← Product Code [MOS](/productcode/MOS) · K222407

# 16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil (K222407)

_Neocoil, LLC · MOS · Oct 14, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K222407

## Device Facts

- **Applicant:** Neocoil, LLC
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Oct 14, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software. The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures. When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.

## Device Story

Phased array, receive-only RF coils for MRI imaging of breast, axilla, and chest wall; available in 1.5T and 3T configurations. Device consists of medial, lateral, and biopsy arrays, baseplate, and accessories (biopsy grids/drapes). Operates by receiving magnetic resonance signals from hydrogen nuclei in anatomy while blocking high-frequency fields from the MRI scanner; signals are amplified and transferred to the MRI scanner for image reconstruction. Used in clinical settings by trained medical professionals. Output is tomographic images (axial, sagittal, coronal oblique) used for diagnostic or interventional planning. Benefits include high-resolution imaging for clinical decision-making.

## Clinical Evidence

No clinical performance testing performed. Substantial equivalence established via non-clinical bench testing, including biocompatibility, electrical safety (IEC 60601-1, 60601-2-33, 60601-1-2), usability, surface temperature, blocking network analysis, B0/B1 field distortion, and NEMA MS 6-2008 image quality metrics (SNR, uniformity).

## Technological Characteristics

Phased array, receive-only RF coils. Materials assessed for biocompatibility. Energy source: MRI scanner (no internal power). Dimensions/form factor: contoured for breast anatomy. Connectivity: system cable to MRI scanner. Tuning: 1.5T or 3T field strengths. Includes antenna fuses, preamplifiers, and decoupling circuitry.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- 3.0T 16ch Breast Coil ([K173377](/device/K173377.md))

## Reference Devices

- 1.5T 16ch Breast Coil ([K182958](/device/K182958.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

October 14, 2022

NeoCoil, LLC % Katie Gonzalez Ouality Systems & Regulatory Manager N27 W23910A Paul Road PEWAUKEE WI 53072

Re: K222407

Trade/Device Name: 16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 9, 2022 Received: September 15, 2022

Dear Katie Gonzalez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K222407

#### Device Name

16ch Breast Coils: 1.5T 16ch Breast Coil / 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil / 3T 16ch Breast Coil

Indications for Use (Describe)

The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 373.0T MRI system software.

The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.

When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:

Image /page/3/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape on the left, with a white curved line inside that forms an arrow pointing to the right. To the right of the shape is the word "NeoCoil" in blue. The font is sans-serif and the letters are bold.

# Special 510(k) Summary

K222407

Applicant

NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA

Contact

Katie Gonzalez Quality Systems and Regulatory Manager 262-522-6124 (office) 262-347-1251 (fax) Katie.Gonzalez@neocoil.com

Preparation Date

August 3, 2022

## Name of Device

- . Trade/Proprietary name(s):
### 16ch Breast Coils

| Submission     | Device Trade/Proprietary Name |
|----------------|-------------------------------|
| Existing model | 3.0T 16ch Breast Coil         |
| Existing model | 1.5T 16ch Breast Coil         |
| New model      | 3T 16ch Breast Coil           |
| New model      | 1.5T 16ch Breast Coil         |

- Magnetic Resonance Specialty Coil ● Common name:
- . Classification name: 21 CFR 892.1000, Magnetic resonance diagnostic device, Product Code MOS

### Predicate Device

3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017, manufactured by NeoCoil, LLC

### Reference Device

1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018, manufactured by NeoCoil, LLC

### Device Description

The NeoCoil 3.0T 16ch Breast Coil is a phased array coil for imaging structures of the breast, axilla and chest wall. The 3.0T 16ch Breast Coil is a three part receive-only coil designed to provide high resolution imaging. The.3.0T 16ch Breast Coil includes a coil support structure, patient support structure, biopsy components, accessories and comfort pads.

{4}------------------------------------------------

N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:

Image /page/4/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The "N" and "e" are in a lighter shade of blue, while the "o" is in a darker shade of blue. The "Coil" part of the word is in black.

The NeoCoil 3.0T 16ch Breast Coil is tuned to receive RF frequency corresponding to the proton precession in a 3 tesla magnetic field, which is governed by the Larmor equation.

The 16ch Breast Coils are intended for use in a manner that is identical to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018. The 16ch Breast Coil consists of the following arrays available in 1.5T, 3.0T and 3T field strengths:

- Medial Array, Breast Coil
- . Lateral Array Left, Breast Coil
- Lateral Array Right, Breast Coil ●
- . Baseplate Assembly, Breast Coil
- Biopsy Array Left, Breast Coil ●
- Biopsy Array Right, Breast Coil ●

The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Head. Neck and Brachial Plexus anatomy while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings. The received signals are amplified before being transferred to the MRI scanner through the coil's system cable. The amplified signals are processed into tomographic images of the breast, axilla and chest wall anatomy by the MRI scanner. Images are typically generated as axial, sagittal, coronal oblique slices. Accessories associated with the 16ch Breast Coils include biopsy grids and biopsy drapes.

The 1.5T Breast Coil, the 3T 16ch Breast Coil, and the 3.0T 16ch Breast Coil are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.

The 16ch Breast Coil is intended for use in a manner that is identical to the predicate device described in this submission.

Proposed labeling is documented in the Instructions for Use manual for the 16ch Breast Coil (NC149IFU-en).

#### Intended Use, including indications

The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software.

The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.

When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.

#### Technological Characteristics

16ch Breast coils are similar in design, material, chemical composition and energy source to the legally marketed predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.

{5}------------------------------------------------

N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:

Image /page/5/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with a white crescent moon shape inside, and a black arrow pointing to the right. To the right of the logo is the word "NeoCoil" in blue and black font. The word "Neo" is in blue, and the word "Coil" is in black.

At a high level, the 16ch Breast Coils included as part of this submission, and the predicate device, 3.0T 16ch Breast Coil (K173377) are based on the following same technological elements:

- Prescription use;
- . Coil designs are receive-only phased array coils;
- . Preamplification methodology
- Decoupling methodology
- Patient contacting materials and chemical composition are known materials that have ● been assessed for compliance with recognized biocompatibility standards;
- Energy source for the coils is the MRI scanner; ●
- Energy is not intended to be supplied by the coils; ●
- . Mechanical designs are contoured for the breast anatomy;
- Facilitates imaging for diagnostic and interventional planning of the breast, axilla and chest wall anatomy:
- Channel count
- Field strength when compared to the predicate device and reference device ●

The following technological differences exist between the subject and predicate and reference devices:

- MRI specific system cabling
- MRI specific interface circuitry .
- MRI specific RF loop and preamplifier tuning ●
- Addition of fuses on antenna elements ●

Testing performed per defined standards demonstrates that the safety and/or effectiveness of the 16ch Breast Coils compared to the predicate device is not adversely affected as a result of the differences.

### Testing

Through NeoCoil's design controls process, a risk assessment leveraging FDA's quidance document decision trees for when to submit a new 510(k) for changes to an existing device, a summary of non-clinical performance data using well-established methodologies is included, referenced, or relied on to demonstrate that the 16ch Breast Coils are safe and effective and performs in a manner that demonstrates substantial equivalence to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.

### Performance testing - Bench:

A Test Report Summary for non-clinical Bench Testing performed, including testing to FDArecognized consensus standards identified as relevant in FDA guidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016, is outlined below:

| Test<br>Performed                                                                  | Objective(s) of the<br>Test                                                                                                                                   | Test Method<br>Description                                                                                                     | Pre-defined<br>pass/fail<br>criteria                                                                   | Results<br>Summary | Discussions /<br>Conclusions                                                                                                           |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Assessment                                                     | Assess potential<br>biological risks                                                                                                                          | Evaluation of<br>data; historical<br>use, biologic<br>testing, where<br>warranted                                              | Acceptable<br>level of risk                                                                            | Pass               | No identified significant<br>risks.                                                                                                    |
| Test<br>Performed                                                                  | Objective(s) of the<br>Test                                                                                                                                   | Test Method<br>Description                                                                                                     | Pre-defined<br>pass/fail<br>criteria                                                                   | Results<br>Summary | Discussions /<br>Conclusions                                                                                                           |
| Electrical<br>Safety                                                               | Basic electrical<br>safety/essential<br>performance, 60601-1                                                                                                  | Test Lab                                                                                                                       | Pre-defined<br>performance<br>standards                                                                | Pass               | Applicable<br>requirements for basic<br>electrical safety and<br>essential performance<br>met.                                         |
| Electrical<br>Safety                                                               | Particular electrical<br>requirements: MR<br>equipment, 60601-2-<br>33                                                                                        | Test Lab                                                                                                                       | Pre-defined<br>performance<br>standards                                                                | Pass               | Applicable<br>requirements of the<br>particular standard<br>were met.                                                                  |
| Electrical<br>Safety                                                               | Collateral electrical<br>safety/essential<br>performance, 60601-<br>1-2                                                                                       | Test Lab /<br>Bench Testing                                                                                                    | Pre-defined<br>performance<br>standards                                                                | Pass               | Applicable<br>requirements of the<br>collateral standard were<br>met.                                                                  |
| Usability<br>Assessment                                                            | Devices meet<br>customer, end user<br>and patient needs                                                                                                       | Actual.<br>simulated or<br>retrospective<br>evaluation of<br>the device<br>and/or data                                         | Pre-defined<br>requirements                                                                            | Pass               | The devices met the<br>needs of the customer,<br>end user and patient.                                                                 |
| Entrapment,<br>Trapping Zone<br>and Cable<br>Looping<br>(assessment w/<br>scanner) | Assess the device for<br>pinch points,<br>entrapment, cable<br>looping - interfacing<br>with MRI scanner                                                      | Evaluation of<br>coil-to-scanner<br>entrapment,<br>trapping and<br>cable looping<br>not covered by<br>test lab<br>assessments. | Requirements<br>based on pre-<br>defined<br>requirements<br>in 60601-1<br>and customer<br>requirements | Pass               | Requirements were<br>met.                                                                                                              |
| Surface<br>Temperature                                                             | Surface temperatures<br>do not exceed limits                                                                                                                  | MRI scanner<br>test                                                                                                            | Pre-defined<br>performance<br>standards                                                                | Pass               | Surface temperatures<br>were within IEC limits.                                                                                        |
| Unplugged<br>Surface<br>Temperature                                                | Devices remain safe in<br>first fault condition                                                                                                               | MRI scanner<br>test                                                                                                            | Acceptable<br>level of risk                                                                            | Pass               | Surface temperatures<br>were within IEC limits<br>when the coil is left<br>unplugged in the MRI<br>scanner.                            |
| Blocking<br>Network<br>Analysis                                                    | Ensures devices are<br>designed with<br>adequate active and<br>passive transmit<br>decoupling                                                                 | Theoretical<br>calculations                                                                                                    | Adequate<br>transmit<br>decoupling                                                                     | Pass               | Blocking network<br>demonstrates adequate<br>active and passive<br>transmit decoupling.                                                |
| B1 Field<br>Distortion                                                             | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner                                                                   | MRI scanner<br>test                                                                                                            | Pre-defined<br>performance<br>standards                                                                | Pass               | B1 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Test<br>Performed                                                                  | Objective(s) of the<br>Test                                                                                                                                   | Test Method<br>Description                                                                                                     | Pre-defined<br>pass/fail<br>criteria                                                                   | Results<br>Summary | Discussions /<br>Conclusions                                                                                                           |
| B0 Filed<br>Distortion                                                             | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner                                                                   | MRI scanner<br>test                                                                                                            | Pre-defined<br>performance<br>standards                                                                | Pass               | B0 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| NEMA MS 6-<br>2008                                                                 | Evaluate single-<br>channel non-volume<br>special purpose<br>radiofrequency (RF)<br>coils for use with<br>magnetic resonance<br>(MR) imaging (MRI)<br>systems | MRI scanner<br>test                                                                                                            | Pre-defined<br>performance<br>standards                                                                | Pass               | SNR and Image<br>Uniformity are<br>consistent with the<br>requirements for<br>indications for use.                                     |

{6}------------------------------------------------

N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 Fax: (262) 347-1251

Image /page/6/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The word "Neo" is in a slightly larger font than the word "Coil".

{7}------------------------------------------------

N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:

Image /page/7/Picture/1 description: The image contains the logo for NeoCoil. To the left of the company name is a black circular logo with an arrow pointing to the right. The company name is to the right of the logo, with "Neo" in blue and "Coil" in black.

## Performance testing - Clinical:

Due to the substantial equivalency of the devices and the evaluation of performance testing against defined standards, no additional clinical performance testing has been performed above that present for the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.

No adverse events have been reported against the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.

### Conclusion

This submission demonstrates by means of nonclinical testing that the 16ch Breast Coils are substantially equivalent to and perform as well as the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.

---

**Source:** [https://fda.innolitics.com/device/K222407](https://fda.innolitics.com/device/K222407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
