CRYOcheck Factor VIII Deficient Plasma with VWF is for clinical laboratory use as a deficient substrate in the quantitative determination of Factor VIII activity in 3.2% citrated human plasma based on the activated partial thromboplastin time (APTT) assay. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use.
Device Story
CRYOcheck Factor VIII Deficient Plasma with VWF is an in vitro diagnostic reagent used in clinical laboratories. It consists of normal human citrated plasma immunodepleted of Factor VIII, supplemented with exogenous human von Willebrand Factor (vWF) and buffered with HEPES. The device serves as a deficient substrate in modified APTT assays performed on automated coagulation analyzers (e.g., IL ACL TOP series). In operation, patient plasma is diluted and mixed with the deficient plasma; the degree of clotting time correction is proportional to the Factor VIII activity in the patient sample, interpolated from a calibration curve. The output is a quantitative percentage of Factor VIII activity. Results assist clinicians in diagnosing Factor VIII deficiency and managing hemophilia A. The device is provided frozen in small-volume aliquots and must be stored frozen until use.
Clinical Evidence
Performance validated via bench testing and method comparison. Precision/reproducibility studies (N=270) demonstrated <10% CV for controls and high-activity samples. Linearity established from 0 to 230% FVIII activity. Method comparison (N=366) against predicate showed high correlation (r=0.96) and equivalent performance via Passing-Bablok regression. Interference studies identified specific drug interferences (e.g., heparin, direct oral anticoagulants). Stability studies support 24-month shelf life at -40°C to -80°C and 24-hour in-use stability.
Technological Characteristics
Reagent: Immunodepleted human citrated plasma with exogenous vWF and HEPES buffer. Form factor: Frozen small-volume aliquots (1.0 mL and 1.5 mL). Sensing principle: Clot-based one-stage APTT assay. Connectivity: Used on automated ACL TOP/ACL TOP 50 Series analyzers. Storage: -40°C to ≤-70°C. Standards: CLSI EP05-A3, EP06, EP25-A, EP07, EP28-A3c.
Indications for Use
Indicated for individuals aged 2 years and older suspected of having Factor VIII deficiency or requiring management of hemophilia A.
K971225 — CRYO CHECK FACTOR VIII DEFICIENT PLASMA · Precision Biologicals, Inc. · Jul 17, 1997
K102851 — NOFACT VIII · R2 Diagnostics, Inc. · Dec 19, 2011
K034007 — HEMOSIL FACTOR VIII DEFICIENT PLASMA · Instrumentation Laboratory CO · Feb 13, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
September 13, 2023
Precision BioLogic Inc. Karen Black VP of Compliance and Product Development 140 Eileen Stubbs Avenue Dartmouth, Nova Scotia B3B 0A9 Canada
Re: K222831
Trade/Device Name: CRYOcheck Factor VIII Deficient Plasma with VWF Regulation Number: 21 CFR 864.7290 Regulation Name: Factor Deficiency Test Regulatory Class: Class II Product Code: GJT Dated: May 19, 2023 Received: May 19, 2023
Dear Karen Black:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the text "Min Wu-S" in a large, sans-serif font. The text is black, and the background is white. The letters are evenly spaced and easy to read. There is a faint watermark in the background.
Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K22831
#### Device Name
CRYOcheck Factor VIII Deficient Plasma with VWF
#### Indications for Use (Describe)
CRYOcheck Factor VIII Deficient Plasma with VWF is for clinical laboratory use as a deficient substrate in the quantitative determination of Factor VIII activity in 3.2% citrated human plasma based on the activated partial thromboplastin time (APTT) assay. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary cryocheck™ Factor VIII Deficient Plasma with VWF
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K222831
| Submitter's | Precision BioLogic Inc. | |
|-------------------------------------|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Information | 140 Eileen Stubbs Ave. | |
| | Dartmouth, Nova Scotia B3B 0A9 | |
| | Canada | |
| Contact Person | Karen M. Black, VP of Compliance & Product Development | |
| | Phone: 902-468-6422, ext. 226, or 902-706-3125 | |
| | E-mail: kblack@precisionbiologic.com | |
| Preparation Date | 25 August 2023 | |
| Device Trade Name | CRYOcheck™ Factor VIII Deficient Plasma with VWF | |
| | Regulation Number and | 21 CFR 864.7290 |
| | Description | Factor Deficiency Test |
| Regulatory | Classification | Class II |
| Information | Product Code | GJT, Plasma, Coagulation Factor<br>Deficient; 21 CFR 864.7290 |
| | Classification Panel | Hematology |
| Predicate Device | HemosIL Factor VIII Deficient Plasma (K110237) | |
| Indication for Use/<br>Intended Use | | CRYOcheck Factor VIII Deficient Plasma with VWF is for clinical<br>laboratory use as a deficient substrate in the quantitative<br>determination of Factor VIII activity in 3.2% citrated human plasma<br>based on the activated partial thromboplastin time (APTT) assay. It is<br>intended to be used in identifying factor VIII deficiency and as an aid in<br>the management of hemophilia A in individuals aged 2 years and<br>older. For in vitro diagnostic use |
| Device Description | | CRYOcheck Factor VIII Deficient Plasma with VWF is normal human<br>citrated plasma which has been immunodepleted of factor VIII and to<br>which an exogenous source of human von Willebrand Factor (vWF) has<br>been added and buffered with HEPES. Factor VIII has been assayed at<br>less than 1% of normal activity levels and vWF antigen and activity are<br>>50%. It will be provided to users frozen in small-volume aliquots (25<br>vials of 1.0 mL, and 25 vials of 1.5 mL). Vials will be packaged into<br>boxes; these will be frozen during the manufacturing process and will<br>be shipped and stored frozen until use to preserve the integrity of the<br>components |
| Comparison to Predicate | | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Predicate | New Device |
| Proprietary and<br>Established Names | HemosIL Factor VIII Deficient<br>Plasma | CRYOCheck Factor VIII Deficient<br>plasma with VWF |
| Manufacturer | Instrumentation Laboratory | Precision BioLogic |
| Similarities | | |
| Measurand | Human Factor VIII | Same |
| Classification Product<br>Code | GJT, Plasma, Coagulation<br>Factor Deficient | Same |
| Regulation Section | 21 CFR 864.7290<br>Factor Deficiency Test | 21 CFR 864.7290<br>Factor Deficiency Test |
| Comparison to Predicate | | |
| Item | Predicate | New Device |
| Classification | Class II | Class II |
| Panel | 81 (Haematology) | 81 (Haematology) |
| Intended Use | HemosIL Factor VIII deficient<br>plasma is human plasma<br>depleted for Factor VIII and<br>intended for the in vitro<br>diagnostic quantitative<br>determination of Factor VIII<br>activity in citrated plasma,<br>based on the activated partial<br>thromboplastin time (APTT)<br>assay, on the ACL TOP Family<br>analyzers. HemosIL Factor VIII<br>deficient plasma is indicated<br>for use on patients who are<br>suspected of congenital or<br>acquired deficiency based on<br>the activated partial<br>thromboplastin time (APTT)<br>assay results. | CRYOcheck Factor VIII Deficient<br>Plasma with VWF is for clinical<br>laboratory use as a deficient<br>substrate in the quantitative<br>determination of Factor VIII<br>activity in 3.2% citrated human<br>plasma based on the activated<br>partial thromboplastin time<br>(APTT) assay. It is intended to<br>be used in identifying factor VIII<br>deficiency and as an aid in the<br>management of hemophilia A in<br>individuals aged 2 years and<br>older. For in vitro diagnostic<br>use. |
| Assay Type | Quantitative (clot-based<br>measurement of FVIII) | Same |
| Methodology | Factor VIII activity in a<br>patient's plasma is determined<br>by performing a modified<br>activated partial<br>thromboplastin time test<br>(APTT). Patient plasma is<br>diluted and added to plasma<br>deficient in Factor VIII.<br>Correction of the clotting time<br>of the deficient plasma is<br>proportional to the<br>concentration (% activity) of<br>that factor in the patient<br>plasma interpolated from a<br>calibration curve. | Same |
| Expression of results | Quantitative; results are<br>expressed as percent activity.<br>interpreted relative to a<br>calibration curve. | Same |
| Instrument(s) | ACL TOP Family/ACL TOP<br>Family 50 Series | Same |
| Differences | | |
| Item | Predicate | New Device |
| Device Description | The HemosIL Factor VIII deficient plasma kit contains 10 x 1 mL vials of lyophilized human plasma that has been artificially depleted of factor VIII containing buffer and stabilizers. The residual factor VIII activity is less than or equal to 1% whereas von Willebrand Factor and the remaining intrinsic pathway factor levels are normal. | CRYOcheck Factor VIII Deficient Plasma with VWF is normal human citrated plasma which has been immunodepleted of factor VIII and to which an exogenous source of human von Willebrand Factor (vWF) has been added and buffered with HEPES. Factor VIII has been assayed at less than 1% of normal activity levels and vWF antigen and activity are >50%. It will be provided to users frozen in small-volume aliquots (25 vials of 1.0 mL, and 25 vials of 1.5 mL). Vials will be packaged into boxes; these will be frozen during the manufacturing process and will be shipped and stored frozen until use to preserve the integrity of the components. |
{4}------------------------------------------------
# PrecisionBioLogic
{5}------------------------------------------------
#### Performance Summary:
All studies were performed using cRYocheck Factor VIII Deficient Plasma with VWF in a modified APTT assay with Instrumentation Laboratories' APTT SynthASil reagent to measure FVIII activity on Instrumentation Laboratories' ACL TOP Series or TOP 50 Series Instruments; the specific instrument(s) used for each study are indicated in the summary reports below.
#### Multi-Reagent Lot Precision
An internal precision study was performed using three (3) lots of cRYocheck Factor VIII Deficient Plasma with VWF in a modified APTT assay by two operators on an IL ACL TOP 700 CTS analyzer (K160276) in accordance with CLSI EP05-A3. The study quantified one normal and two abnormal reference controls and three patient plasma samples representing very low, and high levels of FVIII activity. Each sample was measured with each product lot in duplicate, twice a day for 20 days for a total of 80 replicates per sample per lot. The pooled results demonstrated a precision of <10% CV for all controls and the high FVIII activity plasma sample, and <0.5 SD for the low and very low FVIII plasma samples.
| Aggregated Data (Lots 1, 2 and 3) | | | | | |
|----------------------------------------|------------|----------------------|------|-----------------|------|
| Sample | Mean FVIII | Within-Run Precision | | Total Precision | |
| | (%) | SD | %CV | SD | %CV |
| CRYOcheck Reference Control Normal | 100.4 | 3.5 | 3.4 | 5.7 | 5.7 |
| CRYOcheck Abnormal 1 Reference Control | 37.7 | 1.7 | 4.6 | 2.3 | 6.2 |
| CRYOcheck Abnormal 2 Reference Control | 11.3 | 0.4 | 3.7 | 0.6 | 5.2 |
| High FVIII Plasma Sample | 164 | 6.1 | 3.7 | 9.8 | 6.0 |
| Low FVIII Plasma Sample | 2.0 | 0.2 | 9.3 | 0.2 | 11.1 |
| Very Low FVIII Plasma Sample | 0.1 | 0.0 | 18.2 | 0.0 | 24.1 |
{6}------------------------------------------------
#### Multi-Reagent Lot Site to Site Reproducibility
Reproducibility studies were conducted at three sites (one internal and two external) by two operators per site on IL ACL TOP 500. IL ACL TOP 700 CTS (K160276), and IL ACL TOP 750 CTS (K150877), analyzers using three lots of CRYOcheck Factor VIII Deficient Plasma with VWF in a modified APTT assay in accordance with CLSI EP05-A3. The study quantified one normal and two abnormal reference controls and three patient plasma samples representing very low. low and high levels of FVIII activity. Each sample was measured in triplicate, twice a day for 5 days at each site. The pooled data across three sites demonstrated a reproducibility of <10% CV for all controls and the high FVIII plasma sample, and <0.5 SD for the low and very low FVIII plasma samples.
| Pooled 3-Site Data | | | | | | | | | | | | | | |
|---------------------------------|-----|-------------------|-------------------------------|-------|--------------|------|-------------|-------|-------------|------|--------------|-------|----------------------------|-------|
| Sample | N | Mean<br>FVIII (%) | Repeatability<br>(Within-Run) | | Between- Run | | Between-Day | | Between-Lot | | Between-Site | | Reproducibility<br>(Total) | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Reference Control<br>Normal | 270 | 98.97 | 3.00 | 3.03 | 2.07 | 2.09 | 1.78 | 1.79 | 1.92 | 1.94 | 3.95 | 3.99 | 5.98 | 6.04 |
| Abnormal 1<br>Reference Control | 270 | 37.40 | 1.25 | 3.33 | 0.63 | 1.68 | 0.61 | 1.63 | 0.80 | 2.14 | 0.44 | 1.16 | 1.77 | 4.75 |
| Abnormal 2<br>Reference Control | 270 | 11.87 | 0.41 | 3.48 | 0.18 | 1.55 | 0.28 | 2.43 | 0.33 | 2.81 | 0.66 | 5.54 | 0.91 | 7.68 |
| High FVIII Plasma<br>Sample | 269 | 166.37 | 6.54 | 3.93 | 2.08 | 1.25 | 1.45 | 0.87 | 4.34 | 2.61 | 1.78 | 1.07 | 8.44 | 5.07 |
| Low FVIII Plasma<br>Sample | 270 | 1.93 | 0.11 | 5.84 | 0.06 | 3.08 | 0.03 | 1.81 | 0.08 | 4.01 | 0.09 | 5.01 | 0.18 | 9.39 |
| Very Low FVIII<br>Plasma Sample | 270 | 0.11 | 0.03 | 30.24 | 0.01 | 5.90 | 0.01 | 10.16 | 0.01 | 4.52 | 0.02 | 20.34 | 0.04 | 38.56 |
#### Linearity/Assay Reportable Range
A linearity study was conducted in accordance with CLSI EP06-2m Ed using three lots of cryocheck Factor VIII Deficient Plasma with VWF in a modified APTT assay to quantify FVIII on an IL ACL TOP 700 CTS instrument (K160276). A high FVIII (~260%) plasma was combined with congenital hemophilia A patient plasma (0% FVIII) to create fifteen sample dilutions with estimated FVIII activities in the range of 0 to 260% FVIII. Each level was tested in quadruplicate. The results support the linearity claim described below.
#### Linearity Range: 0 to 230% FVIII activity
#### Reference Interval
A reference interval study was conducted by two operators on two IL ACL TOP 700 CTS (K160276) analyzers in accordance with CLSI EP28-A3c using three lots of CRYOcheck Factor VIII Deficient Plasma with VWF and citrated plasma samples from 136 normal, ostensibly healthy individuals. The reference interval was established by calculating the non-parametric 95% confidence interval (2.5th to 97.5th percentiles).
Reference Interval: 62 to 163 % FVIII activity
{7}------------------------------------------------
#### Stabilitv
#### Shelf-Life Stability
A shelf-life stability study was conducted in accordance with CLSI EP25-A using an IL ACL TOP 700 CTS instrument (K160276). Three lots of CRYOcheck Factor VIII Deficient Plasma with VWF were stored at -40 °C and ≤-70 °C and tested at t=0 and reqular intervals defined by the lotspecific pull schedule up to 37 months (real time study is ongoing). At each timepoint, five replicates of one normal and two abnormal reference controls and two patient plasma samples representing very low and high levels of FVIII activity levels were quantified in a modified APTT assay. The study has been completed up to 25 months and supports a shelf-life stability claim of 24 months when the product is stored at -40 °C to -80 °C.
#### In-Use Stability
An in-use stability study was conducted in accordance with CLSI EP25-A using an IL ACL TOP 700 CTS instrument (K160276). Three lots of CRYOcheck Factor VIII Deficient Plasma with VWF were maintained on board the analyzer (14.5 - 15.5 °C) for up to 25 hours and in a refrigerator (2–8 °C) for up to 25 hours. Each lot was used in a modified APTT assay to quantify five replicates of one normal and two abnormal reference controls and two patient plasma samples representing low and high levels of FVIII activity from each storage condition at defined timepoints. The data support a stability claim of 24 hours on board the instrument and at 2-8 °C.
#### Interferences
Interference studies were conducted according to CLSI EP07-A3 using a single lot of cRYocheck Factor VIII Deficient Plasma with VWF in a modified APTT assay on an IL ACL TOP 700 CTS instrument (K160276). Plasma samples were spiked with possible interferents, and 10 replicates were tested alongside 10 replicates of the corresponding blank matrix control. The following substances showed no interference up to the concentrations indicated:
| Possible Interferent | Concentration |
|--------------------------|------------------|
| Hemoglobin | ≤1000 mg/dL |
| Intralipid | ≤2000 mg/dL |
| Bilirubin (conjugated) | ≤4.0 mg/dL |
| Bilirubin (unconjugated) | 40 mg/dL |
| Lupus Anticoagulant | ≤1.8 dRVVT ratio |
| Warfarin | ≤ INR ratio 2.72 |
Rivaroxaban, fondaparinux, dabigatran, emicizumated heparin, and low molecular weight heparin were shown to interfere with the quantification of FVIII activity.
{8}------------------------------------------------
#### Recovery of FVIII Replacements
A recovery study was conducted using a single lot of CRYOcheck Factor VIII Deficient Plasma with VWF in a modified APTT assay on an IL ACL TOP 700 CTS instrument (K160276). Congenital FVIII deficient plasma was spiked with six FVIII replacement therapies at seven concentrations and percent recovery was determined. CRYOcheck Factor VIII Deficient Plasma with VWF accurately evaluated (100 ± 25%) the potency of FVIII replacements including Advate, Afstyla, Elocate, Jivi, Novoeight, and Wilate at concentrations ranging from 0.05 to 1.0 IU/mL. There was an underestimation of Afstyla*, and a 2x correction factor was applied to the results based on manufacturer's recommendations.
| Product | Mean Percent Recovery (%) |
|-----------|---------------------------|
| Advate | 92.84 |
| Afstyla* | 97.38 |
| Eloctate | 94.90 |
| Jivi | 104.49 |
| Novoeight | 113.33 |
| Wilate | 94.92 |
* Per the manufacturer's recommendations, a chromogenic assay is recommended for measurement of Afstyla and results obtained by a one stage clotting assay will under recover by 50%.
#### Method Comparison
A method comparison study was conducted at four sites (one internal and three external) according to CLSI EP09c to compare the accuracy of factor VIII activity measurement when using cRYocheck Factor VIII Deficient Plasma with VWF in a modified APTT assay relative to a comparator device. Three hundred and sixty-six human plasma samples from normal ostensibly healthy individuals, from patients with congenital or acquired hemophilia A, patients with hemophilia B, patients with von Willebrand disease, hemophilia A patients on FVIII replacement therapies, and patients with other factor deficiencies were distributed across four sites and tested to quantify FVIII activity using a single lot of cRYocheck Factor VIII Deficient Plasma with VWF on IL ACL TOP 500, IL ACL TOP 700 CTS (K160276), and IL ACL TOP 750 CTS (K150877) analyzers. A second aliquot of each sample was tested using a modified APTT assay with HemosIL Factor VIII Deficient Plasma on an IL ACL TOP 750 instrument (K150877).
Results were compared by Passing-Bablok regression statistics show that CRYOcheck Factor VIII Deficient Plasma with VWF performed equivalently to the comparator method.
| | N | Slope | | Intercept | | Pearson Correlation<br>Coefficient (r) |
|---------|-----|-------|--------------|-----------|----------------|----------------------------------------|
| | | Value | 95% CI | Value | 95% CI | |
| Site 1 | 115 | 1.15 | 1.12 to 1.19 | 0.17 | -0.38 to 0.64 | 0.93 |
| Site 2 | 125 | 1.19 | 1.16 to 1.22 | -0.26 | -0.48 to -0.09 | 0.97 |
| Site 3 | 108 | 1.10 | 1.06 to 1.15 | 0.97 | 0.24 to 1.99 | 0.98 |
| Site 4 | 18 | 1.22 | 0.95 to 1.31 | 1.19 | -3.99 to 13.96 | 0.98 |
| Overall | 366 | 1.16 | 1.15 to 1.18 | -0.08 | -0.25 to 0.13 | 0.96 |
Absolute predicted biases at medical decision levels are reported below.
| FVIII activity (%) | Predicted Bias (%) | Lower CI (%) | Upper CI (%) |
|--------------------|--------------------|--------------|--------------|
| 1 | 0.37 | -1.42 | 2.17 |
| 5 | 0.91 | -0.79 | 2.62 |
| 50 | 6.97 | 5.85 | 8.09 |
| 100 | 13.70 | 11.84 | 15.56 |
{9}------------------------------------------------
### Sample Integrity
A sample integrity study was conducted at two external sites to assess the stability of fresh plasma samples at room temperature, when stored frozen at ≤-70 °C and after up to two freeze thaw cycles. The FVIII activity of ninety-four plasma samples was measured using one lot of cryocheck Factor VIII Deficient Plasma with VWF on IL ACL TOP 300 and IL ACL TOP 700 (K160276) analyzers. Results were compared using weighted Deming regression analysis and support a fresh sample stability claim of 2 hours at room temperature and a frozen storage claim of 1.5 months at ≤-70 ℃, including up to two freeze thaw cycles.
#### Conclusion
The performance testing results demonstrate that CRYOcheck Factor VIII Deficient Plasma with VWF is substantially equivalent to the predicate device, HemosIL Factor VIII Deficient Plasma (K110237) and that the assay is effective for its labeled intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.