Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14)
K223100 · Medline Industries, Inc. · CAF · Jun 27, 2023 · Anesthesiology
Device Facts
Record ID
K223100
Device Name
Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14)
Applicant
Medline Industries, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Jun 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hudson RCI AQUATHERM® III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AQUAPAK® system to provide continuous heated aerosol for respiratory applications.
Device Story
The AquaTherm III Plus is an adjustable electronic heater placed between a nebulizer adaptor and a sterile solution reservoir. When gas flow initiates, sterile solution is drawn through a stainless-steel puncture pin in the heater center; the solution is heated as it flows through the pin. The heated solution enters the nebulizer adaptor, where it is aerosolized into the patient gas stream. Used in hospital environments, the device is operated by healthcare professionals. It provides continuous heated aerosol to support respiratory therapy. The device is mains-powered and features adjustable temperature settings (25°C–37°C).
Clinical Evidence
Bench testing only. No clinical data was provided. Non-clinical testing included biocompatibility (ISO 10993-1, -5, -10, -11), gas pathway evaluation (ISO 18562-1, -2, -3), EMC testing (IEC 60601-1-2), and electrical safety (IEC 60601-1-11).
Technological Characteristics
Mains-powered (115 VAC, 60Hz) external adjustable electronic heater. Materials include stainless steel (puncture pin). Dimensions/form factor: external heater unit. Connectivity: standalone. Sterilization: non-sterile. Standards: ISO 10993, ISO 18562, IEC 60601-1-2, IEC 60601-1-11. Temperature range: 25°C–37°C.
Indications for Use
Indicated for use in hospital settings to provide continuous heated aerosol for respiratory applications via the Hudson RCI Nebulizer Adaptor and AQUAPAK system.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2023
Medline Industries, Inc % Joy Gutermuth Senior Specialist (Consultant) Ram+ 2790 Mosside Blvd. Suite 800 Monroeville, Pennsylvania 15146
Re: K223100
Trade/Device Name: Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14) Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: June 13, 2023 Received: June 13, 2023
Dear Joy Gutermuth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Ethan L. Nyberg -S" in a large, sans-serif font. The text is black against a white background. The letters are evenly spaced and clearly legible.
Ethan Nyberg, Ph.D Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K223100
Device Name
Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14)
Indications for Use (Describe)
The Hudson RCI AQUATHERM® III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AQUAPAK® system to provide continuous heated aerosol for respiratory applications.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# DATE PREPARED
June 26, 2023
## MANUFACTURER AND 510(k) OWNER
Medline Industries 1 Three Lakes Drive, Northfield, IL, 60093 Telephone: Official Contact: Nicole Schaffer,
### REPRESENTATIVE/CONSULTANT
Joy Gutermuth RQM+ Telephone: Email: iqutermuth@rgmplus.com
### DEVICE INFORMATION
Proprietary Name/Trade Name: Hudson RCI AquaTherm III Plus External Adjustable
| | Electronic Heather ("AquaTherm Heater") |
|--------------------|-----------------------------------------------------------------------|
| Common Name: | Nebulizer Heater |
| Regulation Number: | 868.5630 |
| Class: | II |
| Product Code: | CAF |
| Premarket Review: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental<br>Devices (OHT1) |
| Review Panel: | Anesthesiology |
# PREDICATE DEVICE IDENTIFICATION
The AquaTherm Heater is substantially equivalent to the following predicates:
| 510(k)<br>Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|------------------|-------------------------------------------------------------|----------------------|
| K880473 | Aero-Mist Nebulizer Heater/ Pegasus Research<br>Corporation | ✓ |
The predicate devices have not been subject to a design related recall.
# DEVICE DESCRIPTION
The AquaTherm devices are adjustable output temperature heaters. The AquaTherm Heater is placed between a nebulizer adaptor and a reservoir of sterile solution. When a gas flow is initiated to the nebulizer, sterile solution is drawn up from the reservoir and through the stainless-steel puncture pin in the center of the AquaTherm Heater. The solution is heated as it flows through the pin. The heated solution is then drawn into the nebulizer adaptor where it is aerosolized into the patient gas stream.
{4}------------------------------------------------
# INDICATIONS FOR USE
The Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AquaPak system to provide continuous heated aerosol for respiratory applications.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Medline Industries believes that the AquaTherm Heater is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design and dimensions, and uses similar or identical materials as the device cleared in K880473. The subject device has the same intended use and similar technological characteristics to the device cleared in K880473. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicate.
| Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater |
|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description | The AquaTherm devices are<br>adjustable output temperature<br>heaters. The AquaTherm Heater is<br>placed between a nebulizer<br>adaptor and a reservoir of sterile<br>solution. When a gas flow is<br>initiated to the nebulizer, sterile<br>solution is drawn up from the<br>reservoir and through the<br>stainless-steel puncture pin in the<br>center of the AquaTherm Heater.<br>The solution is heated as it flows<br>through the pin. The heated<br>solution is then drawn into the<br>nebulizer adaptor where it is<br>aerosolized into the patient gas<br>stream. | The heater warms and evaporates<br>particles impinging on the heater<br>disc which is disposed below the<br>nebulizer. Return water drains<br>through the center tube into the<br>sterile water supply container<br>below the heater. The water is<br>drawn up from the sterile water<br>supply by a draw tube which is part<br>of the nebulizer itself. The draw<br>tube passes through the center<br>tube of the heater in use. |
| Intended Use | To provide continuous heated<br>aerosol for respiratory<br>applications. | "… to provide warm aerosol of<br>water particles into the respiratory<br>tract."*<br>*Intended use not clearly stated in<br>510(k) summary |
| Indications for Use | The Hudson RCI AquaTherm III<br>Plus External Adjustable<br>Electronic Heater is designed for | "Respiratory therapy, in hospital, to<br>provide warm aerosol of water<br>particles into the respiratory tract."* |
| Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater |
| | use with the Hudson RCI<br>Nebulizer Adaptor and the<br>AquaPak system to provide<br>continuous heated aerosol for<br>respiratory applications. | *Indications for use not clearly<br>stated in 510(k) summary |
| Representative<br>Image*<br>*images feature<br>nebulizer adaptors<br>and tubing which<br>are not included<br>within this<br>submission. | Image: Hudson RCI AquaTherm III Plus Nebulizer Heater | Image: Pegasus Research Corporation Aero-Mist Nebulizer Heater |
| Classification | Class II | Class II |
| Product Code | CAF | CAF |
| Regulation<br>Number | §868.5630 | §868.5630 |
| Regulation Name | Nebulizer | Nebulizer |
| Accessories | Aquapak Nebulizer adapter Aquapak prefilled nebulizer reservoir Sold separately and not part of<br>this submission | Nebulizer adapter Prefilled nebulizer reservoir Sold separately |
| Environment | Hospital | Hospital |
| Sterilization<br>Method | Non-sterile | Non-sterile |
| Patient Contacting<br>Materials | Indirect patient contact only:<br>Stainless steel | Unknown- 510(k) does not<br>describe patient-contacting<br>materials |
| Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater |
| Principles of<br>Operation | Adjustable temperature. The<br>AquaTherm Heater is placed<br>between a nebulizer adaptor (not<br>included) and a reservoir of sterile<br>solution (not included). When a<br>gas flow is initiated to the<br>nebulizer, sterile solution is drawn<br>up from the reservoir and through<br>the stainless-steel puncture pin in<br>the center of the AquaTherm<br>Heater. The solution is heated as<br>it flows through the pin. The<br>heated solution is then drawn into<br>the nebulizer adaptor where it is<br>aerosolized into the patient gas<br>stream. Rainout is returned to the<br>reservoir via the nebulizer adapter<br>return tube (rainout return tube is<br>part of the nebulizer adapter, not<br>the subject device). | Adjustable temperature. Warms<br>and evaporates particles on the<br>heater disc below the nebulizer.<br>Return water drains through the<br>center tube into the container<br>below the heater. The water is<br>drawn up from the water supply by<br>a draw tube; the nebulizer draw<br>tube passes through the center<br>tube of the heater. |
| Power Supply | Mains powered | Mains powered |
| Voltage and Power | 115 VAC at 60Hz, 1.5 A | 115 VAC (+/- 10 volts) 1.7 A |
| Storage and<br>Transportation<br>Environment | Temperature: -20 °C to 50 °C<br>Relative Humidity: 0% to 94%<br>(non-condensing) | Temperature: -40 °C to 70 °C<br>Relative Humidity: 10% to 100%<br>(including condensate) |
| Output<br>Temperature<br>Range | 25°C -37°C | 25°C -40°C |
| Heat up Time | 25 minutes (about 20 minutes<br>after each subsequent<br>temperature adjustment) | Approximately 20 minutes |
| Operating<br>Environment | • Temperature: 18°C to 26°C<br>• Relative Humidity: 30% to 75%<br>(non-condensing)<br>• Pressure: 700 to 1060 hPa | • Temperature: 10°C to 40°C<br>• Relative Humidity: 30% to 75%<br>(non-condensing)<br>• Pressure: 700 to 1060 hPa |
| Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater |
| Biocompatibility | External communicating, limited contact device that indirectly contacts tissue/bone/dentin. Indirect gas pathway | Unknown- 510(k) does not<br>describe any biocompatibility. |
| Standards Utilized | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>IEC 60601-1-2<br>IEC 60601-1-11 | UL544 |
| Non-Clinical<br>Testing | Packaging<br>Cleaning process<br>EMC testing | Unknown- 510(k) does not state<br>testing. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
# SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the AquaTherm Heater. The following tests were performed to demonstrate safety based on current industry standards:
- Biocompatibility ●
- Packaging ●
- Cleaning
- EMC Testing ●
The results of these tests indicate that the AquaTherm Heater is substantially equivalent to the predicate devices.
# CONCLUSION
Based on the testing performed, including gas pathway biocompatibility testing, and compliance to the latest edition of IEC 60601-1-2, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed AquaTherm Heater are assessed to be substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.