SILKPRO Titanium Diode Laser System SILKPRO-S20S-TWC
K222862 · Lotuxs Medtech (Suzhou) Co., Ltd. · GEX · Jul 30, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222862
Device Name
SILKPRO Titanium Diode Laser System SILKPRO-S20S-TWC
Applicant
Lotuxs Medtech (Suzhou) Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The indications for use for the SILKPRO-S20S-TWC include: The Super Hair Removal (SHR) Mode are intended for temporary hair reduction. The indications for use for the SILKPRO-S20S-755A include: The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick L-VI), including tanned skin.(HR. SHR Modes) The indications for use for the SILKPRO-S20S-810B include: The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term. stable reduction in the number of airs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and SHR Modes)
Device Story
Diode laser system for hair removal; utilizes selective photothermal effect; long pulse width laser penetrates epidermis to hair follicle; melanin absorbs laser energy; follicle destruction prevents hair regeneration. Features sapphire dynamic cooling device; water-cooled handpiece dissipates heat from light source and TEC block. Operated by clinician in dermatology/clinical setting. User interface via LCD color touchscreen. Output: laser energy delivered to skin. Benefits: permanent hair reduction or temporary hair reduction depending on mode.
Clinical Evidence
Bench testing only. Includes software verification and validation, power output testing, electromagnetic compatibility (IEC 60601-1-2), electrical safety (ANSI AAMI ES60601-1, IEC 60601-1, IEC 60601-2-22, IEC 60825-1), and biocompatibility assessment (ISO 10993-1).
Technological Characteristics
GaAlAs diode laser array; wavelengths 755nm, 810nm, 1064nm; sapphire contact cooling; TEC cooling system; Class IV laser; LCD touchscreen interface; non-sterile, reusable handpiece; electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1; biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for temporary hair reduction (SHR mode) or permanent reduction in hair regrowth (HR/SHR modes) in patients of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
July 30, 2023
Lotuxs Medtech (Suzhou) Co., Ltd. Na Wu Quality Manager Room301,303,304, Building6, Northwest Area of Nanopolis Suzhou,No.99 Jinjihu Avenue Suzhou Industrial Park, Suzhou, Jiangsu 215123 China
Re: K222862
Trade/Device Name: SILKPRO Titanium Diode Laser System SILKPRO-S20S-TWC Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 30, 2023 Received: June 30, 2023
Dear Na Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222862
#### Device Name
SILKPRO Titanium Diode Laser System
Indications for Use (Describe)
The indications for use for the SILKPRO-S20S-TWC include:
The Super Hair Removal (SHR) Mode are intended for temporary hair reduction.
The indications for use for the SILKPRO-S20S-755A include:
The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.
Use on all skin types (Fitzpatrick L-VI), including tanned skin.(HR. SHR Modes)
The indications for use for the SILKPRO-S20S-810B include:
The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term. stable reduction in the number of airs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.
Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and SHR Modes)
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## Section 5 - 510(k) Summary
## K22862 Date of Summary Preparation: September 15, 2022 Date of Modification: July 29, 2023
#### 1. Submitter's Identifications
Submitter's Name: Lotuxs Medtech (Suzhou) Co., Ltd. Address: Room301,303,304, Building6, Northwest Area of Nanopolis Suzhou, No.99 Jinjihu Avenue, Suzhou Industrial Park, Suzhou 215123, China Contact Person: Na Wu Contact Title: Quality Manager Contact Email Address: na.wu@lotuxs.com Telephone: +86-0512-62880553
### 2. Correspondent's Identifications
Correspondent's Name: Lotuxs Medtech (Suzhou) Co., Ltd. Address: Room301,303,304, Building6, Northwest Area of Nanopolis Suzhou, No.99 Jinjihu Avenue, Suzhou Industrial Park, Suzhou 215123, China ZIP Code: 215123 Contact Person: Na Wu Contact Title: Quality Manager Contact E-mail Address: na.wu@lotuxs.com Telephone: +86-0512-62880553
#### 3. Name of the Device
Device Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Product Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Trade Name: SILKPRO Titanium Diode Laser System Model: SILKPRO-S20S-TWC, SILKPRO-S20S-755A, SILKPRO-S20S-810B. Classification Panel: General & Plastic Surgery Product Code: GEX Regulation Number: 21 CFR 878.4810 Device Classification: Class II
#### 4. The Predicate Devices
Primary Predicate device: K222064 Soprano Titanium
#### 5. Device Description
The working principle of SILKPRO Titanium Diode Laser System is selective
{4}------------------------------------------------
photothermal effect. It uses its unique long pulse width laser to penetrate the epidermis to the hair follicle. The energy of the laser is preferentially absorbed by the melanin in the hair, so that the hair follicle produces high heat. Thereby destroying the structure of the hair follicle, making the hair lose the ability to regenerate, and achieving the purpose of hair removal. The treatment head adopts a uniquely designed sapphire dynamic cooling device.
SILKPRO Titanium Diode Laser System is a sophisticated high-tech laser equipment. The machine includes water cooling system. Water flows through the pipes in the handpiece to dissipate heat for the light source and the TEC water cooling block , which is used to cool the skin tissue during treatment.
#### 6. Intended Use of Device
The indications for use for the SILKPRO-S20S-TWC include:
The Super Hair Removal (SHR) Mode are intended for temporary hair reduction.
The indications for use for the SILKPRO-S20S-755A include:
The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.
Use on all skin types (Fitzpatrick L-VI), including tanned skin.(HR. SHR Modes)
The indications for use for the SILKPRO-S20S-810B include:
The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term. stable reduction in the number of airs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.
Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and SHR Modes)
{5}------------------------------------------------
# 7. Summary of Substantial Equivalence
# Table 1
| | Proposed device | Primary predicate device | Comparison |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510k Number | K222862 | K222064 | -------- |
| Product Code | GEX | GEX, ILY | Same |
| Proprietary Name | SILKPRO Titanium Diode Laser System | Soprano Titanium | -------- |
| Model | SILKPRO-S20S-TWC,<br>SILKPRO-S20S-755A,<br>SILKPRO-S20S-810B, | / | -------- |
| Manufacturer | Lotuxs Medtech (Suzhou) Co., Ltd. | Alma Lasers Inc. | -------- |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Regulatory Class | Class II | Class II | Same |
| Controls | Footswitch or handpiece | Footswitch or handpiece | Same |
| Indications for use | The indications for use for the<br>SILKPRO-S20S-TWC include:<br>The super hair removal (SHR) Mode are<br>intended for temporary hair reduction.<br>The indications for use for the<br>SILKPRO-S20S-755A include:<br>The hair removal (HR) and super hair<br>removal (SHR) Mode are intended for<br>permanent reduction in hair regrowth | The Soprano Titanium diode laser module is<br>intended for use in dermatology procedures<br>requiring coagulation. The indications for<br>use for the Soprano Trio diode laser module<br>include:<br>The Super Hair Removal ( SHR ) Mode is<br>intended for temporary hair reduction.<br>The Soprano Trio diode laser module HR<br>mode is intended for use in dermatology | Same |
{6}------------------------------------------------
| defined as a long term, stable reduction in<br>the number of hairs re-growing when<br>measured at 6,9 and 12 months after the<br>completion of a treatment regimen.<br>Use on all skin types (Fitzpatrick L-VI),<br>including tanned skin.(HR. SHR Modes)<br>The indications for use for the<br>SILKPRO-S20S-810B include:<br>The hair removal (HR) and super hair<br>removal (SHR) Mode are intended for<br>permanent reduction in hair regrowth<br>defined as a long term. stable reduction in<br>the number of hairs re-growing when<br>measured at 6,9 and 12 months after the<br>completion of a treatment regimen.<br>Use on all skin types (Fitzpatrick I-VI),<br>including tanned skin.(HR, and SHR<br>Modes) | procedures requiring coagulation indications<br>for use for the Soprano Trio diode laser HR<br>module include : Benign vascular and<br>vascular dependent lesions.<br>810nm Applicator<br>Soprano Titanium 810 nm applicator<br>intended use and indications for use:<br>The indications for use for the 810nm<br>Modified Diode Laser Module 2 cm 2<br>include:<br>The Hair Removal ( HR ) and Super Hair<br>Removal ( SHR ) Mode are intended for<br>permanent reduction in hair regrowth<br>defined as a long term , stable reduction in<br>the number of hairs re-growing when<br>measured at 6 , 9 and 12 months after the<br>completion of a treatment regimen<br>Use on all skin types ( Fitzpatrick I-VI ) ,<br>including tanned skin . HR , and SHR<br>Modes.<br>755nm Applicator<br>Soprano Titanium 755 nm applicator<br>intended use and indications for use:<br>The indications for use for the 755nm Diode<br>Laser Module include:<br>The Hair Removal ( HR ) and Super Hair |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{7}------------------------------------------------
| | Removal ( SHR ) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen Use on all skin types (Fitzpatrick I-VI), including tanned skin . HR, and SHR Modes. NIR Applicator: NIR Applicator intended use and indications for use: The Alma Lasers NIR Modules intended use is to emit energy in the near infrared ( NIR ) spectrum to provide topical heating. The indications for use for NIR Modules are: Elevating the tissue temperature for the temporary relief of minor muscle pain and joint pain and stiffness . The temporary relief of minor joint pain associated with arthritis The temporary increase in local circulation where applied , and The relaxation of muscles may also help muscle spasms , minor sprains and strains , and minor |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{8}------------------------------------------------
| | | muscular back pain. | |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating theory | The working principle of SILKPRO Titanium Diode Laser System is selective photothermal effect. It uses its unique long pulse width laser to penetrate the epidermis to the hair follicle. The energy of the laser is preferentially absorbed by the melanin in the hair, so that the hair follicle produces high heat. Thereby destroying the structure of the hair follicle, making the hair lose the ability to regenerate, and achieving the purpose of hair removal. The treatment head adopts a uniquely designed sapphire dynamic cooling device. | The Soprano trio Diode Laser Module is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI); The working principle of Soprano trio Diode Laser Module is selective photothermal effect. It uses its unique long pulse width laser to penetrate the epidermis to the hair follicle. The energy of the laser is preferentially absorbed by the melanin in the hair, so that the hair follicle produces high heat. Thereby destroying the structure of the hair follicle, making the hair lose the ability to regenerate, and achieving the purpose of hair removal. The treatment head adopts a uniquely designed sapphire dynamic cooling device. | Same |
| Laser Type | GaAlAs Diode Laser array | GaAlAs Diode Laser array | Same |
| Handpiece<br>tipmaterial | Sapphire | Sapphire | Same |
| Wavelength | SILKPRO-S20S-TWC:<br>755nm/810nm/1064nm Three in one<br>combination wavelength | 755nm/810nm/1064nm Three in one<br>combination wavelength | Same |
| | SILKPRO-S20S-755A:755 nm | 755 nm | Same |
| | SILKPRO-S20S-810B:810 nm | 810 nm | Same |
| | SILKPRO-S20S-TWC: 10×40 mm² | Soprano Trio: 4cm² | Same |
| Spot size | SILKPRO-S20S-755A: 12×12 mm² | 755nm Applicator: 1.5 cm² | Similar<br>The spot size of<br>the proposed<br>device is<br>similar to the<br>predicate<br>device, so this<br>definition does<br>not affect the<br>safety and<br>effectiveness. |
| | SILKPRO-S20S-810B: 10×20mm² | 810nm Applicator :2cm² | Same |
| | SILKPRO-S20S-TWC: Up to 200 ms | Soprano Trio: Up to 200 ms | Same |
| Pulse Width (ms) | SILKPRO-S20S-755A : 3.3-200 ms | 755nm Applicator: 3.3-200 ms | Same |
| | SILKPRO-S20S-810B: 3.3-200 ms | 810nm Applicator : 3.3-200 ms | Same |
| Pulse Repetition<br>Rate | SILKPRO-S20S-TWC:<br>SHR: Up to 10 Hz | Soprano Trio:<br>HR: Up to10 Hz<br>SHR: Up to10 Hz | Same |
| | SILKPRO-S20S-755A: | 755nm Applicator: | Same |
| | | | |
| | HR: 0.5-3Hz<br>SHR: 5-10Hz | HR: 0.5-3 Hz<br>SHR: 5-10 Hz | |
| | SILKPRO-S20S-810B:<br>HR: 0.5-3Hz<br>SHR: 5-10Hz | 810nm Applicator:<br>HR: 0.5-3 Hz<br>SHR: 5-10 Hz | Same |
| Energy Density<br>(Fluence) | SILKPRO-S20S-TWC<br>SHR: 2-8 J/cm² | Soprano Trio:<br>HR:2- 120 J/ cm²<br>SHR: 2-20 J/ cm² | Same |
| | SILKPRO-S20S-755A<br>HR: Up to 120J/ cm²<br>SHR: Up to 20J/ cm² | 755nm Applicator:<br>HR: Up to 120J/ cm²<br>SHR: Up to 20J/ cm² | Same |
| | SILKPRO-S20S-810B<br>HR: 2-120J/ cm²<br>SHR: 2-20J/ cm² | 810nm Applicator:<br>HR:2-120J/ cm²<br>SHR:2-20J/ cm² | Same |
| User Interface | LCD Color Touchscreen | LCD Color Touchscreen | Same |
| Laser classification | Class IV | Class IV | Same |
| Cooling system | TEC | TEC | Same |
| Software | Yes | Yes | Same |
| Delivery Devices<br>(How supplied) | Non -Sterile. reusable. cleanable | Non -Sterile. reusable. cleanable | Same |
| Electromagnetic<br>compatibility and<br>electrical safety<br>compliance | IEC 60601-1 | IEC 60601-1…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.