K222597 · Owlet Babycare, Inc. · DQA · Jun 16, 2023 · Cardiovascular
Device Facts
Record ID
K222597
Device Name
BabySat 3
Applicant
Owlet Babycare, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jun 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222597 · Jun 16, 2023
BabySat 3
Owlet Babycare, Inc.
Observational study in home-use population
An observational study was conducted in the intended patient population to estimate the incidence of skin reactions during routine use and to collect signal quality data across various patient demographics, including Fitzpatrick skin tones.
Skin irritation; Signal quality; Home-use; Observational study
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Observational study for skin irritation; Observational study
Intended patient population (infants 1 month to 18 months, 6-30 lbs)
Not applicable for this study
Incidence of clinically important skin reactions; time for recovery; incidence of skin reactions requiring medical intervention; signal quality data.
Indications for Use
The BabySat 3 pulse oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse rate. It is indicated for spot-checking and/or continuous monitoring of well-perfused patients greater than one month old up to 18 months old and weighing between 6 and 30 lbs., in the home environment
Device Story
BabySat 3 is a home-use pulse oximeter for infants. System components: fabric sock with integrated sensor, base station, and mobile app. Sensor uses red (653nm) and infrared (931nm) LEDs to measure light absorption through tissue; photodiode detects pulsatile arterial blood flow to calculate SpO2 and pulse rate. Sensor transmits data via Bluetooth to base station; base station provides audible/visual alarms for out-of-range readings and relays data to mobile app/cloud. Operated by parents/caregivers in home settings. Output displayed on mobile app. Device assists caregivers in monitoring infant vitals; provides alerts for potential desaturation or abnormal pulse rates.
Clinical Evidence
Clinical evidence includes a desaturation study per ISO 80601-2-61 across the full Fitzpatrick skin tone scale, confirming SpO2 accuracy in motion and non-motion conditions. An observational study in the intended population (infants) evaluated skin irritation, reporting no clinically significant events. Human factors study with lay users confirmed successful execution of critical tasks.
Technological Characteristics
Materials: Biocompatible (ISO 10993). Sensing: Red (653nm) and IR (931nm) LEDs, photodiode detector. Connectivity: Bluetooth (sensor to base), Wi-Fi (base to app/cloud). Power: Battery-operated sensor, AC-powered base station with 10-min battery backup. Standards: AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-8, ISO 80601-2-61.
Indications for Use
Indicated for spot-checking and/or continuous monitoring of SpO2 and pulse rate in well-perfused infants aged >1 month to 18 months, weighing 6-30 lbs, in home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Guardian Angel Rx GA 1001 Digital Vital Sign Monitoring System (K182882)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 16, 2023
Owlet Baby Care, Inc. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE St. Petersburg, Florida 33704
Re: K222597
Trade/Device Name: BabySat 3 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: May 25, 2023 Received: May 25, 2023
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James J. Lee -S
James J. Lee, PhD Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K22597
Device Name
BabySat 3
Indications for Use (Describe)
The BabySat 3 pulse oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse rate. It is indicated for spot-checking and/or continuous monitoring of well-perfused patients greater than one month old up to 18 months old and weighing between 6 and 30 lbs., in the home environment
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 PP
{3}------------------------------------------------
| Date Prepared: | 08-Jun-2023 |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Owlet Baby Care Inc.<br>3300 Ashton Blvd.<br>Suite 300<br>Lehi, UT 84043<br>Tel - 801-901-4303 | |
| Official Contact: | Tammy Lavery<br>Sr. Director, Quality Engineering & Regulatory Affairs |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr NE<br>St. Petersburg, FL 33704 |
| Proprietary or Trade Name:<br>Common/Usual Name:<br>Classification: | BabySat 3<br>Oximeter<br>21 CFR 870.2700<br>DQA - Oximeter |
| Predicate Devices: | K182882 - Taiwan Aulisa Medical Devices<br>Technologies, Inc. - Guardian Angel Rx GA 1001<br>Digital Vital Sign Monitoring System |
| Common/Usual Name:<br>Classification: | Oximeter<br>21 CFR 870.2700<br>DQA - Oximeter |
#### Device Description:
The Owlet BabySat 3 is a Pulse Oximeter for use in the home while the baby is resting or sleeping. The Sock secures the Sensor to the baby's foot. The Sensor measures the baby's readings (SpO2 and Pulse Rate) and transmits the baby's readings to the Base Station directly via Bluetooth. The Base Station monitors the baby's readings and will alarm if the readings are outside prescribed limits. The Base Station also relays the data to the App and to the Owlet servers. The Owlet BabySat 3 mobile device application displays the baby's readings.
The device is stand-alone and not part of a multiparameter module. The device is non-sterile, single patient use reusable and cleanable. The device is not reprocessed.
The BabySat 3 Pulse Oximeter is comprised of several components:
- . Sock
- This holds the sensor in the appropriate place O
- There are 2 size of socks available O
- It is cleanable O
- Sensor
- This is a battery-operated sensor with LED and collects data that is transmitted to O the Base Station
- . Base Station
- This is a local unit that has audible and visual alarms which follow the prescribed O and pre-set alarm limits for Pulse rate and SpO2
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 11
- It can transmit data to the Cloud or a mobile device with the downloaded app o
- Mobile App
- o This connects with the Base Station via a Wi-Fi local network and allows the individual to monitor and receive messages and alarms remotely
## Principle of Operation:
The BabySat 3 sock contains a sensor which uses red and infrared light which is transmitted through oxyhemoglobin and are sensed in the photoelectric cell. The red and infrared light is absorbed in different amounts depending on the oxygenation of the blood. This data is then used to calculate SpO2. This is the standard principle of almost all pulse oximeters.
The BabySat 3 Pulse Oximeter measures functional Oxygen Saturation of arterial Hemoglobin (SpO2) and Pulse Rate. Pulse oximetry measurements are based on two physiological principles: The differences in optical absorbance properties between oxyhemoglobin (HbO2, oxygenated blood) and deoxyhemoglobin (Hb, non-oxygenated blood). The volume (and therefore light absorption) of arterial blood changes due to heart pulsation as it is distributed in tissue throughout the body.
A Pulse Oximeter determines the saturation of arterially Perfused tissue by sequential illumination with the red (653nm) and infrared (931nm) wavelengths of light via light emitting diodes (LED's). The time-varying signal passing through the tissue is measured from a photodiode detector many times per second, and the differences between the maximum and minimum absorbance due to pulsation are used to determine the saturation, and the pulsation time period is used to calculate the Pulse Rate.
## Indications for Use:
The BabySat 3 pulse oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse rate. It is indicated for spot-checking and/or continuous monitoring of well-perfused patients greater than one month old up to 18 months old and weighing between 6 and 30 lbs., in the home environment.
{5}------------------------------------------------
# 510(k) Summary
Page 3 of 11
# Table 1 - Comparison - Subject vs. Predicate
| Subject Device | Predicate | Comparison | Does the difference raise<br>different questions of safety<br>and effectiveness? | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model: | BabySat 3 | GA 1001 | | |
| Manufacturer: | Owlet, Inc. | Taiwan Aulisa Medical Devices<br>Technologies | | |
| K# | K222597 | K182822 | | |
| ProCode | DQA - Oximeter<br>CFR 870.2700 | DQA - Oximeter<br>CFR 870.2700 | Same | No |
| General Specifications | | | | |
| Use Environment | Home setting | hospitals, medical facilities, home<br>care, and subacute environments | Similar, with the subject<br>restricted to home use<br>only | No, Subject device has a more<br>limited use environment than<br>predicate. |
| Patient population | Infant at least 1 month old and up to<br>18 months old and 6-30 lbs. | Infants and pediatrics | Similar | No, Subject device has a more<br>limited patient population. |
| Indications for use: | The BabySat 3 pulse oximeter is<br>indicated for use in measuring and<br>displaying functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and Pulse rate. It is<br>indicated for spot-checking and/or<br>continuous monitoring of well-<br>perfused patients greater than one<br>month old up to 18 months old and<br>weighing between 6 and 30 lbs., in<br>the home environment | The Guardian Angel Rx GA1001<br>Digital Vital Sign Monitoring<br>System is indicated for use in<br>measuring and displaying<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate. It is indicated for spot-<br>checking and/or continuous<br>monitoring of pediatrics and<br>infants during non-motion and<br>under well-perfused conditions.<br>The intended environments of use<br>are hospitals, medical facilities,<br>home care, and subacute<br>environments. This system is a<br>reusable device. | Similar | No, Equivalent with exception<br>of the more limited patient<br>population for the subject<br>device. Also, the subject device<br>is not limited to non-motion<br>conditions. |
| | Subject Device | Predicate | Comparison | Does the difference raise<br>different questions of safety<br>and effectiveness? |
| Contraindications: | - Not for patients under 1<br>month old or weighing less than 6<br>pounds<br>- Not for patients older than 18<br>months or weighing more than 30<br>pounds<br>- Not a substitute for a caregiver<br>- Not for use as an apnea monitor | -Do not use any part of this<br>system in an MRI environment.<br>-Explosion Hazard: Do not use<br>this system in an explosive<br>atmosphere or in the presence of<br>flammable anesthetics or gasses.<br>-This device is not a replacement<br>for a caregiver. | Similar | No<br>Subject device is for home use<br>only. MRI environments do not<br>need to be contraindicated<br>because they are not found in<br>home environments.<br>Explosive environments are<br>uncommon in the home<br>environment. |
| Application sight: | Foot | Foot | Same | No |
| Accessory | Sock with sensor | Sock with sensor | Similar | No |
| Materials in patient<br>contact | ISO 10993 tested | Unknown | Similar | No<br>Same patient contact |
| Operating<br>Temperature<br>Range: | 5 to 40°C | 5 to 40°C | Same | No<br>Equivalent. Specified by<br>60601-1-11:2015, 4.2.3.1 |
| Storage/Transport<br>Temperature<br>Range: | -25 to 60°C | -25 to 60°C | Similar | No<br>Equivalent. Specified by<br>60601-1-11:2015, 4.2.3.1 |
| Storage/Transport<br>Atmospheric<br>Pressure: | 500 to 1060 hPa | Not specified | Similar | No<br>Based on 60601-1 subclause<br>7.9.3.1 rationale |
| Operating Altitude: | ≤ 3000 m | ≤ 3000 m | Same | No |
| Operating<br>Atmospheric<br>Pressure: | 700 hPa to 1013 hPa | 700 hPa to 1013 hPa | Same | No, Equivalent. Specified by<br>60601-1-11:2015, 4.2.3.1 |
| Operating<br>Humidity: | 15 to 90% RH non-condensing | 15 to 90% RH non-condensing | Same | No, Equivalent. Specified by<br>60601-1-11:2015, 4.2.3.1 |
| | Subject Device | Predicate | Comparison | Does the difference raise<br>different questions of<br>safety and effectiveness? |
| Storage/Transport<br>Humidity: | 0% to 90 %, non-condensing.<br>Water vapor pressure not to<br>exceed 50hPa. | 10% to 93% RH non-condensing | Similar | No<br>Subject device meets<br>requirements specified by<br>60601-1-11:2015, subclause<br>4.2.2 |
| Ingress rating: | Sensor: IP35<br>Base: IP22 | IP22 | Similar | No<br>Sensor meets more stringent<br>ingress standard. |
| Alarms | | | | |
| SpO2 Upper Limit | Default: Off, Range: 85 to 100 | Default: Off, Range: 85 to 100 | Same | No |
| SpO2 Lower Limit: | Default: 85%, Range: 50 to 95 | Default: 85%, Range: 50 to 95 | Same | No |
| Pulse Upper Limit: | Default: 200 BPM<br>Range: 75 to 275 | Default: 200 BPM<br>Range: 75 to 275 | Same | No |
| Pulse Lower Limit: | Default: 75 bpm<br>Range: 30 to 110 | Default: 75 bpm<br>Range: 30 to 110 | Similar | No<br>Min range of 30 bpm is<br>adequate for the patient<br>population of the subject device. |
| Alarm Silence<br>Duration: | 2 min | 2 min | Same | No |
| Alarm Sound<br>Pressure: | 55-60 dBA at 1m | 60 db | Similar | No |
| Permanent Alarm<br>Silence: | No permanent alarm silence | Alarm can be permanently<br>silenced | Similar | No |
| Classification (IEC 60601-1) | | | | |
| Type of Protection: | Class II | Class II | Same | No |
| Degree of Protection: | Type BF-Applied Part | Type BF-Applied Part | Same | No |
| Mode of Operation: | Continuous | Continuous | Same | No |
| | Subject Device | Predicate | Comparison | Does the difference raise<br>different questions of<br>safety and effectiveness? |
| Wireless Communication | | | | |
| Transmitter<br>Bluetooth<br>compliance: | BLE 4.0 | BLE 4.0 | Same | No |
| Transmitter<br>Operating Range: | 30 feet | 10 meters | Similar | No |
| Accuracy and Range | | | | |
| Oxygen Saturation<br>Range: | 1% to 100% | 1% to 100% | Same | No |
| Pulse Rate Range: | 30 to 300 bpm | 30 to 300 bpm | Same | No |
| SpO2 Accuracy: | Non-motion and motion<br>70-100% ±3.0% Arms | Non-motion<br>70-100% ± 3 digits | Same | No<br>Adding motion data |
| Pulse Rate Accuracy: | Non-motion<br>± 3 bpm Arms<br>Motion<br>± 5 bpm Arms | ± 3 digits | Same | No<br>Predicate not suitable for motion |
| Sensor Specifications | | | | |
| Measurement<br>Wavelength/Power<br>(Red): | 653 nm, <10 mW max. | 660 nm @ 9.8mW | Similar | No<br>Validated by clinical study |
| Measurement<br>Wavelength/Power<br>(Infrared): | 931 nm, <10 mW max | 880 nm @ 6.5mW | Similar | No<br>Validated by clinical study |
| Sensor Internal<br>Battery: | 3.8 V | 3.7 V | Similar | No |
| | Subject Device | Predicate | Comparison | Does the difference raise<br>different questions of<br>safety and effectiveness? |
| Sensor Battery Life: | 16 hrs | 22 hrs | Similar | No risks associated with 12 hour<br>sensor battery life. Base station<br>provides a low priority alarm<br>when sensor battery is getting<br>low. Base station gives a<br>medium priority alarm if the<br>sensor runs out of battery and<br>powers off. |
| Sensor Weight: | 14 grams (with Large sock) | 17.5 grams | Similar | No |
| Sensor Dimensions: | 2.2" L x 1.4" W x 0.5" H | 5.6" D x 1.2" W x 0.6" H | Similar | No…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.