TIB Abutments

K222469 · Southern Implants (Pty), Ltd. · NHA · Apr 17, 2023 · Dental

Device Facts

Record IDK222469
Device NameTIB Abutments
ApplicantSouthern Implants (Pty), Ltd.
Product CodeNHA · Dental
Decision DateApr 17, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TIB Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The TIB abutments consist of two major parts. Specifically, the titanium base and mesostructure components make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: Scan files from desktop scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The intended use for the TiB narrow abutments used with the Ø3.0 mm External hex and Ø3.0 mm Deep Conical implants is limited to replacement of maxillary lateral incisors and mandibular lateral and central incisors.

Device Story

TIB Abutments are two-piece prosthetic components; titanium base interfaces with endosseous dental implant; zirconia mesostructure restoration bonded to base. System integrates digital dentistry workflow: desktop scanners, CAD/CAM software, milling machines. Used in dental clinics by clinicians. Workflow: scan patient anatomy; design custom restoration using 3Shape Abutment Designer; mill zirconia restoration; bond to titanium base; secure to implant via screw. Provides custom-fit dental restoration to restore chewing function. Narrow versions restricted to specific incisor replacement; straight placement only for Ø3.0mm Deep Conical; occlusal forces must align with central longitudinal axis.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing: static and dynamic compression-bending fatigue testing per ISO 14801 on worst-case constructs, validated sterilization instructions (ISO 17665), MR safety assessment, and software validation.

Technological Characteristics

Two-piece abutment; Titanium Grade 5 alloy (ASTM F136) base and screw; Zirconia mesostructure. Machined and anodized surface. Non-sterile. CAD/CAM workflow integration. Mechanical fatigue testing per ISO 14801.

Indications for Use

Indicated for prosthetic rehabilitation via endosseous dental implants. Narrow TIB abutments (Ø3.0 mm External hex and Ø3.0 mm Deep Conical) are limited to replacement of maxillary lateral incisors and mandibular lateral and central incisors.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Southern Implants (Pty) Ltd Leith Cumming Acting Head of Regulatory Affairs and Quality 1 Albert Road Irene, Gauteng 0062 SOUTH AFRICA Re: K222469 Trade/Device Name: TIB Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA. PNP Dated: March 14, 2023 Received: March 14, 2023 Dear Leith Cumming: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Andrew I. Steen -S Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K222469 Device Name TIB Abutments #### Indications for Use (Describe) The TIB Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The TIB abutments consist of two major parts. Specifically, the titanium base and mesostructure components make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: Scan files from desktop scanners, CAM software, ceramic material, milling machine and associated tooling and accessories. The intended use for the TiB narrow abutnents used with the Ø3.0 mm External hex and Ø3.0 mm Deep Conical implants is limited to replacement of maxillary lateral incisors and mandibular lateral and central incisors. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ ## 510(k) Summary ## TIB Abutment System ## Southern Implants (Pty) Ltd April 17, 2023 #### ADMINISTRATIVE INFORMATION Manufacturer Name Official Contact Southern Implants (Pty) Ltd 1 Albert Road Irene, Gauteng, 0062 South Africa Telephone +27 12 667 1046 Fax +27 12 667 1029 Leith C. Cumming Acting Head of Regulatory Affairs and Quality Email: leith.c@southernimplants.com ## DEVICE NAME AND CLASSIFICATION Trade/Proprietary name Common name Classification name Classification regulation Product Code Classification Panel Reviewing Branch TIB Abutment System Dental Abutment Endosseous Dental Implant Abutment 21 CFR 872.3630, Class II NHA, PNP Dental Products Panel Dental Devices Branch ## PREDICATE DEVICE INFORMATION The primary predicate device for the subject device Narrow TIB Abutments is K193084. The reference devices are K163060. K173706 and K220841. {4}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT The TIB Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The TIB abutments consist of two major parts. Specifically, the titanium base and mesostructure components make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: Scan files from desktop scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The intended use for the TiB narrow abutments used with the Ø3.0 mm External hex and Ø3.0 mm Deep Conical implants is limited to replacement of maxillary lateral incisors and mandibular lateral and central incisors. #### SUBJECT DEVICE DESCRIPTION The Narrow TIB Abutment bases are standard premanufactured titanium allov abutments for supporting a dental restoration and mesostructure. The dental laboratory is to fabricate the mesostructure restoration by CAD/CAM technique out of zirconia. The TIB abutment base then serves as the interface between the endosseous implant and the Zirconia restoration. The TIB Abutment Base is designed to support the restoration on an endosseous implant in order to restore chewing function for the patient. The DC Ø3.0mm Narrow TiB abutments are only to be placed straight in the patient. Occlusal loading forces are only to be applied through the central longitudinal axis of the implant body. The mesostructured restoration is a CAD/CAM designed prosthesis milled out of Zirconia, which is designed to fit the abutment base in order to restore chewing for the patient. Each restoration is custom designed using 3Shape Abutment Designer Software in order to meet the requirements of each patient on a case-by-case basis. Limitations have been put in place in 3Shape Abutment Designer in order to prevent malfunctioning of the restoration. The subject Narrow TIB Abutments are compatible with the Southern Implants' Deep Conical and External Hex implants and screws. The subject TIB abutment bases are manufactured from Titanium alloy conforming to ASTM F136 and are color coded by gold anodizing. The anodization process is the same as used for previously cleared anodized titanium alloy devices in K193084. The mesostructured restoration is to be manufactured from Zirconia - Sage Max NexxZr which has been previously cleared for use in K130991. The digital workflow includes the following products (not subject devices to this submission): - Ceramic material: Sage Max NexxZr Zirconia Restorative material (K130991) - . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436) - Intra-oral scanner ● - . Lab scanner: 3Shape E3 Desktop Scanner (3Shape A/S) - Abutment design software: 3Shape Abutment Designer Software (K151455) ● - Milling machine: Roland DXW51D ● {5}------------------------------------------------ #### PERFORMANCE DATA Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K193084); static and dynamic compression-bending to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants, validated sterilization instructions per ISO 17665-1 and ISO 17665-2; MR safety testing as per the recommendations of the FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", software validation testing per the FDA Guidance Document for Off-The-Shelf Software Use in Medical Devices (referenced from K193084); and scanning and milling validation (referenced from K193084). No clinical data was included in this submission. #### EQUIVALENCE TO MARKETED DEVICE Southern Implants (Pty) Ltd submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical device is substantially equivalent in indications and design principles to the following legally marketed predicate devices: K193084, TIB Abutment System, Southern Implants (Pty) Ltd. K163060, Deep Conical (DC) Implants and Accessories, Southern Implants (Pty) Ltd. K173706, Piccolo Implants and Accessories, Southern Implants (Pty) Ltd. K220841, PrimeTaper EV Dental Implants Ø3.0, DS Implants abutments with EV connection (XS) The primary predicate device for the subject device Narrow TIB Abutment is K193084. The reference devices are K163060, K173706 and K220841. A comparison of the technological characteristics of the subject devices and the predicate devices is provided in the following table. | | Subject Device | Primary Predicate Device | |----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison | TIB Abutment System<br>Southern Implants (Pty) Ltd<br>(External Hex Narrow TIB Abutment) | K193084<br>TIB Abutment System<br>Southern Implants (Pty) Ltd | | Indications for Use<br>Statement | The TIB Abutments are premanufactured prosthetic<br>components directly connected to endosseous dental<br>implants and are intended for use as an aid in prosthetic<br>rehabilitation. The TIB abutments consists of two<br>major parts. Specifically, the titanium base and<br>mesostructure components make up a two-piece<br>abutment. The system integrates multiple components<br>of the digital dentistry workflow: Scan files from<br>desktop scanners, CAD software, CAM software,<br>ceramic material, milling machine and associated<br>tooling and accessories.<br>The intended use for the TiB narrow abutments used<br>with the Ø3.0 mm External hex and Ø3.0 mm Deep<br>Conical implants is limited to replacement of maxillary<br>lateral incisors and mandibular lateral and<br>central incisors. | The TIB Abutments are premanufactured prosthetic<br>components directly connected to endosseous dental<br>implants and are intended for use as an aid in prosthetic<br>rehabilitation. The TIB abutments consists of two<br>major parts. Specifically, the titanium base and<br>mesostructure components make up a two-piece<br>abutment. The system integrates multiple components<br>of the digital dentistry workflow: Scan files from<br>desktop scanners, CAD software, CAM software,<br>ceramic material, milling machine and associated<br>tooling and accessories. | | Product Code | NHA, PNP | NHA, PNP | | Item Code | TIBS-EX-30-C1.5<br>TIBS-EX-30-C3 | TIB-EX-34<br>TIB-EX-34-C1.5 | ## Table of Substantial Equivalence – External Hex Narrow TIB Abutments {6}------------------------------------------------ | | | TIB-EX-34-C3 | |----------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | Variations available for all External Hex connection<br>sizes. | | Reason for<br>Predicate/Reference | n/a | Abutment (general design and functioning)<br>Software Compatibility | | | Abutment Design | 2 Piece – Premanufactured titanium abutment, mounted onto the implant and fixed with a screw.<br>SageMaxx Zirconia (K130991) hybrid/crown restoration milled and bonded to the titanium abutment. | | Collar Height | 1.5 or 3mm | 0.6, 1.5 or 3mm | | Maximum Abutment<br>Top-Cap Angulation | 20° | 20° | | Implant Connection | External Hex | External Hex | | Post Height | Minimum 4.5mm | Minimum 4.5mm | | Abutment<br>Restorative Platform<br>Diameter | 3.85mm | 4.3 mm | | Abutment Material | Titanium Grade 5 Alloy (ASTM F136) | Titanium Grade 5 Alloy (ASTM F136) | | Abutment Surface | Machined and anodized | Machined and anodized | | Abutment Screw<br>Material | Titanium Grade 5 Alloy (ASTM F136) | Titanium Grade 5 Alloy (ASTM F136) | | Restoration Material | Zirconia - Sage Maxx NexxZr (K130991) | Zirconia - Sage Maxx NexxZr (K130991) | | CAD/CAM Design<br>Workflow | 3Shape E3 Desktop Scanner (3Shape A/S)<br>3Shape Abutment Designer Software | 3Shape E3 Desktop Scanner (3Shape A/S)<br>3Shape Abutment Designer Software | | CAD/CAM<br>Manufacturing<br>Workflow | WorkNC CAM software, Roland DWX51D milling unit | WorkNC CAM software, Roland DWX51D milling unit | | Mechanical Fatigue<br>Testing | Dynamic Fatigue Testing per ISO 14801 | Dynamic Fatigue Testing per ISO 14801 | | Sterility | Provided non-sterile | Provided non-sterile | | Usage | Single-patient single-use | Single-patient single-use | {7}------------------------------------------------ | | Subject Device | Primary Predicate Device | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | TIB Abutment System | K193084 | | Comparison | Southern Implants (Pty) Ltd | TIB Abutment System | | | (Deep Conical Narrow TIB Abutment) | Southern Implants (Pty) Ltd | | | | | | Indications for Use<br>Statement | The TIB Abutments are premanufactured prosthetic<br>components directly connected to endosseous dental<br>implants and are intended for use as an aid in prosthetic<br>rehabilitation. The TIB abutments consists of two<br>major parts. Specifically, the titanium base and<br>mesostructure components make up a two-piece<br>abutment. The system integrates multiple components<br>of the digital dentistry workflow: Scan files from<br>desktop scanners, CAD software, CAM software,<br>ceramic material, milling machine and associated<br>tooling and accessories.<br><br>The intended use for the TiB narrow abutments used<br>with the Ø3.0 mm External hex and Ø3.0 mm Deep<br>Conical implants is limited to replacement of maxillary<br>lateral incisors and mandibular lateral and | The TIB Abutments are premanufactured prosthetic<br>components directly connected to endosseous dental<br>implants and are intended for use as an aid in prosthetic<br>rehabilitation. The TIB abutments consists of two<br>major parts. Specifically, the titanium base and<br>mesostructure components make up a two-piece<br>abutment. The system integrates multiple components<br>of the digital dentistry workflow: Scan files from<br>desktop scanners, CAD software, CAM software,<br>ceramic material, milling machine and associated<br>tooling and accessories. | | | central incisors. | | | Product Code | NHA, PNP | NHA, PNP | | Item Code | TIBS-DC3-C1.5 | TIB-DC4 | | | TIBS-DC3-C3 | TIB-DC4-CSET | | | TIBS-DC4-C1.5 | Variations available for all Deep Conical connection | | | TIBS-DC4-C3 | sizes. | | Reason<br>for<br>Predicate/Reference | n/a | Abutment (general design and functioning)<br>Software Compatibility | | Abutment Design | 2 Piece – Premanufactured titanium abutment,<br>mounted onto the implant and fixed with a screw.<br>SageMaxx Zirconia (K130991) hybrid/crown<br>restoration milled and bonded to the titanium<br>abutment. | 2 Piece - Premanufactured titanium abutment,<br>mounted onto the implant and fixed with a screw.<br>SageMaxx Zirconia (K130991) hybrid/crown<br>restoration milled and bonded to the titanium<br>abutment. | | Collar Height | 1.5 or 3mm | 0.6, 1.5 or 3mm | | Maximum Abutment<br>Top-Cap Angulation | 0° (for Ø3mm implant) and 20° | 20° | | Implant Connection | Deep Conical | Deep Conical | | Post Height | Minimum 4.5mm | Minimum 4.5mm | | Abutment<br>Restorative Platform<br>Diameter | 3.85mm | 4.5 mm | | Abutment Material | Titanium Grade 5 Alloy (ASTM F136) | Titanium Grade 5 Alloy (ASTM F136) | | Abutment Surface | Machined and anodized | Machined and anodized | | Abutment<br>Screw<br>Material | Titanium Grade 5 Alloy (ASTM F136) | Titanium Grade 5 Alloy (ASTM F136) | | Restoration Material | Zirconia - Sage Maxx NexxZr (K130991) | Zirconia - Sage Maxx NexxZr (K130991) | | CAD/CAM Design<br>Workflow | 3Shape E3 Desktop Scanner (3Shape A/S)<br>3Shape Abutment Designer Software | 3Shape E3 Desktop Scanner (3Shape A/S)<br>3Shape Abutment Designer Software | | CAD/CAM<br>Manufacturing<br>Workflow | WorkNC CAM software, Roland DWX51D milling<br>unit | WorkNC CAM software, Roland DWX51D milling<br>unit | | Mechanical Fatigue<br>Testing | Dynamic Fatigue Testing per ISO 14801 | Dynamic Fatigue Testing per ISO 14801 | | Sterility | Provided non-sterile | Provided non-sterile | | Usage | Single-patient, single-use | Single-patient, single-use | # Table of Substantial Equivalence – Deep Conical Narrow TIB Abutments {8}------------------------------------------------ #### Subject device Narrow TIB Abutments The primary predicate device K193084 is for substantial equivalence of the subject device Narrow TIB abutment designs. The subject device abutments have similar indication for use, equivalent design, are both two-piece abutments with a premanufactured titanium base making up the first piece, and a ceramic composite intended to be bonded to the titanium base making up the second piece of the abutment. The primary predicate and subject device have comparable designs differing only in dimension. All the predicate device design features for example restorative platform, anodizing, post and grooves are present. The Indications for Use Statement for the subject device is similar to the reference device K193084 only with restrictions imposed on the Ø3.0 Deep Conical and Ø3.0 External Hex abutments. The abutment design and indications for use are thus substantially equivalent to that of the primary predicate. The primary predicate and the subject device have the identical maximum angulation for the ceramic restoration and follow an equivalent design workflow. With the only exception being the Ø3.0 Deep Conical Narrow TiB abutments which are not intended for angled placement. The DC Ø3.0mm Narrow TiB abutments are only to be placed straight in the patient. Occlusal loading forces are only to be applied through the central longitudinal axis of the implant body. In the design workflow the primary predicate and subject devices use the same scanners and software to design the restoration. The subject device is also substantially equivalent to the primary predicate with reference to the abutments restoration material with the subject and the primary predicate both making use of a ceramic composite material (K130991). The titanium abutment material for both the subject device and primary predicate is a Titanium Alloy (ASTM F136), the screw material is Titanium for both the subject and primary predicate. Both the subject and the primary predicate devices are provided non-sterile. Thus, the abutment materials and method provided of the subject device are substantially equivalent to that of the primary predicate. Substantial equivalence of the subject device and primary predicate components in terms of biocompatibility is supported by the fact that materials are identical in formulation, processing, component interactions, and storage conditions to the predicate device in K193084. The reference devices K173706 and K163060 are for substantial equivalence of the abutment connection types namely External Hex and Deep Conical. The reference devices contain the identical connection types and sizes to those of the subject devices. In support of substantial equivalence in terms of mechanical performance, dynamic compression-bending testing according to ISO 14801 Dentistry – Implants – Dynamic fatigue test for endosseous dental implants was performed. Dynamic testing was performed on worst-case subject device constructs. The reference devices K193084 and K220841 are included as reference devices for substantial equivalence in terms of fatigue strength of the worst-case combinations. ## CONCLUSION The subject devices and the primary predicates have similar indications for use, operating principles, have similar technological characteristics and are materials. The subject devices and the predicate devices encompass similar physical designs. The subject devices and the primary predicate devices are packaged identically and are all provided non-sterile. The subject device and the primary predicate follow the same design and manufacturing CAD/CAM workflows. {9}------------------------------------------------ The data included in this submission demonstrates substantial equivalence to the predicate devices listed above
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...