K222481 · 3-D Matrix, Inc. · QAU · Feb 28, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222481
Device Name
PuraStat
Applicant
3-D Matrix, Inc.
Product Code
QAU · General, Plastic Surgery
Decision Date
Feb 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4456
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222481 · Feb 28, 2023
PuraStat
3-D Matrix, Inc.
Published clinical literature (Branchi et al., 2021); Multi-center observational clinical study data
The sponsor utilized a prospective, multi-center observational study from the literature to support the safety and effectiveness of PuraStat for the additional indication of primary non-variceal gastrointestinal bleeding.
Non-variceal GI bleeding; Observational study; Hemostasis; Prospective cohort
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
PuraStat in gastrointestinal bleeding: results of a prospective multicentre observational pilot study; Prospective, consecutive open-label, multi-center observational study; Follow-up/Duration: 30 days post procedure; Study Period: July 2017 - December 2018
111 patients with acute upper or lower non-variceal gastrointestinal bleeding; Sample Size: 111; Number of Sites: 15
Not applicable for this study
Primary: achievement of hemostasis during the procedure. Secondary: prevention of rebleeding at 3 and 7 days, risk and side effect profiles.
Indications for Use
PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.
Device Story
Sterile, synthetic peptide hydrogel (RADA-16) in water; provided in prefilled cyclo-olefin polymer syringes. Delivered endoscopically via catheter to bleeding site. Upon contact with physiological fluids (blood), peptides self-assemble into transparent scaffold/mechanical barrier; blocks blood flow without volume expansion. Used in clinical endoscopy settings by physicians. Acts as primary, adjunct, bridge, or rescue hemostat. Benefits include stabilization of bleeding sites, reduction of rebleeding risk, and potential avoidance of surgery.
Clinical Evidence
Prospective, multicenter, open-label study (Branchi et al., 2021) of 111 patients (age >18) with acute non-variceal GI bleeding. Primary endpoint: procedural hemostasis success. Results: 94% initial hemostatic success for primary application; 75% success as secondary therapy. 30-day rebleeding rate 16%. No adverse events reported. Bench and biocompatibility data referenced from predicate.
Technological Characteristics
Synthetic 16-amino acid (RADA-16) oligopeptide in sterile water. Cyclo-olefin polymer syringe, HDPE plunger, butyl rubber cap/gasket. Self-assembling hydrogel scaffold. Delivered endoscopically via polypropylene adapter and catheter. Terminally sterilized. Non-animal/non-plant derived. No preservatives.
Indications for Use
Indicated for hemostasis of mild/moderate bleeding post ESD/EMR, intraprocedural venous bleeding, and primary non-variceal GI bleeding in adult patients (>18 years). Contraindicated for arterial Forrest 1a bleeding.
Regulatory Classification
Identification
A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2023
3-D Matrix, Inc. % Stephen Rhodes Principal Streamline Regulatory 3502 Dundee Driveway Chevy Chase, Maryland 20815
Re: K222481
Trade/Device Name: PuraStat Regulation Number: 21 CFR 878.4456 Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use Regulatory Class: Class II Product Code: QAU Dated: February 15, 2023 Received: February 15, 2023
Dear Stephen Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tek N. Tek N. l amichhane -S Date: 2023.02.27 22:27:19 Lamichhane -S for Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222481
Device Name PuraStat
### Indications for Use (Describe)
PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary is provided below.
#### SUBMITTER 1.
3-D Matrix. Inc. 1234 Chestnut St., Suite 205 Newton, MA 02464
Contact Person: Lisa Spirio, PhD Phone: 617-875-6204 Email: lisa@3dmatrix.com
Date Prepared: February 28, 2023
#### 2. DEVICE
Name of Device: PuraStat Common Name: Hemostatic device for intraluminal gastrointestinal use Classification Regulation: 21 CFR 878.4456 Regulatory Class: II Product Code: OAU Panel: General & Plastic Surgery
#### PREDICATE DEVICE 3.
Predicate Device: 3-D Matrix, Inc.'s PuraStat-GI (K210098)
#### DEVICE DESCRIPTION 4.
PuraStat is a sterile gel composed of a synthetic peptide and sterile water for injection. It is provided as a prefilled syringe (2.5% peptide content) ready for use as a hemostat. The gel is delivered to the intended application site(s) via a commercially available endoscopic catheter that is attached to the gel syringe via the polypropylene adapter.
PuraStat is completely non-animal and non-plant derived and contains no preservatives that might present a risk of allergic reaction or skin irritation.
Exposure to physiological fluids such as blood causes the peptide solution to quickly form a transparent gel without expansion in volume. PuraStat achieves hemostatic effects by forming a hydrogel matrix barrier that blocks the flow of blood at the site of application.
#### INDICATIONS FOR USE ട്.
PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic
{4}------------------------------------------------
therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.
The subject device is the identical product as the PuraStat-GI predicate (K210098).
PuraStat is comprised of a synthetic repeating 16-amino acid (RADA-16) oligopeptide in sterile water for injection. In the presence of a bleeding wound, the peptides assemble into a scaffold structure that forms a mechanical barrier over the bleeding site. The hemostatic solution is sterilefiltered and filled into 5-ml syringes made of cyclo-olefin polymers with a high-density polyethylene plunger and a butyl rubber head cap and gasket. Each syringe is filled with either 1, 3, or 5 ml of gel. The device is terminally sterilized, and the resorbable gel is delivered to the intended application site(s) via a polypropylene adapter connect to a commercially available catheter. The device is delivered endoscopically via a catheter to the bleeding site.
#### 7. PERFORMANCE DATA
The subject device is identical to the predicate device. The intended use of the subject device is identical to the intended use of the predicate device. The indications for use of the subject device are identical to the predicate's indications for use, with the exception that the subject device has the following statement not found in the predicate device: "... and for primary non-variceal gastrointestinal (GI) bleeding."
Because the subject device is identical to the predicate device and has the same intended use, the bench testing, animal testing and biocompatibility testing provided in the predicate device (K210098) was referenced for the subject device.
The additional indication for primary non-variceal gastrointestinal (GI) bleeding for the subject device was addressed by clinical data in an OUS study reported in the literature.
## Branchi et al Study (2021) *
The safety and effectiveness of the use of PuraStat as a primary hemostat to control primary bleeding in patients with acute non-variceal gastrointestinal bleeding (upper and lower) was evaluated in 111 patients in a prospective, consecutive open-label, multi-center study between July 2017 and December 2018 at 15 endoscopy departments in Germany. The effectiveness of PuraStat was assessed during the procedure, at 3 days and 1 week after application.
The primary endpoint was the achievement of hemostasis during the procedure with PuraStat (procedure success). Secondary end points were the prevention of rebleeding (defined as therapy success), the documentation of risk and side effect profiles of PuraStat. Rebleeding was defined as the presence of clinical signs of gastrointestinal bleeding (hematemesis, melena, hematochezia) in association with cardiovascular instability or Hemoglobin drop. A stable clinical condition without signs of rebleeding assessed at 3 and 7 days after the application of PuraStat was defined as therapy success.
Inclusion criteria were age > 18 years, acute upper or lower gastrointestinal bleeding including active bleeding and lesions with signs of recent hemorrhage such as visible vessels (Forrest Classification IIA or adherent clots or pigmented spots (Forrest Classification IIB) according to the Forrest classification. Adverse events were recorded for 30 days post procedure.
{5}------------------------------------------------
## i. Demographics and Patient Accounting
A total of 111 patients with gastrointestinal bleeding were included and treated with PuraStat; 68 were men and 43 were women. The bleeding activity was in most cases classified as oozing bleeding (76/111, 69%), in seven cases spurting hemorrhage (6%), in 16 cases a visible vessel (14%) and in 6 cases each (5%) an adherent clot or hematin on the ground of a lesion were observed. In all seven patients with spurting bleeding, PuraStat was applied as a secondary therapy to standard techniques, thus utilizing a combination of submucosal injection, hemoclips, over-thescope clips (OTSC) and gel to reach final hemostasis. After endoscopy therapy, none of these patients required surgery which was an overwhelming positive outcome for these patients.
No patients were lost to follow-up.
After primary application of PuraStat, initial hemostatic success was achieved in 94% of patients (74/79, 95% CI 88-99%), and in 75% of the patients when used as a secondary hemostatic product, following failure of established techniques (24/32, 95% CI 59-91%). The therapeutic success rates (absence of rebleeding) after 3 and 7 days were 91% and 87% after primary use, and 87% and 81% in all study patients. Overall rebleeding rate at 30-day follow-up was 16% (18/11). In the 5 patients who finally required surgery (4.5%), PuraStat allowed temporary hemostasis and stabilization.
No adverse events due to application of PuraStat or technical failures were reported.
The largest proportion of patients in the study were patients with active oozing bleeding, in whom typically a rebleeding rate of up to 50% is reported in the literature: however, in this study we found a rebleeding rate of 12% at 7 days and of 16% at 30 days in our whole patients' collective. After primary application of PuraStat, initial hemostatic success was achieved in 94% of patients (74/79, 95% CI 88-99%). The therapeutic success rates (absence of rebleeding) after 3 and 7 days were 91% and 87% after primary use. PuraStat was safely applied and administered without complications as a primary or secondary therapy.
* Branchi et al. PuraStat in gastrointestinal bleeding: results of a prospective multicentre observational pilot study. Surgical Endoscopy. 2021 June 15: 9 Feb; 7(1): 155–162.
#### 8. CONCLUSIONS
The intended use of the subject device is identical to the predicate device's intended use. The indications for use for the subject PuraStat are modifications of the predicate device, with the difference being that the subject device has the additional indication of primary non-variceal gastrointestinal (GI) bleeding and the additional clarification that PuraStat is not indicated for arterial Forrest 1a bleeding. The addition of non-variceal primary bleeding does not change the intended use of hemostasis. Thus, PuraStat does not have a new intended use. The subject device is the identical product as the PuraStat-GI predicate (K210098).
PuraStat complies with the special controls for a hemostatic device for intraluminal gastrointestinal internal use.
In conclusion. PuraStat is substantially equivalent to the predicate PuraStat-GI (K210098).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.