K222280 · Bigfoot Biomedical, Inc. · QLG · Mar 1, 2023 · Clinical Chemistry
Device Facts
Record ID
K222280
Device Name
Bigfoot Unity® Diabetes Management System
Applicant
Bigfoot Biomedical, Inc.
Product Code
QLG · Clinical Chemistry
Decision Date
Mar 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1355
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bigfoot Unity® Diabetes Management System is indicated for the management of diabetes in persons age 12 years and older. Bigfoot Unity® provides glucose monitoring data via the Abbott FreeStyle Libre 2 Flash Glucose Monitoring sensor. The system incorporates real time alarm capabilities and is designed to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated. The device is intended to provide insulin dose information using the available glucose data to assist persons with diabetes mellitus who use disposable pen-injectors for the self-injection of insulin in implementing health care provider recommended insulin dose reqimens. The device is intended for single patient use only and requires a prescription. Bigfoot Unity® is also intended to communicate autonomously with digitally connected medical devices where the user manually controls therapy decisions.
Device Story
System integrates continuous glucose monitoring (CGM) with insulin dose recommendations for diabetes management; supports users of disposable insulin pens. Inputs: interstitial glucose data from Abbott FreeStyle Libre 2 sensor. Components: two reusable insulin pen caps (rapid-acting and long-acting) and mobile application. Operation: pen caps communicate via NFC/Bluetooth; rapid-acting cap displays glucose value, trend, and calculated correction/meal insulin doses; long-acting cap displays prescribed dose. Mobile app manages system alerts, stores provider-prescribed regimens, and enables cloud data transfer. Used in home environment by patients; users manually select and administer insulin based on device recommendations and clinical context. Benefits: provides accessible, real-time insulin dose guidance to assist in manual therapy decisions.
Clinical Evidence
No new clinical data provided. Clinical performance established in previous submissions (K193371, K202145). Human factors and usability testing conducted per FDA guidance to validate safe use of new features.
Technological Characteristics
System includes Abbott FreeStyle Libre 2 iCGM sensor, two reusable insulin pen caps, and mobile app. Connectivity via NFC and Bluetooth. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare), and RTCA/DO-160 (airworthiness). Software performs data processing, alert management, and dose calculation based on provider-prescribed regimens.
Indications for Use
Indicated for management of diabetes in persons age 12 years and older who use disposable insulin pen-injectors for self-injection and require insulin dose assistance based on healthcare provider-prescribed regimens.
Regulatory Classification
Identification
An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.
Predicate Devices
Bigfoot Unity® Diabetes Management System (K202145)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K222280
B Applicant
Bigfoot Biomedical, Inc.
C Proprietary and Established Names
Bigfoot Unity® Diabetes Management System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QLG | Class II | 21 CFR 862.1355 - Integrated Continuous Glucose Monitoring System | CH - Clinical Chemistry |
| QOG | Class II | 21 CFR 880.5860 - Piston Syringe | General Hospital |
## II Submission/Device Overview:
A Purpose for Submission:
Modification to an existing device to add the Android mobile app as an alternate primary display and to add the silence all alarms feature
B Measurand:
Glucose in interstitial fluid
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
C Type of Test:
Quantitative, amperometric assay via glucose oxidase
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Bigfoot Unity® Diabetes Management System is indicated for the management of diabetes in persons age 12 years and older.
Bigfoot Unity® provides glucose monitoring data via the Abbott FreeStyle Libre 2 Flash Glucose Monitoring sensor. The system incorporates real time alarm capabilities and is designed to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated. The device is intended to provide insulin dose information using the available glucose data to assist persons with diabetes mellitus who use disposable pen-injectors for the self-injection of insulin in implementing health care provider recommended insulin dose regimens. The device is intended for single patient use only and requires a prescription.
Bigfoot Unity® is also intended to communicate autonomously with digitally connected medical devices where the user manually controls therapy decisions.
C Contraindications
The System must not be used with automated insulin dosing (AID) systems, including closed loop and insulin suspend systems.
The System must be removed prior to Magnetic Resonance Imaging (MRI), Computed Tomography (CT) scan, or high-frequency electrical heat (diathermy) treatment. The effect of MRI, CT scans, or diathermy on the performance of the Sensor has not been evaluated
D Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Taking ascorbic acid (vitamin C) supplements while wearing the Sensor may falsely raise Sensor glucose readings. Taking more than 500 mg of ascorbic acid per day may affect the Sensor readings which could result in a severe low glucose event.
Do not use the System in people less than 12 years of age.
Do not use the System if you are pregnant, on dialysis or critically ill. The System is not cleared for use in these groups and it is not known how different conditions or medications common to these populations may affect performance of the System.
Performance of the System has not been evaluated when used with other implanted medical devices, such as pacemakers.
K222280 - Page 2 of 10
{3}
The System does not support insulin doses in half-unit increments. Patients that require half doses should not use the System.
The System supports once daily dosing of long-acting insulin. Patients who take more than one daily dose of long-acting insulin should not use the System.
Do not ignore symptoms that may be due to low or high blood glucose: if you are experiencing symptoms that are not consistent with the glucose readings, consult your health care professional.
Use your blood glucose meter to make diabetes treatment decisions when you see the "check blood glucose" symbol during the first 12 hours of wearing a Sensor, if your Sensor glucose reading does not match how you feel, or if the reading does not include a number.
Wash application site on the back of your upper arm using a plain soap, dry, and then clean with an alcohol wipe. This will help remove any oily residue that may prevent the sensor from sticking properly. Allow site to air dry before proceeding. Carefully preparing the site according to these instructions will help the Sensor stay on your body for the full 14-day wear period and help prevent it from falling off early.
Store the Sensor Kit between 36°F and 82°F. If you suspect that the temperature may exceed 82°F (e.g., an un-airconditioned home in the summer), you should refrigerate your Sensor Kit. Do not freeze your Sensor Kit.
You must scan the Sensor to get your current glucose level as the Bigfoot Unity App will not provide this information.
Take standard precautions for transmission of blood borne pathogens to avoid contamination.
Use of the Sensor with devices, apps, and software that are not listed by the manufacturer as compatible with the System may cause inaccurate glucose readings.
If a Sensor breaks inside your body, call your health care professional.
If you use an Android phone, do not lower the notification vibration. If you use an Android phone, do not lower the notification vibration intensity to the lowest level in your phone's Vibration Intensity menu. At the lowest level, you may not detect vibrations with your alerts, including the Very Low Glucose Alert.
If you use an iPhone, do not turn off vibration in the iOS Sound & Haptics or Accessibility menus. If you turn off any vibration setting, you will not get any vibrations with your alerts, including the Very Low Glucose Alert.
E Special Instrument Requirements:
Not Applicable
K222280 - Page 3 of 10
{4}
K222280 - Page 4 of 10
## IV Device/System Characteristics:
### A Device Description:
The Bigfoot Unity utilizes continuous glucose monitoring to support people with diabetes mellitus who use disposable insulin pens for self-injection of insulin. The system consists of the Abbott FreeStyle Libre 2 integrated continuous glucose monitor (iCGM), two reusable insulin pen caps (one each for rapid-acting and long-acting insulin pens) and a mobile application. The components communicate via near field communication (NFC) and Bluetooth Low Energy (BLE).
The Bigfoot Unity contains a digital lookup table that displays insulin dose recommendations on the Bigfoot Unity pen caps based on healthcare provider (HCP)-prescribed regimens. The System also provides a set of meal dose recommendations based on meal size, where the meal doses are set by the healthcare provider (HCP) and entered in the lookup table.
The device generates glucose data using the FreeStyle Libre 2 sensor and displays the data (value and trend) on the rapid-acting insulin pen cap (the White Cap) when the user scans the sensor using the White Cap. The rapid-acting pen cap also displays correction and meal insulin doses based upon the lookup table and available glucose data. The long-acting pen cap (the Black Cap) displays the long-acting insulin dose prescribed by the user's healthcare provider. From the dose recommendations on the pen caps, as well as other contextually relevant information such as glucose trend arrows, users manually select an insulin dose and administer it using the pens according to the insulin manufacturers' instructions. In addition to displaying dose information, both pen caps track the time of insulin doses.
The mobile app allows user entry, in consultation with their HCP, of a prescribed insulin dosing regimen into the digital lookup table, as well as provides system alerts and historical information. The mobile app has a built-in, mandatory low-glucose alarm at 55 md/dL and an optional low-glucose alarm at 70 mg/dL. The alarms are able to be silenced for up to 4 hours through the mobile app, but alarms will still be pushed as notifications and vibrate as they occur. The mobile app receives glucose data from the sensor via BLE and will alert if the sensor glucose reading is below the threshold(s). However, in order to obtain the glucose sensor value, the user must scan the sensor with the White Cap. Additionally, the mobile app manages the secure wireless communication between the system components and enables the transfer of the system data to the cloud.
### B Principle of Operation:
Bigfoot Unity incorporates the FreeStyle Libre 2 continuous glucose sensor and glucose translation library for calculating glucose values and trend information. Additionally, Bigfoot Unity uses the FreeStyle Libre 2 wireless communication protocols to (1) obtain glucose values and trend information on the White Cap, and (2) provide real-time alerts in the mobile app.
{5}
K222280 - Page 5 of 10
C Instrument Description Information:
1. Instrument Name:
Bigfoot Unity
2. Specimen Identification:
Not Applicable
3. Specimen Sampling and Handling:
Not Applicable
4. Calibration:
The sensor is factory calibrated and does not require calibration from the user/operator.
5. Quality Control:
Not Applicable
V Substantial Equivalence Information:
A Predicate Device Name(s):
Bigfoot Unity Diabetes Management System
B Predicate 510(k) Number(s):
K202145
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K222280 | K202145 |
| --- | --- | --- |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Bigfoot Unity Diabetes Management System provides glucose monitoring data via the Abbott FreeStyle Libre 2 Flash Glucose Monitoring sensor. The system incorporates real time alarm capabilities and is designed to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated. | Same |
{6}
| Primary Principle of Operation | Amperometric measurement of current proportional to glucose concentration in interstitial fluid via glucose oxidase chemical reaction. | Same |
| --- | --- | --- |
| Sample Type | Interstitial fluid | Same |
| Enzyme | Glucose oxidase | Same |
| Clinical Setting/Sites of Use | Home use | Same |
| Glucose Trend Arrow | ↑, more than +2 mg/dL/min
☑, +1 and +2 mg/dL/min
→, -1 to +1 mg/dL/min
☐, -2 to -1 mg/dL/min
↓, more than -2 mg/dL/min | Same |
| Sensor Calibration | Factory Calibrated | Same |
| Sensor Warm-up Time | 1 hour | Same |
| Restriction on Non-adjunctive Use | Not for use to make treatment decisions in the first 12 hours | Same |
| Sensor Glucose Reading Range | 40-400 mg/dL | Same |
| General Device Characteristic Differences | | |
| Mobile Phone Operating System | Android and Apple iOS | Apple iOS |
| Mandatory Alerts | Able to silence all alerts and alarms (includes very low glucose alarm <55 mg/dL) for up to 4 hours while maintaining vibration annunciation.
User can manually turn off specific alerts and alarms in the mobile application except for very low glucose alarm when not in silence mode. | Mandatory low glucose alarm (<55 mg/dL) cannot be silenced.
Users cannot turn off specific alerts and alarms. |
VI Standards/Guidance Documents Referenced:
- 21 CFR 862.1355 (integrated continuous glucose monitoring system (iCGM)) special controls
- FDA Guidance document Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016)
- FDA Guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- FDA Guidance document Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices (September 6, 2017)
K222280 - Page 6 of 10
{7}
- ISO 10993-1 (5th Edition): 2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5 (3rd Edition): 2009 – Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ANSI/AAMI ES60601-1: 2005/(R)2012 and A1: 2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 (Ed. 4.0): 2014 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements and Tests
- IEC 60601-1-11 (Ed. 2.0): 2015 - Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ANSI IEEE C63.27-2017 - American National Standard For Evaluation Of Wireless Coexistence
- ASTM D4332-14 Standard Practice For Conditioning Containers, Packages, Or Packaging Components For Testing
- ASTM D4169-16 Standard Practice For Performance Testing Of Shipping Containers And Systems
- ISO 14971 (2nd Edition): 2007 - Medical Devices – Application of risk management to medical devices
- ISO 15223-1 (3rd Edition): 2016 - Medical devices -- Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements 2016
- IEC 62366-1 (1.0 Edition): 2015 - Medical devices – Part 1: Application of usability engineering to medical devices
- IEC 62304 (Edition 1.1): 2015 Consolidated Version - Medical device software - Software life cycle processes
- AAMI TIR57: 2016 - Principles for medical device security- Risk Management
- ANSI UL 2900-1 (1st Edition): 2017 - Standard for Software Cybersecurity Network-Connectable Products, Part 1: General Requirements
- ANSI IEEE C63.18-2014 –American National Standard Recommended Practice for an On-site, Ad Hoc Test Method for Estimating Electromagnetic Immunity of Medical Devices to Radiated Radio-Frequency (RF) Emissions from RF Transmitters
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. Precision/Reproducibility:
Precision was established in K202145 and there are no changes in this submission.
K222280 - Page 7 of 10
{8}
2. Linearity:
Linearity was established in K202145 and there are no changes in this submission.
3. Analytical Specificity/Interference:
Analytical specificity and interference parameters were established in K202145 and there are no changes in this submission.
4. Assay Reportable Range:
Assay reportable range was established in K202145 and there are no changes in this submission.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Same as in K193371.
6. Detection Limit:
Unchanged since K193371.
7. Assay Cut-Off:
Not Applicable
8. Accuracy (Instrument):
Not Applicable
9. Carry-Over:
Not Applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not Applicable. Accuracy is determined by comparing device values to an FDA cleared laboratory grade glucose measurement method and was previously assessed in K193371.
2. Matrix Comparison:
Not Applicable. Interstitial fluid is the only indicated matrix.
K222280 - Page 8 of 10
{9}
K222280 - Page 9 of 10
C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable
2. Clinical Specificity:
Not Applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Clinical performance of the device was established in K193371 and K202145.
D Clinical Cut-Off:
Not Applicable
E Expected Values/Reference Range:
Not Applicable
F Other Supportive Instrument Performance Characteristics Data:
Human Factors
Human factors and usability testing of the System was conducted in accordance with the FDA Guidance, Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016) to evaluate the safe and effective use of the Bigfoot Unity for the intended users, uses, and use environment.
Validation data demonstrated that use-related risk was brought to an acceptable level and that no serious use errors or problems associated with high levels of residual risk remain.
Software Verification and Validation
Software verification and validation testing was conducted to confirm that the software used in the Bigfoot Unity performed in accordance with established specifications and followed FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)."
Electromagnetic Compatibility
Electromagnetic compatibility (EMC) testing was performed for the Bigfoot Unity to verify that the system demonstrated compliance with requirements of IEC 60601-1-2:2014 (ed. 4) and Federal Communication Commission Regulations Part 15, Subpart B. In the context of electromagnetic emissions and immunity, testing and the accompanying documents (or labeling) support the safe use of the system in the home healthcare environment. Bigfoot performed testing to AAMI TIR69:2017: Risk Management of Radio-Frequency Wireless Coexistence For Medical Devices And Systems and ANSI/IEEE C63.27:2017: American National Standard for Evaluation of Wireless Coexistence and demonstrated wireless coexistence for their device.
{10}
K222280 - Page 10 of 10
## Electrical Safety
The basic safety and essential performance of the Bigfoot Unity was evaluated in accordance with ANSI/AAMI ES6060-1: 2005/(R)2012 and A1: 2012, C1:2009/(R)2012 and A2:2010/(R)2012, as well as IEC 60601-1-11 (ed. 2.0): 2015 and demonstrated compliance to the requirements.
## Cybersecurity
The sponsor has provided cybersecurity risk management documentation for the Bigfoot Unity that includes the cybersecurity mitigations of the Abbott Diabetes Care FreeStyle Libre 2 CGM. Bigfoot has provided their analysis of threats to confidentiality, integrity, and availability for data and software in the System. Bigfoot identified vulnerabilities and completed a risk assessment for the impact to confidentiality, integrity, and availability. Risk mitigation controls have been implemented and verified.
In addition, Bigfoot has controls and processes in place to ensure continued support for keeping the device secure and to ensure that the device firmware, software and components are malware free. Bigfoot has provided a Software Bill of Materials (SBOM), which captures relevant details for each of the software components utilized in the system, including a review of the National Vulnerability Database (NVD) to assess if each software component has any open vulnerabilities. Additional controls are in place in manufacturing through distribution to ensure that the medical device firmware and software remain malware free from point of origin to the hands of the end user.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.