← Product Code [FAJ](/productcode/FAJ) · K223040

# Endo SmartCap (K223040)

_Medivators, Inc. · FAJ · Mar 1, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223040

## Device Facts

- **Applicant:** Medivators, Inc.
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Mar 1, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

## Device Story

Endo SmartCap is a sterile water bottle adapter used during endoscopic procedures. It connects to a user-provided sterile water container and an air or CO2 source (e.g., endoscope processor or insufflator). The device utilizes a coaxial tube system; pressurized air or CO2 enters the outer tube to pressurize the water bottle, forcing sterile water through the inner tube to the endoscope. It features a bottle cap, a weighted distal tip for submersion, and a pinch clip to prevent dripping when detached. The device is provided sterile (EtO), intended for 24-hour multi-patient use, and discarded daily. It does not contact the patient directly. It facilitates irrigation and insufflation, aiding visualization and procedure performance.

## Clinical Evidence

Bench testing only. Performance data included flow rate testing and backflow prevention testing to ensure the device meets or exceeds manufacturer requirements and prevents water backflow into the insufflator. All tests passed.

## Technological Characteristics

Coaxial tube system; materials include metals, plastics, and elastomers. Sterilized via EtO. Standalone accessory; no electronic components. Operates via pneumatic pressure from external insufflator or endoscope processor.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ENDO SMARTCAP ([K093665](/device/K093665.md))

## Reference Devices

- CO2Compact ([K111648](/device/K111648.md))
- Co2Co-efficient ([K053008](/device/K053008.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 1, 2023

Medivators, Inc. Disha Kabrawala Senior Regulatory Affairs Specialist 14605 28th Ave North Minneapolis, MN 55447

Re: K223040

> Trade/Device Name: Endo SmartCapTM Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: January 26, 2023 Received: January 30, 2023

Dear Disha Kabrawala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223040

Device Name Endo SmartCap™

Indications for Use (Describe)

The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cantel Medical. The logo features a blue and green globe-like design on the left, followed by the word "CANTEL" in large, bold, blue letters. The globe design has a layered appearance, with alternating blue and green sections that curve around the sphere. 

## 510(k) — K2203040 Summary For Endo SmartCap™

Medivators Inc 14605 28th Ave N, Minneapolis, MN 55447 Fax No: 844-348-5637

| Contact Person: | Disha Kabrawala<br>Senior Regulatory Affairs Specialist<br>Phone: 732-319-7766<br>Email: Disha.Kabrawala@cantel.com |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| Summary Date:   | October 19, 2022                                                                                                    |

{4}------------------------------------------------

#### 1. Device Name

| Trade Name:            | Endo SmartCap™               |
|------------------------|------------------------------|
| 510(K)                 | K2203040                     |
| Device Class:          | II                           |
| Common/usual Name:     | Sterile Water Bottle Adapter |
| Classification Name:   | Endoscopes and Accessories   |
| Classification Number: | 21 CFR 876.1500              |
| Product Code:          | FAJ                          |

#### 2. Predicate Devices

## K093665: ENDO SMARTCAP

#### 3. Description of Devices

The Endo SmartCap™ supplies sterile water and either air or CO2 to an endoscope during endoscopic procedures when connected to a sterile water source and an air or CO2 source. Two tubes are arranged coaxially; air or CO2 supplied through the outer tube pressurizes the sterile water container, forcing water up through the inner tube and to the endoscope. The proximal end of the tube set includes a bottle cap for attachment to a user-provided sterile water container; the Endo SmartCap™ is compatible with all major brands of sterile water containers. The proximal end of the tube set terminates in a tube weight which keeps the tip of the tubing submerged. A pinch clip prevents water from dripping from the tube's distal end when the device is detached from the endoscope. The distal end of the tube set terminates in a connector which mates with the corresponding brand of endoscope; Endo SmartCap™ models are available for all major brands of endoscopes. The device is provided sterile (EtO). The device is fabricated from metals, plastics, and elastomers. The Endo SmartCap™ is intended for 24 hour multi-patient use and should be discarded daily. No portion of the device comes in direct contact with patients.

{5}------------------------------------------------

#### 4. Intended Use / Indications for Use

The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

#### 5. Comparison of Technological Characteristics with the Predicate Device

The Endo SmartCap™ is the same as the predicate device with the exception of the proposed modification. A comparison between the proposed and predicate devices is included in Table 1. Since there are no technological differences between the proposed and predicate devices, there are no new concerns of safety and effectiveness.

## Table 1.: Physical Description and Technological Properties vs the Predicate Device

| Feature                           | Proposed<br>Endo SmartCap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate (K093665)<br>Endo SmartCap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Comparison |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications<br>for Use            | The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                           | The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical  |
| Feature                           | Proposed Endo SmartCap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate (K093665) Endo SmartCap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison |
| Device Use/Principle of operation | The Endo SmartCap tubesets' mode of operation is the same as the device provided by the endoscope manufacturer when installed to commonly used sterile water bottles used in endoscopic procedures. The threaded cap is installed onto commonly used, readily available sterile water bottles and uses a coaxial tube to partition water and air/CO2 which is connected to the endoscope-interfacing connector. Pressurized air or CO2 provided by the endoscope's processor/light source or a CO2 insufflator is fed into the endoscope (either directly or first through the Endo SmartCap & bottle system via luer connection). | The Endo SmartCap tubesets' mode of operation is the same as the device provided by the endoscope manufacturer when installed to commonly used sterile water bottles used in endoscopic procedures. The threaded cap is installed onto commonly used, readily available sterile water bottles and uses a coaxial tube to partition water and air/CO2 which is connected to the endoscope-interfacing connector. Pressurized air or CO2 provided by the endoscope's processor/light source or a CO2 insufflator is fed into the endoscope (either directly or first through the Endo SmartCap & bottle system via luer connection). | Identical  |
| Intended Use                      | Daily (24hr) multi-patient use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Daily (24hr) multi-patient use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical  |
| Reprocessing                      | None; Daily disposal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | None; Daily disposal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical  |
| Patient Contact                   | No direct contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | No direct contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical  |
| Water Bottles                     | Supplied by user; Disposed of daily                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Supplied by user; Disposed of daily                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical  |
| Materials                         | Metals, plastics, and elastomers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Metals, plastics, and elastomers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical  |
| Water Flow Rate Performance       | Meet or exceed the endoscope manufacturer's requirements in the average of total volume of water flowed through the endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Meet or exceed the endoscope manufacturer's requirements in the average of total volume of water flowed through the endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical  |
| Gas Flow Rate performance         | Meet or exceed the endoscope manufacturer's requirements in the average of total volume of gas flowed through the endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Meet or exceed the endoscope manufacturer's requirements in the average of total volume of gas flowed through the endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical  |

{6}------------------------------------------------

## Cantel Medical Special 510(k) PREMARKET NOTIFICATION Modification to Endo SmartCapTM

{7}------------------------------------------------

#### 6. Summary of Changes

The changes that are the subject of this submission are as follows:

- The use of Endo SmartCap source tubing, PN 100551 with the ● CO2Compact (K111648) and Co2Co-efficient (K053008) insufflators. There is no impact on device design. The change has been initiated to accommodate with customer need.
- The use of the Endo SmartCap CO2 source tubing, PN 100551 with Endo SmartCap- CO2 extension tubing as tubing kits. The change has been initiated to accommodate with customer preferences. There is no impact to the device design, safety and efficacy.

#### 7. Summary of the Performance Data

The subject Endo SmartCap is equivalent to the predicate Endo SmartCap, aside from the modifications subject of this submission.

The flowrate testing and backflow prevention testing were conducted using the same methodology as that indicated in the predicate 510(k), K093665, to confirm the safety and effectiveness of the subject device has not been compromised with the device modification.

The following testing has been conducted to demonstrate that the use of a Endo SmartCap is safe and effective for its intended use:

| Test                      | Acceptance Criteria                                                                                                                        | Result |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Confirmation of flow rate | The average mean flow of the 100551-<br>production sample must be no lower than<br>10% of the flow of the same tubeset<br>without a filter | PASS   |
| Prevention of backflow    | The 100551 shall not allow water to<br>backflow into the CO2EFFICIENT<br>insufflator                                                       | PASS   |

#### 7. Conclusion

Based on the intended use, technological characteristics and for the non-clinical performance data, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device (K093665), Class II (21 CFR 876.1500), product code FAJ.

---

**Source:** [https://fda.innolitics.com/device/K223040](https://fda.innolitics.com/device/K223040)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com