← Product Code [JDB](/productcode/JDB) · K222807

# TEMA Elbow system - Line extension (K222807)

_Lima Corporate S.P.A. · JDB · Oct 26, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K222807

## Device Facts

- **Applicant:** Lima Corporate S.P.A.
- **Product Code:** [JDB](/productcode/JDB.md)
- **Decision Date:** Oct 26, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3160
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

TEMA Elbow system is intended for use in cemented applications for patients suffering from disability due to: 1. Elbow joint destruction which significantly compromises the activities of daily living 2. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis with hemophilia. 3. Rheumatoid arthritis or degenerative arthritis with incapacitating pain 4. Revision where other devices or treatments have failed. 5. Correction of severe functional deformity. 6. Treatment of acute or chronic fractures with distal humerus epicondyle involvement. 7. Post-traumatic lesions or bone loss contributing to elbow instability or loss of motion The unlinked, semi-constrained version is for use in cases where there are functioning medial and lateral elbow ligaments of good quality that provide immediate stability, resisting any tendency of disengagement between ulnar and humeral components, when intraoperative manual testing is performed.

## Device Story

Cemented total elbow prosthesis; modular humeral and ulnar assemblies. Humeral assembly: metallic stem, metallic body, polyethylene bushings. Ulnar assembly: metallic stem, metallic body, polyethylene liner. Constrained (linked) configuration uses axle to connect assemblies; unlinked (semi-constrained) configuration omits axle. Used in orthopedic surgery for joint replacement. Surgeon implants device to restore joint function, reduce pain, and correct deformity. Benefits include improved mobility and stability for patients with severe elbow pathology.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via mechanical bench testing, including range of motion, stability, fatigue, fretting-corrosion, and wear evaluation.

## Technological Characteristics

Cemented total elbow prosthesis. Modular metallic humeral/ulnar stems and bodies; polyethylene bushings/liners. Constrained (linked) or semi-constrained (unlinked) configurations. Design includes screw-secured taper couplings. Mechanical assembly.

## Regulatory Identification

An elbow joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a radial resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- TEMA Elbow system ([K181362](/device/K181362.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 26, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

LimaCorporate S.p.A. % Kenneth Newman Regulatory Affairs Associate Lima USA Inc. 2001 NE Green Oaks Blvd. Ste. 100 Arlington. Texas 76006

Re: K222807

Trade/Device Name: TEMA Elbow system - Line extension Regulation Number: 21 CFR 888.3160 Regulation Name: Elbow joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDB, JDC Dated: August 30, 2022 Received: September 16, 2022

Dear Kenneth Newman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 

{1}------------------------------------------------

801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic provisions (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Victoria Lilling, M.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K222807

Device Name TEMA Elbow system - Line extension

#### Indications for Use (Describe)

TEMA Elbow system is intended for use in cemented applications for patients suffering from disability due to:

- 1. Elbow joint destruction which significantly compromises the activities of daily living
- 2. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis with hemophilia.
- 3. Rheumatoid arthritis or degenerative arthritis with incapacitating pain
- 4. Revision where other devices or treatments have failed.
- 5. Correction of severe functional deformity.
- 6. Treatment of acute or chronic fractures with distal humerus epicondyle involvement.
- 7. Post-traumatic lesions or bone loss contributing to elbow instability or loss of motion

The unlinked, semi-constrained version is for use in cases where there are functioning medial and lateral elbow ligaments of good quality that provide immediate stability, resisting any tendency of disengagement between ulnar and humeral components, when intraoperative manual testing is performed.

| Type of Use (Select one or both, as applicable)                                                 |                                             |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Date: October 21, 2022

Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy

U.S. Contact Person: Kenneth Newman Regulatory Affairs Associate kenneth.newman@limacorporate.comLima USA Inc. 2001 NE Green Oaks Blvd. Ste.100 Arlington, Texas 76006, USA www.limacorporate.com Cell Phone: 682.597.3381 Office Phone: 817.385.0777 FAX: 817.385.0377

| Product                               | Product<br>Code | Regulation and Classification Name                                                |
|---------------------------------------|-----------------|-----------------------------------------------------------------------------------|
| TEMA Elbow System<br>– Line extension | JDC             | 21 CFR 888.3150 Elbow joint metal/polymer<br>constrained cemented prosthesis      |
|                                       | JDB             | 21 CFR 888.3160 Elbow joint metal/polymer<br>semi-constrained cemented prosthesis |

## Description:

TEMA Elbow System is a cemented total elbow prosthesis. The system provides both linked (constrained) and unlinked (semi-constrained) configurations.

TEMA Elbow System consists of modular humeral and ulnar assemblies. The humeral assembly consists of a metallic humeral stem and a metallic humeral body component with pre-assembled polyethylene bushings. The ulnar assembly is composed of a metallic ulnar stem, a metallic ulnar body and a polyethylene ulnar liner. Screws are used to secure the taper couplings between the stems and bodies of the humeral and ulnar assemblies.

The constrained (linked) version of the elbow consists of the modular humeral assembly and the modular ulnar assembly with an axle component that is inserted across the joint to connect the humeral assembly to the ulnar assembly.

The semi-constrained (unlinked) version consists of the same components without the use of the axle to link the humeral and ulnar assemblies.

TEMA Elbow system - Line extension introduces two new sizes of Ulnar stems to the ulnar stem range already cleared; moreover, a design change is introduced for the axle component, ulnar body component and for the humeral body compared to the original cleared version.

{4}------------------------------------------------

#### Indication for Use:

TEMA Elbow system is intended for use in cemented applications for patients suffering from disability due to:

- 1. Elbow joint destruction which significantly compromises the activities of daily living
- 2. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis with hemophilia.
- 3. Rheumatoid arthritis or degenerative arthritis with incapacitating pain
- 4. Revision where other devices or treatments have failed.
- 5. Correction of severe functional deformity.
- Treatment of acute or chronic fractures with distal humerus epicondyle 6. involvement.
- 7. Post-traumatic lesions or bone loss contributing to elbow instability or loss of motion

The unlinked, semi-constrained version is for use in cases where there are functioning medial and lateral elbow ligaments of good quality that provide immediate stability, resisting any tendency of dislocation or disengagement between ulnar and humeral components, when intraoperative manual testing is performed.

#### Predicate Devices:

| Company       | Device name       | Cleared via |
|---------------|-------------------|-------------|
| LIMACORPORATE | TEMA Elbow system | K181362     |

#### Summarv of technology comparison:

TEMA Elbow system - line extension and TEMA Elbow system (predicate device) have extensive and significant similarities in terms of intended use, indications, technological characteristics, materials, and principles of operation. The differences between these two systems are related to a design improvement of axle, humeral body and ulnar body components aimed to improve the overall mechanical performance and laxity of the system plus the addition of two new sizes of Ulnar stems.

A comparison of technological characteristics and performance testing demonstrate that the TEMA Elbow system-line extension is substantially equivalent to the predicate devices.

## Non-clinical testing

Mechanical evaluations (both bench tests and rationales) demonstrated the device's ability to perform in a substantially equivalent manner to the predicate devices in terms of:

- Range of motion:
- Stability, fatigue and fretting-corrosion evaluation post-fatigue of modular ● connections;
- Wear evaluation. ●

{5}------------------------------------------------

## Clinical testing

Clinical testing was not necessary to demonstrate substantial equivalence of the TEMA Elbow system – Line extension components to the predicate device.

## Conclusion

Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, the TEMA Elbow system - Line extension components are substantially equivalent to the predicate device components identified in this premarket notification.

---

**Source:** [https://fda.innolitics.com/device/K222807](https://fda.innolitics.com/device/K222807)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
