K222511 · Southmedic, Inc. · CCK · Nov 21, 2023 · Anesthesiology
Device Facts
Record ID
K222511
Device Name
Oxy2Pro
Applicant
Southmedic, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Nov 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Indications for Use
The Oxy2Pro is a single patient, disposable procedural mask with access for the insertion of oral scopes, probes, or tubes, delivery of supplemental oxygen and monitoring breathing by providing a means to sample exhaled CO2. It is for non-intubated, adult patients who are breathing spontaneously.
Device Story
Oxy2Pro is a single-patient, disposable procedural mask for non-intubated, spontaneously breathing adults. Device features a flexible, slitted membrane allowing oral access for scopes, probes, or tubes while maintaining mask functionality. Oxygen is delivered via a rotatable elbow and diffuser sleeve; exhaled CO2 is sampled through the same diffuser. Used in hospitals, clinics, and pre-hospital settings by clinicians. Device connects to standard oxygen supplies and external FDA-cleared capnographs. By providing supplemental oxygen and CO2 monitoring during procedures, it supports patient respiratory stability and enables clinician monitoring of breathing status.
Clinical Evidence
Bench testing only. Comparative performance testing against the predicate device (K172365) evaluated FiO2 and EtCO2 sampling accuracy at oxygen flow rates from 5 to 15 L/min under simulated conditions (RR 12, Tidal Volume 500mL, I:E 1:1). Results demonstrated Oxy2Pro performance was not statistically significantly less than the predicate.
Technological Characteristics
Materials: Clear polyvinylchloride (mask), acrylic (elbow/sleeve), polypropylene (insert). Sensing: Gas sampling via male/female luer connection to external capnograph. Oxygen delivery: 5-15 L/min via diffuser sleeve. Form factor: Over-nose/mouth mask with slitted oral access membrane. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 18562-2, ISO 18562-3 compliant.
Indications for Use
Indicated for non-intubated, adult patients who are breathing spontaneously, requiring supplemental oxygen delivery and exhaled CO2 monitoring during procedures involving oral scope, probe, or tube insertion.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2023
Southmedic Inc. Tish Whitehead Vice President of Regulatory Affairs & New Product Development 50 Alliance Blvd. Barrie, Ontario L4M 5K3 Canada
Re: K222511
Trade/Device Name: Oxy2Pro Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: September 29, 2023 Received: September 29, 2023
Dear Tish Whitehead:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{3}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222511
Device Name Oxy2Pro
Indications for Use (Describe)
Oxy2Pro is a single patient, disposable procedural mask with access for the insertion of oral scopes, probes, delivery of supplemental oxygen and monitoring breathing by providing a means to sample exhaled CO2. It is for nonintubated, adult patients who are breathing spontaneously.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Southmedic. The text "Southmedic" is written in a dark blue sans-serif font. To the right of the text is a blue heartbeat line with an arrow at the end, and the number 1983 is written below the line.
## 510(k) Summary
| Company: | Southmedic Inc. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 50 Alliance Blvd. |
| | Barrie, Ontario, Canada |
| | L4M 5K3 |
| Official Contact: | Tish Whitehead |
| | P: (705) 720-1902 |
| | F: (705) 728-6537 |
| Date Prepared: | 11/20/2023 |
| Trade Name: | Oxy2Pro |
| Common Name: | Gas sampling oxygen mask |
| Classification: | Product Code: CCK |
| | Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase |
| | Class II per 21 CFR 868.1400 |
| | Anesthesiology |
| Predicate Device: | Panoramic Oxygen Mask (POM) – K172365 |
| | POM Medical, LLC |
| Device Description: | The Southmedic Oxy2Pro is designed to have a flexible, thin membrane that is intended to be breached when requiring oral access for scope entry. This slitted membrane is intended to allow the mask to return to a closed- mask functionality after having been used with a scope. This device is to be used in conjunction with FDA cleared capnographs. |
| Intended Use: | The Oxy2Pro is a single patient, disposable procedural mask with access for the insertion of oral scopes, probes, or tubes, delivery of supplemental oxygen and monitoring breathing by providing a means to sample exhaled CO2. It is for non-intubated, adult patients who are breathing spontaneously. |
Image /page/4/Picture/3 description: The image shows a graphic with the text "Our Passion" at the top. Below that is a stylized image of the earth. To the right of the earth image is the text "Innovative Manufacturing & Distribution, Profitably". Below the earth image is the text "Exceptional Customer Service & Quality".
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 www.southmedic.com ISO 13485
Image /page/4/Picture/5 description: The image shows the word "Deloitte." in a bold, sans-serif font. The word is black, and there is a small gray dot to the right of the word. The background is white.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Southmedic" in a bold, sans-serif font. To the right of the word is a teal-colored graphic that resembles a heart rate monitor. The year "1983" is printed in a small font below the heart rate monitor graphic.
## Predicate Device Comparison:
## Table 1: Predicate Device Comparison Table
| Description | Subject Device | Predicate Device (K172365) | Comments |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Oxy2Pro is a single patient,<br>disposable procedural mask with<br>access for the insertion of oral scopes,<br>probes, or tubes, delivery of<br>supplemental oxygen and monitoring<br>breathing by providing a means to<br>sample exhaled CO2. It is for non-<br>intubated, adult patients who are<br>breathing spontaneously. | The Panoramic Oxygen Mask (POM)<br>is a single patient, disposable device<br>intended for delivering supplemental<br>oxygen and monitoring expired gases<br>from the patient, with ports to allow<br>the clinician to insert scopes, probes,<br>or tubes. It is for non-intubated,<br>spontaneously breathing patients<br>greater than 30kg. | Same |
| Patient<br>Population | Non-intubated, adult patients who are<br>breathing spontaneously. | Non-intubated spontaneously<br>breathing patients.<br>Greater than 30 kg | Similar; the predicate device is<br>available in adult and pediatric sizes.<br>This comparison is only made to the<br>adult size. |
| Environments of<br>Use | Locations where procedures are<br>performed where the patient requires<br>supplemental oxygen, monitoring of<br>exhaled gases and scope access.<br><br>Hospital, sub-acute, clinic, physician<br>offices, pre-hospital | Locations where procedures are<br>performed where the patient requires<br>supplemental oxygen, monitoring of<br>exhaled gases and scope access.<br><br>Hospital, sub-acute, clinic, physician<br>offices, pre-hospital | Same |
| Duration of Use | Single patient, disposable | Single patient, disposable | Same |
| Prescriptive | Yes | Yes | Same |
| Mode of<br>Operation | Oxygen delivery through standard<br>oxygen supply tubing and<br>simultaneous exhaled gas monitoring<br>via a gas sampling line connected<br>from mask to capnography or oxygen<br>only delivery. Slit access membrane | Oxygen delivery through standard<br>oxygen supply tubing and<br>simultaneous exhaled gas monitoring<br>via a gas sampling line connected<br>from mask to capnography or oxygen | Similar; Oxy2Pro does not feature a<br>universal size inlet but it achieves the<br>same performance through the diffuser<br>sleeve and insert without the need for<br>additional adaptors. |
| Description | Subject Device | Predicate Device (K172365) | Comments |
| | allowing for introduction of a scope.<br>Oxygen and sampling tube are<br>connected to mask via a rotatable<br>elbow positioned in front of the face. | only delivery. Slit access ports<br>allowing for introduction of a scope.<br>Universal size inlet allows for removal<br>or insertion of high or medium<br>concentration adaptor, located to the<br>side of the front of the mask. The<br>sampling line is connected on the<br>opposite side. | |
| Components<br>which may be<br>supplied or<br>packaged with<br>the mask | Standard oxygen tubing (connected to<br>mask) with universal connector<br>Gas sampling line | Standard oxygen tubing with universal<br>connector<br>Gas sampling line<br>Oxygen rebreather bag | Similar; Oxy2Pro does not feature a<br>universal size inlet with a rebreather<br>bag but it achieves the same<br>performance through the diffuser<br>sleeve and insert without the need for<br>additional adaptors. |
| Gas sampling<br>connections | Male Luer and Female Luer | Luer slip fit | Similar; Southmedic shall offer both a<br>female and male luer sampling line to<br>accommodate different capnography<br>equipment. |
| Sizes | Adult | Adult | Oxy2Pro is only available in adult<br>size; comparison only made to adult<br>predicate. |
| Description | Subject Device | Predicate Device (K172365) | Comments |
| Profile | Over the nose/mouth | Over the nose/mouth; aluminum nose strip | Similar; Oxy₂Pro does not feature an aluminum nose strip but instead is shaped to sit on the nose. This difference does not impact performance. |
| Face Strap | Yes | Yes | Same |
| Internal Volume | Adult – 230 ml | Adult – 198 ml | Similar; Oxy₂Pro has a somewhat larger volume but it is not expected to affect the intended use. |
| Access Membranes/ Ports and Maximum Instrument size | Oral Access Membrane – 58 mm x 42 mm | Oral port – 60 mm ID / 20 mm OD<br>Nasal – 36 mm ID / 12 mm OD | Similar; Oxy₂Pro contains an oral access membrane to allow for the introduction of instruments while delivering oxygen and sampling expired gases. The absence of a nasal port does not impact the intended use. |
| Vents | Open ports allow room entrainment to prevent rebreathing | One-way valves to prevent rebreathing | Similar; the difference in methodology does not affect performance or intended use |
| Oxygen Flow Rates | Flow rates from 5 liters per minute to 15 liters per minute | Medium Concentration: 6-8 liters<br>High Concentration: 10-15 liters | Same; devices offer the same flow rate range |
| % FiO2 | Testing was completed under simulated conditions at various flow rates | Testing was completed under simulated conditions at various flow rates | Same; Oxy₂Pro was not significantly less than the predicate. |
| Description | Subject Device | Predicate Device (K172365) | Comments |
| % CO2<br>Accuracy | Testing was completed under<br>simulated conditions with true<br>baseline EtCO2 at 5%, at various flow<br>rates. | Testing was completed under<br>simulated conditions with true<br>baseline EtCO2 at 5%, at various flow<br>rates. | Same; Oxy2Pro was not significantly<br>less than the predicate. |
| Materials | Mask: Clear Polyvinylchloride<br>Elbow and Sleeve: Acrylic<br>Insert: Polypropylene | Mask: Clear Polyvinylchloride | Similar |
| Biocompatibility | Surface contacting for a limited<br>duration (<24 hours). Gas pathway<br>contacting for a limited duration (<24<br>hours).<br><br>Testing includes:<br>Cytotoxicity (ISO 10993-5:2009)<br>Sensitization (ISO 10993-10:2010<br>Intracutaneous/Irritation (ISO 10993-<br>10:2010)<br>Emissions of particulate matter (ISO<br>18562-2:2017)<br>Emissions of volatile compounds<br>(VOCs) (ISO 18562-3:2017) | Externally communicating: Tissue<br>contact, surface communicating, and<br>skin contact for a limited duration of<br>use (<24 hours).<br><br>Testing included:<br>Cytotoxicity (ISO 10993-5:2009)<br>Sensitization (ISO 10993-10:2010<br>Intracutaneous/Irritation (ISO 10993-<br>10:2010)<br>Acute Systemic Toxicity (ISO 10993-<br>11:2017) | Similar; devices have a similar<br>biological categorization. Differences<br>in testing endpoints may be attributed<br>to the application of ISO 10993-1 and<br>ISO 18562. |
| Storage | Standard warehouse conditions | -20° to +50°C | Similar |
| Device<br>Classification | Class II<br>Product code - CCK | Class II<br>Product code - CCK | Same |
Our Passion
Image /page/5/Picture/5 description: The image shows a logo with a globe and text. The globe is a light blue color and has a wireframe design. To the right of the globe, the text reads "Innovative Manufacturing & Distribution, Profitably". Below the globe, the text reads "Exceptional Customer Service & Quality".
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 www.southmedic.com ISO 13485
Image /page/5/Picture/8 description: The image is a logo for "Canada's Best Managed Companies". To the left of the text is a grey maple leaf. Below the text "Best Managed Companies" is the text "Platinum member".
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Southmedic. The logo has the word "Southmedic" in a dark blue sans-serif font. To the right of the word is a teal-colored line that resembles a heart rate monitor, and the year "1983" is printed in a small font below the line.
Our Passion
Image /page/6/Picture/3 description: The image shows a light blue globe with the continents outlined. To the right of the globe, the text "Innovative Manufacturing & Distribution, Profitably" is displayed. Below the globe, the text "Exceptional Customer Service & Quality" is written in a curved manner.
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 www.southmedic.com ISO 13485
Image /page/6/Picture/6 description: The image is a logo for "Canada's Best Managed Companies". The logo features a stylized maple leaf on the left, with the text "CANADA'S" vertically aligned along the leaf's edge. To the right of the leaf, the words "BEST MANAGED COMPANIES" are stacked in bold, sans-serif font. Below this, the words "Platinum member" are written in a smaller, regular font.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Southmedic. The text "Southmedic" is in a dark blue sans-serif font. To the right of the text is a teal-colored heartbeat line that ends in an arrow, with the number "1983" below the line.
Our Passion
Image /page/7/Picture/3 description: The image shows a logo with a world map in light blue. The text "Innovative Manufacturing & Distribution, Profitably" is on the right side of the map. The text "Exceptional Customer Service & Quality" is at the bottom of the map.
Image /page/7/Picture/6 description: The image shows a logo for "Canada's Best Managed Companies". To the left of the text is a grey maple leaf. Below the text "Best Managed Companies" is the text "Platinum member".
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Southmedic. The text "Southmedic" is in a dark blue sans-serif font. To the right of the text is a teal-colored heartbeat line that ends with an arrow and the number 1983.
Our Passion
Image /page/8/Picture/3 description: The image shows a world map with the text "Innovative Manufacturing & Distribution, Profitably" on the right side of the map. The map is light blue and has a grid of lines over it. Below the map, the text "Exceptional Customer Service & Quality" is written in a smaller font.
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 www.southmedic.com ISO 13485
Image /page/8/Picture/6 description: The image is a logo for "Canada's Best Managed Companies Platinum Member". On the left side of the logo is a gray maple leaf, which is a symbol of Canada. To the right of the maple leaf is the text "BEST MANAGED COMPANIES" in bold, black letters. Below the text is the phrase "Platinum member" in a smaller, non-bold font.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Southmedic. The logo consists of the word "Southmedic" in a dark blue sans-serif font. To the right of the word is a teal-colored heartbeat line that ends in an arrow. Below the arrow is the number 1983.
## Substantial Equivalence Discussion:
The Oxy2Pro Mask is considered substantially equivalent to the predicate Panoramic Oxygen Mask (POM), by POM Medical (510k holder, K172365) based on the following discussion. In making this determination, Southmedic followed the FDA's Guidance entitled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" (Document issued on: July 28, 2014) and the detailed flowchart provided in Appendix A. 510(k) Decision-Making Flowchart.
Intended Use: The Oxy2Pro is a single patient, disposable procedural mask with access for the insertion of oral scopes, probes, or tubes, delivery of supplemental oxygen and monitoring breathing by providing a means to sample exhaled CO2. It is for non-intubated, adult patients who are breathing spontaneously. Both devices provide supplemental oxygen and monitoring of exhaled gases, with oral scope access and are to be used in similar environments of use.
The predicate is available in pediatric and adult sizes and features an additional nasal access port. The adult version of the predicate was used for comparison testing. Regardless of model, both sizes have the same intended use. The nasal access port in the predicate facilitates nasal scope/probe entry. This may lead to additional indications provided by the predicate, however, the intended use and indications per the predicate labeling does not differentiate between the nasal or oral option. Per the labeling, the Oxy2Pro and predicate device have the same intended use and the same environments of use when compared within the same indication. The difference of the predicate device having a nasal access port should not affect oxygen delivery and sampling performance.
Technology: The devices have similar sampling line luer and oxygen tubing technology, and feature oral access ports/membranes for scope entry. The oxygen tubing is connected to the oxygen source using a universal connector, and the sampling line is connected to the capnography equipment. The Oxy2Pro mask is made of a clear polyvinylchloride material which is the same material as the predicate.
Oxy2Pro's oxygen and sampling lines are connected to the mask via a rotatable elbow positioned at the front of the face. Oxygen is delivered through a diffuser sleeve and insert that directs the oxygen towards the nose and mouth. Carbon dioxide is sampled through this same diffuser.
The predicate device features a universal size inlet to allow for the removal or insertion of a high or medium concentration adaptor, which is located to the front of the mask. Oxygen is delivered through this adaptor. The sampling line is connected to the opposite side of the mask.
Image /page/9/Picture/8 description: The image shows a graphic with the text "Our Passion" at the top and a stylized globe in the center. The globe is light blue with white lines indicating continents. To the right of the globe, the text "Innovative Manufacturing & Distribution, Profitably" is displayed. At the bottom, the text "Exceptional Customer Service & Quality" is curved along the bottom edge of the graphic.
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 www.southmedic.com ISO 13485
**Deloitte.**
BEST
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Southmedic. The word "Southmedic" is written in a dark blue sans-serif font. To the right of the word is a teal-colored graphic that resembles a heart rate monitor. The year "1983" is written in small font to the right of the graphic.
The safety and performance of the Oxy2Pro and predicate device are similar, as shown in the testing summary below. The differences in technology do not affect device performance and safety with respect to oxygen delivery and EtCO2 sampling.
## Testing summary:
## Biocompatibilitv and Materials:
The Oxy>Pro and predicate have similar biocompatibility categorizations. The Oxy>Pro has been assessed as a limited contact duration mask against the following requirements and endpoints: ISO 18562-2 Emissions of particulate matter: ISO 18562-3 Emissions of volatile compounds (VOCs); ISO 10993-5 Cytotoxicity; ISO 10993-10 Sensitization; ISO 10993-10:2010 Intracutaneous/Irritation. Differences in testing endpoints are attributed to the application of ISO 10993-1 and ISO 18562. The primary material for both devices is clear polyvinylchloride.
## Performance Testing:
Accredited third-party testing was completed to evaluate the ability to deliver FiO2 and sample EtCO2 relative to the predicate device.
| | Oxy2Pro | | Predicate (K172365) | |
|------------------|---------|-------|---------------------|-------|
| Oxygen Flow Rate | FiO2 | ETCO2 | FiO2 | ETCO2 |
| 5 l/min | 53.3% | 6.13% | N/A | N/A |
| 8 l/min | 64.0% | 4.87% | 52.3% | 4.03% |
| 10 l/min | 64.7% | 4.70% | 48.3% | 4.17% |
| 12 l/min | 67.3% | 3.77% | 54.7% | 3.30% |
| 15 l/min | 72.3% | 3.70% | 52.0% | 3.53% |
Test conditions: Respiratory Rate: 12, Tidal Volume: 500, Inspiratory:Expiratory Ratio: 1:1, Baseline true EtCO2 5%. Acronyms: I/min is liters per minute, FiO2 is fraction of inspired oxygen, EtCO2 is end tidal carbon dioxide
The results demonstrated that the Oxy2Pro mask was not statistically significantly less than the predicate for FiO2 or EtCO2 at the tested flow rates, meeting the acceptance criteria.
In conclusion, substantial equivalence is demonstrated between the Oxy2Pro mask and the predicate device. The differences observed do not impact performance and safety per the intended use of supplemental oxygen delivery and EtCO2 sampling.
Image /page/10/Picture/11 description: The image shows a logo with a globe and text. The text at the top reads "Our Passion". The text on the right side of the globe reads "Innovative Manufacturing & Distribution, Profitably". The text at the bottom of the globe reads "Exceptional Customer Service & Quality".
Southmedic Incorporated Head Office: 50 Alliance Blvd., Barrie, Ontario, Canada L4M 5K3 1-705-726-9383 1-800-463-7146 Fax: 1-705-728-9537 ISO 13485 www.southmedic.com
Image /page/10/Picture/13 description: The image shows the word "Deloitte." in a bold, sans-serif font. The word is black, and there is a small gray circle to the right of the word. The background is white.
Platinum member
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.