← Product Code [LKF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF) · K223064

# ALLY II Uterine Positioning System™ (UPS) (K223064)

_CooperSurgical, Inc. · LKF · Oct 28, 2022 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K223064

## Device Facts

- **Applicant:** CooperSurgical, Inc.
- **Product Code:** [LKF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF.md)
- **Decision Date:** Oct 28, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The ALLY II UPS™ (Uterine Positioning System) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.

## Device Story

Electromechanical uterine positioning system; attaches to OR table rail; supports uterine manipulator during laparoscopic surgery. Consists of single, multi-segmented, articulated arm and sterile disposable adapter drape. User positions manipulator manually when arm is unlocked; foot pedal activates linear actuator to apply tension to internal cable, locking arm segments in desired position. Provides lateral/medial movement from single point. Used by surgical staff in OR. Enhances surgical access and manipulator stability; improves ergonomics and device durability compared to predicate.

## Clinical Evidence

Bench testing only. Verification and validation activities included durability (useful life) testing, holding force testing, range of motion testing, link soak testing, and foot pedal cable retention analysis. Device demonstrated compliance with IEC 60601-1 and IEC 60601-1-2 standards. All pre-determined acceptance criteria were met.

## Technological Characteristics

Electromechanical articulated arm; 17-4 stainless steel links; PEEK protective inserts for actuator cable; linear actuator energy source; detachable foot pedal; integrated grounding post; non-sterile; IEC 60601-1 and IEC 60601-1-2 compliant.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- ALLY Uterine Positioning System ([K141523](/device/K141523.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 28, 2022

CooperSurgical, Inc. Yin Huang Regulatory Affairs Specialist 95 Corporate Drive Trumbull, CT 06611

Re: K223064

> Trade/Device Name: ALLY II UPSTM (Uterine Positioning System) Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: September 28, 2022 Received: September 30, 2022

Dear Yin Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223064

Device Name ALLY II UPSTM (Uterine Positioning System)

#### Indications for Use (Describe)

The ALLY II UPS™ (Uterine Positioning System) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary – K223064 ALLY II UPS™ (Uterine Positioning System)

# I. Submitter Information

| Company Name:           | CooperSurgical Inc.                                                 |
|-------------------------|---------------------------------------------------------------------|
| Company Address:        | 95 Corporate Drive<br>Trumbull, CT 06611<br>Telephone: 203-601-5200 |
| Contact Person:         | Yin Huang<br>Regulatory Affairs Specialist                          |
| Date Prepared:          | October 27, 2022                                                    |
| II. Device Information: |                                                                     |

| Trade Name:        | ALLY II UPS™ (Uterine Positioning System)           |
|--------------------|-----------------------------------------------------|
| Common Name:       | Uterine positioning system                          |
| Regulation Name:   | Obstetric-gynecologic specialized manual instrument |
| Regulation Number: | 884.4530                                            |
| Regulatory Class:  | II                                                  |
| Product Code:      | LKF (Cannula, Manipulator/Injector, Uterine)        |

## III. Predicate Device Information:

ALLY Uterine Positioning System™ (K141523), manufactured by CooperSurgical Inc. The predicate device has not been subject to a design-related recall

# IV. Device Description:

{4}------------------------------------------------

The ALLY II UPS™ (Uterine Positioning System) attaches to the operating room table and enables the bed-side assistant to readily mount, hold, and position the manipulator during laparoscopic surgical procedures. The ALLY II UPS enables access and provides the ability to maneuver and maintain the manipulator in a desired position. The ALLY II UPS consists of the ALLY II UPS and a manipulator adapter with built-in sterile drape, known as the adapter drape.

The CooperSurgical ALLY II UPS is a non-patient contacting, electromechanical device that consists of a single, multi-segmented, articulated arm. The ALLY II UPS can be attached to the standard operating room bed rail, and a separate, sterile, disposable Adapter Drape that is used to attach a uterine manipulator to the ALLY II UPS. When unlocked, the flexible arm allows the attached manipulator to be positioned by the user. The arm can then be locked in the desired position by releasing a foot pedal, activating a linear actuator that applies tension to an internal cable, drawing the segments together and thus locking the arm. The segmented design of the arm allows lateral/medial movement from a single point to position the uterine manipulator.

The purpose of this submission is to gain clearance for minor modifications made to the ALL Y Uterine Positioning System to meet user needs and allow for user convenience. The overall functionality and interface of the ALLY II UPS for the user remains the same.

#### V. Indications for Use:

The ALLY II UPS™ (Uterine Positioning System) is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.

### VI. Comparison of Technological Characteristics with the Predicate Device

A comparison of the intended use and technological characteristics of the subject and predicate device are included in the table below.

{5}------------------------------------------------

| Attribute                              | Predicate                                                                                                                                                                                                                                                                                                                                     | Subject                                                                                                                                                                                                                                                                                                                                                                                                                      | Comparison                                                                                                                                                                                                                                                                 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                        | ALLY Uterine                                                                                                                                                                                                                                                                                                                                  | ALLY II UPSTM                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                            |
|                                        | Positioning System                                                                                                                                                                                                                                                                                                                            | (Uterine Positioning                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                            |
|                                        | (UPS) - K141523                                                                                                                                                                                                                                                                                                                               | System)                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                            |
| Principles of Operation                | The ALLY II UPSTM<br>(Uterine Positioning<br>System) is intended to<br>assist the surgical staff in<br>mounting, positioning,<br>and holding uterine<br>manipulators during<br>gynecological<br>laparoscopic surgical<br>procedures. It is intended<br>for use by trained<br>operating room personnel<br>in an operating room<br>environment. | The ALLY II UPSTM<br>(Uterine Positioning<br>System) is intended to<br>assist the surgical staff in<br>mounting, positioning,<br>and holding uterine<br>manipulators during<br>gynecological<br>laparoscopic surgical<br>procedures. It is intended<br>for use by trained<br>operating room personnel<br>in an operating room<br>environment.                                                                                | Identical                                                                                                                                                                                                                                                                  |
| Sterilization                          | Non-sterile                                                                                                                                                                                                                                                                                                                                   | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                  |
| Range of motion                        | When the device is<br>attached to the operating<br>table bed rail:<br><br>Anteverted hand position<br>- The distal tip of the<br>uterine manipulator<br>attached to the device can<br>reach the:<br>Anteverted hand<br>position can<br>manipulate to<br>designated 6<br>points of<br>extensionRetroverted hand position<br>- Not Applicable   | When the device is<br>attached to the operating<br>table bed rail:<br><br>Anteverted hand position<br>- The distal tip of the<br>uterine manipulator<br>attached to the device can<br>reach the:<br>Anteverted hand<br>position can<br>manipulate to<br>designated 6<br>points of<br>extensionRetroverted hand position<br>=<br>Retroverted hand position<br>is able to manipulate to<br>designated 5 points of<br>extension | Different:<br><br>The subject device has<br>greater Range of Motion<br>than the predicate. The<br>Range of Motion<br>difference does not raise<br>different questions of<br>S&E.                                                                                           |
| Rotation angle of the<br>Elbow         | None – elbow is fixed at<br>50° toward sagittal plane<br>of patient                                                                                                                                                                                                                                                                           | Rotate 70° away from the<br>sagittal plane of patient                                                                                                                                                                                                                                                                                                                                                                        | Different:<br><br>The subject device has<br>greater Range of Motion<br>than the predicate. The<br>Elbow rotation difference<br>does not raise different<br>questions of S&E.                                                                                               |
| Material(s) of the arm<br>LINKS        | 303 stainless Steel                                                                                                                                                                                                                                                                                                                           | 17-4 stainless steel                                                                                                                                                                                                                                                                                                                                                                                                         | Different:                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                              | This modification allows<br>for a longer useful life of<br>the Flexible Arm for the<br>subject device than the<br>predicate The material<br>change does not raise<br>different questions of<br>S&E                                                                         |
| PEEK Inserts in the<br>LINKS and Elbow | No protective inserts over<br>actuator cable                                                                                                                                                                                                                                                                                                  | Additional inserts to<br>provide a protective<br>barrier for the cable                                                                                                                                                                                                                                                                                                                                                       | Different:                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                              | This modification allows<br>for a longer useful life of<br>the Flexible Arm for the<br>subject device than the<br>predicate. The PEEK<br>insert in the LINKs does<br>not raise different<br>questions of S&E                                                               |
| Foot pedal connection to<br>device     | Hard wired to device -<br>undetachable                                                                                                                                                                                                                                                                                                        | Connector can be<br>detachable                                                                                                                                                                                                                                                                                                                                                                                               | Different:                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                              | The subject device<br>provides more<br>convenience for users<br>than the predicate by<br>allowing ease of mobility<br>and packing of the device<br>in moving the unit around<br>the hospital. The<br>detachable connection<br>does not raise different<br>questions of S&E |
| Grounding Post                         | None                                                                                                                                                                                                                                                                                                                                          | Attached to bottom of<br>device as a built-in feature<br>for grounding the device                                                                                                                                                                                                                                                                                                                                            | Different:                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                              | The Grounding Post is<br>added as an additional<br>grounding option for<br>hospitals when they have<br>concerns over electrical<br>grounding within the<br>hospital setting.<br>The Grounding Post does<br>not raise different<br>questions of S&E                         |
| Handle                                 | Flat plate with finger<br>holding                                                                                                                                                                                                                                                                                                             | Whole hand handle                                                                                                                                                                                                                                                                                                                                                                                                            | Different:                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                              | This modification was<br>made as an ergonomic<br>change to provide more<br>user convenience                                                                                                                                                                                |

{6}------------------------------------------------

{7}------------------------------------------------

The subject ALLY II UPS has the same Intended Use, fundamental scientific technology, and similar materials as the predicate device. The subject device has an identical principle of operation and similar design to the predicate device.

The minor design changes for user convenience consist of:

- 1) Improvement to the range of motion of the flexible arm to give users more flexibility in positioning uterine manipulators during gynecological laparoscopic surgical procedures.
- 2) Minor modifications to the design features of the flexible arm to aid in flexibility, durability, and cleaning of the device.
- 3) Addition of a detachable foot pedal connection where the predicate device was hard wired into the unit.

The changes in technological characteristics do not raise different questions of safety and effectiveness.

## VII. Performance Data

The support the proposed modifications to the subject device, design control activities and a risk analysis (depicting device change, risk associated, verification method, acceptance criteria and summary of results) was performed. The ALLY II UPS was tested to identified verification and validation activities including tests comparing the subject device to the predicate ALL Y UPS -K141523.

- Durability (Useful Life) testing .
- . Holding force testing
- Range of motion testing ●
- . Link Soak testing
- . Foot Pedal Cable Retention Justification

The device is compliant with the following standards:

IEC 60601-1: 2005+AMD1:2012+AMD2:2020 CSV Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance

{8}------------------------------------------------

IEC 60601-1-2: 2014 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance

The subject ALLY II UPS met the pre-determined acceptance criteria for each intended output. Therefore, design verification testing determined that the subject ALLY II UPS is substantially equivalent to the identified predicate device.

#### VIII. Conclusion

The results of the testing described above demonstrate that the ALLY II UPS™ (Uterine Positioning System) is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K223064](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K223064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
