← Product Code [LLZ](/productcode/LLZ) · K222987

# Akunah REFLECT (K222987)

_Akunah Pty, Ltd. · LLZ · May 10, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K222987

## Device Facts

- **Applicant:** Akunah Pty, Ltd.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** May 10, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.

## Device Story

Akunah REFLECT is a software-only preoperative planning tool for shoulder surgery. Input consists of anatomic data (medical images); software allows surgeons to visualize, measure, reconstruct, and annotate anatomy; select and position shoulder implants; and generate pre-surgical planning reports. Used in clinical settings by orthopedic surgeons to assist in surgical planning and component positioning. Output is a planning report used by surgeons to guide clinical decision-making. Benefits include improved surgical preparation and implant placement accuracy. Device is used in conjunction with expert clinical judgment.

## Clinical Evidence

Bench testing only. Software verification and validation performed per FDA guidance. Functional requirements met. Geometric accuracy of virtual models (auto-segmentation and premorbid reconstruction) and measurement tools verified as equivalent to predicate. Usability testing validated desktop software for surgical planning.

## Technological Characteristics

Software-only medical image management and processing system. Features auto-segmentation, premorbid reconstruction, and measurement tools. Operates as a desktop application connected to an online management system (OMS).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Mimics Medical ([K183105](/device/K183105.md))

## Reference Devices

- BLUEPRINT™ Patient Specific Instrumentation ([K203315](/device/K203315.md))
- Materialise Shoulder System, Materialise Shoulder Guide and Models, SurgiCase Shoulder Planner ([K193560](/device/K193560.md))
- Surgicase Shoulder Planner ([K212569](/device/K212569.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Akunah PTY LTD Shanthan Pather Coo 131 Warriewood St QUEENSLAND, 4151 AUSTRALIA

Re: K222987

May 10, 2023

Trade/Device Name: Akunah REFLECT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: April 6, 2023 Received: April 6, 2023

Dear Shanthan Pather:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K22987

Device Name Akunah REFLECT

#### Indications for Use (Describe)

Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Akunah. The logo consists of a circular design made up of many small orange circles. To the right of the circular design is the word "akunah" in orange lowercase letters.

# 510(k) Summary

#### K222987

# Traditional 510(k) – Akunah REFLECT

May 10th 2023

| Device Name           |                                                                                                                     |
|-----------------------|---------------------------------------------------------------------------------------------------------------------|
| Trade name:           | Akunah REFLECT                                                                                                      |
| Common Name:          | Image processing system                                                                                             |
| Classification Name:  | Medical image management and processing system                                                                      |
| Regulation Number:    | §892.2050                                                                                                           |
| Submitter             |                                                                                                                     |
| Name:                 | Akunah Pty Ltd                                                                                                      |
| Address:              | Suite 307, Nicholson Street Specialist Centre<br>Level 9, 121 Newdegate St, Greenslopes 4120<br>Brisbane, Australia |
| Company Contact       |                                                                                                                     |
| Contact Name:         | Akunah Pty Ltd                                                                                                      |
| Contact Person:       | Dr. Shanthan Pather                                                                                                 |
| Title:                | COO, Design & Regulatory Affairs Engineer                                                                           |
| Address:              | Suite 307, Nicholson Street Specialist Centre<br>Level 9, 121 Newdegate St, Greenslopes 4120<br>Brisbane, Australia |
| Tel:                  | +61 3445 1591                                                                                                       |
| Fax:                  | +61 3445 1592                                                                                                       |
| Email:                | sha@akunah.com                                                                                                      |
| Classification        |                                                                                                                     |
| Device                | Class: Class II                                                                                                     |
| Classification Panel: | Radiology                                                                                                           |
| Product Code:         | QIH, LLZ                                                                                                            |

21 CFR section 892.2050, Class II Radiology Devices – Panel: Radiology Office of Radiological Health (OHT8)

### Equivalent / Predicate Device

| Trade Name                                                                                                          | 510(k) number | Decision Date  | Applicant      |
|---------------------------------------------------------------------------------------------------------------------|---------------|----------------|----------------|
| Mimics Medical (PRIMARY PREDICATE)                                                                                  | K183105       | March 27, 2019 | Materialise NV |
| BLUEPRINT™ Patient Specific<br>Instrumentation (REFERENCE)                                                          | K203315       | April 15, 2021 | Tornier SAS    |
| Materialise Shoulder System,<br>Materialise Shoulder Guide and<br>Models, SurgiCase Shoulder Planner<br>(REFERENCE) | K193560       | March 20, 2020 | Materialise NV |

Suite 307 Nicholson Street Specialist Centre Level 9, 121 Newdegate Street Greenslopes Private Hospital, QLD, 4127 AUSTRALIA

E: info@akunah.com P: +61 7 3445 1591 F: +61 7 3445 1592

akunah.com ABN 58 646 020 322

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Akunah. On the left side of the logo is a circular design made up of many small circles. The circles are arranged in a spiral pattern, with the circles getting smaller as they move towards the center. To the right of the circular design is the word "akunah" in lowercase letters. The word is written in a sans-serif font and is colored in a gradient of orange and yellow.

# Device Description

Akunah REFLECT Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The user interface software is connected to an online management system (OMS), and is intended to be used by orthopaedic surgeons as a preoperative planning software for shoulder surgery. The software is intended to allow surgeons to:

- Visualize, measure, reconstruct, and annotate anatomic data,
- Select and position shoulder implant devices,
- . Output pre-surgical planning reports

The software is to be used for adult patients only and should not be used for diagnostic purpose.

#### Intended Use

Akunah REFLECT Software is intended to be used as a medical software to assist in pre-operative surgical planning for shoulder surgery.

#### Indication For Use

Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.

### Comparison to Predicate Device

Akunah Reflect has been compared to legally marketed predicate device Materialise – Mimics Medical (K183105), and references devices:- Tornier - BLUEPRINT Patient Specific Instrumentation (K203315), and Materialise - Surgicase Shoulder Planner (K212569). The predicate device shares the same intended use for the software, specific to treatment planning for surgical procedures, allowing visualisation, reconstruction, measurements, annotations and editing of anatomic data. To achieve these features Akunah REFLECT employs similar fundamental technologies as the predicates.

The following technological differences exists between the subject device and the predicates:

- Akunah REFLECT does not offer a hardware component – Patient Specific Guide (PSI) or Anatomic Models, as it is a software-only device
- . Akunah REFLECT possess a unique user interface to optimise workflow and user experience.

The differences outlined above do not raise new questions of safety and effectiveness over the predicate device as demonstrated in verification and validation testing.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Akunah. The logo consists of two parts: a circular pattern of orange dots on the left and the word "akunah" in orange on the right. The dots are arranged in concentric circles, creating a visually appealing design. The word "akunah" is written in a sans-serif font, with a gradient of orange color that transitions from a darker shade to a lighter shade.

# Performance Data

Software verification and validation testing were performed, and documentation provided as per the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005). Verification testing was performed to address all functional requirements, and usability testing was performed to demonstrate overall validation of the software against the user needs.

- Akunah REFLECT met the acceptance criteria stated for all functional verification testing across all systems and modules.
- . Geometric accuracy of the virtual models created in the subject device were assessed. Both Akunah Auto-segmentation and Premorbid Reconstruction demonstrated equivalent accuracy as established by the predicate devices.
- . Measurement Tools were verified to achieve an acceptable accuracy
- . Usability testing demonstrated the validation of the software on the Desktop for surgical planning

In conclusion, all performance testing conducted demonstrated device performance and substantial equivalence to the predicate device.

# Substantial Equivalence Conclusion

The results of the verification and validation testing raise no concerns regarding the safety and effectiveness of the product. Minor differences in the non-fundamental technological characteristics raise no question of safety and effectiveness of the software. Performance data support the demonstration of substantial equivalence and usability. Therefore, the intended use, fundamental technological characteristics, and principle of operation of the Akunah REFLECT software are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K222987](https://fda.innolitics.com/device/K222987)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
