Retrospective evaluation of K. pneumoniae ATCC BAA-2814 at three laboratories to establish and validate quality control MIC ranges for the VITEK 2 System.
Retrospective analysis; Quality control validation; Laboratory records
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective evaluation of K. pneumoniae ATCC BAA-2814; Retrospective laboratory data review
K. pneumoniae ATCC BAA-2814; Number of Sites: 3
Not applicable for this study
Establishment of expected QC MIC range (16-64 μg/mL)
Indications for Use
VITEK® 2 AST-Gram Negative Fosfomycin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Fosfomycin in is a quantitative test. Fosfomycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Escherichia coli The VITEK® 2 Gram-Negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
Device Story
VITEK 2 AST-Gram Negative Fosfomycin is an automated, quantitative antimicrobial susceptibility test (AST) card for use with VITEK 2 and VITEK 2 Compact systems. The device utilizes a miniaturized, 64-well card containing premeasured fosfomycin concentrations (2, 8, 16, 64 µg/mL) and nutrient media. Clinical laboratory technicians prepare a standardized saline suspension of the bacterial isolate (Escherichia coli) via auto-dilution or manual dilution. The system fills, seals, and incubates the card, monitoring bacterial growth every 15 minutes via optical scanner light attenuation. The system's analysis program calculates the Minimum Inhibitory Concentration (MIC) based on growth patterns. Results are generated as an automated report providing MIC values and interpretive categories (Susceptible, Intermediate, Resistant). This information assists clinicians in selecting appropriate antimicrobial therapy for patients with Gram-negative infections.
Clinical Evidence
Performance evaluated via external study comparing VITEK 2 AST-GN Fosfomycin to CLSI agar dilution reference method. Study used 380 clinical isolates of E. coli. Results: 99.2% Essential Agreement (EA), 99.7% Category Agreement (CA), 0.0% Very Major Errors (VME), 0.0% Major Errors (ME), and 0.3% minor errors (mE). Reproducibility and quality control testing were acceptable.
Indicated for antimicrobial susceptibility testing of clinically significant aerobic Gram-negative bacilli, specifically Escherichia coli, using the VITEK® 2 and VITEK® 2 Compact Systems in clinical laboratories.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K222430
B Applicant
bioMérieux SA, Inc.
C Proprietary and Established Names
VITEK 2 AST-Gram Negative Fosfomycin (≤4 - ≥256 μg/mL), VITEK 2 AST-GN Fosfomycin (≤ - ≥256 μg/mL), VITEK 2 AST-GN Fosfomycin
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
| LTW | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTT | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for fosfomycin for testing of Gram-negative bacilli on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems
### B Measurand:
Fosfomycin (≤4 - ≥256 μg/mL)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
C Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test.
III Intended Use/Indications for Use:
A Intended Use(s):
The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
B Indication(s) for Use:
VITEK 2 AST-Gram Negative Fosfomycin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Fosfomycin is a quantitative test. Fosfomycin has been shown to be active against most strains of the microorganism listed below, according to the FDA label for this antimicrobial.
Active in vitro and in clinical infections:
Escherichia coli
The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.04 software
IV Device/System Characteristics:
A Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically (or allow operator to manually) dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then, the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and
K222430 - Page 2 of 11
{3}
loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 AST-Gram Negative Fosfomycin has the following concentrations in the card: 2, 8, 16, and 64 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The fosfomycin MIC result range for the VITEK 2 Gram Negative Fosfomycin is ≤ 4 to ≥ 256 µg/mL.
## B Principle of Operation:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems use visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well determine organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
## V Substantial Equivalence Information:
A Predicate Device Name(s):
VITEK 2 AST-Gram Negative Eravacycline (<=0.12 - >=4 µg/mL)
B Predicate 510(k) Number(s):
K191766
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device K222430 | Predicate K191766 |
| --- | --- | --- |
| Device Trade Name | VITEK 2 AST-Gram Negative Fosfomycin
(≤ 4 - ≥ 256 µg/ml) | VITEK 2 AST-GN Eravacycline
(≤0.12 - ≥4 µg/mL) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically | Same |
K222430 - Page 3 of 11
{4}
K222430 - Page 4 of 11
| Device & Predicate Device(s): | Device K222430 | Predicate K191766 |
| --- | --- | --- |
| | significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed. | |
| Test Methodology | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of microorganisms. | Same |
| Inoculum | Saline suspension of organism | Same |
| Test Card | Gram Negative (AST-GN) Susceptibility Card | Same |
| Instrument | VITEK 2 and VITEK 2 Compact Systems | Same |
| Analysis Algorithm | Growth Pattern Analysis | Same |
| Type of Test | Quantitative | Same |
| General Device Characteristic Differences | | |
| Antimicrobial Agent | Fosfomycin | Eravacycline |
| Concentrations of antimicrobial on card | 2, 8, 16, 64 µg/ml | 0.25, 1, 2, 4 µg/ml |
| Indications for Use | VITEK 2 AST-Gram Negative Fosfomycin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK 2 AST-Gram Negative Fosfomycin is a quantitative test.
Fosfomycin has been shown to be active against most strains of the microorganism listed below, according to the | VITEK 2 AST-Gram Negative Eravacycline is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK 2 AST-Gram Negative Eravacycline is a quantitative test.
Eravacycline has been shown to be active against most strains of the microorganisms listed below, according to the |
{5}
| Device & Predicate Device(s): | Device K222430 | Predicate K191766 |
| --- | --- | --- |
| | below, according to the FDA label for this antimicrobial.
*Active in vitro* and in clinical infections:
*Escherichia coli* | FDA label for this antimicrobial.
*Active in vitro* and in clinical infections:
*Citrobacter freundii*
*Enterobacter cloacae*
*Escherichia coli*
*K. oxytoca*
*Klebsiella pneumoniae*
*In vitro* data are available but clinical significance is unknown:
*Citrobacter koseri*
*Klebsiella (Enterobacter) aerogenes* |
VI Standards/Guidance Documents Referenced:
- FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- 7-279: M07, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard – Eleventh Edition”, (January 2018)
- 7-294: M100, “Performance Standards for Antimicrobial Susceptibility Testing; 32nd Edition (January 2022)
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Reproducibility testing for the VITEK 2 AST-Gram Negative Fosfomycin was conducted at three external clinical sites using a panel of 10 E. coli isolates. Each isolate was tested in triplicate, using separate inocula, over three days for a total of 270 data points. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. In addition, inocula were prepared by the manual dilution method for use with the VITEK 2 Compact. The mode of MIC values was determined for each isolate and the reproducibility was calculated based on the number of MIC values that fell within ±1 doubling dilution of the mode. The data were analyzed taking into consideration best-case and worst-case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems.
K222430 - Page 5 of 11
{6}
The reproducibility data for both of the VITEK 2 System card inoculation options and the manual inoculum for VITEK 2 Compact are summarized in Table 1 below. The results are acceptable.
Table 1. Reproducibility Data for Each Inoculum Method
| VITEK 2 System Card Inoculum Method | Best Case | Worst Case |
| --- | --- | --- |
| Automatic Dilution | 100% | 100% |
| Manual Dilution | 99.6% | 99.3% |
| VITEK 2 Compact Card Inoculum Method | Best Case | Worst Case |
| --- | --- | --- |
| Manual Dilution | 99.3% | 99.3% |
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality Control (QC) Testing:
The CLSI recommends four QC strains for testing fosfomycin, namely Escherichia coli ATCC 25922, Pseudomonas aeruginosa ATCC 27853, Enterococcus faecalis ATCC 29212 and Staphylococcus aureus ATCC 29213.
One organism, E. coli ATCC 25922, was tested a sufficient number of times (i.e., at least 20/site) at each testing site using both the VITEK 2 card and agar dilution (AD) reference method. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact. The QC organism was tested by the automatic dilution method at the four clinical sites and at the sponsor's laboratory while the manual VITEK 2 and COMPACT testing was performed at only the four clinical sites. This QC organism yielded off-scale VITEK 2 fosfomycin MICs of ≤ 4 μg/mL (Table 2). A VITEK result of ≤ 4 μg/mL is considered acceptable for this organism.
K222430 - Page 6 of 11
{7}
Table 2. Fosfomycin VITEK 2 Quality Control Data
| QC Organism | VITEK Result Range (μg/mL) | BMD Result Range (μg/mL) | VITEK 2 Auto Dilution | Reference Result | VITEK 2 Manual Dilution | Reference Result | VITEK 2 COMPACT Manual Dilution | Reference Result |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| E. coli ATCC 25922 Expected Result 0.5-2 (μg/mL) | | ≤ 0.25 | | | | | | |
| | | 0.5 | | 84 | | 35 | | 37 |
| | | 1 | | 86 | | 50 | | 52 |
| | | 2 | | 16 | | 11 | | 11 |
| | ≤4* | 4 | 186 | 0 | 96 | 0 | 100 | 0 |
* ≤ 4 μg/mL (i.e. lowest reporting value for VITEK 2 and VITEK 2 COMPACT Gram-Negative (AST-GN) Susceptibility Card
The off-scale results necessitated identifying an additional Gram-negative QC organism. K. pneumoniae ATCC BAA-2814 for use specifically for quality control testing of Fosfomycin with the VITEK 2 System. This isolate was evaluated retrospectively at three laboratories and yielded an expected QC range of 16-64 μg/mL, as noted in the device labeling.
The two Gram-positive QC organisms, E. faecalis ATCC 29212 and S. aureus ATCC 29213, were used as ancillary QC organisms for only the agar dilution reference method. This was done to perform further quality control of the agar dilution plates. QC results for the agar dilution method were within the expected result range 100% of the time for both organisms (Table 3).
Table 3. Ancillary Quality Control Data for Fosfomycin
| ATCC QC Strain | MIC (μg/mL) | Reference Result |
| --- | --- | --- |
| S. aureus ATCC 29213
Expected Result 0.5-4 (μg/mL) | ≤ 0.25 | |
| | 0.5 | 2 |
| | 1 | 26 |
| | 2 | 89 |
| | 4 | 72 |
| Total | | 189 |
| | | |
| E. faecalis ATCC 29212
Expected Result 32-128 (μg/mL) | ≤ 0.16 | |
| | 32 | 76 |
| | 64 | 113 |
| | 128 | |
| Total | | 189 |
## Inoculum Density Control
The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.
## Purity Check
A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.
K222430 - Page 7 of 11
{8}
K222430 - Page 8 of 11
# Device Failure
Device failures occurred at each of the test sites on 12 different days. These involved failure of an instrument to fill cards properly, pipettor errors, failure of an instrument to connect to the computer, failure of an instrument to recognize a card, an instrument rejecting cards, a reader error, and a card being dropped on the floor before testing. All failures were resolved by one of the following: repeat testing with the original inoculum; card replacement; straightening of kinked tubing; technician intervention, or BioMérieux intervention. There was no impact on testing or results.
# Growth Failure Rate
A total of 380 clinical and challenge isolates were tested by VITEK 2 Fosfomycin. No growth failure was recorded and all 380 isolates have VITEK 2 AST results available.
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Not applicable
# B Comparison Studies:
1. Method Comparison with Predicate Device:
Testing of fosfomycin on the VITEK 2 AST-Gram Negative card was performed at four clinical sites. There were 300 clinical isolates and 80 challenge isolates tested for a total of 380 isolates. Results obtained with the VITEK 2 AST-Gram Negative card with fosfomycin were compared to results obtained with the CLSI recommended agar dilution testing conditions. The MIC result range for the VITEK 2 AST-Gram Negative Fosfomycin is ≤4 to ≥256 μg/mL for Escherichia coli. The reference method testing consisted of two-fold serial dilutions of fosfomycin with a range of ≤0.25 to ≥1024 μg/mL. The testing conditions for the reference method consisted of the following:
- Medium – Mueller Hinton agar with addition of 25 μg/mL of glucose-6-phosphate and appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C; 16-24 hours
The VITEK 2 AST cards were inoculated with test organisms using the auto-dilution method (VITEK 2) and manual dilution method (VITEK 2 and VITEK 2 Compact). All test inocula used for the VITEK 2 AST cards and the reference method were standardized using the DensiCHEK Plus instrument.
A total of 300 clinical isolates were evaluated at four sites: the majority were recently isolated from clinical specimens and the remainder were stock isolates. Complete test results
{9}
are available for all 300 clinical isolates. All clinical isolates were tested with the autodilution option of the VITEK 2.
A total of 80 E. coli challenge isolates were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 and with the manual dilution method on the VITEK 2 Compact.
At the time of comparative testing, 14 resistant isolates and 1 intermediate isolate were available for E. coli. The 14 resistant isolates comprise 3.6% of 380 total isolates. However, since the resistant breakpoint is at the upper limit of the reporting range, an upward shift in MIC values may not be easily detected. Since resistance rate to Fosfomycin is relatively low, the occurrence of such shifts may be rare. However, the following footnote is included in the device labeling to inform users of this issue:
An MIC shift in the resistance range (≥256 μg/mL) will be unknown since the resistant breakpoint and the upper limit of the reporting range is the same.
## Clinical and Challenge Data – VITEK 2 Auto-Dilution
VITEK 2 AST-GP Fosfomycin performance was determined with 380 isolates (300 clinical isolates and 80 challenge isolates). The 80 challenge isolates were also tested at one site with the manual dilution option for the VITEK 2 Compact system. The data are summarized in Table 4. The overall performance of using the VITEK 2 auto-dilution is acceptable with an EA of 99.2% and a CA of 99.7%. There were no major or very major errors. The overall performance of the manual dilution option of the VITEK 2 and VITEK 2 Compact systems is acceptable with an EA of 100% and a CA of 98.8% with the VITEK 2 Compact system. There were no major or very major errors.
Table 4. Essential and Category Agreement for VITEK 2 Combined Challenge and Clinical E. coli Isolates
| Organism Group | Total Tested | # EA | % EA | Total Eval | # EA of Eval | % EA of Eval | # CA | % CA | # R | # S | # min | # maj | # vmaj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Inoculation Method Auto - Read Method VITEK 2 | | | | | | | | | | | | | |
| Clinical | 300 | 297 | 99 | 7 | 4 | 57.1 | 300 | 100 | 0 | 300 | 0 | 0 | 0 |
| Challenge | 80 | 80 | 100 | 11 | 11 | 100 | 79 | 98.8 | 14 | 65 | 1 | 0 | 0 |
| Combined | 380 | 377 | 99.2 | 18 | 15 | 83.3 | 379 | 99.7 | 14 | 365 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| Inoculation Method Manual – Read Method VITEK 2 | | | | | | | | | | | | | |
| Challenge | 80 | 79 | 98.8 | 11 | 10 | 90.9 | 79 | 98.8 | 14 | 65 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| Inoculation Method Manual – Read Method VITEK 2 Compact | | | | | | | | | | | | | |
| Challenge | 80 | 80 | 100 | 11 | 11 | 100 | 79 | 98.8 | 14 | 65 | 1 | 0 | 0 |
EA – Essential Agreement
CA – Category Agreement
Eval – Evaluable isolates
R – Resistant isolates
S – Susceptible isolates
min – minor errors
maj – major errors
vmaj – very major errors
K222430 - Page 9 of 11
{10}
Essential agreement (EA) occurs when the result of the reference method and that of the VITEK 2 AST-Gram Negative Fosfomycin are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on-scale for both the reference method and the VITEK 2 AST-Gram Negative Fosfomycin or results in which an off-scale result is at least two doubling dilutions from the on scale result. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the VITEK 2 AST-Gram Negative Fosfomycin.
## Resistance Mechanism Characterization
The FDA drug label for Fosfomycin does not indicate any specific mechanisms of resistance.
As required under 511A(b)(2)(C)(ii)(I) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the Precautions section of the device labeling to address testing and reporting of non-indicated species:
Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
## MIC Trends
A trending analysis was conducted using the combined data (clinical and challenge) obtained for each inoculation method (Table 5). This trending calculation analyzes device MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method. MIC values that are off-scale for both the reference and device are not considered in the trending analysis. A difference between the percentage of isolates with higher or lower MIC values $\geq 30\%$ with a statistically significant confidence interval are considered to have evidence of trending. No trending was detected for 22 on-scale results with VITEK 2 auto dilution.
Table 5. Trending (clinical and challenge isolates)
| Inoculation/Read Method | Organism | Total On Scale for Trending | ≥ 1 Dilution Lower # (%) | Exact #(%) | ≥ 1 Dilution Higher # (%) | Percent Difference (95% CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Auto\VITEK 2 | E. coli | 22 | 6, (27.3) | 8 | 8, (36.4) | 9% , (-18%, 34%) | No |
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
K222430 - Page 10 of 11
{11}
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
The FDA-identified recognized interpretive criteria for fosfomycin are listed in Table 6
Table 6. FDA-Recognized Interpretive Criteria for Fosfomycin (μg/mL)ᵃ
| Organisms | S | I | R |
| --- | --- | --- | --- |
| Escherichia coli | ≤64 | 128 | ≥256 |
ᵃAccording to FDA STIC Website.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 AST-GN Fosfomycin when revised breakpoints for fosfomycin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the fosfomycin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K222430 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.