← Product Code [OTN](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTN) · K222293

# ArcSP Suprapubic Sling System (K222293)

_Urocure, LLC · OTN · Sep 21, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTN/K222293

## Device Facts

- **Applicant:** Urocure, LLC
- **Product Code:** [OTN](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTN.md)
- **Decision Date:** Sep 21, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Implantable Sling intended to treat female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

## Device Story

ArcSP Suprapubic Sling System is a surgical mesh device for treating female stress urinary incontinence. System comprises a knitted polypropylene mesh with integrated bioresorbable suture, plastic insertion sheaths, and two stainless-steel needle delivery assemblies. Device is used in a clinical setting by a physician. During surgery, needles facilitate placement of the sling mid-urethra via a suprapubic approach. The mesh and suture are permanently implanted to provide support to the urethra, addressing hypermobility or intrinsic sphincter deficiency. The delivery needles are single-use tools and are not implanted. The device functions by providing mechanical support to the urethra, which helps restore continence and improve patient quality of life.

## Clinical Evidence

No clinical data. Bench testing only. Performance specifications including dimensional, needle functional, sling functional, fatigue, tensile, packaging, usability, distribution, and biocompatibility testing were conducted or leveraged from the predicate.

## Technological Characteristics

Materials: knitted polypropylene mesh, bioresorbable suture, stainless-steel needles, plastic sheaths. Energy: None (mechanical). Connectivity: None. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Form factor: Single-use sterile kit with delivery needles and sheathed mesh. Biocompatibility: ISO 10993-1:2009 compliant.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- ArcTV Transvaginal Sling System ([K183134](/device/K183134.md))

## Reference Devices

- SPARC Sling System ([K081613](/device/K081613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

September 21, 2022

UroCure LLC % Denise Lenz Regulatory Consultant Libra Medial, Inc. 8401 73rd Ave N. Suite 63 Brooklyn Park, MN 55428

K222293 Re:

> Trade/Device Name: ArcSP Suprapubic Sling System Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTN Dated: August 31, 2022 Received: September 6, 2022

Dear Denise Lenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K22293

Device Name ArcSP Suprapubic Sling System

Indications for Use (Describe)

The ArcSP Suprapubic Sling System is indicated to be placed at the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 7 510(K) SUMMARY

### 7.1 ADMINISTRATIVE INFORMATION

Date of Summary Preparation: July 29, 2022

#### 7.2 SUBMITTER INFORMATION

John Nealon Submitter UroCure, LLC 701 North 3rd Street, Suite 110, Minneapolis, MN 55401 Phone: 612-850-6814 Email: john.nealon(@urocure.com

#### 7.3 CONTACT INFORMATION

| Primary Submission Contact   | Denise Lenz               |
|------------------------------|---------------------------|
|                              | Regulatory Consultant,    |
|                              | Libra Medical Inc.        |
|                              | Phone: 763-232-3701       |
|                              | Email: dlenz@libramed.com |
| Secondary Submission Contact | Sew-Wah Tay, PhD          |

| Secondary Submission Contact | Sew-Wah Tay, PhD<br>Regulatory Consultant,<br>Libra Medical Inc.<br>Phone: 612-801-6782<br>Email: swtay@libramed.com |
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
|------------------------------|----------------------------------------------------------------------------------------------------------------------|

#### DEVICE INFORMATION 7.4

| Trade Name                                 | ArcSP Suprapubic Sling System |
|--------------------------------------------|-------------------------------|
| Common Name                                | Urinary sling                 |
| Classification Name                        | Surgical Mesh                 |
| Classification Regulation                  | 878.3300                      |
| Class                                      | II                            |
| Panel                                      | Gastroenterology/Urology      |
| Product Code                               | OTN                           |
| Prior Regulatory Submissions<br>for Device | None                          |

### 7.5 510(K) TYPE AND REASON FOR SUBMISSION

This 510(k) is a Special 510(k) and is submitted to obtain marketing clearance for the ArcSP Suprapubic Sling System, a modification to the delivery device for the predicate ArcTV

{4}------------------------------------------------

Transyaginal Sling. The implantable portion of the device is identical to the predicate (ArcTV), with the delivery tool being the only difference from the predicate. ArcTV.

### 7.6 PREDICATE DEVICE

The predicate is ArcTV Transvaginal Sling System (K183134).

#### DEVICE DESCRIPTION 7.7

The ArcSP Suprapubic Sling System consists of the following:

- 1 ArcSP Sling with bio-resorbable Suture with plastic sheath ●
- 2 ArcSP Handle and Needle Delivery Assemblies .

The Handle and Needle delivery assembly is a sterile, single-use system, consisting of a two curved, stainless-steel Needles with attached Handles. The conical tip of each delivery Needle is designed to allow for passage through tissue. The Needles are a tool and together with the Sling Connectors and Sheaths, facilitate placement of the Sling and are not implanted.

The ArcSP Sling assembly is identical to the predicate. ArcTV Sling assembly and includes one knitted polypropylene mesh with an integrated bioresorbable Suture, two removable plastic insertion Sheaths, and two Sling Connectors. The Suture is an integral feature of the Mesh. The integrated Suture helps to minimize deformation of the mesh during sling insertion and placement and allows for adjustment of the Sling after removal of the Sheaths. The two plastic Sheaths cover and facilitate placement of the Sling. The Mesh and the bioresorbable Suture are permanently implanted.

### 7.8 INTENDED USE

Implantable Sling intended to treat female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

#### 7.9 INDICATIONS FOR USE

The ArcSP Suprapubic Sling System is indicated to be placed mid-urethra for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

# 7.10 PERFORMANCE DATA

The ArcSP Suprapubic Sling System has been tested to meet the device's intended use and to ensure conformance to the product specifications.

The ArcSP meets all its physical and performance specifications including:

- Dimensional .
- Needle Functional Testing ●
- *Sling Functional Testing ●
- *Fatigue
- *Tensile
- *Packaging Testing

{5}------------------------------------------------

- Usability Testing ●
- *Distribution Testing ●
- *Biocompatibility

* No testing was repeated for this submission since the Sling implant assembly and packaging are identical to the predicate, ArcTV.

# 7.11 DEVICE COMPARISON

The ArcSP Suprapubic Sling System consists of the following:

- 1 ArcSP Sling with bio-resorbable Suture and plastic sheath
- 2 ArcSP Handle and Needle Delivery Assemblies .

The ArcSP Suprapubic Sling is identical to the predicate sling, ArcTV. The delivery approach of the sling is suprapubic for the subject device (ArcSP) versus transvaginal for the predicate device (ArcTV); therefore, the shape of the needle has changed to facilitate a suprapubic delivery. The final placement of the sling is the same for both. The two different Needles are to accommodate physician's preference of approach to implant.

The following table compares the subject device, ArcSP to the predicate device, ArcTV and the reference device, SPARC.

| Device<br>Characteristics                | Subject Device<br>ArcSP Suprapubic Sling<br>System                                                                                                                                                                                                          | Predicate Device<br>ArcTV Transvaginal<br>Sling System (K183134)                                                                                                                                                                       | Reference Device<br>SPARC Sling System<br>(K081613)                                                                                                                                                                 |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                             | Same as predicate<br>Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD).                                                         | Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD).                                                         | Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD).                                      |
| Device<br>Characteristics                | Subject Device<br>ArcSP Suprapubic Sling<br>System                                                                                                                                                                                                          | Predicate Device<br>ArcTV Transvaginal<br>Sling System (K183134)                                                                                                                                                                       | Reference Device<br>SPARC Sling System<br>(K081613)                                                                                                                                                                 |
| Indication for<br>Use                    | Same as predicate<br>The polypropylene sling<br>is indicated to be placed<br>mid-urethra for the<br>treatment of female stress<br>urinary incontinence<br>(SUI) resulting from<br>urethral hypermobility<br>and/or intrinsic sphincter<br>deficiency (ISD). | The polypropylene sling<br>is indicated to be placed<br>mid-urethra for the<br>treatment of female stress<br>urinary incontinence<br>(SUI) resulting from<br>urethral hypermobility<br>and/or intrinsic sphincter<br>deficiency (ISD). | Intended for the<br>placement of pubourethral<br>sling for the treatment of<br>female stress urinary<br>incontinence (SUI)<br>resulting from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency. |
| Regulation<br>Number and<br>Product Code | Same as predicate                                                                                                                                                                                                                                           | 878.3300<br>OTN                                                                                                                                                                                                                        | 878.3300<br>OTN                                                                                                                                                                                                     |
| Regulatory<br>Classification<br>Name     | Same as predicate                                                                                                                                                                                                                                           | Surgical mesh                                                                                                                                                                                                                          | Surgical mesh                                                                                                                                                                                                       |
| Device<br>Configuration                  | Same as predicate                                                                                                                                                                                                                                           | Single use, sterile device                                                                                                                                                                                                             | Single use, sterile device                                                                                                                                                                                          |
| Entry Method                             | Same as technological<br>reference                                                                                                                                                                                                                          | Transvaginal Retropubic                                                                                                                                                                                                                | Suprapubic Retropubic                                                                                                                                                                                               |
| Delivery System                          | Same as technological<br>reference                                                                                                                                                                                                                          | Handle with detachable<br>curved Needle                                                                                                                                                                                                | Curved needle with<br>attached handle                                                                                                                                                                               |
| Sling                                    | Same as predicate                                                                                                                                                                                                                                           | Sheathed polypropylene<br>mesh with adjustment<br>bioresorbable suture                                                                                                                                                                 | Sheathed polypropylene<br>mesh with adjustment<br>bioresorbable suture                                                                                                                                              |
| Sterilization                            | Same as predicate                                                                                                                                                                                                                                           | EO, SAL 10-6-, single use                                                                                                                                                                                                              | EO, SAL 10-6-, single use                                                                                                                                                                                           |
| Biocompatibility                         | Same as predicate                                                                                                                                                                                                                                           | Meets ISO 10993-1:2009,<br>Biological evaluation of<br>medical devices<br>requirements                                                                                                                                                 | Meets ISO 10993-1:2009,<br>Biological evaluation of<br>medical devices<br>requirements                                                                                                                              |
| Packaging                                | Same as predicate                                                                                                                                                                                                                                           | Molded tray with Tyvek<br>lid and separate pouch for<br>Sling                                                                                                                                                                          | Molded tray with Tyvek<br>lid and separate pouch for<br>Sling                                                                                                                                                       |

Table 7-1: Comparison of Device Characteristics to Marketed Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

### 7.12 SUBSTANTIAL EQUIVALENCE

The ArcSP Suprapubic Sling System covered by this substantially equivalent to the predicate ArcTV Transvaginal Sling System device (K183134). The minor difference in the shape and size of the needle to optimize the ArcSP's top-down needle passage and technique do not raise new safety and efficacy as the shape, size and stiffness of the needle is the same as the technological reference.

## 7.13 CONCLUSION

Since the implantable sling assembly system is identical to the predicate, ArcTV and the intended use is the same, the ArcSP is substantially equivalent to its predicate. The minor difference in the shape, size and stiffness of the needle does not raise any new questions of safety or efficacy.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTN/K222293](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTN/K222293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
