The BioClean Implant Restorative System is designed to support fixed/detachable partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed/detachable restorations that can be attached with a friction-based engagement system. All digitally designed copings for use with the BioClean Implant Restorative System are to be sent to BioClean Dental or a validated milling center for manufacture. The BioClean Implant Restorative System is compatible with the following: Bone Level Straumann® Multi-Unit Abutments NC/RC Straumann Multi-Unit Abutment Occlusal Screw 4.6mm Platform Diameter, All Gingival Heights
Device Story
System provides rigid connection of partial/full arch restorations to endosseous dental implants; consists of titanium alloy copings, denture housings, and nylon retention inserts. Copings attach to OEM multi-unit abutments via friction-based engagement system. Designed for use in dental clinics; operated by dental professionals. Copings available in pre-manufactured (0°-9° angulation) or patient-specific versions designed via CAD software and fabricated at validated milling centers. System allows parallel multi-unit prosthesis creation by combining angled abutments and copings (total assembly angulation up to 30°). Enables patient-removable fixed/detachable restorations for cleaning; benefits patient by restoring masticatory function.
Clinical Evidence
Bench testing only. Includes sterilization validation (ISO 17665), biocompatibility (ISO 10993), mechanical dynamic loading (ISO 14801), reverse engineering for OEM compatibility, retention strength testing compared to K033699, and non-clinical MRI safety review.
Technological Characteristics
Materials: Ti 6Al-4V ELI (copings, denture housings), Nylon (retention inserts). Sensing/Actuation: Mechanical friction-based engagement. Dimensions: 4.6mm platform diameter. Connectivity: CAD/CAM digital design workflow. Sterilization: Non-sterile, end-user steam sterilization. Software: CAD design software used for patient-specific copings.
Indications for Use
Indicated for patients requiring fixed/detachable partial or full arch restorations on endosseous dental implants in the mandible or maxilla to restore masticatory function. Compatible with specific Straumann Bone Level Multi-Unit Abutments.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
High Retention Attachment System, Zest Anchors, LLC (K213391)
Locator Bar Attachment System, Zest Anchors, LLC (K033699)
Straumann Variobase Abutment, IPS e.max CAD MO Coping, Straumann USA, LLC (K142890)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
12/29/22
BioClean Dental, LLC % Melissa Burbage Senior Regulatory Specialist PaxMed International, LLC 12264 EL Camino Real, Suite 400 San Diego, California 92130
Re: K222984
Trade/Device Name: BioClean Implant Restorative System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 30, 2022 Received: December 1, 2022
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222984
Device Name
BioClean Implant Restorative System
Indications for Use (Describe)
The BioClean Implant Restorative System is designed to support fixed/detachable partial or full arch restorations on endosseous dental implants in the maxilla for the purpose of restoring masticatory function. It is used in fixed/detachable restorations that can be attached with a friction-based engagement system.
All digitally designed copings for use with the BioClean Implant Restorative System are to BioClean Dental or a validated milling center for manufacture.
The BioClean Implant Restorative System is compatible with the following:
| BioClean Implant Restorative System Compatibility | |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Implant Components | Configurations |
| Bone Level Straumann® Multi-Unit<br>Abutments NC/RC<br>Straumann Multi-Unit Abutment Occlusal<br>Screw | 4.6mm Platform Diameter, All Gingival<br>Heights |
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
# K222984
## BioClean Implant Restorative System
## BioClean Dental, LLC
December 29, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | BioClean Dental, LLC<br>6088 Eagles Nest Drive<br>Jupiter, FL 33458<br>Telephone: +1 561-373-8238 | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Official Contact | Russ Bonafede, Chief Commercial Officer | | |
| Representative/Consultant | Melissa Burbage<br>Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: mburbage@paxmed.com<br>kthomas@paxmed.com; flarson@paxmed.com | | |
### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | BioClean Implant Restorative System |
|----------------------|-------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
### PREDICATE DEVICE INFORMATION
Primary Predicate Device K191123, Multi-unit Abutments, Medentika GmbH
Reference Devices K213391, High Retention Attachment System, Zest Anchors, LLC K033699, Locator Bar Attachment System, Zest Anchors, LLC K142890, Straumann Variobase Abutment, IPS e.max CAD MO Coping, Straumann USA, LLC
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
The BioClean Implant Restorative System is designed to support fixed/detachable partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed/detachable restorations that can be attached with a friction-based engagement system.
All digitally designed copings for use with the BioClean Implant Restorative System are to be sent to BioClean Dental or a validated milling center for manufacture.
The BioClean Implant Restorative System is compatible with the following:
| BioClean Implant Restorative System Compatibility | |
|--------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Implant Components | Configurations |
| Bone Level Straumann® Multi-Unit Abutments NC/RC<br>Straumann Multi-Unit Abutment Occlusal Screw | 4.6mm Platform Diameter, All Gingival Heights |
### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for the BioClean Implant Restorative System. The BioClean Implant Restorative System provides rigid connection of partial and full arch restorations to endosseous dental implants in the mandible or maxilla.
The BioClean Implant Restorative System consists of copings, denture housings, and retention inserts that are intended to allow for a fixed/detachable prosthesis to be connected to OEM multiunit abutments for stable attachment of the prosthesis to endosseous dental implants. All BioClean copings are made of titanium alloy and have the same coronal ridge retention design that attaches to the prosthesis component by a friction-based engagement system. The subject device coping connects to the OEM multi-unit abutment and is specific to each compatible abutment system and diameter.
The subject device copings are offered with angulation from 0° to 9° in 0.5° increments and also in a patient specific version in which the dental laboratory technician designs the coping in CAD software and then sends the design to BioClean Dental or a validated milling center for fabrication from a coping blank. By use of a combination of pre-manufactured and patientspecific versions of the copings, a multi-unit prosthesis can be created in which the copings are parallel.
{6}------------------------------------------------
## PERFORMANCE DATA
Non-clinical testing data submitted to demonstrate substantial equivalence included:
- Sterilization validation according to ISO 17665-1 Sterilization of health care products -● Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1
- Biocompatibility testing according to ISO 10993-5 Biological evaluation of medical ● devices - Part 5: Tests for in vitro cytotoxicity and ISO 10993-12 Biological evaluation of medical devices – Part 12: Sample preparation and reference materials
- . Mechanical testing according to ISO 14801 Dentistry - Implants - Dynamic loading test for endosseous dental implants
- . Reverse engineering to verify OEM compatibility to the OEM abutment-implant system. Key dimensions were measured from OEM implants, abutments and screws, from which dimensions and tolerances were established for corresponding BioClean copings, to ensure that the subject device components are compatible with the corresponding OEM system connections. When determining manufacturing tolerances, suitable adjustments in dimensions to avoid interference at maximum material conditions for the abutment and for the coping are taken into consideration.
- . Retention testing was conducted to ensure that retention is maintained throughout the expected use of the product. Retention strength of the subject devices was tested and compared to the retention strength of reference device K033699.
- Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI ● environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
### EQUIVALENCE TO MARKETED DEVICES
#### Indications for Use
The subject device and primary predicate K191123 include indications for partial or full arch restorations, referred to as bridges and bars in the predicate device indications. The slight difference in terminology does not affect the intended use; a bridge is considered a partial restoration and a bar is considered a full arch restoration. Both devices have indications for use with dental implants in the mandible or maxilla. Both devices are compatible with OEM systems.
Reference devices were identified to bridge noted technological or verbiage-based differences between the primary predicate and subject devices. The subject device indications are different from primary predicate K191123 in that the primary predicate device indications do not include "fixed/detachable" restorations. The primary predicate K191123 is only for fixed or "screw retained" restorations; this difference is addressed with reference device K033699. The subject
{7}------------------------------------------------
device indications include the fact that the restoration can be attached with a friction-based engagement system; this difference is addressed with reference device K213391. The subject device also allows for the copings to be digitally designed and manufactured at a validated milling center; this difference is addressed with reference device K142890.
The subject device and reference device K213391 include indications for partial or full arch restorations. Indications for both devices include use with dental implants in the mandible or maxilla for purpose of restoring masticatory function. Indications for both devices include attachment of restorations with a friction-based engagement system; this is referred to as a snapin system for reference device K213391. The slight difference in terminology does not affect the intended use because a snap-in system is another name for a friction-based engagement system. Both devices are compatible with OEM systems.
The subject device indications are different from those for reference deviceK213391 in that the reference device indications do not contain "fixed/detachable" restorations. The reference device K213391 is only for fixed restorations (meaning that it is not removable by the patient); this difference is addressed with reference device K033699. The subject device allows for the copings to be digitally designed and manufactured at a validated milling center; this difference is addressed with reference deviceK142890.
The subject device and reference device K033699 include indications for partial or full arch restorations, referred to as overdentures or partial dentures in the predicate device. The slight difference in terminology does not affect the intended use because a partial denture is considered a partial restoration and an overdenture is considered a full arch restoration. Both devices have indications for use with dental implants in the mandible or maxilla.
The subject device indications are different from reference device K033699 in that the reference device K033699 indications statement does not contain "fixed/detachable" restorations. However, even though it is not listed in the indications, the reference device K033699 device is intended to be removed by the patient. A fixed/detachable restoration is intended to be removed by the patient for cleaning, whereas a fixed restoration is intended to be removed by the clinician for cleaning. The subject device indications include the attachment of the restoration with a friction-based engagement system. However, even though it is not listed in the indications, the reference device K033699 device is attached with a friction-based system. The subject device also allows for the copings to be digitally designed and manufactured at a validated milling center; this difference is addressed with reference device K142890.
The subject device and reference device K142890 include indications for partial restorations, referred to as bridges in predicate device indications. The slight difference in terminology does not affect the intended use because a bridge is considered a partial restoration. Both devices allow for customized restorations using digitally designed copings. Digitally designed subject device copings are to be sent to BioClean Dental or a validated milling center for manufacture. This is similar to reference device K142890. The subject device states "digitally designed copings" whereas the reference device states "digitally designed copings and/or crowns." Both devices are attached to the abutment that is used within a customized prosthetic restoration, which does not affect the intended use.
{8}------------------------------------------------
The subject device indications are different from reference device K142890 in that the predicate device indications do not include "fixed/detachable" restorations. The reference device K142890 is only for screw-retained or cement-retained fixed restorations; this difference is addressed with reference device K033699. The subject device has limits on the type of restoration, as it does not include single unit. This difference is addressed in primary predicate K191123, reference device K142890, and reference device K213391.
#### Technological Characteristics
The subject device consists of copings similar to those of the primary predicate K191123 multiunit titanium base device. Both of these devices attach directly to multi-unit abutments that are angled up to 30°. A difference is that the subject device coping is offered in straight and angled versions up to 9°, whereas the primary predicate K191123 is only offered straight. The total angulation of the assembly does not extend past 30° for either the subject device or primary predicate K191123.
The subject device also consists of retention inserts and denture housings that connect the dental implant/abutment assembly to the restorations by means of a friction-based engagement system (also referred to as snap-in system), similar to the reference device K213391 and reference device K033699. The reference device K213391 is a high retention system that requires the restoration to be removed by the clinician for cleaning (known as a fixed restoration). However, the reference device K033699 has a lower retention strength that allows the patient to remove the restoration for cleaning (known as a fixed/detachable restoration).
The subject device retention insert and denture housing attach to a coping, whereas the reference device K213391 and reference device K033699 device are attached to the abutment. The subject device retention insert and denture housing do not allow for any divergence/convergence, unlike the reference device K213391 and reference device K033699, which allow divergence/ convergence up to 20° per implant or 40° between implants by action of the retention insert and denture housing. The reference device K213391 and reference device K033699 devices are placed on a straight abutment where there is a need for additional angulation correction to allow for the prosthetic restoration connections to be parallel. This additional angulation correction is not needed with the subject devices because they are placed on an angled abutment with an angled coping (with the total not exceeding 30°) in order to obtain a parallel prosthetic restoration connection.
{9}------------------------------------------------
## 510(k) Summary Page 6 of 8
| Comparison | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BioClean Implant Restorative<br>System<br>BioClean Dental, LLC | K191123<br>Multi-unit Abutments<br>Medentika GmbH | K213391<br>High Retention Attachment<br>System<br>Zest Anchors, LLC | K033699<br>Locator Bar Attachment System<br>Zest Anchors, LLC | K142890<br>Straumann Variobase Abutment<br>IPS e.max CAD MO Coping<br>Straumann USA, LLC |
| Indications | The BioClean Implant<br>Restorative System is designed<br>to support fixed/detachable<br>partial or full arch restorations<br>on endosseous dental implants<br>in the mandible or maxilla for<br>the purpose of restoring<br>masticatory function. It is used<br>in fixed/detachable restorations<br>that can be attached with a<br>friction-based engagement<br>system.<br>All digitally designed copings<br>for use with the BioClean<br>Implant Restorative System are<br>to be sent to BioClean Dental<br>or a validated milling center for<br>manufacture.<br>The BioClean Implant<br>Restorative System is<br>compatible with the following:<br>For complete Indications for<br>Use statement on OEM<br>Compatibility see Section 4. | Multi-unit abutments are<br>indicated for use with dental<br>implants as a support for multi-<br>unit screw retained bridges and<br>bars in the maxilla or mandible<br>of a partially or fully edentulous<br>patient.<br>Multi-unit Abutments are used<br>for the restoration of the<br>following dental implant<br>systems:<br>For complete Indications for<br>Use statement on OEM<br>Compatibility see 510(k)<br>Summary for K191123 in<br>Section 12. | The High Retention Attachment<br>System is designed to support<br>fixed, partial or full arch<br>restorations on endosseous<br>dental implants in the mandible<br>or maxilla for the purpose of<br>restoring masticatory function.<br>It is used in fixed hybrid<br>restorations that can be attached<br>with a snap-in system.<br>The High Retention Attachment<br>System is compatible with the<br>following implants:<br>For complete Indications for<br>Use statement on OEM<br>Compatibility see 510(k)<br>Summary for K213391 in<br>Section 12. | The Locator Bar Attachment<br>System is designed for use with<br>overdentures or partial dentures<br>retained in whole or in part by<br>bar splinted endosseous<br>implants in the mandible or<br>maxilla. | The Straumann® Variobase™<br>Abutment is a titanium base<br>placed onto Straumann dental<br>implants to provide support for<br>customized prosthetic<br>restorations. Straumann®<br>Variobase™ Abutments are<br>indicated for screw-retained<br>single tooth or cement-retained<br>single tooth and bridge<br>restorations.<br>All digitally designed copings<br>and/or crowns for use with the<br>Straumann® Variobase™<br>Abutment system are intended<br>to be sent to Straumann for<br>manufacture at a validated<br>milling center. |
| Comparison | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| | BioClean Implant Restorative<br>System<br>BioClean Dental, LLC | K191123<br>Multi-unit Abutments<br>Medentika GmbH | K213391<br>High Retention Attachment<br>System<br>Zest Anchors, LLC | K033699<br>Locator Bar Attachment System<br>Zest Anchors, LLC | K142890<br>Straumann Variobase Abutment<br>IPS e.max CAD MO Coping<br>Straumann USA, LLC |
| | | Multi-unit titanium base with<br>multi-unit abutment | | | Variobase for Bridge/Bar |
| Reason for<br>Predicate | n/a | Coping | Friction-based system | Patient removable, retention<br>performance | Patient-specific (CAD/CAM)<br>manufactured at validated<br>milling center |
| System<br>components | Coping, retention insert, and<br>denture housing<br>Note: multi-unit abutment and<br>screw are provided by OEM | Multi-unit abutment, multi-unit<br>titanium base (coping), and<br>screw | Abutment, retention insert, and<br>denture housing | Bar abutment, retention insert,<br>and denture housing | Abutment, coping, crown, and<br>screw |
| Restoration | Multi-unit | Multi-unit | Multi-unit | Multi-unit | Single and multi-unit |
| Abutment<br>angulation | 0°, 17°, 30° | 0°, 17°, 30° | 0° | 0° | 0° |
| Coping<br>angulation | 0 - 9° | 0° | n/a | n/a | 30° |
| Divergence<br>allowance<br>(attachment<br>mechanism) | 0° | n/a | 20° / 40° | 20° / 40° | n/a |
| Comparison | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| | BioClean Implant Restorative<br>System<br>BioClean Dental, LLC | K191123<br>Multi-unit Abutments<br>Medentika GmbH | K213391<br>High Retention Attachment<br>System<br>Zest Anchors, LLC | K033699<br>Locator Bar Attachment System<br>Zest Anchors, LLC | K142890<br>Straumann Variobase Abutment<br>IPS e.max CAD MO Coping<br>Straumann USA, LLC |
| Assembly<br>angulation | Up to 30° | Up to 30° | Up to 20° | Up to 20° | Up to 30° |
| Abutment<br>Platform<br>Diameter | 4.6 mm | 3.3 - 6.5 mm<br>Multiple systems | 2.3 - 7.0 mm<br>Multiple systems | 2.3 – 7.0 mm<br>Multiple systems | 3.5 - 4.6 mm |
| Prosthetic<br>attachment<br>mechanism | Nylon Insert retained in<br>Denture Housing | Screw retrained | PEEK Insert retained in<br>Denture Attachment Housing | Nylon Insert retained in<br>Denture Attachment Housing | Screw or cement |
| Material | | | | | |
| Abutment | n/a | Ti 6Al-4V ELI | Ti 6Al-4V ELI with titanium<br>nitride coating | Ti 6Al-4V ELI or Stainless<br>steel | Ti 6Al-7Nb |
| Coping | Ti 6Al-4V ELI | Ti 6Al-4V ELI | n/a | n/a | IPS e.max® CAD Ceramic,<br>coron® |
| Retention<br>inserts | Nylon | n/a | PEEK | Nylon | n/a |
| Denture<br>housing | Ti 6Al-4V ELI | n/a | Ti 6Al-4V ELI | Ti 6Al-4V ELI | n/a |
| How provided |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.