K222351 · Vincent Healthcare Products Limited · BTT · Dec 21, 2022 · Anesthesiology
Device Facts
Record ID
K222351
Device Name
inspired™ VHB20 Heated Humidifier
Applicant
Vincent Healthcare Products Limited
Product Code
BTT · Anesthesiology
Decision Date
Dec 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The inspired™ VHB20 Heated Humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases. The inspired TM VHB20 Heated Humidifier is indicated for use by trained personnel only within a hospital/institutional environment. It is compatible with the Fisher and Paykel MR290 Humidification Chamber (Single Use), RT380 Dual Heated Breathing Circuit and RT265 Dual Heated Infant Breathing Circuit.
Device Story
The inspired™ VHB20 Heated Humidifier is a mains-powered respiratory device used in hospital settings by trained personnel. It receives dry medical gases and passes them through a water chamber heated by an electronically controlled plate. A microprocessor manages the heating element and monitors temperature via probes at the chamber outlet and patient end. The device provides electrical energy to heated-wire breathing circuits to maintain gas temperature and humidity during delivery. It supports flow rates of 5-60 lpm (invasive) and 5-120 lpm (non-invasive). Output is displayed on an LCD screen. Safety features include hardware-based over-temperature cutouts and software-controlled temperature limits, plus over-current protection via an internal switched-mode power supply (SMPS). The device helps prevent airway drying and thermal loss for patients on respiratory support.
Clinical Evidence
Bench testing only. Performance testing demonstrated compliance with ISO 80601-2-74, ANSI AAMI ES60601-1, and IEC 60601-1-2. Testing included temperature and humidity output across specified flow ranges, electrical safety, EMC, and software verification/validation. Biocompatibility was assessed per ISO 10993 and ISO 18562 for externally communicating, permanent duration devices.
Technological Characteristics
Mains-powered respiratory humidifier; pass-over heating method. Materials: Externally communicating, tissue-contacting, permanent duration. Connectivity: Standalone with heater-wire adaptor and sensor cable. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, ISO 80601-2-74, AIM 7351731. Software: Microprocessor-controlled heating with hardware/software over-temperature protection. Dimensions: 159mm x 204mm x 150mm. Weight: 1.9kg.
Indications for Use
Indicated for infant to adult patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. For use by trained personnel in hospital/institutional environments only.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
December 21, 2022
Vincent Healthcare Products Limited % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704
Re: K222351
Trade/Device Name: inspired™ VHB20 Heated Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: August 3, 2022 Received: August 4, 2022
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ethan L. Nyberg -S
for James Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222351
Device Name inspired™ VHB20 Heated Humidifier
Indications for Use (Describe)
The inspired™ VHB20 Heated Humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases.
The inspired TM VHB20 Heated Humidifier is indicated for use by trained personnel only within a hospital/institutional environment. It is compatible with the Fisher and Paykel MR290 Humidification Chamber (Single Use), RT380 Dual Heated Breathing Circuit and RT265 Dual Heated Infant Breathing Circuit.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summarv Page 1 of 6
| Date Prepared: | 19-Dec-22 |
|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Vincent Healthcare Products Limited<br>Flat/RM B2, 7/F., Hang Fung Industrial Building, Phase 2, 2G Hok Yuen Street<br>Hung Hom, Kowloon Hong Kong | |
| Official Contact: | Patrick Dennis, QA/RA Supervisor |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr NE<br>St. Petersburg, FL 33704 |
| Proprietary or Trade Name: | inspired™ VHB20 Heated Humidifier |
| Common/Usual Name: | Respiratory gas humidifier |
| Classification Name: | 21CFR 868.5450 |
| Product Code: | BTT |
| Predicate Device: | K073706 - Fisher & Paykel MR850 Respiratory<br>Humidifier |
| Common/Usual Name: | Respiratory gas humidifier |
| Classification Name: | 21CFR 868.5450 |
| Product Code: | BTT |
#### Device Description:
The subject device, inspired™ VHB20 Heated Humidifier, is indicated to add humidity and heat to the breathing gasses delivered to patients requiring mechanical ventilation or positive pressure breathing assistance.
It consists of an electrically powered heat controlled with a microprocessor to provide software control of the heating element that transfers heat to the water in a chamber.
Breathing tubes enable the humidified gas to be transported to the patient which may also be electrically heated (if connected), by means of a heater-wire placed internally to the tubes, to minimize the loss of humidity. A passive electrical adaptor provides electrical energy from the humidifier to the heater-wire in the breathing circuit. Over-current protection is provided by the SMPS inside the inspired™ VHB20 Heated Humidifier, which incorporates an over-current protection that will be triggered (cut-out) if output power of the SMPS reaches 110% of the rating (100W). The SMPS will recover automatically after fault condition is removed, while the humidifier will restart. This safety feature protects the humidifier from short-circuits in the DC part of the circuit, and prevents voltage and current transients on the heater-wire.
The inspired™ VHB20 Heated Humidifier provides heated and humidified gas with flow rates of 5 to 60 lpm (invasive) and 5 to 120 lpm (non-invasive).
The inspired™ VHB20 Heated Humidifier is a mains powered device that complies with ES60601-1, IEC 60601-1-2, ISO 80601-2-74 and AIM 7351731.
{5}------------------------------------------------
## Principle of Operation:
The inspired™ VHB20 Heated Humidifier is a respiratory humidifier that provides a heat source and temperature control to warm a water container, referred as a water chamber, to heat and humidify dry respiratory gases.
#### Indications for Use:
The inspired™ VHB20 Heated Humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases.
The inspired™ VHB20 Heated Humidifier is indicated for use by trained personnel only within a hospital/institutional environment. It is compatible with the Fisher and Paykel MR290 Humidification Chamber (Single Use), RT380 Dual Heated Breathing Circuit and RT265 Dual Heated Infant Breathing Circuit.
#### Patient Population:
Infant to adult patients.
#### Environments of use:
Hospital/institutional environment
We present the proposed device vs. the predicate in Table 1.
As part of the comparison we will present and discuss the:
- Indications for Use ●
- . Technology and Principle of Operation
- Performance and Specifications .
{6}------------------------------------------------
## 510(k) Summary Page 3 of 6
## Table 1 - Comparison - Subject vs. Predicate
| 510(k) Number | Subject Device | K073706 Predicate | Comment |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device Name | Vincent Medical inspiredTM VHB20 Heated Humidifier | Fisher & Paykel MR850 Respiratory Humidifier | |
| Classification No. & Product code | 868.5450 / BTT | 868.5450 / BTT | Similar |
| Indications for Use | The inspiredTM VHB20 Heated Humidifier is intended to be used<br>to warm and add humidity to gases delivered to patients requiring<br>mechanical ventilation or positive pressure breathing assistance or<br>general medical gases. | The Fisher & Paykel Healthcare MR850 humidifier is intended<br>to be used to warm and add humidity to gases delivered to<br>patients requiring mechanical ventilation or positive pressure<br>breathing assistance or general medical gases. | Similar |
| | The inspiredTM VHB20 Heated Humidifier is indicated for use by<br>trained personnel only within a hospital/institutional environment.<br>It is compatible with the Fisher and Paykel MR290<br>Humidification Chamber (Single Use), RT380 Dual Heated<br>Breathing Circuit and RT265 Dual Heated Infant Breathing<br>Circuit. | | |
| Principle of operation | The device has two heating control units and two temperature<br>sensors respectively. Water within a humidification chamber is<br>heated by the device's heating plate and this temperature is<br>controlled by the device with the use of temperature probes. Dry<br>medical gases passing through the chamber gain increased humidity<br>and heat. The heater wire adaptor supplies current to heater wires<br>within breathing tubes that maintain gas temperature travelling to<br>patient. Temperature probes measurement temperature and device<br>controls chamber temperature to achieve desired gas warmth and<br>humidify for patient. | The device has two heating control units and two temperature<br>sensors respectively. Water within a humidification chamber is<br>heated by the device's heating plate and this temperature is<br>controlled by the device with the use of temperature probes. Dry<br>medical gases passing through the chamber gain increased<br>humidity and heat. The heater wire adaptor supplies current to<br>heater wires within breathing tubes that maintain gas temperature<br>travelling to patient. Temperature probes measurement<br>temperature and device controls chamber temperature to achieve<br>desired gas warmth and humidify for patient. | Similar |
| Components | The device consists of VHB20 Heated Humidifier, Heater wire<br>adaptor and Temperature-humidity data/sensor cable. | The device consists of MR850 Respiratory Humidifier, Heater<br>wire adaptor and Temperature probe. | Similar |
| Electrical Protection Classification | Class I | Class I | Similar |
| Applied part | Type BF | Type BF | Similar |
| Drip proof | IPX1 | IPX1 | Similar |
| Target population | Infant to adult | Any patient requiring active humidification | Similar |
| Heating Method | Pass over | Pass over | Similar |
| Dimensions (H x D x W) | 159mm x 204mm x 150mm | 140mm × 173mm × 135mm | Similar |
| Weight (without humidification<br>chamber) | 1.9kg | 2.8kg | Similar |
| Supply voltage | 110 - 127V | 115V | Similar |
| 510(k) Number | Subject Device | K073706 Predicate | Comment |
| Device Name | Vincent Medical inspired™ VHB20 Heated Humidifier | Fisher & Paykel MR850 Respiratory Humidifier | |
| Supply frequency | 50/60 Hz | 50/60 Hz | Similar |
| Supply current | 3.0A max at 110V | 2.0 A max at 115V | Similar |
| Heater plate power | 150W | 150 W | Similar |
| Heater plate over temperature<br>cutout (hardware) | $118 ± 7°C$ | $118 ± 6°C$ | Similar |
| Safety cutoff temperature<br>(software control) | $105°C$ | $110°C$ | Similar |
| Heater Wires power supply | 24VDC, 80VA | 22V~, 2.73A max, 60W, 50/60Hz | Similar |
| | Temperature control settings | | |
| | Airway | | |
| Invasive | Default: $39°C$ | Default: $40°C$ | Similar |
| | Range: $35-40°C$ | Range: $35-40°C$ , (Versions 7.22) | |
| Non-invasive | Default: $34°C$ | Default: $34°C$ | Similar |
| | Range: $30-37°C$ | Range: $28-34°C$ , (Versions 7.22) | |
| | Chamber outlet | | |
| Invasive | Default: $36°C$ | Default: | Similar |
| | Range: $35-43°C$ | Range: $35.5-42°C$ , (Versions 7.22) | |
| Non-Invasive | Default: $31°C$ | Default: | Similar |
| | Range: $30-32°C$ | Range: $31-36°C$ , (Versions 7.22) | |
| Display | 2.8-inch - LCD display | 3 digit 14 mm - 7 segment LED | Similar |
| Display Range | 10 - 70°C | 10 - 70°C | Similar |
| Accuracy | $± 2°C$ | $± 0.3°C$ | Similar |
| | | (in 25 to $45°C$ temperature range) | |
| | Alarms | | |
| High temperature alarm | Invasive mode: Patient-side temperature exceeds $41°C$ | Invasive and Non-invasive mode: | Similar |
| | Non-invasive mode: Patient-side temperature exceeds $38°C$ | Displayed temperature exceeds $41°C$ | |
| | | or<br>Airway temperature exceeds $43°C$ | |
| | Invasive mode: Patient-side temperature < $34°C$ for 22 minutes | Invasive mode only: | Similar |
| Low temperature alarm - | Non-invasive mode: Patient-side temperature < $29°C$ for 22 | After 10 minutes @ $29.5°C$ | |
| | minutes | or | |
| | | After 60 minutes @ $34.5°C$ | |
| Sound Pressure Level | Alarms exceed 50 dBA @ 1m | Alarms exceed 50 dBA @ 1m | Similar |
| 510(k) Number | Subject Device | K073706 Predicate | Comment |
| Device Name | Vincent Medical inspiredTM VHB20 Heated Humidifier | Fisher & Paykel MR850 Respiratory Humidifier | |
| Performance | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.