CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. It is not meant to be used in digital mammography.
Device Story
32-inch 8MP Color LCD monitor; displays medical images for diagnostic review by trained practitioners. Input: digital video signals (DVI-D, DisplayPort). Transformation: utilizes IPS LCD panel with LED backlight; integrated luminance sensor and control circuit maintain stable luminance; lookup table applies DICOM GSDF grayscale characteristics; luminance mura correction ensures screen uniformity. Output: high-resolution visual display of medical images. Used in clinical environments; operated by physicians or certified personnel. Benefits: provides high-fidelity, calibrated visualization of diagnostic images to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical or animal data. Performance verified against FDA guidance for diagnostic radiology displays, including spatial resolution (MTF), pixel defects (ISO 13406-2), artifacts (AAPM-TG18), temporal response, luminance stability, DICOM GSDF conformance, color tracking, and gray tracking.
Indicated for trained medical practitioners or certified personnel to display and view medical images for diagnostic purposes. Not indicated for digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
21.3 inch (54.0cm) Color LCD Monitor CL-R211 (K182539)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
JVCKENWOOD Corporation Hideki Tengeiji Senior Manager 3-12, Moriya-cho, Kanagawa-ku Yokohama, Kanagawa 221-0022 Japan
December 21, 2022
Re: K222864
Trade/Device Name: 8MP Color LCD Monitor CL-R813 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: December 14, 2022 Received: December 15, 2022
Dear Hideki Tengeiji:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Low
Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K222864
Device Name 8MP Color LCD Monitor CL-R813
Indications for Use (Describe)
CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel.
It is not meant to be used in digital mammography.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15" x="0" y="0"></rect> <path d="M2 2 L13 13 M2 13 L13 2" stroke="#000000" stroke-width="1"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial;"> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K222864
# 510(k) SUMMARY
## I. Submitter
| Submitter: | JVCKENWOOD Corporation |
|-----------------|----------------------------------------|
| | 3-12, Moriya-cho, Kanagawa-ku, |
| | Yokohama-shi, Kanagawa, 221-0022 Japan |
| Contact Person: | Hideki Tengeiji, Senior Manager |
| | Email: tengeiji.hideki@jvckenwood.com |
| | Tel: +81-45-450-2715 |
| | Fax: +81-45-450-1926 |
| Date Prepared: | September 22, 2022 |
#### II. Device
| Name of Device: | 8MP Color LCD Monitor CL-R813 |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Display, diagnostic radiology |
| Classification Name: | Medical Image Management and Processing System<br>(Part 892 Radiology Devices Sec. 892.2050) |
| Regulatory Class | II |
| Product Code | PGY |
#### III. Predicate Device
| Name of Device: | 21.3 inch (54.0cm) Color LCD Monitor CL-R211 (K182539) |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Display, diagnostic radiology |
| Classification Name: | Medical Image Management and Processing System<br>(Part 892 Radiology Devices Sec. 892.2050) |
| Regulatory Class | II |
| Product Code | PGY |
# JVCKENWOOD
{4}------------------------------------------------
## IV. Device Description
32-inch 8MP Color LCD Monitor CL-R813, 3840 x 2160 (landscape)
- Color LCD panel, which has wide view angle, is used for this product. It is designed for ■ medical image display.
- Luminance stabilization function composed with luminance sensor and luminance control circuit always observes the luminance and makes it stable.
- Images are faithfully displayed along grayscale characteristics (DICOM GSDF) based on the calibrated data stored to the lookup table of the monitor.
- Luminance mura correction functions will help achieve uniformity on the whole screen.
#### V. Indications for use
CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. It is not meant to be used in digital mammography.
#### VI. Cybersecurity
FDA quidance located at https://www.fda.gov/media/86174/download, is followed for cybersecurity concerns.
{5}------------------------------------------------
### VII. Comparison of Technological Characteristics with the predicate device
The comparison table below list information obtained from the product catalogs and measurements and different technological characteristics are discussed in it.
| | Predicate device<br>LCD Monitor CL-R211 | Proposed device<br>LCD Monitor CL-R813 | Explanation of<br>Differences |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K182539 | - | - |
| Indication for use | CL-R211 is intended to be<br>used in displaying and viewing<br>medical images for diagnosis<br>by trained medical<br>practitioners or certified<br>personnel. It is not meant to<br>be used in digital<br>mammography. | CL-R813 is intended to be<br>used in displaying and<br>viewing medical images for<br>diagnosis by trained medical<br>practitioners or certified<br>personnel. It is not meant to<br>be used in digital<br>mammography. | - |
| Display Technology | IPS LCD panel with TFT<br>active-matrix array with LED<br>backlight | IPS LCD panel with TFT<br>active-matrix array with LED<br>backlight | - |
| Screen size | Diagonal: 21.3"<br>Aspect ratio: 4:3 | Diagonal: 32"<br>Aspect ratio: 16:9 | The display area of the<br>proposed device is larger<br>than that of the predicate<br>device. |
| Backlight type | LED | LED | - |
| Frame rate and refresh rate | 60Hz | 60Hz | - |
| Resolution / Pixel array | 2MP (1200 x 1600) | 8MP (3840 × 2160) | The resolution of the<br>proposed device is higher<br>than that of the predicate<br>device. |
| Pixel Pitch | Horizontal: 0.270mm<br>Vertical: 0.270mm | Horizontal: 0.1845mm<br>Vertical: 0.1845mm | Pixel pitch of the<br>proposed device are<br>smaller than those of the<br>predicate device. |
| Subpixel pattern | Stripe RGB | Stripe RGB | - |
| Pixel aperture ratio | 57 % | 57 % | - |
| Subpixel driving (spatial and<br>temporal dithering) | N/A | N/A | - |
| Display Interface | Input: DVI-D x1<br>DisplayPort x1<br>Output: DisplayPort x1 | Input: DVI-D x1<br>DisplayPort x2<br>Output: DisplayPort x1 | - |
| Video bandwidth | Dot clock: 162 MHz | Dot clock: 533.25 MHz | So the resolution of the<br>proposed device is higher<br>than that of the predicate<br>device, the dot clock of<br>the proposed device is<br>higher than that of the<br>predicate device |
| User control | DICOM CONFORMANCE TEST<br>Configuration switch<br>Input signal switch<br>Dynamic gamma<br>AUTO TEXT<br>Human presence sensor<br>EDID switch<br>Test pattern<br>USB Power | Config Select<br>Backlight<br>Contrast<br>Source Select<br>Human Sensor | Although the user<br>controls are different, the<br>performance of the<br>proposed device is the<br>same as the predicate<br>device. |
| Ambient light sensing | Built-in Sensor (For correction<br>during calibration) | Built-in Sensor (For<br>correction during calibration) | - |
| Touch-screen technology | N/A | N/A | - |
| Luminance calibration tools /<br>Quality-control procedures | Hardware:<br>Integrated sensor<br>External sensor<br>Software:<br>QA Medivisor/Medivisor NX<br>FCAL | Hardware:<br>Integrated sensor<br>External sensor<br>Software:<br>QA Medivisor Agent<br>FCAL | - |
| Additional Software/Firmware | N/A | N/A | - |
The above differences in technical characteristics do not affect the safety and the effectiveness of the CL-R813.
#### JVCKE OD
{6}------------------------------------------------
#### VIII. Performance Test
The following performance tests were performed on the CL-R813 in accordance with the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on October 2, 2017.
| Performance test items<br>in the guidance | Test method(s) |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| a. Spatial resolution | The bar pattern is displayed and captured by a digital<br>camera equipped with a macro lens. The MTF is calculated<br>with the captured data. |
| b. Pixel defects (maximum<br>counts, allowed defect<br>types, and locations) | ISO 13406-2<br>IDMS 1.03, 7.6 DEFECTIVE PIXELS<br>Pixel defects are counted based on the ISO13406-2, 3.4.13<br>table 3. |
| c. Artifacts | AAPM-TG18, 4.9 Miscellaneous Tests<br>IDMS 1.03, 4.6 ARTIFACTS & IRREGULARITIES |
| d. Temporal response | IDMS 1.03, 10.2.3 GRAY-TO-GRAY RESPONSE TIME<br>Rise and fall time constants at four grayscale intervals (0-<br>100%, 5-95%, 10-90%, 40-60%) are provided by the<br>panel manufacturer. |
| e. Luminance (maximum,<br>minimum, achievable, and<br>recommended | Lmin and Lmax on the calibrated luminance are confirmed. |
| f. Conformance to a grayscale<br>to luminance function (e.g.,<br>DICOM GSDF) | AAPM-TG18, 4.3.5 Advanced Luminance Response<br>Luminance response for 256 levels are measured. |
| (For color displays)<br>m. Color tracking (primary<br>colors and color gamut) | Color scale:<br>IDMS 1.03, 6. Gray- and Color-Scale Measurement<br>IDMS 1.03, 5.4 Color-Signal White<br>Color gamut volume:<br>IDMS 1.03, 5.31 Volume-Color-Reproduction Capability |
| (For color displays)<br>n. Gray tracking (gray shades<br>and white point) | AAPM-TG196 Gray Tracking<br>IEC 62563-1: 2009+AMD1:2016+AMD2:2021, 7.4.9<br>Greyscale chromaticity evaluation |
As a result of the performance test, it was confirmed that the CL-R813 has the equivalent display characteristics as the predicate device, CL-R211. The display characteristics of the CL-R813 meet the predefined criteria.
The CL-R813 has not been tested on animal or clinical testing.
#### IX. Conclusions
As shown above, the intended use of the subject device and the predicate device is identical, the technical characteristics are similar, and any differences in the characteristics do not affect the safety or effectiveness of the device. The results of the performance testing and the verification and validation demonstrate that the subject device is substantially equivalent to the predicate device currently on the market.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.