K223076 · Care Fusion · ONB · Mar 24, 2023 · General Hospital
Device Facts
Record ID
K223076
Device Name
BD Texium™ Closed Male Luer
Applicant
Care Fusion
Product Code
ONB · General Hospital
Decision Date
Mar 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BD Texium™ Closed Male Luer (CML) is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of hazardous drugs when paired with the SmartSite™ NFC. The BD Texium™ Closed Male Luer (CML) is indicated for use by trained healthcare professionals within healthcare facilities who prepare and/or administer non-hazardous drugs for adults, pediatrics and neonates.
Device Story
BD Texium™ CML is a sterile, single-use closed system drug transfer device (CSTD). Used in healthcare facilities by trained professionals for drug reconstitution, transfer, and administration. Operates by pairing with SmartSite™ Needle-Free Connector (NFC) to create an airtight, leak-free, and drip-free connection. Mechanically prevents environmental contaminant ingress and drug vapor escape, minimizing exposure to hazardous drugs. Output is a secure fluid path for drug delivery. Benefits include reduced risk of leaks, spills, and hazardous drug exposure for clinicians and the environment.
Clinical Evidence
No clinical data included. Evidence is based on non-clinical bench testing, including biocompatibility (ISO 10993), particulate matter (USP <788>), sterilization validation (ISO 11137), packaging integrity (ASTM F88, F2096, D6653), and performance testing (leakage, flow rate, stress cracking, and microbial ingress).
Technological Characteristics
Materials: Polycarbonate (Makrolon RX1805), Polyvel Erucamide (RE 20), Silicone, Fluorosilicone fluid, Polypropylene. Sterilization: Irradiation. Connectivity: Mechanical luer connection (ISO 80369-7). Form factor: Single-use CSTD. No software/algorithm.
Indications for Use
Indicated for trained healthcare professionals in healthcare facilities to reconstitute, transfer, and administer hazardous and non-hazardous drugs for adult, pediatric, and neonatal patients.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 24, 2023
CareFusion Elgin Oktay Staff Regulatory Affairs Specialist 10020 Pacific Mesa Blvd San Diego, California 92121
Re: K223076
Trade/Device Name: BD Texium™ Closed Male Luer Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: February 24, 2023 Received: February 24, 2023
Dear Elgin Oktay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223076
Device Name BD Texium™ Closed Male Luer
#### Indications for Use (Describe)
The BD Texium™ Closed Male Luer (CML) is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of hazardous drugs when paired with the SmartSite™ NFC.
The BD Texium™ Closed Male Luer (CML) is indicated for use by trained healthcare professionals within healthcare facilities who prepare and/or administer non-hazardous drugs for adults, pediatrics and neonates.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CareFusion. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller font, it says "has joined BD."
# K223076-10(k)Summary
Submitter Information
Company Name: Company Address:
Name of contact Person: Company Phone: Email: Date Prepared:
CareFusion 10020 Pacific Mesa Blvd. San Diego, CA 92121, USA Elgin Oktay, Staff Regulatory Affairs Specialist 801-304-3908 Elgin.Oktay@bd.com March 24, 2023
#### Subject Device Identification
BD Texium™ Closed Male Luer Trade/Proprietary Name: Common Name: Closed System Drug Transfer Device (CSTD) Classification Name: Closed Antineoplastic and Drug Reconstitution and Transfer System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Classification Panel: General Hospital
#### Predicate Device Identification
| Trade/Proprietary Name: | |
|-------------------------|--|
| Common Name: | |
| Classification Name: | |
| Regulation Number: | |
| Regulation Name: | |
| Regulatory Class: | |
| Product Code: | |
| Classification Panel: | |
| Premarket Notification: | |
Alaris® Safety Male Luer Intravascular Administration Set Intravascular Administration Set 21 CFR 880.5440 Intravascular Administration Set Class II FPA General Hospital K053049
#### Reason for the Submission
The reason for this submission is to incorporate the following changes:
- New sterilization claim to "content sterile"
- Indications use updated to align with ONB product code
- Design modifications and material changes
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
- Packaging material changes ●
#### Device Description
The BD Texium™ CML is an airtight, leak-free and drip-free closed system drug transfer device (CSTD). When paired with devices containing a SmartSite™ NFC, the BD Texium™ CML mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the BD Texium™ CML/SmartSite™ NFC connection, thereby minimizing individual and environmental exposure to drugs, leaks and spills. The BD Texium™ CML is a sterile single-use CSTD intended for the reconstitution, transfer and administration of hazardous and non-hazardous drugs when paired with the SmartSite™ NFC.
#### Indication for Use
The BD Texium™ Closed Male Luer (CML) is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of hazardous and non-hazardous drugs when paired with the SmartSite™ Needle-Free Connector (NFC).
The BD Texium™ Closed Male Luer (CML) is indicated for use by trained healthcare professionals within healthcare facilities who prepare and or administer non-hazardous drugs for adults, pediatrics and neonates.
## Technological Characteristics and Substantial Equivalence
The following table presents an overview of comparisons between the subject device and the predicate device.
| | SUBJECT<br>BD Texium™ CML<br>(Subject Device) | PREDICATE<br>Alaris® Safety Male Luer<br>(K053049) | Substantial<br>Equivalence |
|--------------------------|-----------------------------------------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| FDA Regulation<br>Number | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| FDA Regulation<br>Name | Intravascular Administration<br>Set | Intravascular Administration<br>Set | Same |
| FDA Class | Class II | Class II | Same |
| FDA Product Code | ONB | FPA | Different –<br>product code<br>“ONB” was not<br>available at the<br>time of<br>predicate<br>submission |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
| Traditional 510(k) Premarket Notification |
|-------------------------------------------|
| BD Texium™ Closed Male Luer |
| | SUBJECT<br>BD Texium™ CML<br>(Subject Device) | PREDICATE<br>Alaris® Safety Male Luer<br>(K053049) | Substantial<br>Equivalence |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>operation/mechanism<br>of operation | The BD Texium™ Closed Male Luer (CML) is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of non-hazardous and hazardous drugs when paired with the SmartSite™ Needle-Free Connector (NFC). When paired with devices containing a SmartSite™ NFC the BD Texium™ CML mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the BD Texium™ CML/SmartSite™ NFC connection, thereby minimizing individual and environmental exposure to drugs, leaks, and spills. (e.g., airtight, leak-free and drip-free). | The Safety Male Luer is indicated for use when reconstituting, dispensing/transferring, administering, and disposal of potential hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products, as well as non-hazardous fluids. The Safety Male Luer is intended for use with the SmartSite Needle Free Valve port or standard open female luers. | Equivalent (updates to align with ONB product code) – air leakage, vacuum leakage, fluid leakage and residual fluid testing was conducted to verify new claims |
| Indication for Use | The BD Texium™ Closed Male Luer is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of hazardous and non-hazardous drugs when paired with the SmartSite™ NFC.<br><br>The BD Texium™ Closed Male Luer is indicated for use by trained healthcare professionals within | The Safety Male Luer is indicated for use when reconstituting, dispensing/transferring, administering, and disposal of potential hazardous fluids, such as chemotherapy, radioactive isotopes, and blood products, as well as non-hazardous fluids. The Safety Male Luer is intended for use with the SmartSite Needle Free Valve port or standard open female luers. | Different – clarified the medical setting and user |
| | SUBJECT<br>BD Texium™ CML<br>(Subject Device) | PREDICATE<br>Alaris® Safety Male Luer<br>(K053049) | Substantial<br>Equivalence |
| Device<br>Compatibility | prepare and/or administer<br>non-hazardous and<br>hazardous drugs for adults,<br>pediatrics and neonates.<br>SmartSite™ Needle-Free<br>Connector | SmartSite™ Needle Free<br>Valve port or standard open<br>female luers. | Equivalent –<br>device to be<br>used with<br>SmartSite™<br>Needle-Free<br>Connector |
| Method of<br>Administration | Closed system drug transfer<br>device (CSTD) | Closed system drug transfer<br>device (CSTD) | Same |
| NON-DEHP | Yes | Yes | Same |
| Device Components<br>/ Materials | CML<br>Male & Female Luer:<br>Polycarbonate (Makrolon<br>RX1805)<br>Actuator: Polyvel Erucamide<br>(RE 20)<br>Piston: Silicone<br>Seal Lubricant:<br>Fluorosilicone Fluid<br>Cap<br>Polypropylene | SML<br>Male & Female Luer:<br>Polycarbonate<br>Actuator: Polypropylene,<br>TPE<br>Piston: Silicone<br>Seal Lubricant:<br>Fluorosilicone Fluid<br>Priming Cap<br>Seal: Polypropylene<br>Actuator: Acetal<br>Membrane/Filter: PTFE | Different –<br>biocompatibility<br>testing was<br>conducted to<br>assess new<br>materials;<br>female luer<br>body design<br>changes were<br>verified through<br>ISO 80369-7<br>testing; actuator<br>and male luer<br>body design<br>changes were<br>verified through<br>air leakage,<br>vacuumed<br>leakage and<br>fluid leakage |
| Packaging | Individual device in peelable<br>pouch. 50 pouches placed in<br>dispenser box. 2 dispenser<br>boxes in each shipper with<br>one (1) Directions for Use.<br>New black ink and webs<br>included with primary<br>packaging. | Individual device in peelable<br>pouch with instructions. 50<br>pouches placed in dispenser<br>box. 2 dispenser boxes in<br>each shipper. | Different –<br>packaging<br>validation<br>verifies new<br>webs |
| No natural rubber<br>latex | Yes | Yes | Same |
| Sterilization Method | Irradiation | Irradiation | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller font, it says "has joined BD".
# Traditional 510(k) Premarket Notification
Traditional 510(k) Premarket Notification
The BD Texium™ Closed Male Luer
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
| Traditional 510(k) Premarket Notification | |
|-------------------------------------------|-----------------------------|
| | BD Texium™ Closed Male Luer |
| | SUBJECT<br>BD Texium™ CML<br>(Subject Device) | PREDICATE<br>Alaris® Safety Male Luer<br>(K053049) | Substantial<br>Equivalence |
|-----------------------------------------|----------------------------------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization Claim | Content Sterile | Fluid Path Sterile | Different -<br>package<br>integrity testing<br>including seal<br>strength, corner<br>thickness, seal<br>width, air<br>volume.<br>microbial<br>barrier, dye test,<br>and bubble leak<br>testing was<br>conducted to<br>verify sterile<br>barrier claim. |
| Biocompatibility | Biocompatible for the<br>intended use per ISO<br>10993-1 | Biocompatible for the<br>intended use per ISO 10993-<br>l | Same |
| Non-Pyrogenic | Yes | Yes | Same |
| Duration of Use | 7 days | 7 days | Same |
| Disinfect with 70%<br>Isopropyl Alcohol | Disinfect with 70%<br>Isopropyl Alcohol | Disinfect with 70% Isopropyl<br>Alcohol | Same |
| Priming Volume | 0.17 ml | 0.2 ml maximum | Equivalent -<br>specification is<br>within predicate<br>specification |
| Flow Rate | >4280 ml/hr, when activated<br>with a minimum 3.2 mm<br>(0.125") insertion depth | >4756 ml/hr | Different - flow<br>rate was<br>conducted to<br>verify rate |
| Shelf Life | 3 Years | 3 Years | Same |
## Substantial Equivalence Discussion:
Design verification testing was performed to demonstrate that the subject device is equivalent to the predicate device. All test results met their acceptance criteria and support that the BD Texium™ CML is safe and effective and is substantially equivalent to the predicate Alaris® Safety Male Luer. The subject device and the predicate devices are sterilized via irradiation and are single patient use devices. Any differences in materials between the two products have been evaluated through ISO 10993 testing, which demonstrate material safety.
Both the subject and predicate devices have the same principle of operation. The primary technological differences between the subject device and the predicate are geometry and raw material changes to the male
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" are the words "has joined BD" in a smaller font.
and female luers, as well as the actuator with seal. These differences are verified through testing of air leakage, vacuum leakage, fluid leakage and ISO 80369-7 testing. The BD Texium™ CML is claiming content sterile and the predicate claims fluid path sterile. Sterile barrier testing was performed to verify the claim. The BD Texium™ CML has changes to the primary packing in that an Instructions for Use is being provided with each shipper box.
### Discussion of Non-Clinical Tests:
The BD Texium™ CML, like the predicate device, was evaluated for biocompatibility appropriate to the contact characterization (tissue contacting (indirect) prolonged (> 24 hours to 30 days)). Testing is performed in accordance with the requirements of ISO 10993-1 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process, and the FDA Guidance for Industry - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Specific testing included:
- ISO 10993-3:2014 "Biological evaluation of medical device Part 3: Tests for genotoxicity, . carcinogenicity, and reproductive toxicity"
- ISO 10993-4:2017 "Biological evaluation of medical device Part 4: Selection of tests for ● interactions with blood"
- . ISO 10993-5:2009 "Biological evaluation of medical device – Part 5: Tests for in vitro cytotoxicity"
- ISO 10993-10:2021 "Biological evaluation of medical device Part 10: Tests for skin sensitization" ●
- ISO 10993-11:2017 "Biological evaluation of medical device - Part 11: Tests for systemic toxicity"
- ISO 10993-17:2002 "Biological evaluation of medical device Part 17: Establishment of allowable ● limits for leachable substances"
- ISO 10993-18:2020 "Biological evaluation of medical device - Part 18: Chemical characterization of materials"
- ISO 10993-23:2021 "Biological evaluation of medical device Part 23: Test for irritation" .
Other standards followed included:
- ISO 8536-4:2019 "Infusion equipment for medical use Part 4: Infusion Sets for Single Use . Gravity Feed, Section 8 Chemical Requirements and Biological Requirements"
## Particulate Testing:
The BD Texium™ CML was tested to demonstrate the product meets particulate requirements of United States Pharmacopeia, National Formulary (USP), General Chapter <788>, Particulate Matter in Injections (Current Standard).
#### Sterilization and Shelf Life
The subject device is radiation sterilized and data supports a shelf-life claim of 3 years. Sterilization and shelf-life testing were completed in accordance with the following FDA recognized guidelines:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
Sterilization:
- . ISO 11137-1:2006/AMD 1:2013 "Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices - Amendment 1"
- ISO 11137-2:2013 "Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose"
- United States Pharmacopeia, National Formulary (USP), General Chapter <85>, Bacterial . Endotoxins Test
- . United States Pharmacopeia, National Formulary (USP), General Chapter <161>, Medical Devices - Bacterial Endotoxin and Pyrogen Tests (2015)
- ANSI/AAMI ST72:2011 R:2016 Bacterial endotoxins Test methods, routine monitoring and ● alternatives to batch testing.
Shelf-Life:
- ISO 11607-1 First Edition 2006-04-15 Packaging for Terminally Sterilized Medical Devices Part . 1: Requirements for Materials, Sterile Barrier Systems And Packaging Systems [Including: Amendment 1 (2014)]
- . ISO 11607-2 First Edition 2006-04-15 Packaging for Terminally Sterilized Medical Devices – Part 2: Validation Requirements for Forming, Sealing and Assembly Processes [Including: Amendment 1 (2014)].
- Package testing included: ●
- O Standard Test Method for Seal Strength of Flexible Barrier Materials: ASTM F88/F88M-21
- O Standard Test Method for Thickness Measurement of Flexible Packaging Material: ASTM F2251-13
- Seal Transfer Width: internal testing O
- Standard Test Methods for Determining the Effects of High Altitude on Packaging Systems O by Vacuum Method: ASTM D6653/D6653M-13
- Bubble Leak Detection Test: ASTM F2096 and ASTM F2096-11 O
- Microbial Barrier: internal testing o
- Toluidine Blue Leak Detection Test: ASTM 3039-15 O
## Performance Testing:
The BD Texium™ CML was tested to verify compliance with the relevant sections of the following standards:
- ISO 8536-4:2020 "Infusion equipment for medical use Part 4: Infusion sets for single use, . gravity feed"
- ISO 80369-1:2018 "Small-bore connections for liquids and gases in healthcare applications ● - Part 1: General requirements"
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in a gray sans-serif font. Below the text, in a smaller font, it says "has joined BD".
- ISO 80369-7:2021 "Small-bore connections for liquids and gases in healthcare applications ● - Part 7: Connectors for intravascular or hypodermic applications"
- ISO 80369-20:2015 "Small-bore connections for liquids and gases in healthcare applications ● - Part 20: Common test methods"
Performance testing was conducted per below:
- Leakage by pressure decay
- Sub-Atmospheric Air Pressure Leakage ●
- Stress Cracking
- Resistance to Separation from Axial Load ●
- Resistance to Separation from Unscrewing
- Resistance to Overriding
- Air leakage
- Vacuum leakage ●
- Fluid leakage ●
- Residual fluid
- Flow rate
#### Microbial Ingress Testing:
Microbial ingress was performed based on the following FDA guidance document:
- . Guidance for Industry and FDA staff; Intravascular Administration Sets Premarket Notification Submissions [510(k)], July 11, 2008
Additional testing was conducted to demonstrate:
- . Harsh Infusates testing: Device tests for multiple days with worst case infusates
## Clinical Data:
There are no clinical data included in this submission.
#### Conclusion:
The information in this submission supports the safety and efficacy of the subject device for its intended use and demonstrates substantial equivalence with the predicate device. The BD Texium™ CML differences in geometry, materials and sterilization claim do not raise new questions about safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.