical applications that involve endoscopes, medical instruments or microscopes.
Device Story
Light-guide cables transmit light from external halogen, xenon, or LED sources to endoscopes or medical instruments. Device consists of fused optical fiber bundles (GOF85) encased in silicone tubing with stainless steel connectors. Two variants (WA03300A, WA03310A) accommodate different endoscope shaft diameters. Used in clinical settings by physicians or trained staff during endoscopic diagnosis or therapy. Cables are non-patient contacting; they provide illumination to the surgical field to assist visualization. Device is delivered unsterile and requires manual or automated cleaning followed by steam or hydrogen peroxide plasma sterilization before use. Service life is 400 reprocessing cycles.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (ANSI AAMI ES60601-1, IEC 60601-2-18) and reprocessing/sterilization validation.
Indicated for transmission of light energy from a light source to an optical instrument during endoscopic procedures or other medical illumination applications involving endoscopes, medical instruments, or microscopes.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 13, 2023
Olympus Winter & Ibe GmbH % Christina Flores Manager Regulatory Affairs Olympus Surgical Technologies of the Americas 800 West Park Drive Westborough, Massachusetts 01581
Re: K223183
Trade/Device Name: Light-Guide Cables Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: Class II Product Code: HBI Dated: October 10, 2022 Received: October 12, 2022
Dear Christina Flores:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K223183
Device Name
Light-Guide Cables (WA03300A, WA03310A)
#### Indications for Use (Describe)
Transmission of light energy from the light source to an optical instrument.
The light-guide cable is used to transmit light during endoscopic procedures or for other medical illumination application when the properties and operating instructions are complied with.
The light-guide cable is designed for use with halogen, xenon or LED based light which are utilized in medical applications that involve endoscopes, medical instruments or microscopes.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# K223183
# 510(k) Summary of Safety and Effectiveness September 21, 2022
## 1. General Information
| Manufacturer: | Olympus Winter & Ibe GmbH<br>Kuehnstr. 61<br>22045 Hamburg<br>Germany<br>Establishment Registration Number: 9610773 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Christina Flores, RAC<br>Manager Regulatory Affairs<br>Olympus Surgical Technologies of the Americas<br>800 West Park Drive Westborough, MA 01581<br>Phone: (508) 808-3341<br>Email: Christina.Flores@olympus.com<br>Establishment Registration No. 3003790304 |
#### 2. Device Identification
| Common Name: | Light-Guide Cables |
|------------------------|----------------------------------------------------------|
| Regulation Number | 21 CFR 878.4580 |
| Classification: | Illuminator, Fiberoptic, Surgical Field<br>Surgical Lamp |
| Device Class | II |
| Product Code | HBI |
| Review Panel | General & Plastic Surgery |
| Proprietary/Trade Name | Light-Guide Cables |
Model numbers: WA03300A, WA03310A
## 3. Predicate Devices
The predicate device was chosen from the following predicate 510(k):
| 510(k)<br>No. | Name | Predicate Model No. | Product code /<br>Reg No. |
|---------------|-------------------------------------------|-----------------------|---------------------------|
| K111342 | Light Guide Cable<br>Schott North America | not stated in K111342 | HBI / 21 CFR<br>878.4580 |
These predicates have not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
### 4. Product Description
The Olympus Light-Guide Cables that are subject to this submission are intended to transmit light from the light source to an optical instrument. For that purpose the Light-Guide Cable includes a bundle of optical fibers as transmission medium. The light guide adapters are intended for the mechanical connection of the Light-Guide Cables to light sources or to endoscopes.
There are two variants of Light-Guide Cables which differ according to the shaft diameter of the connected endoscope. Cable WA03300A (2.8 mm, 3 m, CF type) can be combined with endoscopes, which have a diameter ≤ 4.1mm. Endoscopes with a diameter > 4.1mm are combined with WA03310A (4.25 mm, 3 m, CF type). Otherwise the Light-Guide Cables are identical.
#### 5. Indications for Use
light energy Transmission from the light of source to an optical instrument. The light-guide cable is used to transmit light during endoscopic procedures or or for other medical illumination application when the properties and operating instructions are complied with. The light-guide cable is designed for use with halogen, xenon or LED based light sources of cold light which are utilized in medical applications that involve endoscopes, medical instruments or microscopes.
#### 6. Comparison of Technological characteristics
The subject and predicate device K111342 are based on the same technological principle with the same elements:
- I In both cases an illumination of the operation field by transmitting of visual light is indicated during surgical procedure or endoscopic diagnosis/therapy.
- Dimensions differ. Subject Device: WA03300A: 2.8 mm, 3 m, CF type; and WA03310A: 4.25 mm, 3 m, CF type. Predicate Device: 1,8m - 3m depending on device model.
- 트 Both devices consist of fused optical fibers: Glass fibers fused into fenules on both ends. The ends are polished to transmit light.
- I The subject and predicate device both use the same type of fiber: GOF85
- Fiber cross section of subject and predicate device. The fiber cross-section results in a different amount of brightness.
{5}------------------------------------------------
- The Light-Guide Cables use the same materials for the outer tube: Silicone
- . The Light-Guide Cables use the same materials for the connector: stainless steel.
- . Both devices are not have patient contact and no patient contacting materials.
- 트 Both devices stating different combination information. The subject device is provided in a "ready to use" state in combination with Olympus products.
- . The subject device has an additional Olympus connector and additional inner stainless steel wire for better mechanical protection
- 트 Design changes of the cables are minor and do not negatively impact safety or effectiveness of the subject devices
As stated above, the subject and predicate devices have similar design characteristics and performance specifications, with the exception of the cross-sections, variations of the cables in length and dimension and an additional stainless steel wire in the subject device. These minor differences, however, do not raise different questions of safety or effectiveness. As demonstrated in the non-clinical testing (e.g. electrical safety, Bench Testing, reprocessing validation) and the Substantial Evidence Discussion (Section 6), the different technological characteristics do not affect the safety and effectiveness of the subject devices.
Based on the information presented above, Olympus Winter & Ibe GmbH believes that the subject device Light-Guide Cables are substantially equivalent to the cited predicate device (K111342) and that the differences between devices are minor and raise no new issues of safety or effectiveness.
#### 7. Performance Data
The Light-Guide Cables comply with all applicable requirements/standards as listed in Appendix IIIc of this submission.
The design of the subject device Light-Guide Cables is identical to that of the predicate devices. As described in detail in Section 6 of this submission, the only differences between the Light-Guide Cables subject to this submission and the predicate Light-Guide Cables from K111342 are minor design modifications. Furthermore this premarket submission is submitted by Olympus to clear a new sterilization method. Currently the subject devices in this submission are covered in K111342 submitted by Schott. Because of that, Olympus Light-Guide Cables such as the predicate devices have been used safely and effectively for years.
In view of the above, it was not considered necessary to re-test the performance of the Light-Guide Cables concerned. However, complete evidence of the performance test records is included in this submission.
{6}------------------------------------------------
Respective test records are referenced and are provided in Appendix 12a-e.
## 8. Electrical safety and electromagnetic compatibility (EMC)
Electrical Safety was tested according to
| ANSI AAMI ES60601-<br>1:2005/(R)2012 and<br>A1:2012, C1:2009/(R)2012<br>and A2:2010/(R)2012<br>(Consolidated Text) | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance (IEC 60601-1:2005, MOD) |
|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-18: Edition 3.0<br>2009-08 | Medical electrical equipment - Part 2-18:<br>Particular requirements for the basic safety and<br>essential performance of endoscopic equipment |
## 9. Reprocessing, Sterilization and Shelf Life
The Light-Guide Cables are delivered in unsterile condition and thus have no shelf life, but rather a defined lifetime. This lifetime can be specified as a certain number of cycles for reprocessing. For the Light-Guide Cables, the expected service lifetime is 400 reprocessing cycles.
The products are delivered in unsterile condition and should be reprocessed before first and each subsequent use. The product should be cleaned manually or automatically, before being inspected and then sterilized with steam or hydrogen peroxide plasma. Validated Sterilization methods for Light-Guide Cables WA03300A and WA03310A are Hydrogen peroxide plasma sterilization (Sterrad 100S, Sterrad NX, Sterrad 100NX), Steam Sterilization (Autoclave, prevacuum) and Vaporized hydrogen peroxide (Steris V-PRO maX, Steris V-PRO maX 2, Steris V-PRO s2 and Steris V-PRO 60).
Olympus recommends automated cleaning and steam sterilization for this product.
#### 10. Conclusion
The performance data support the safety and effectiveness of the subject device and demonstrate that the subject device is substantially equivalent to the predicate device.
In conclusion, the Light-Guide Cables are substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.