# EVOLUTION SPINE Interbody System (K223146)

_Evolution Spine · OVD · Feb 7, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223146

## Device Facts

- **Applicant:** Evolution Spine
- **Product Code:** [OVD](/productcode/OVD.md)
- **Decision Date:** Feb 7, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Evolution Spine Static Interbody Devices are thoraco-lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12- L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. Evolution Spine Integrated Interbody Devices are integrated lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Devices are to be filled with autograff bone and/or allogenic bone graft composed of cancellous bone. These devices are intended to be used with three screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). When used without screws or anchors, these devices are with supplemental fixation (e.g. facet screws or posterior fixation). When used for multilevel degenerative scolical deformity, supplemental fixation that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems) must be used.

## Device Story

Interbody fusion system providing structural stability during spinal fusion; static and integrated interbody devices; placed via anterior approach. Static devices used with supplemental fixation; integrated devices function as stand-alone when used with three screws or with supplemental fixation when used with anchors or without screws. Implants filled with autograft or allogenic bone graft. Used by surgeons in clinical settings to treat spinal pathologies. Benefits include stabilization of spinal segments to facilitate fusion. System components include various sizes to accommodate patient anatomy.

## Clinical Evidence

Bench testing only. Performed static/dynamic compression and shear (ASTM F2077), wear debris analysis (ASTM F1877), and expulsion/subsidence testing (ASTM F2267). No clinical data provided.

## Technological Characteristics

Interbody devices additively manufactured from titanium alloy powder (ASTM F3001). Integrated system screws subtractively manufactured from titanium alloy (ASTM F136). Static and integrated form factors. No software or energy source.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Globus HEDRON Lumbar Spacers ([K191391](/device/K191391.md))
- Genesys Spine Apache® Interbody Fusion System ([K103034](/device/K103034.md))

## Reference Devices

- PISCES-SA Standalone ALIF Interbody System ([K213935](/device/K213935.md))
- TransLoc 3D ([K211496](/device/K211496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG", and then "ADMINISTRATION" in a smaller font size.

February 7, 2023

Evolution Spine Mr. Todd Wallenstein VP Regulatory/Quality 2300 N. Haskell Ave Dallas, Texas 75204

Re: K223146

Trade/Device Name: EVOLUTION SPINE Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX, PHM Dated: December 15, 2022 Received: December 16, 2022

Dear Mr. Wallenstein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

## Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223146

Device Name Evolution Spine Interbody System

#### Indications for Use (Describe)

Evolution Spine Static Interbody Devices are thoraco-lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12- L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

Evolution Spine Integrated Interbody Devices are integrated lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Devices are to be filled with autograff bone and/or allogenic bone graft composed of cancellous bone. These devices are intended to be used with three screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). When used without screws or anchors, these devices are with supplemental fixation (e.g. facet screws or posterior fixation).

When used for multilevel degenerative scolical deformity, supplemental fixation that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems) must be used.

| Type of Use (Select one or both, as applicable)                   |                                                                  |
|-------------------------------------------------------------------|------------------------------------------------------------------|
| <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> | <span>     Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary - K

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

| Company:               | Evolution Spine<br>2300 N Haskell Rd<br>Dallas, TX 75204                                                                                |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact:               | Todd Wallenstein<br>Evolution Spine<br>2300 N Haskell Ave<br>Dallas, TX 75204<br>Phone: 571-594-7409<br>twallenstein@evolutionspine.com |
| Date Prepared:         | February 1, 2023                                                                                                                        |
| Device Trade Name:     | EVOLUTION SPINE Interbody System                                                                                                        |
| Common Name:           | Intervertebral Body Fusion Device                                                                                                       |
| Classification:        | 21 CFR §888.3080                                                                                                                        |
| Class:                 | II                                                                                                                                      |
| Product Code:          | OVD, MAX, PHM                                                                                                                           |
| Primary Predicate:     | Globus HEDRON Lumbar Spacers (K191391)                                                                                                  |
| Additional Predicates: | Genesys Spine Apache® Interbody Fusion System (K103034)                                                                                 |
| Reference Devices:     | PISCES-SA Standalone ALIF Interbody System (K213935)<br>TransLoc 3D (K211496)                                                           |

#### Device Description:

The EVOLUTION SPINE Interbody System provides interbody fusion devices designed to provide structural stability during spinal fusion. The EVOLUTION SPINE Interbody System consists of interbodies offered in various sizes to accommodate surgical needs and anatomic requirements. The EVOLUTION SPINE Interbodies were designed to be placed via an anterior approach. All interbodies in the system are additively manufactured from titanium alloy powder, per ASTM F3001. The System offers both Static and Integrated Interbodies. The Integrated version is to be used in conjunction with three (3) screws that are subtractively manufactured from titanium alloy, per ASTM F136.

#### Indications For Use:

Evolution Spine Static Interbody Devices are thoraco-lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12- L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity

{5}------------------------------------------------

(degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

Evolution Spine Integrated Interbody Devices are integrated lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with three screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). When used without screws or anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).

When used for multilevel degenerative scoliosis or sagittal deformity, supplemental fixation that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems) must be used.

#### Substantial Equivalence:

The subject EVOLUTION SPINE Interbody System has been demonstrated to be substantially equivalent with respect to indications, design, materials, function, manufacturing, and/or performance as compared to the predicate devices.

#### Performance Data:

Testing on the EVOLUTION SPINE Interbody System included static compression, static shear, dynamic compression, and dynamic shear per ASTM F2077. Additional testing included wear debris analysis per ASTM F1877, expulsion, and subsidence per ASTM F2267. The results demonstrate that the EVOLUTION SPINE Interbody System is substantially equivalent to predicate devices.

#### Conclusion:

Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject EVOLUTION SPINE Interbody System has been shown to be substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K223146](https://fda.innolitics.com/device/K223146)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).