K222970 · Dia Imaging Analysis, Ltd. · QIH · Feb 1, 2023 · Radiology
Device Facts
Record ID
K222970
Device Name
LVivo IQS
Applicant
Dia Imaging Analysis, Ltd.
Product Code
QIH · Radiology
Decision Date
Feb 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Left Ventricle Image Quality Score
—
Overall agreement > 75%
81%
—
—
Validation study: 100 patient examinations (22,663 frames) referred to routine Echo examination.
>1 (experienced sonographers)
Left Ventricle Image Quality Score
—
80% of saved exams with visual quality 3-5 receive at least 'Medium' quality score; 90% of these cases are clinically interpretable.
90% received at least 'Medium' quality score; 93% were clinically interpretable.
—
—
POCUS environment study: 64 patients with POCUS indication and visual image quality 3-5.
>1 (expert echocardiologists)
Indications for Use
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.
Device Story
LVivo IQS is a software module extension to the LVivo Software Application. It processes ultrasound images of the heart (4-chamber apical view) to provide real-time Quality Score feedback. The device is intended for integration into legally marketed ultrasound imaging systems. It acts as an accessory to assist clinicians or point-of-care (POC) users during scanning by evaluating image quality before clips are saved. By providing immediate feedback on whether an image is of sufficient quality for clinical interpretation, the device helps ensure that diagnostic-quality data is captured, potentially reducing the need for repeat scans and improving clinical decision-making efficiency.
Clinical Evidence
Performance validated via two studies. Study 1: 100 patients (22,663 frames) compared LVivo IQS against expert sonographer quality tagging; achieved 81% overall agreement (success criteria 75%). Study 2: 64 patients in a POCUS environment; 90% of images rated 3-5 by visual estimation received at least 'Medium' quality score from LVivo IQS (success criteria 80%); 93% of these clips were clinically interpretable by expert echocardiologists (success criteria 90%).
Technological Characteristics
Software-based medical device accessory. Operates on Windows/Linux platforms. Provides real-time image quality assessment for 4-chamber apical view ultrasound clips. Integrates via API into third-party ultrasound imaging systems. Uses automated image processing algorithms to calculate quality scores.
Indications for Use
Indicated for patients aged 18+ with suspected disease undergoing ultrasound examination. Used for non-invasive processing of ultrasound images to detect, measure, and calculate cardiac structures and function, and to provide real-time image quality feedback.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
DiA Imaging Analysis Ltd. % George Hattub Senior Project Manager Medicsense USA LLC 291 Hillside Avenue SOMERSET MA 02726
February 1, 2023
Re: K222970
Trade/Device Name: LVivo IQS Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: December 20, 2022 Received: December 21, 2022
Dear George Hattub:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222970
Device Name LVivo IOS
Indications for Use (Describe)
L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.
Type of Use (Select one or both, as applicable)
| <span style="font-size:120%;">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:120%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
K222970
Pursuant to CFR 807.92, the following 510(k) Summary is provided:
| 1. (a) | Submitter<br>Address: | George J. Hattub<br>Medicsense USA LLC<br>291 Hillside Avenue<br>Somerset, MA 02726<br>ghattub@comcast.net |
|--------|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. (b) | Manufacturer<br>Address: | DiA Imaging Analysis Ltd<br>HaEnergia Street 77<br>Beer-Sheva, Israel 8470912 |
| | Mfg. Phone: | Tel.: +972 77 7648318 |
| | Contact Person: | Mrs. Michal Yaacobi |
| | Date: | December 20, 2022 |
| 2. | Device &<br>Classification<br>Name: | Medical Image Management and Processing System –<br>classified as Class 2 QIH, Regulation Number 21 CFR 892.2050<br>LVivo IQS |
| 3. | Predicate Devices: | LVivo Software Application K210053 |
| 4. | Description: | The LVivio IQS is an extension to the LVivio Software Application<br>(K210053), as an additional Algorithm with API that will be able to provide a<br>Quality Score in real time to the Left Ventricle from the 4 chamber apical<br>view of the heart. In addition, the LVivo IQS will be provided as a software<br>component to be integrated by another computer programmer into their<br>legally marketed ultrasound imaging device. Essentially, the Algorithm and<br>API, which is a module, will be a medical device accessory. The QIS Meter<br>concept is similar to the legally marketed Caption Guidance Software<br>Device. |
| 5. | Indications for<br>Use: | LVivo platform is intended for non-invasive processing of ultrasound<br>images to detect, measure, and calculate relevant medical parameters of<br>structures and function of patients with suspected disease. In addition, it has<br>the ability to provide Quality Score feedback |
| 6. | Comparison of<br>Technological<br>Characteristics: | With respect to technology and intended use, DiA's LVivo IQS is<br>substantially equivalent to its predicate devices. Based upon the outcomes<br>from the risk analysis and Performance Testing Evaluation, DiA believes<br>that the extension of LVivo IQS module to the LVivo Software Application<br>predicate device does not raise additional safety of efficacy concerns. The<br>following comparison table depicts the changes. |
{4}------------------------------------------------
| | Submitted Device | Predicate Device<br>K210053 |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features/ Characteristics | LVivo IQS as an addition to<br>the LVivo Software<br>Application | LVivo Software<br>Application |
| Product Code | same | QIH |
| Indication for Use | LVivo platform is<br>intended for non-<br>invasive processing<br>of ultrasound<br>images to detect,<br>measure, and<br>calculate relevant<br>medical parameters of<br>structures and<br>function of patients<br>with suspected<br>disease. In addition, it has the<br>ability to provide Quality Score<br>feedback | LVivo platform is intended<br>for non-invasive<br>processing of ultrasound<br>images to detect,<br>measure, and<br>calculate relevant medical<br>parameters of structures<br>and function of patients<br>with suspected disease. |
| Modules | LVivo IQS, LVivo EF, LVivo<br>SG, LVivo SAX, LVivo RV &<br>LVivo Bladder | LVivo EF, LVivo SG, LVivo<br>SAX, LVivo RV & LVivo<br>Bladder |
| Automation | same | yes |
| Manual Adjustment | same | yes |
| Bi plane EF evaluation | same | yes |
| Simultaneous 2CH and<br>4CH evaluation | same | yes |
| Off-line LV RV and<br>Bladder evaluation using<br>DICOM clips of any<br>vendor | same | yes |
| Automated ED and ES<br>frames selection | same | yes |
| Manual editing by user<br>capability | same | yes |
| Visually confirm<br>results | same | yes |
| Automated rejection<br>of false results | same | yes |
| Volume calculation<br>By standard<br>Simpson's<br>method of discs for EF | same | yes |
| Volume\Area curve<br>Presentation | same | yes |
| EF, Strain, SWM, RV,<br>SAX, Bladder results<br>presentation | same | yes |
| Enables presentation<br>cardiac function results<br>for different cycle | same | yes |
| Algorithm | same | yes |
| Calculation speed | same | yes |
| Capability or a part of a<br>bigger package (device)<br>for LV function<br>evaluation and Bladder | same | yes |
| Segmental Longitudinal<br>Strain Measure | same | yes |
| Global Longitudinal<br>Strain Measure | same | yes |
| Segmental wall motion<br>evaluation | same | yes |
| Operating System | same | Windows/Linux<br>(with Android option<br>for LVivo EF) |
| Quality score feedback | Yes | No |
| 510(k) # | Pending | K219953 |
510k Notification: LVivo IQS
{5}------------------------------------------------
{6}------------------------------------------------
- 7. Performance A summary of the Performance Evaluation, which was based upon well-Evaluation: established test methods, demonstrated conformity to the intended use:
1. The performance of the LVivo IQS system was validated using already data acquired with different ultrasound devices and various cardiac pathologies, compared to quality tagging by experienced sonographer.
Success criteria:
Overall agreement of 75% between the LVivo IQS results and the data tagging by experienced sonographers
100 patient examinations were used for the validation. Inclusion criteria: Age>18, patients referred to routine Echo examination. Exclusion criteria: Subjects who fail to meet any inclusion criteria. Total of 22,663 frames were analvzed.
The overall agreement was agreement between the LVivo IQS and quality tagging by the experienced sonographers was 81%
2. The performance of the LVivo IQS system was validated using data acquired after using the LVivo IQS in real time while scanning the LV from the 4CH apical view. The scans were done by POC interns in POC environment. The obtained quality score before saving a clip was recorded.
Success criteria: a. 80% of the saved Exams with image quality 3-5 by visual estimation, received at least "Medium" image quality by LVivo IQS. b. 90% of these cases are clinically interpretable by expert echocardiologist
64 patients were included in the study and the saved scans were analyzed. Inclusion criteria: Age ≥18. indication for POCUS, image quality 3-5 according to ACEP Guidelines, Exclusion criteria: subjects who fail to meet any inclusion criteria, image quality 1-2 according to ACEP Guidelines. The results are summarized in Table-1:
| End point number | Results |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| a. | In 90% of the patients with image quality 3-5<br>by visual estimation it was possible to obtain<br>at least “Medium” quality score by LVivo IQS |
| b. | 93% of the above saved clips were clinically<br>interpretable |
Table-1: Results summary
{7}------------------------------------------------
- 8. Conclusion: The Intended Use and the technological characteristics in the current device are the same as those in the LVivo software application predicate device, including the addition of the IQS Module, do not affect the safety and effectiveness of the device. The performance tests have been completed and successfully support the device performance. Therefore, DiA Imaging Analysis concludes the LVivo IQS is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.