K222549 · Purgo Biologics, Inc. · NPK · May 11, 2023 · Dental
Device Facts
Record ID
K222549
Device Name
OpenTex
Applicant
Purgo Biologics, Inc.
Product Code
NPK · Dental
Decision Date
May 11, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OpenTex is a temporarily implantable material (non-resorbable) for as a space-making barrier in the treatment of periodontal defects.
Device Story
OpenTex is a non-resorbable PTFE membrane used as a physical barrier in periodontal surgery. It prevents gingival cell invasion into the defect site, facilitating neovascularization and bone cell repopulation. The membrane is designed to be sufficiently stiff to maintain space for bone regeneration while remaining supple enough to conform to tissue contours. It is supplied sterile for single-use in various sizes. The device is intended for use by clinicians in dental surgical settings. By maintaining the necessary space and preventing soft tissue ingrowth, the membrane supports the repair of periodontal defects.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence was demonstrated through bench testing, including pH measurement, evaporation residue (solubility), tensile strength (ASTM D882), tear resistance (ISO 6383-1), and suture retention strength (ANSI/AAMI/ISO 7198). Biocompatibility was evaluated per ISO 10993 series (cytotoxicity, irritation, sensitization, genotoxicity, acute/subacute toxicity, implantation, and material-mediated pyrogenicity).
Technological Characteristics
Material: 100% Polytetrafluoroethylene (PTFE). Form: Membrane (plain or textured). Sterilization: Ethylene Oxide (ETO) to SAL 10^-6. Dimensions: Various sizes, 0.16 mm thickness. Mechanical properties verified via ASTM D882 and ISO 6383-1. Single-use, prescription-only device.
Indications for Use
Indicated for use as a space-making barrier in the treatment of periodontal defects in patients requiring temporary, non-resorbable implantable material.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Purgo Biologics Inc. Byungsun Kim RA Team Manager #812, 27 Dunchon-daero 457beon-gil, Jungwon-gu Seongnam-si, Gyeonggi-do 13219 Korea. South
5/11/23
### Re: K222549
Trade/Device Name: OpenTex Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPK Dated: April 10, 2023 Received: April 11, 2023
### Dear Byungsun Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image contains a logo for "Purgo Dental Biologics Solution". The logo features a gray circular icon with a smaller circle inside and a line extending upwards, resembling a stylized atom or molecule. To the right of the icon, the word "Purgo" is written in a bold, purple font. Further to the right, the words "Dental Biologics Solution" are stacked vertically, also in purple, but in a smaller font size.
# Indication for use
510(k) Number: K222549
Device Name: OpenTex
Indication for use:
OpenTex is a temporarily implantable material (non-resorbable) for as a space-making barrier in the treatment of periodontal defects.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _____________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Indication for use
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a bold, purple font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked on top of each other, also in purple.
# 510(k) Summary
05/10/2023
## 1. Company
| | Submitter |
|----------------|------------------------------------------------------------------------------------------|
| Name | Purgo Biologics Inc. |
| Address | #812, 27 Dunchon-daero 457beon-gil, Jungwon-gu, Seongnam-si,<br>Gyeonggi-do, Korea 13219 |
| Phone/Fax | Tel. +82-70-4827-0451, Fax. +82-70-8673-0660 |
| Contact person | Byungsun Kim / RA<br>kimbs@purgobio.com |
| Summary Date | 05/10/2023 |
## 2. Device Name
| Proprietary name | : | OpenTex |
|------------------------|---|------------------------------|
| Regulation number | : | 21 CFR 872.3930 |
| Regulation Description | : | Bone Grafting Material |
| Product code | : | NPK |
| Classification name | : | Barrier, Synthetic, Intraora |
| Device class | : | Class II |
| Classification Panel | : | Dental |
### 3. Predicate Device
| Primary predicate device | |
|--------------------------|-----------------------------------------------|
| K160493 | Salvin CytoSurg™ Non-Resorbable PTFE Membrane |
Reference device K964342 Cytoplast GBR
### 510(k) Summary
{5}------------------------------------------------
### 4. Indication for use
OpenTex is a temporarily implantable material (non-resorbable) for as a space-making barrier in the treatment of periodontal defects.
### 5. Description
OpenTex is a non-resorbable PTFE membrane composed of proprietary 100% polytetrafluoroethylene (PTFE) sheet with inert biological features and predictable barrier effect.
OpenTex is designed to function as a physical barrier to avoid gingival cell invasion, thus providing a favorable environment for neovascularization and bone derived cells to repopulate and repair the defect. Since space-making is critical to this procedure, the membrane is sufficiently stiff to prevent spontaneous collapse but supple enough to conform easily to tissue contours.
OpenTex is supplied sterile and intended for single use only. It is available in various sizes as shown below.
| Size | Thickness | Packaging |
|------------|-----------|------------|
| 12 x 20 mm | 0.16 mm | |
| 14 x 24 mm | | |
| 17 x 25 mm | | 1ea / box |
| 20 x 25 mm | | 5ea / box |
| 24 x 30 mm | | 10ea / box |
| 30 x 40 mm | | |
### 6. Performance Data
Non-clinical testing was performed in accordance with FDA recognized consensus standards and FDA guidance as applicable.
· Performance testing was performed to demonstrate substantial equivalence of the subject device to its primary predicate device as table below. The test results of the subject device met the criteria and was equal or higher than that of the predicate device.
### 510(k) Summary
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray circular icon with a smaller circle inside, followed by the word "Purgo" in a purple, bold font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked on top of each other in a smaller, purple font.
| Test item | Test method | Criteria | Results |
|------------------------------|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| pH | pH measurement test | Difference between the blank<br>and the extracts<br><br>< 1.5 (Internal)<br>* pH measured<br>- Blank: 6.23<br>- Subject device: 6.23<br>- Predicate device: 6.21 | The results of both devices met<br>the criteria. |
| Dissolution<br>/ Solubility | Evaporation residue test | Evaporation residue of the<br>extraction liquid<br>$\leq$ 1.0mg | The results of both devices met<br>the criteria. |
| Tensile strength | ASTM D882 | ≥ 34 MPa (Internal) | The result of the subject device<br>met the criteria and was higher<br>than that of predicate device. |
| Tear resistance | ISO 6383-1 | ≥ 1.5 kgf/mm (Internal) | The result of the subject device<br>met the criteria and was higher<br>than that of predicate device. |
| Suture retention<br>strength | ANSI/AAMI/ISO 7198 | Criteria for the subject device<br>was established as equivalent<br>to result of predicate device. | The result of the subject device<br>was higher than that of predicate<br>device. |
· Biocompatibility was evaluated in accordance with ISO 10993 series as followings.
- Biocompatibility risk assessment per ISO 10993-1
- Cytotoxicity per ISO 10993-5
- Irritation per ISO 10093-10
- Sensitization per ISO 10993-10
- Genotoxicity per ISO 10993-3
- Acute toxicity per ISO 10993-11
- Subacute toxicity per ISO 10993-11
- Implantation per ISO 10993-6
- Material mediated pyrogenicity per ISO 10993-11
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a purple sans-serif font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in purple.
· EO(Ethylene Oxide) gas sterilization process validation was performed accordance with ISO 11135 demonstrating a sterility assurance level (SAL) of 10-6.
No clinical data were included in this submission.
### 7. Technological Characteristics
The following comparison table of the technological characteristics of the subject device and the predicate device outlines and provides the substantial equivalency of the subject device and the predicate.
| | Subject device | Predicate device | | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Primary predicate<br>device | Reference device | Discussion |
| Device name | OpenTex | Salvin CytoSurg™<br>Non-Resorbable PTFE<br>Membrane | Cytoplast GBR | - |
| Manufacturer | Purgo Biologics Inc. | Salvin Dental<br>Specialties | Osteogenics<br>Biomedical, Inc. | - |
| 510(k) Number | New Device | K160493 | K964342 | - |
| Indication for<br>use | OpenTex is a temporarily<br>implantable material (non-<br>resorbable) for as a space-<br>making barrier in the<br>treatment of periodontal<br>defects. | The Salvin CytoSurg™<br>Non-Resorbable PTFE<br>Membrane is<br>temporarily implantable<br>material<br>(non-resorbable) for as<br>a space-making barrier<br>in the treatment of<br>periodontal defects. | A temporarily<br>implantable material<br>(non-resorbable) for use<br>as a space-making<br>barrier in the treatment<br>of periodontal defects. | Substantially equivalent |
| Materials | PTFE | PTFE | PTFE | Equivalent<br>The subject and predicate<br>devices are composed of<br>PTFE. |
| Form | Membrane<br>(Plain, Textured type) | Membrane<br>(Textured type) | Membrane<br>(Textured type) | Equivalent<br>The subject device is<br>provided in plain and<br>textured type while the<br>predicate devices are<br>provided in textured type.<br>Textured type is<br>manufactured by giving |
| | | | | PTFE sheet a textured<br>surface through the 'patten<br>input' process. Because this<br>process does not affect the<br>properties of PTFE sheet,<br>the property of plain and<br>textured type is same. The<br>equivalence of performance<br>to the predicate device was<br>verified trough the bench<br>testing.<br>Equivalent |
| Size | 12 x 20 mm, 14 x 24 mm<br>17 x 25 mm, 20 x 25 mm<br>24 x 30 mm, 30 x 40 mm | 12 x 24 mm, 25 x 30<br>mm | 12 x 24 mm, 25 x 30<br>mm | Equivalent<br>The subject and predicate<br>devices are provided in<br>various sizes for intra-oral<br>surgical procedures. |
| Thickness | 0.16 mm | 0.25 mm | 0.25 mm | Equivalent<br>The subject is thinner than<br>the predicate, but the<br>equivalence of performance<br>was verified trough the<br>bench testing. |
| Sterilization | Sterile<br>(ETO, SAL 10-6) | Sterile<br>(ETO, SAL 10-6) | Sterile<br>(ETO, SAL 10-6) | Identical |
| Shelf-life | 5 years | Unknown | 5 years | Equivalent<br>The shelf-life of the subject<br>device is equivalent to the<br>referemce device. |
| pH | 6.23 | 6.21 | Unknown | Equivalent |
| Dissolution<br>/ Solubility | 0 mg | 0 mg | Unknown | Equivalent<br>The performance test results<br>of the subject device met the<br>criteria and was equal or<br>higher than that of the<br>predicate device. |
| Tensile strength | 72.1524 MPa | 33.8444 MPa | Unknown | |
| Tear resistance | 5.9254 kgf/mm | 1.288 kgf/mm | Unknown | |
| Suture retention<br>strength | 1.083 MPa | 0.842 MPa | Unknown | |
| Use | Prescription | Prescription | Prescription | |
| Single Use Only | Yes | Yes | Yes | Identical |
| Duration of<br>implantation | Less than 30 days | Less than 30 days | Less than 30 days | Equivalent<br>The subject and predicate<br>devices are temporarily<br>implantable materials and<br>intended to be removed<br>within 30 days. |
### Comparison of Characteristics
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains a logo for a dental biologics solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a purple, bold font. To the right of "Purgo", the words "Dental Biologics Solution" are stacked vertically, also in purple.
The subject device is substantially equivalent to the primary predicate device K160493 in indication for use, material, design(form), sterilization and duration of implantation. Both devices are provided
### 510(k) Summary
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a purple sans-serif font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in purple.
in various sizes for intra-oral surgical procedures. The subject device is thinner than the predicate devices, but the equivalence of performance was verified through tensile strength and tear resistance tests.
## 8. Conclusion
Based on the information provided, the subject device is substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.