← Product Code [DPW](/productcode/DPW) · K222242

# FloPatch FP120 (K222242)

_Flosonics Medical · DPW · Dec 9, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K222242

## Device Facts

- **Applicant:** Flosonics Medical
- **Product Code:** [DPW](/productcode/DPW.md)
- **Decision Date:** Dec 9, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode. The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

## Device Story

Non-invasive blood flow detection device; uses ultrasound and Doppler effect to assess carotid artery blood flow. Consists of signal processing unit and adhesive strap; transmits ultrasonic waves to peripheral vessel; receives reflected Doppler-shifted waves. Outputs signal wirelessly to mobile medical application; application processes signal to display Max Velocity trace, Max VTI, and Corrected Flow Time. Used in hospital/professional settings by physicians/nurses. Provides real-time hemodynamic data to assist clinical decision-making; benefits patient through non-invasive monitoring.

## Clinical Evidence

Bench testing only. Testing included EMC (CISPR B), biocompatibility (ISO 10993), packaging (ISTA3), electrical safety (IEC 60601-1), software functional verification, dust ingress (IP6X), transport/shock/vibration testing, and acoustic output verification (ISPTA ≤720 mW/cm2, MI ≤1.9).

## Technological Characteristics

Surface device for intact skin; 4 MHz center frequency; CW Doppler mode. LiPo battery powered. Wireless connectivity to mobile app. Dimensions: 54x35x18mm (without adhesive). Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1 (Type BF, defibrillation protected). EMC per IEC 60601-1-2. Ultrasound safety per IEC 60601-2-37. IP67 ingress protection.

## Regulatory Identification

A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.

## Predicate Devices

- FloPatch FP120 ([K200337](/device/K200337.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 9, 2022

Flosonics Medical Caleb Chin Director of Quality and Regulatory 325 Front St W, Floor 4 Toronto. Ontario M5V 2Y1 Canada

Re: K222242

Trade/Device Name: FloPatch FP120 Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: October 20, 2022 Received: November 7, 2022

Dear Mr. Chin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Robert T. Kazmierski - S

for

Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Device Name FloPatch FP120

#### Indications for Use (Describe)

The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode.

The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Flosonics Medical. The logo features a stylized blue and black icon above the company name. The icon resembles a stylized letter 'f' with curved lines emanating from it, suggesting a wave or signal. Below the icon, the word "FLOSONICS" is printed in a bold, sans-serif font, with the word "MEDICAL" underneath in a smaller, lighter font.

## SECTION 5 – 510(k) Summary

#### 1. Submitter Information

| Submitter:     | Flosonics Medical                                        |
|----------------|----------------------------------------------------------|
| Address:       | 325 Front St W, Floor 4<br>Toronto, ON<br>Canada M5V 2Y1 |
| Telephone:     | 1-289-998-2982                                           |
| Contact:       | Caleb Chin                                               |
| Date Prepared: | July 19, 2022                                            |

#### 2. Device Information

| Trade Name:          | FloPatch FP120                 |
|----------------------|--------------------------------|
| Common Name:         | Cardiovascular Blood Flowmeter |
| Classification:      | Class II per CFR 870.2100      |
| Classification Name: | Cardiovascular blood flowmeter |
| Product Code:        | DPW                            |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for product modifications to the previously cleared device.

Product modifications include:

- Adhesive material ●
- . Device packaging
- . Software including but not limited to:
	- Addition of a Guided Assessment workflow with smart window algorithm O
	- New software features o
	- Signal Processing Algorithm Changes o
- . Instructions for use to update acoustic output values

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Flosonics Medical. The logo features a stylized letter 'f' in blue, with three curved lines emanating from it, resembling radio waves or sound waves. Below the symbol, the word "FLOSONICS" is written in a bold, sans-serif font, with the word "MEDICAL" underneath in a smaller font, with each letter separated by a space.

#### 4. Predicate Device Information

| 510(k) No. | Device         | Manufacturer      |
|------------|----------------|-------------------|
| K200337    | FloPatch FP120 | Flosonics Medical |

#### 5. Device Description

The FloPatch (FP120) is a non-invasive blood flow detection device to be used in a medical/hospital setting for use by a medical professional. The device uses ultrasound and the Doppler effect to assess the flow of blood. The device consists of a signal processing unit and an adhesive strap. The device transmits ultrasonic waves from the ultrasonic transducer to a peripheral vessel such as the carotid artery. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing unit which outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time.

#### 6. Intended Use

The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode.

The device is intended to be used by medical professionals, such as physicians and nurses, in hospital and professional environments The device is intended for prescription use on adults only.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Flosonics Medical. The logo features a stylized letter 'f' in blue, with three curved lines emanating from the top right of the 'f', resembling radio waves. Below the symbol, the word "FLOSONICS" is written in a bold, sans-serif font, with the word "MEDICAL" in a smaller font size and a lighter blue color underneath.

#### 7. Comparison to Predicate Device

| Feature/<br>Characteristic                                    | FloPatch (FP120)<br>[Subject Device]                                                                                                                                                                                                                                                                                                                                                                                                                                                 | FloPatch (FP120)<br>Primary Predicate [K200337]                                                                              |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Class/Classification/Product<br>Code                          | Class II/DPW (21 CFR 870.2100<br>Cardiovascular blood<br>flowmeter)                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Intended Use                                                  | The FloPatch FP120 is indicated<br>for use for the noninvasive<br>assessment of blood flow in the<br>carotid artery. FloPatch FP120<br>operates in a single mode, the<br>Continuous Wave (CW) mode,<br>and is not capable of operating<br>in any other mode.<br>The device is intended to be<br>used by medical professionals,<br>such as physicians and nurses,<br>in hospitals and professional<br>environments. The device is<br>intended for prescription use on<br>adults only. | Same                                                                                                                         |
| Indications for Use                                           | Identical to Intended Use                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                         |
| Intended Users                                                | Medical professionals such as<br>Physicians and Nurses                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                         |
| Use environment                                               | Hospitals and professional<br>environments such as clinics<br>and doctor's offices.                                                                                                                                                                                                                                                                                                                                                                                                  | Same.                                                                                                                        |
| Patient Population                                            | Adults, ages 18 years and older                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                         |
| Intended for Prescription Use                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Installation and Use                                          | Body Worn                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                         |
| Theory of Operation                                           | Use of the Doppler effect to<br>evaluate the flow velocity of<br>blood in peripheral vasculature.                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                         |
| Center Frequency                                              | 4 MHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                         |
| Global Maximum<br>Outputs/Worst Case Setting                  | Max ISPTA.3 (mW/cm2) - 172.05<br>Max MI - 5.22E-02                                                                                                                                                                                                                                                                                                                                                                                                                                   | Max ISPTA.3 (mW/cm2) - 15.78<br>Max MI - 1.11E-02                                                                            |
| Modes of Operation                                            | One mode, continuous                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                         |
| Reusable                                                      | No, the device is single use for a<br>single patient.                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                         |
| Dimensions                                                    | With adhesive<br>Height 114 mm<br>Width 70 mm<br>Depth 32 mm<br>Without Adhesive<br>Height 54mm<br>Width 35 mm<br>Depth 18 mm                                                                                                                                                                                                                                                                                                                                                        | With adhesive<br>Height 200 mm<br>Width 65 mm<br>Depth 30mm<br>Without Adhesive<br>Height 54mm<br>Width 35 mm<br>Depth 18 mm |
| Weight                                                        | 22 gms                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                         |
| The degree of protection<br>against harmful ingress of liquid | IP67                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | IPX7                                                                                                                         |
| Type of Power Source                                          | LiPo Battery (IEC 62133<br>certified)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                         |
| Battery Operating Voltage                                     | 4.2 V for the battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                         |
| Battery Chemistry                                             | Lithium Polymer                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                         |
| The degree of protection<br>against electric shock            | Type BF (Defibrillation<br>Protected)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Type B (Defibrillation Protected)                                                                                            |
| Buttons                                                       | One Power Button on FloPatch<br>FP120 hardware                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                         |
| Status LED                                                    | One, power and battery<br>Indicator                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Onboard Screen                                                | None - Multi Touch Mobile<br>Medical Application screen                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                         |
| Displays Doppler Waveform                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Displays Max Velocity<br>Waveform                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Displays VTI Calculation                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Displays Corrected Flow Time<br>Calculation                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Displays Peak Systolic Velocity                               | Yes – in a different form                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                          |
| Wireless Mobile Application                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Calibration Required                                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                         |
| Maintenance                                                   | Single-use device                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                         |
| Contact Classification                                        | Surface Device, Intact Skin<br>Contacting, Contact Duration:<br><24 hrs                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                         |
| Electrical Safety                                             | IEC 60601-1:2005 (Third Edition)<br>+ CORR. 1:2006 + CORR.2:2007<br>+ A1:2012                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                         |
| EMC                                                           | IEC 60601-1-2:2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                         |
| Ultrasound Basic Safety and<br>Essential Performance          | IEC 60601-2-37:2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                         |
| Biocompatibility                                              | ISO 10993-1, -5, -10, -12                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                         |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Flosonics Medical. The logo consists of a stylized blue and black symbol resembling a curved sound wave or radio wave above the company name. The company name "FLOSONICS" is written in a bold, sans-serif font, with the word "MEDICAL" in a smaller, lighter blue font below it.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Flosonics Medical. The logo features a stylized letter 'f' in blue, with three curved lines emanating from the top right of the 'f', resembling radio waves or sound waves. Below the symbol, the word "FLOSONICS" is written in a sans-serif font, with the word "MEDICAL" written in a smaller font size below it.

#### 8. Performance Data

This submission is for modifications to the FloPatch FP120 cleared in 510(k) K200337.

The following tests were performed to demonstrate the substantial equivalence of the modified device to its predicate.

| Test                         | Brief Description                                                         | Result |
|------------------------------|---------------------------------------------------------------------------|--------|
| EMC                          | Verified that the device meets Class B requirements.<br>CISPR B           | Pass   |
| Biocompatibility             | Verified that the adhesive met biocompatibility<br>requirements           | Pass   |
| Material and Shelf Life Test | Verified and validated that the adhesive satisfies<br>performance testing | Pass   |
| Packaging Test               | Verified that the packaging met ISTA3 and<br>performance testing          | Pass   |
| Electrical Safety Testing    | Verified that the device meets applied part BF<br>requirements            | Pass   |

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for Flosonics Medical. The logo features a stylized letter 'f' in blue, with three curved lines emanating from it, resembling radio waves or sound waves. Below the symbol, the word "FLOSONICS" is written in a bold, sans-serif font, with the word "MEDICAL" underneath in a smaller, lighter font.

| Software Testing  Display Metrics Flow Accuracy and Depth Velocity Accuracy Saved Measurement Unit Appearance Session Length | Verified that the software application satisfies functional requirements and performs as intended. Algorithms and calculations were also verified. | Pass |
|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Dust Ingress Testing                                                                                                         | Verified that the device met IP6X dust ingress requirements                                                                                        | Pass |
| Emergency Services Equipment Testing                                                                                         | Verified the device met transport medical equipment requirements.  Shock testing Vibration testing Free fall testing                               | Pass |
| Acoustic Output (Track 3)                                                                                                    | Verified that global maximum derated ISPTA ≤720 mW/cm2 and the global maximum MI should be ≤ 1.9                                                   | Pass |
| Ultrasound Safety                                                                                                            | Verified that the device meets IEC 60601-2-37                                                                                                      | Pass |

#### 9. Conclusion

Based on the indications for use, technological characteristics, performance testing and comparison to the predicate device, the FloPatch FP120 has been shown to be substantially equivalent to the legally marketed predicate device identified in this submission and does not present any changes to safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K222242](https://fda.innolitics.com/device/K222242)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
