Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
Device Story
Akunah REFLECT is a software-only preoperative planning tool for shoulder surgery. Input consists of anatomic data (medical images); software allows surgeons to visualize, measure, reconstruct, and annotate anatomy; select and position shoulder implants; and generate pre-surgical planning reports. Used in clinical settings by orthopedic surgeons to assist in surgical planning and component positioning. Output is a planning report used by surgeons to guide clinical decision-making. Benefits include improved surgical preparation and implant placement accuracy. Device is used in conjunction with expert clinical judgment.
Clinical Evidence
Bench testing only. Software verification and validation performed per FDA guidance. Functional requirements met. Geometric accuracy of virtual models (auto-segmentation and premorbid reconstruction) and measurement tools verified as equivalent to predicate. Usability testing validated desktop software for surgical planning.
Technological Characteristics
Software-only medical image management and processing system. Features auto-segmentation, premorbid reconstruction, and measurement tools. Operates as a desktop application connected to an online management system (OMS).
Indications for Use
Indicated for adult patients requiring pre-surgical planning for shoulder arthroplasty, fracture, deformity correction, and stabilization. Not for diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Akunah PTY LTD Shanthan Pather Coo 131 Warriewood St QUEENSLAND, 4151 AUSTRALIA
Re: K222987
May 10, 2023
Trade/Device Name: Akunah REFLECT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: April 6, 2023 Received: April 6, 2023
Dear Shanthan Pather:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22987
Device Name Akunah REFLECT
#### Indications for Use (Describe)
Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Akunah. The logo consists of a circular design made up of many small orange circles. To the right of the circular design is the word "akunah" in orange lowercase letters.
# 510(k) Summary
#### K222987
# Traditional 510(k) – Akunah REFLECT
May 10th 2023
| Device Name | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------|
| Trade name: | Akunah REFLECT |
| Common Name: | Image processing system |
| Classification Name: | Medical image management and processing system |
| Regulation Number: | §892.2050 |
| Submitter | |
| Name: | Akunah Pty Ltd |
| Address: | Suite 307, Nicholson Street Specialist Centre<br>Level 9, 121 Newdegate St, Greenslopes 4120<br>Brisbane, Australia |
| Company Contact | |
| Contact Name: | Akunah Pty Ltd |
| Contact Person: | Dr. Shanthan Pather |
| Title: | COO, Design & Regulatory Affairs Engineer |
| Address: | Suite 307, Nicholson Street Specialist Centre<br>Level 9, 121 Newdegate St, Greenslopes 4120<br>Brisbane, Australia |
| Tel: | +61 3445 1591 |
| Fax: | +61 3445 1592 |
| Email: | sha@akunah.com |
| Classification | |
| Device | Class: Class II |
| Classification Panel: | Radiology |
| Product Code: | QIH, LLZ |
21 CFR section 892.2050, Class II Radiology Devices – Panel: Radiology Office of Radiological Health (OHT8)
### Equivalent / Predicate Device
| Trade Name | 510(k) number | Decision Date | Applicant |
|---------------------------------------------------------------------------------------------------------------------|---------------|----------------|----------------|
| Mimics Medical (PRIMARY PREDICATE) | K183105 | March 27, 2019 | Materialise NV |
| BLUEPRINT™ Patient Specific<br>Instrumentation (REFERENCE) | K203315 | April 15, 2021 | Tornier SAS |
| Materialise Shoulder System,<br>Materialise Shoulder Guide and<br>Models, SurgiCase Shoulder Planner<br>(REFERENCE) | K193560 | March 20, 2020 | Materialise NV |
Suite 307 Nicholson Street Specialist Centre Level 9, 121 Newdegate Street Greenslopes Private Hospital, QLD, 4127 AUSTRALIA
E: info@akunah.com P: +61 7 3445 1591 F: +61 7 3445 1592
akunah.com ABN 58 646 020 322
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Akunah. On the left side of the logo is a circular design made up of many small circles. The circles are arranged in a spiral pattern, with the circles getting smaller as they move towards the center. To the right of the circular design is the word "akunah" in lowercase letters. The word is written in a sans-serif font and is colored in a gradient of orange and yellow.
# Device Description
Akunah REFLECT Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The user interface software is connected to an online management system (OMS), and is intended to be used by orthopaedic surgeons as a preoperative planning software for shoulder surgery. The software is intended to allow surgeons to:
- Visualize, measure, reconstruct, and annotate anatomic data,
- Select and position shoulder implant devices,
- . Output pre-surgical planning reports
The software is to be used for adult patients only and should not be used for diagnostic purpose.
#### Intended Use
Akunah REFLECT Software is intended to be used as a medical software to assist in pre-operative surgical planning for shoulder surgery.
#### Indication For Use
Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
### Comparison to Predicate Device
Akunah Reflect has been compared to legally marketed predicate device Materialise – Mimics Medical (K183105), and references devices:- Tornier - BLUEPRINT Patient Specific Instrumentation (K203315), and Materialise - Surgicase Shoulder Planner (K212569). The predicate device shares the same intended use for the software, specific to treatment planning for surgical procedures, allowing visualisation, reconstruction, measurements, annotations and editing of anatomic data. To achieve these features Akunah REFLECT employs similar fundamental technologies as the predicates.
The following technological differences exists between the subject device and the predicates:
- Akunah REFLECT does not offer a hardware component – Patient Specific Guide (PSI) or Anatomic Models, as it is a software-only device
- . Akunah REFLECT possess a unique user interface to optimise workflow and user experience.
The differences outlined above do not raise new questions of safety and effectiveness over the predicate device as demonstrated in verification and validation testing.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Akunah. The logo consists of two parts: a circular pattern of orange dots on the left and the word "akunah" in orange on the right. The dots are arranged in concentric circles, creating a visually appealing design. The word "akunah" is written in a sans-serif font, with a gradient of orange color that transitions from a darker shade to a lighter shade.
# Performance Data
Software verification and validation testing were performed, and documentation provided as per the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005). Verification testing was performed to address all functional requirements, and usability testing was performed to demonstrate overall validation of the software against the user needs.
- Akunah REFLECT met the acceptance criteria stated for all functional verification testing across all systems and modules.
- . Geometric accuracy of the virtual models created in the subject device were assessed. Both Akunah Auto-segmentation and Premorbid Reconstruction demonstrated equivalent accuracy as established by the predicate devices.
- . Measurement Tools were verified to achieve an acceptable accuracy
- . Usability testing demonstrated the validation of the software on the Desktop for surgical planning
In conclusion, all performance testing conducted demonstrated device performance and substantial equivalence to the predicate device.
# Substantial Equivalence Conclusion
The results of the verification and validation testing raise no concerns regarding the safety and effectiveness of the product. Minor differences in the non-fundamental technological characteristics raise no question of safety and effectiveness of the software. Performance data support the demonstration of substantial equivalence and usability. Therefore, the intended use, fundamental technological characteristics, and principle of operation of the Akunah REFLECT software are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.