← Product Code [NEY](/productcode/NEY) · K223138

# AB1 Electrosurgical Instrument, Creo Electrosurgical System (K223138)

_Creo Medical, Ltd. · NEY · Jun 26, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223138

## Device Facts

- **Applicant:** Creo Medical, Ltd.
- **Product Code:** [NEY](/productcode/NEY.md)
- **Decision Date:** Jun 26, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The AB1 Instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue. The AB1 Instrument is not intended for use in cardiac procedures. The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories. The System is not intended for use in cardiac procedures.

## Device Story

AB1 electrosurgical instrument; microwave ablation device; used with Creo Electrosurgical System. Input: microwave energy (5.8 GHz, 25 W) from generator. Operation: antenna at distal tip couples microwave energy to soft tissue for coagulation. Configuration: single-use, sterile; user may remove outer sheath and distal handle to reduce shaft diameter from 4.3 mm to 1.9 mm for minimally invasive procedures via trocar. Used in open or minimally invasive surgery; operated by physician. Output: thermal ablation of target tissue. Benefit: enables soft tissue coagulation in varied surgical access scenarios.

## Clinical Evidence

Bench testing and preclinical evaluation performed. Bench tests included pull/tensile testing, shaft bending, VNA return loss, thermal profiling, and simulated use through introducers. Preclinical evaluation conducted in a pig model to assess safety and performance. Results demonstrate the device meets design specifications and is substantially equivalent to the predicate.

## Technological Characteristics

Microwave ablation instrument; 5.8 GHz energy source; 25 W power. Materials: FEP shaft, silicone elastomer handle, Pebax sleeve tube. Dimensions: 0.95 m length; 4.3 mm (with sheath) or 1.9 mm (without sheath) diameter. Single-use; Ethylene Oxide sterilization. Non-pyrogenic.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Creo Electrosurgical System with AB1 Accessory ([K200298](/device/K200298.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 26, 2023

Creo Medical Ltd. % Phil Triolo President and Senior Consultant Phil Triolo and Associates LC 86 Skvcrest Ln Salt Lake City, Utah 84108

Re: K223138

Trade/Device Name: AB1 Electrosurgical Instrument, Creo Electrosurgical System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: September 30, 2022 Received: May 26, 2023

Dear Phil Triolo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2013.06.26
13:08:24 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223138

Device Name

Creo Electrosurgical System with AB1 Electrosurgical Instrument

Indications for Use (Describe)

The AB1 Instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.

The AB1 Instrument is not intended for use in cardiac procedures.

The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories.

The System is not intended for use in cardiac procedures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Creo Electrosurqical System with AB1 Instrument

# 510(k) Summary

Date Prepared: June 23, 2023

#### 1.SUBMITTER

- 510(k) Sponsor: Creo Medical Ltd Creo House, Unit 2, Beaufort Park Way Chepstow NP16 5UH United Kingdom
- Contact Persons: Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801 699 9846 / Email: philt@philt.com

Tiffany Powell Global RA/QA Director, Creo Medical Ltd. Phone +44 (0) 7760 150775 / Email Tiffany.Powell@CreoMedical.com Fax +44 (0) 1291 606015

- Phil Triolo and Associates, LC Prepared By: 86 Skycrest Lane SLC UT 84108 801 699 9846
### 2.DEVICE

| Trade/Device Name:   | Creo Electrosurgical System with AB1 Instrument                |
|----------------------|----------------------------------------------------------------|
| Common Name:         | Electrosurgical System                                         |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulation number:   | CFR 878.4400                                                   |
| Product Code:        | NEY (Class II)                                                 |

### 3.PREDICATE DEVICE

K200298 - Creo Electrosurgical System with AB1 Accessory

{4}------------------------------------------------

#### Creo Electrosurqical System with AB1 Instrument

## 4.DEVICE DESCRIPTION

The Creo Medical (Creo) AB1 soft tissue microwave ablation instrument (electrosurgical instrument, accessory or instrument, also referred to as AB1) is designed for use with the Creo electrosurgical system, previously cleared under K200298. No changes have been made to the Creo generator, footswitch, or interface cable with sterile sheaths that comprise the electrosurgical system, nor are any changes proposed to the AB1 instrument in this 510k.

AB1 is provided as a sterile, single-use instrument that is 0.95 m long with a 4.3 mm diameter shaft that terminates in a distal antenna that is 38 mm long and less than 2.00 mm in diameter (nominally, 1.9 mm) with a conical tip. The AB1 instrument is provided with a flexible, removable (distal) handle. The outer sheath of the flexible shaft may be removed.

This 510k provides instructions to the end user on how to remove the outer extrusion and distal handle to create a configuration that has an outer diameter less than 2.00 mm. The instrument may be used to access target tissue directly, e.q. in open surgical procedures, or, with the handle and outer sheath removed, through a compatible trocar in minimally invasive procedures. The antenna couples microwave energy to target tissue when AB1 is powered by the Creo Electrosurqical System. AB1 is only compatible for use with the Creo Electrosurgical System, and the Creo Electrosurgical System is only intended for use with compatible Creo instruments and accessories.

# 5.INDICATIONS FOR USE

The AB1 Instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.

The AB1 Instrument is not intended for use in cardiac procedures.

The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories.

The System is not intended for use in cardiac procedures.

## 6.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS OF THE AB1 WITH MODIFIED IFU WITH THE CLEARED INSTRUMENT

The only change addressed in this 510k is a change to the IFU that instructs the user how to remove the outer sheath and distal handle of the cleared AB1 instrument to create a configuration that has a diameter less than 2.00 mm which can be used with a compatible trocar in minimally invasive procedures. No other changes have been made to the design, materials, or processing of the cleared AB1 instrument.

{5}------------------------------------------------

### Creo Electrosurqical System with AB1 Instrument

| Attribute                       | Subject Device                                                                                                                                                        | Predicate Device (K200298)                                                                           |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Instrument dimensions           | Shaft (with outer sheath): 4.3 mm diameter x 0.95 m length<br>Shaft (without outer sheath): 1.9 mm diameter x 0.95 m length<br>Antenna: 1.9 mm diameter x 38mm length | Shaft (with outer sheath): 4.3 mm diameter x 0.95 m length<br>Antenna: 1.9 mm diameter x 38mm length |
| Patient contacting<br>materials | Shaft: FEP (Fluorinated Ethylene Propylene)<br>Handle: Silicone Elastomer<br>Sleeve tube: Pebax                                                                       | Shaft: FEP (Fluorinated Ethylene Propylene)<br>Handle: Silicone Elastomer<br>Sleeve tube: Pebax      |
| Microwave energy<br>supplied    | 25 W microwave power at 5.8 GHz                                                                                                                                       | 25 W microwave power at 5.8 GHz                                                                      |
| Device life and sterility       | Single-use, delivered in a sterile condition.<br>Sterilized using Ethylene Oxide gas.                                                                                 | Single-use, delivered in a sterile condition.<br>Sterilized using Ethylene Oxide gas.                |
| Pyrogenicity                    | Non-pyrogenic                                                                                                                                                         | Non-pyrogenic                                                                                        |

### 7.PERFORMANCE DATA

In order to address the change in labeling to the device, a risk assessment was conducted that identified that the following tests needed to be control any new, or increased risks and demonstrate substantially equivalent safety and effectiveness of the configuration created by the user following the revised labeling, to that of the predicate configuration.

- Pull (tensile) test
- . Removal of outer extrusion test (effects on underlying cable, cosmetics, electrical performance)
- Insertion and retraction into and from an introducer test
- Shaft bending test
- VNA (vector network analyzer) return loss test
- Thermal profile with introducer test
- Comparison of dimensions of ablations created by new and predicate device in ex-vivo ● tissue
- Preclinical evaluation of safety and performance in a pig model
- Simulated use testing of deployment through introducers
- . Electrical safety testing
- . Usability engineering validation

{6}------------------------------------------------

The results of the tests conducted and leveraged demonstrate that the AB1 instrument in the proposed configuration with outer sheath and distal handle removed meets all design specifications and applicable medical device standard requirements, and that it is substantially equivalent to the predicate AB1 instrument with the handle and outer sheath in place.

### 8.CONCLUSION

Results of the bench and preclinical tests conducted demonstrate that the potential risks associated with the configuration of the AB1 instrument created by following the change to the IFU have been reduced to low and acceptable levels. The results further demonstrate that there are no new or different questions of safety or effectiveness raised by the modification to the IFU, and the AB1 instrument in the configuration created following the revised labeling is as safe and effective as the predicate device. Consequently, The Creo Medical AB1 electrosurgical instrument (Accessory) in the configuration created by following the modified IFU is substantially equivalent to the predicate instrument.

---

**Source:** [https://fda.innolitics.com/device/K223138](https://fda.innolitics.com/device/K223138)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
