The Dental Desensitizer is a colorless transparent gel that is applied to the sensitive part of the tooth to form a film, sealing the exposed dentin tubules and relieving dentin allergy. The product is used either by a dental professional in the dental office or provided to the patient for home treatment of dentin sensitivity.
Device Story
Dental Desensitizer is a viscous, colorless, transparent gel containing 3% potassium nitrate and 0.11% fluoride ion; functions as a cavity varnish. Applied to sensitive tooth areas via custom-fabricated dental tray to form a film; physically blocks exposed dentin tubules to eliminate sensitivity. Used in dental offices by professionals or at home by patients. Provides sustained release of active ingredients; intended to relieve dentin allergy. Non-sterile; shelf life of 24 months at room temperature.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including appearance, loading capacity, chemical content (potassium, fluoride, heavy metals, arsenic), pH, microbial limits, and dentin tubule occlusion comparative tests. Biocompatibility verified per ISO 7405, ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-11.
Technological Characteristics
Viscous gel containing 3% potassium nitrate and 0.11% w/w fluoride ion. pH 6-7.5. Non-sterile. Biocompatibility compliant with ISO 7405, ISO 10993-1, -5, -10, -11. Designed for use with custom-fabricated dental trays. No electronic or software components.
Indications for Use
Indicated for patients experiencing dentin sensitivity. Used by dental professionals or patients for home treatment to seal exposed dentin tubules and relieve dentin allergy.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
May 25, 2023
Guangzhou Beogene Biotech Co., Ltd % Tracy Che Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center No. 3101-90. Oianhai Road ShenZhen, GuangDong 518052 CHINA
Re: K222891
Trade/Device Name: Dental Desensitizer Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: March 29, 2023 Received: March 29, 2023
Dear Tracy Che:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22891
Device Name Dental Desensitizer
Indications for Use (Describe)
The Dental Desensitizer is a colorless transparent gel that is applied to the sensitive part of the tooth to form a film, sealing the exposed dentin tubules and relieving dentin allergy. The product is used either by a dental professional in the dental office or provided to the patient for home treatment of dentin sensitivity.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) Summary
This "510(k) Summary" of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.
# (1) Applicant information
| 510 (k) owner's name: | Guangzhou Beogene Biotech CO., Ltd |
|----------------------------|--------------------------------------------------------------------------------------------------------------|
| Address: | Second floor, Building C, Self-compiled (1) Building, No.2 Ruita<br>Road, Huangpu District, Guangzhou, China |
| Contact person: | Duan Qiangqiang |
| Phone number: | +86-16626708885 |
| Fax number: | +86-020-32029939 |
| Email: | 2885151672@qq.com |
| Date of summary prepared: | 2023-3-29 |
| Reason for the submission: | New device, there were no prior submissions for the device. |
# (2) Proprietary name of the device
| Trade name/Model: | Dental Desensitizer |
|--------------------|---------------------|
| Common name | Tooth Desensitizer |
| Regulation name: | Varnish, Cavity |
| Regulation number: | 21 CFR 872.3260 |
| Product code | LBH |
| Review panel: | Dental |
| Regulation class: | Class II |
## (3) Predicate device
| Sponsor | Ultradent Products, Inc. |
|-----------------------|---------------------------|
| Device Name and Model | UltraEZ Desensitizing Gel |
| 510(k) Number | K061438 |
| Product Code | LBH |
| Regulation Number | 21 CFR 872.3260 |
| Regulation Class | II |
# (4) Description/ Design of device
Dental Desensitizer is a sustained release, 3% potassium nitrate and 0.11% weight fluoride ion, viscous gel. This product is designed to be used together with a custom-fabricated tray.
{5}------------------------------------------------
## (5) Intended use/ Indications for use
The Dental Desensitizer is a colorless transparent gel that is applied to the sensitive part of the tooth to form a film, sealing the exposed dentin tubules and relieving dentin allergy. The product is used either by a dental professional in the dental office or provided to the patient for home treatment of dentin sensitivity.
### (6) Materials
| Component name | Body Contact Category<br>(ISO 10993-1) | Contact Duration<br>(ISO 10993-1) |
|---------------------|------------------------------------------------------|-----------------------------------|
| Dental Desensitizer | External communicating device:<br>tissue/bone/dentin | Permanent contact >30 d |
The Dental Desensitizer has passed biocompatibility tests. Details can be seen in "Biocompatibility Discussion".
## (7) Technological characteristics and substantial equivalence
| Item | Subject device | Predicate device | Remark |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Trade name | Dental Desensitizer | UltraEZ Desensitizing Gel | / |
| 510 (k) number | / | K061438 | / |
| Regulation<br>number | 21 CFR 872.3260 | 21 CFR 872.3260 | Same |
| Regulation name | Varnish, Cavity | Varnish, Cavity | Same |
| Product code | LBH | LBH | Same |
| Class | II | II | Same |
| Indications for<br>use/ Intended use | The Dental Desensitizer is a<br>colorless transparent gel that is<br>applied to the sensitive part of<br>the tooth to form a film, sealing<br>the exposed dentin tubules and<br>relieving dentin allergy. The<br>product is used either by a dental<br>professional in the dental office<br>or provided to the patient for<br>home treatment of dentin<br>sensitivity | The UltraEZ Desensitizing<br>Gel provides a film-like<br>varnish for sensitive teeth,<br>sealing dentinal tubules of<br>over exposed dentin and other<br>exposed areas where<br>post-operative or other dentin<br>sensitivity is a concern. The<br>product is used either by a<br>dental professional in the<br>dental office or provided to the<br>patient for home treatment of<br>dentin sensitivity | The<br>descripti<br>on is<br>different<br>compare<br>d to the<br>predicate<br>device, it<br>is<br>actually<br>the same. |
| Prescription or<br>OTC | Prescription Use | Prescription Use | Same |
| Scope of use | Hospital, dental office, or home<br>treatment | Dental office or home<br>treatment | Same |
{6}------------------------------------------------
| Design | Designed to be used together<br>with a custom-fabricated dental<br>tray. | Designed to be used in a<br>custom-fabricated dental tray. | Same |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------|
| Materials | 3% Potassium Nitrate;<br>0.11% w/w Fluoride Ion | 3% Potassium Nitrate;<br>0.11% w/w Fluoride Ion | Same |
| Appearance | Colorless, odourless,<br>transparent gel, no visible<br>impurity | Colourless, Odourless, Gel | Same |
| PH-value | 6-7.5 | 6-7.5 | Same |
| Recommended<br>contact time | Treatment 15-60 mins | Treatment 15-60 mins | Same |
| Device<br>description | Immediate physical blockage<br>of dentinal tubules to eliminate<br>painful sensitivity. | Immediate physical blockage<br>of dentinal tubules to eliminate<br>painful sensitivity. | Same |
| Biocompatibility | Passed the tests as per ISO<br>10993-5, ISO 10993-10 and<br>ISO 10993-11<br>(Cytotoxicity, sensitization,<br>irritation, acute systemic<br>toxicity) | Passed the tests as per ISO<br>10993-5 and ISO 10993-10<br>(Cytotoxicity, sensitization,<br>irritation) | Similar |
| Sterility | Non-sterile | Non-sterile | Same |
### Conclusion:
Based on the above analysis, the Dental Desensitizer is substantial equivalent to the predicate device.
### (8) Non-clinical studies and tests performed
Non-clinical testings have been conducted to verify that the Dental Desensitizer meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device.
#### Performance Test Report
A performance verification test was performed to analyze treatment effect of Dental Desensitizer during patient treatment, and to validate that the output meets design specifications and demonstrate that the treatment effect of Dental Desensitizer is expected compared to the predicate device. Properties evaluated include appearance, loading capacity, potassium content, fluoride content, total heavy metal content, arsenic content, PH value, microbial limit, and dentin tubule occlusion comparative test.
#### Shelf Life Test Report
Shelf life testing was conducted by evaluating the physical properties of the device to confirm a shelf life at room temperature of 24 months.
#### Biocompatibility
{7}------------------------------------------------
A biocompatibility discussion was conducted. The Dental Desensitizer has been tested and shown to be compliant with the following standards:
- > ISO 7405, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
- A ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- A ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For InVitro Cytotoxicity
- A ISO 10993-10, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- > ISO 10993-11, Biological Evaluation of Medical Devices - Part 11: Tests For Systemic Toxicity
## (9) Clinical studies and tests performed
Clinical studies and tests were not conducted.
## (10) Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the Dental Desensitizer is to be concluded substantial equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.