Sleepiz One+
K223163 · Sleepiz AG · DRT · Aug 18, 2023 · Cardiovascular
Device Facts
| Record ID | K223163 |
| Device Name | Sleepiz One+ |
| Applicant | Sleepiz AG |
| Product Code | DRT · Cardiovascular |
| Decision Date | Aug 18, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2300 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
Sleepiz One+ is a contactless medical device intended to measure heart rate and respiration rate in adult patients, at rest or during sleep (in non-motion condition). The Sleepiz One+ hardware unit is intended to be used by a healthcare professional when the recordings are performed in a clinical setting, or by patients or their caregivers when the recordings are performed in a home environment. The Sleepiz One+ web application is intended for use by healthcare professionals. Sleepiz One+ device can also detect the presence of patients and their body movements at rest or during sleep. This device is not indicated for active patient monitoring, as it does not provide alarms for timely response in life-threatening situations. It is not indicated for use on pregnant women or patients with active implantable devices.
Device Story
Sleepiz One+ is a contactless bedside monitor using continuous-wave radar to detect chest displacements from heartbeats and breathing. The hardware unit captures signals, which are transmitted via Wi-Fi to a cloud-based 'Sleep Analytics Software' for processing. The system outputs respiration rate, heart rate, presence detection, body movement, and time-in-bed statistics. Used in clinical or home settings by healthcare professionals, patients, or caregivers. Data is accessed via a web application for retrospective analysis, report generation, and device management. The device aids healthcare providers in monitoring physiological status; it does not provide real-time alarms for life-threatening events.
Clinical Evidence
Two clinical studies (n=199 total) compared Sleepiz One+ against ECG, pulse oximetry, respiratory effort belts, nasal cannulas, and capnography (etCO2). Heart rate accuracy was 94-96% (95% LOA -2.64 to 5.82 bpm). Respiration rate accuracy was 93.7-99% (95% LOA -2.51 to 2.04 breaths/min). Comparator devices included Somnotouch RESP and Nox A1. Results demonstrate performance across diverse age, BMI, and health conditions.
Technological Characteristics
Contactless continuous-wave radar sensor; battery-operated with DC jack. Dimensions: bedside form factor. Connectivity: Wi-Fi to cloud, tablet accessory for data transfer. Software: proprietary signal processing algorithms. Standards: ANSI AAMI ES60601-1, IEC 60601-1-11, IEC 62304, IEC 62366-1, ISO 14971, ANSI IEEE C63.27.
Indications for Use
Indicated for adult patients to measure heart rate and respiration rate at rest or during sleep (non-motion). Not for active patient monitoring or life-threatening situations. Contraindicated for pregnant women and patients with active implantable devices.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
- Vital Sign Monitoring Sensor (Model: XK300) (K202464)
Reference Devices
- Somnotouch RESP (K140861)
- Nox A1 (K192469)
Related Devices
- K253388 — Sleepiz One+ (Model 2.5) · Sleepiz AG · Jan 28, 2026
- K251364 — Sleepiz One+ (2.5) · Sleepiz AG · Jul 29, 2025
- K234003 — The Circadia C200 System · Circadia Technologies, Ltd. · May 30, 2024
- K092062 — EVERON 1.0 · Earlysense, Ltd. · May 24, 2010
- K252676 — The Circadia C300 System (C300) · Circadia Health, Inc. · Feb 3, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 18, 2023
Sleepiz AG % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K223163
Trade/Device Name: Sleepiz One+ Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, BZQ Dated: May 7, 2023 Received: May 8, 2023
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shruti N. Mistry -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K223163
Device Name Sleepiz One+
#### Indications for Use (Describe)
Sleepiz One- is a contactless medical to measure heart rate and respiration rate in adult patients, at rest or during sheep (in nor-motion condition).
The Sleepiz One+ hartware unit is intended to be healthcare professional when the recordings are performed in a clinical setting, or by patients or their caregivers when the recordings are performed in a home environment.
The Sleepiz One+ web application is intended for use by healthcare professionals.
Sleepiz One+ device can also detect the presence of patients and their body movements at rest or during sleep.
This device is not indicated for active patient monite alarms for timely response in life-threatening situations. It is not indicated for use on pregnant women or patients with active implantable devices.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# 1.1 Applicant/Submitter
| Company Name | : Sleepiz AG |
|----------------|---------------------|
| Street Address | : Hornbachstrasse 2 |
| City | : Zurich |
| State | : Zurich |
| Zip Code | : 8008 |
| Phone Number | : 0767837350 |
| Fax Number | : Not Applicable |
# 1.2 Contact Person
| Full Name | : Marta Stepien |
|-----------|--------------------|
| Phone | : +41 76 783 73 50 |
# 1.3 Date of Preparation
Date of Preparation
: 08/08/2023
# 1.4 Subject Device Information
Table - 1.1 The table below provides details on the subject device.
| Trade Name | Sleepiz One+ |
|----------------------|--------------------------------------------------------|
| Common or Usual Name | Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) |
| Classification Name | 21 CFR 870.2300 |
| Regulatory Class | 2 |
| Product Code | DRT, BZQ |
{4}------------------------------------------------
| Predicate<br>Type | 510(k)<br>Number | Device Name | Manufacturer | Product<br>Code |
|-------------------|------------------|--------------------------------------------|------------------------|-----------------|
| Primary | K202464 | Vital Sign Monitoring Sensor (Model:XK300) | Xandar<br>Kardian Inc. | DRT, BZQ |
Table - 1.2 The table below provides details on the chosen predicate device.
## 1.6 Device Description
Sleepiz One+ is a contactless medical device that uses radar technology to measure respiration rate and heart rate. The device is placed on a bedside table or a stand, mounted slightly higher than the mattress level, from where it detects the presence of a patient and their physiological signals. From that position, distance changes between the device and the patient's body are captured by Doppler radar. The recorded signals are then transmitted to the cloud software where these are analyzed by the signal processing software ("Sleep Analytics Software") to obtain respiration rate, heart rate and facilitate the monitoring of the presence of the patient and their body movement. These outputs are then displayed on the web application to allow the annotation of the data, compilation of results into reports, and the management of the hardware units.
### Outputs
The following parameters are estimated:
- . Breathing pattern
- Instantaneous breathing rate [breaths per minute]
- Breathing rate statistics (10th, 50th, and 90th quantiles) [breaths per ● minute]
- . Body movement
- . Time in bed [hours]
- Presence detection .
- . Heart rate [beats per minute]
- Heart rate statistics (10th, 50th, and 90th quantiles) [beats per minute] ●
### Major Components
- 1. Sleepiz One+ Hardware unit: The Sleepiz One+ hardware unit uses radar technology that allows contactless detection of heart rate and respiration rate by sensing chest displacements originating from heartbeats and breathing activity of patients at rest or during sleep. The hardware unit can be used by a healthcare professional when the recordings are performed in a clinical setting, or by patient or their caregiver when the recordings are performed in home environment. The signals recorded by Sleepiz One+ hardware unit is then transferred via Wi-Fi to the cloud software where it is interpreted and displayed on a web application.
- 2. Sleepiz One+ Web Application: The Sleepiz One+ web application is intended to display the vital signs that have been interpreted from the derived signals by the Sleep Analytics Software on the secured Sleepiz Cloud software. The displayed data can then be accessed by a healthcare professional for retrospective analyses to remotely monitor the patient's physiological status. The intended users of the Sleepiz One+ web application are physicians and other healthcare professionals in a clinical setting. The web application allows the data, compilation of results into reports, the management of the Sleepiz One+ hardware units and to perform other administrative functions that do not involve vital signs.
#### Accessories
Sleepiz hardware unit is supplied with the following components which are intended to be used in combination with it:
- Power supply
- . Tablet with pre-installed application to enable data transfer via Wi-Fi and display the connectivity status Support structure to allow easy positioning of the device
{5}------------------------------------------------
#### Sleepiz One+ Features Summary
- 1. Processes the raw signals from the device and presents the data in a graphical and readable form.
- 2. Facilitates monitoring of patients remotely.
- 3. Creation of a patient on the frontend which contains patient information.
- 4. The ability for healthcare professionals to add annotations to the patient's graphical data.
- 5. Storage and display of historical patients' data.
- 6. The ability for healthcare professionals to manage Sleepiz hardware units remotely.
- 7. The ability for healthcare professionals to download the medical report in a PDF format.
### 1.7 Intended Use/Indications for Use
Sleepiz One+ is a contactless medical device intended to measure heart rate and respiration rate in adult patients, at rest or during sleep (in non-motion condition).
The Sleepiz One+ hardware unit is intended to be used by a healthcare professional when the recordings are performed in a clinical setting, or by patients or their caregivers when the recordings are performed in a home environment. The Sleepiz One+ web application is Intended for use by healthcare professionals.
Sleepiz One+ device can also detect the presence of patients and their body movements at rest or during sleep. This device is not indicated for active patient monitoring, as it does not provide alarms for timely response in life-threatening situations. It is not indicated for use on pregnant women or patients with active implantable devices.
{6}------------------------------------------------
## 1.8 Comparison of Technological Characteristics with Predicate Device
The table below describes the similarities and differences between the subject device and the predicate device with regard to the intended use, principal of operation and technological characteristics.
| | Table - 1.3 Substantial Equivalence Discussion | |
|--|------------------------------------------------|--|
| | | |
| Features/Technical<br>Information | Subject Device<br>Sleepiz One+<br>(K223163) | Predicate Device<br>Vital Sign Monitoring<br>Sensor (Model:<br>XK300)<br>(K202464) | Differences |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | II | II | - |
| Product code | DRT and BZQ | DRT and BZQ | - |
| Regulation number | 870.2300 and 868.2375 | 870.2300 and 868.2375 | - |
| Indication of use | Sleepiz One+ is a contactless<br>medical device intended to<br>measure heart rate and<br>respiration rate in adult<br>patients, at rest or during<br>sleep (in non-motion<br>condition). The Sleepiz One+<br>hardware unit is intended to<br>be used by a healthcare<br>professional when the<br>recordings are performed in a<br>clinical setting, or by patients<br>or their caregivers when the<br>recordings are performed in a<br>home environment. The<br>Sleepiz One+ web application<br>is intended for use by<br>healthcare professionals.<br>Sleepiz One+ device also<br>detects the presence of<br>patients and their body<br>movements at rest or during<br>sleep. This device is not<br>indicated for active patient<br>monitoring, as it does not<br>provide alarms for timely<br>response in life- threatening<br>situations. It is not indicated<br>for use on pregnant women or<br>patients with active<br>implantable devices. | The Vital Sign Monitoring<br>Sensor (Model XK300) is<br>intended to measure heart<br>rate and respiration rate in<br>adult patients in a general<br>care hospital environment<br>including nursing homes.<br>The Vital Sign Monitoring<br>Sensor can be used for<br>home healthcare for data<br>collection to inform patient<br>care but not to acutely<br>treat a patient.<br>XK300 monitors presence<br>or absence of a patient in<br>a detection area of within<br>7 meters. The XK300 also<br>monitors the length of<br>continuous patient motion<br>or absence of patient<br>motion. | Same |
| Intended user | Patients and healthcare<br>professionals | Patients and healthcare<br>professionals | Same |
| Site of use | Clinical and home<br>environment | General Care Hospital<br>Environment, Nursing<br>Homes and at home | Same |
| OTC/Rx | Rx | Rx | Same |
| Patient population | Adult | Adult | Same |
| Accessories | Hardware unit<br>(sensor) Tablet | Sensor, USB micro b<br>cable | Different |
| K223163 | | | 510(k) Submiss |
| Device application | Contactless | Contactless | Same |
| Power source | Battery operated, DC<br>jack adapter | Wall plug-ins, USB micro<br>b cable | Different |
| Data collection method | Continuous-wave<br>radar technology | Ultra-Wide Band (UWB)<br>radar technology | Different |
| Data storage | ARM Cortex-A7<br>Processor Chip | Arm Cortex M7 MCU chip | Different |
| Data transfer method | Wireless (Wi-Fi, LTE<br>via accessory tablet) | Wireless (LTE) | Different |
| Data display method | Web application | Web application | Same |
| Data display interface | Personal computer | Personal computer | Same |
| Operating System | Internet browser | Internet browser | Same |
| Outputs | Respiration Rate (5-40<br>breaths/min)<br>Heart Rate (40-120 beats/min)<br>Presence Detection | Respiration Rate (6-55<br>breaths/min)<br>Heart Rate (60-<br>120 beats/min)<br>Presence<br>Detection | The parameters<br>being measured<br>by both the<br>devices are the<br>same. The<br>difference lies in<br>the measuring<br>ranges covered<br>by both the<br>devices. See<br>below for an<br>explanation. |
{7}------------------------------------------------
## 1.9 Substantial Equivalence Summary
The subject and the predicate devices have the same intended use, which is to measure patients' heart rate and respiration rate and make them available to healthcare professionals for monitoring patients remotely. Like the predicate device, the subject device is intended for healthcare professionals and adult users who can use it upon the prescription and instructions of a registered healthcare professional. The subject device is intended to be used in the same environment as the predicate device. Both devices detect the presence of patients and their movement, and additionally, the subject device also indicates breathing patterns, instantaneous breathing rate, time in bed, and breathing rate statistics by processing the obtained signals using the proprietary algorithms.
The subject device like the predicate uses radar technology to collect the signal originating from the patient. The difference lies in the subject device uses continuous-wave radar technology while the predicate uses UWB radar technology. The difference in technology does not raise any new question of effectiveness and safety because both serve the purpose of collecting raw data using radars. UWB Radars can detect the exact distance while the CW radars can measure the target velocity without ambiguity. Distance or Velocity, any of these parameters can be used to monitor the vital signs which are inferred using chest displacement. Both the devices use a Cortex processor to collect and store the received data locally. Cortex-A7 processor chips in the subject device make use of the rich operating system while Cortex-M7 MCU Chip in the predicate is based on MCUs. This difference does not raise question on the safety and effectiveness of subject device because both chips serve the intended use of collecting and storing the data locally. The collected data is then transmitted via the internet to the backend server in both devices. However, the difference lies in the fact that the predicate can work on POE, Wi-Fi, LTE-M, NB-IoT to transmit the data while the subject device transmits them via the datahub (tablet with LTE) or user's own Wi-Fi network. While the subject device is battery operated and requires to be charged every 12 hours, the predicate device has plug-in power/POE but both essentially require a power source to perform.
In the subject device, the data collection method, the data display interface, and the OS required to access the processed data/ vital sign data are the same as the predicate device. Any difference in the technological characteristics has been tested through verification testing, performance testing of the device and clinical study.
{8}------------------------------------------------
### 1.10.1 Non-Clinical
Sleepiz One+ has been verified and validated using a robust and detailed verification and validation plan. The subject device has also been subjected to multiple performance tests to verify that the device characteristics meet industry standards and relevant international standards for product quality and functionality. The objective was to ensure that the subject device performs its intended functions correctly, and to confirm its quality, safety, and reliability.
The hardware-related testing verified proper assembly of the device components, battery indication functionality, HW certifications, device operations, power consumption, battery performance etc. It has also been subjected to transportation, packaging, drop and slide tests to verify the robustness of the device and its functionality under simulated conditions of the most common use case scenarios. The subject device has been subjected to electrical and mechanical safety tests as per ANSI AAMI ES60601- 1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text), Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005 MOD) and IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, thus verifying that the subject device is electrically and mechanically safe during usage in the home and clinical-based environments.
The subject device has also passed all the emission tests performed as per IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, Medical device software - Software life cycle processes and Federal Register CFR 47 Part 15 subpart B:2017 Class B: and the Coexistence Immunity and Wireless Crosstalk tests performed as per 27701:2019, Security techniques - Extension to ISO/IEC 27001 and ISO/IEC 27002 for privacy information management - Requirements and guidelines. and ANSI IEEE C63.27-2017, American National Standard for Evaluation of Wireless Coexistence. To verify the operational ambient temperature, transportation, and storage ambient environmental conditions, the subject device has been tested under extreme temperature, pressure, and humidity conditions, as per clauses 4.2.2 and 4.2.3 of IEC 62366-1 Edition 1.0 2015-02, Medical devices - Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)] . All tests were successfully passed.
All the obtained results comply with the applicable consensus standards. The bench tests performed for the subject device addressed the parameters potentially affecting its performance and safety, and thus confirmed that the subject device is acceptable for the intended use in home and clinicalbased environments.
#### Software
All software components have been developed as recommended by IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, Medical device software - Software life cycle processes. All software components have been verified against the System Requirements Specifications which was followed by system-level validation against the user needs. All tests have passed that ensures the subject device performs as intended. The performance tests of the subject device in estimating vital parameters such as the breathing rate and heart rate, toss and turn, chest displacement, and presence detection conducted on subjects at rest or in sleep were found to have passed thus evidencing the subject device's ability in meeting the acceptance criteria and thus can be considered substantially equivalent for the intended use.
#### Risk Analysis
Risk Analysis was performed on the subject device as recommended by ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices and risk controls were implemented to mitigate all identified hazards through suitable modifications to each components' functionality, user interfaces, labeling and packaging, IFU and User Manual. All foreseeable cybersecurity risks associated with the subject device have been identified in addition
to finding timely errors through penetration testing. This ensures the IT security of the subject device and the protection of private data (PHI and PII).
#### Usability
An extensive Human Factor Engineering/Usability Engineering was performed as recommended by
- 1. IEC 62366-1 Edition 1.0 2015-02, Medical devices Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
- 2. Applying Human Factors and Usability Engineering to Medical Devices Guidance for Industry and Food and Drug Administration Staff
to identify critical tasks and associated use errors with the medical device's use and was found to be substantially equivalent for the intended users, uses, and use environments.
{9}------------------------------------------------
## 1.11 Biocompatibility Testing
As Sleepiz One+ is a contactless device, this testing is not applicable for the subject device.
## 1.12 Clinical Testing
To evaluate the performance of the Sleepiz One+, the subject device outputs were compared to electrocardiography and pulse oximetry measurements to assess the heart rate estimation accuracy, and respiratory effort belt and nasal cannula to assess the respiration accuracy in two clinical studies where patients were continuously monitored overnight. Polysomnography devices (Somnotouch RESP (K140861), Nox A1 (K192469)) have been used in 2 clinical studies as a comparator device, with the aforementioned subset of channels for the clinical performance assessment.
The accuracy was found to be +/ - 3 breaths per minute (accuracy rate 99%) and the 95% limits of agreements were -1.42 to 0.97 breaths/min compared to the respiratory effort belt; +/ - 5 beats per minute (accuracy rate 94%) and the 95% limits of agreements was -2.64 to 5.82 beats/min compared to ECG in 59 and 32 neurorehabilitation ward patients, respectively. Similar results were achieved in patients suspected of suffering from sleep apnea, with a slightly narrower confidence bounds of -1.3 to 0.8 breaths per minute (n=105) and 96% heart rate accuracy (n=73).
In addition, the performance of Sleepiz One+ in respiration was evaluated against end-tidal CO2 (etCO2) recorded with the FDA-cleared device, manually scored by a healthcare professional. The performance of the Sleepiz One+ in estimating respiration rate was found to be +/ - 2 breaths/minute (accuracy 93.7%) and 95% limit of agreement was -2.51 to 2.04 breaths/minute for instantaneous breathing rate, and the average breathing rate having a mean absolute error of 0.79 breaths/minute and 95% limit of agreement of -2.63 to 2.01 breaths/minute compared to the manually scored capnography measurements from 35 participants.
Thus, the performance of Sleepiz One+ has been demonstrated on 199 subjects with wide range of age, body- mass index, gender and health conditions to reflect intended user profile. The presented results support substantial equivalence to the predicate device.
## 1.13 Conclusion
Sleepiz AG considers the subject device, Sleepiz One+ substantially equivalent to the predicate device. This conclusion is based upon similarities in the indication for use, principle of operation and technological characteristics. The performance and clinical data support the safety of the subject device while the verification and system validation tests results suggest that Sleepiz One+ should perform as intended in the specified use conditions and comparably to the predicate device. Thus, any technological difference between the subject and predicate devices do not pose any new questions about the safety and effectiveness for Sleepiz One+.