Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly
K222232 · Bard Access Systems, Inc. · LJS · Aug 24, 2022 · General Hospital
Device Facts
Record ID
K222232
Device Name
Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly
Applicant
Bard Access Systems, Inc.
Product Code
LJS · General Hospital
Decision Date
Aug 24, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The Sherlock 3CG® TPS Stylet/T-Lock Assembly provides real time catheter tip location information through the use of passive magnet and cardiac electrical signal detection.
Device Story
Sterile, single-use stylet (0.49 mm OD x 78.5 cm) provides internal reinforcement for PICC placement; used with Sherlock 3CG® Tip Confirmation System (TCS). Input: passive magnetic field and cardiac electrical signals detected via intravascular sensing wire. Operation: stylet acts as an intravascular ECG electrode; connects to TCS sensor via fin assembly (3-electrode system). Output: real-time catheter tip location/orientation feedback to clinician. Used in clinical settings during catheter insertion to ensure tip placement in superior vena cava near cavoatrial junction. Benefits: aids accurate placement, reduces need for post-procedure confirmation imaging.
Clinical Evidence
Bench testing only. Performance evaluated via ISO 594-2 (leakage, force, torque, stress cracking), biocompatibility (ISO 10993-4, -5, -10, -11), and USP <788> particulate matter testing. No clinical data presented.
Technological Characteristics
Stylet: stainless steel core, polyimide casing, hydrophilic coating, Neodymium-Iron-Boron magnets, conductive epoxy. T-Lock: PVC, MABS, polypropylene, polyisoprene. Sensing: passive magnetic and 3-electrode ECG (intravascular + 2 body electrodes). Connectivity: wired connection to TCS sensor. Sterilization: sterile.
Indications for Use
Indicated for catheter stylet internal reinforcement to aid in catheter placement. When used with the Sherlock 3CG® Tip Confirmation System (TCS), it provides rapid feedback on catheter tip location and orientation via passive magnets and cardiac electrical signal detection.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly (K142267)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 24, 2022
Bard Access Systems, Inc. Teresa Do-Mccage Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
## Re: K222232
Trade/Device Name: Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: July 22, 2022 Received: July 25, 2022
## Dear Teresa Do-Mccage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22232
#### Device Name
Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly
Indications for Use (Describe)
Catheter stylets provide internal reinforcement to aid in catheter placement. When used with the Sherlock 3 CG® Tip Confirmation System (TCS), the Sherlock 3CG® TPS Stylet/T-Lock Assembly also provides the placer rapid feedback on catheter tip location and orientation through the use of passive magnets and cardiac electrical signal detection.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|----------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpa |
he-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K222232 510(k) Summary 21 CFR 807.92(a)
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
| General Provisions | Submitter Name: | Bard Access Systems, Inc. (Bard has joined BD) |
|---------------------------------|----------------------|-------------------------------------------------------------------|
| | Submitter Address: | 605 North 5600 West<br>Salt Lake City, UT 84116 |
| | Contact Person: | Teresa Do-McCage<br>Regulatory Affairs Specialist |
| | Telephone Number: | 801.522.5959 |
| | Email: | teresa.do-mccage@bd.com |
| | Date of Preparation: | 8/24/2022 |
| | Subject Device | Trade Name: |
| Common Name: | | Long-Term Greater Than 30 Days Therapeutic Intravascular Catheter |
| Regulation Number: | | 21 CFR §880.5970 |
| Regulation Classification Name: | | Percutaneous, Implanted Long-Term Intravascular Catheter |
| Regulatory Class: | | II |
| Product Code: | | LJS |
| Classification Panel: | | General Hospital |
| Predicate Device | Trade Name: | Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly |
| | Common Name: | Long-Term Greater Than 30 Days Therapeutic Intravascular Catheter |
| | Regulation Number: | 21 CFR §880.5970 |
{4}------------------------------------------------
| Regulation | Classification Name: | Percutaneous, Implanted Long-Term Intravascular Catheter | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| | Regulatory Class: | II | | |
| | Product Code: | LJS | | |
| | Classification Panel: | General Hospital | | |
| | 510(k) Status: | K142267 (Concurrence date October 17, 2014) | | |
| Device Description | Bard Access Systems, Inc.'s Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly is a sterile, single<br>use device 0.49 mm (0.019 in) outer diameter x 78.5 cm, made of specially-formulated materials designed to aid in the<br>placement of specific Bard catheters, as well as any open ended, non-valved, polyurethane, peripherally inserted central<br>catheters (PICCs) that meet the dimensional specifications of the stylet. The Sherlock 3CG® TPS Stylet/T-Lock<br>Assembly is designed to work with catheters containing a minimum lumen diameter of 0.51mm (0.020 in). The stylet<br>provides internal reinforcement to aid in catheter placement. The Sherlock 3CG® TPS Stylet/T-Lock Assembly may be<br>used with the Sherlock 3CG® Tip Confirmation System (TCS) to provide catheter tip placement information during the<br>procedure. | | | |
| Intended Use | The Sherlock 3CG® TPS Stylet/T-Lock Assembly provides real time catheter tip location information through the use of<br>passive magnet and cardiac electrical signal detection. | | | |
| Indications for Use | Catheter stylets provide internal reinforcement to aid in catheter placement. When used with the Sherlock 3CG® Tip<br>Confirmation System (TCS), the Sherlock 3CG® TPS Stylet/T-Lock Assembly also provides the placer rapid feedback<br>on catheter tip location and orientation through the use of passive magnets and cardiac electrical signal detection. | | | |
| Technological<br>Characteristics | Technological characteristics of the subject Sherlock 3CG® Tip Positioning System (TPS) Stylet/T-Lock Assembly<br>device are substantially equivalent to those of the cited predicate device with respect to intended use, indications for<br>use, target patient population, operating principle, fundamental scientific technology, packaging configurations, sterility<br>assurance level, and method of sterilization. The differences of the subject device from the predicate device are limited<br>to the T-Lock Extension Set Assembly, a primary device component used in the subject/predicate device.<br>The following tables provides a summary comparison between the subject and predicate device component (T-Lock<br>Extension Set Assembly): | | | |
| | Attribute | Subject Device | Predicate device | Testing Conducted to<br>Demonstrate<br>Substantial<br>Equivalence |
| Owner | Bard Access Systems, Inc. | Same | | |
| 510(k) status | Subject of this 510(k) | K142267 | | |
| Device<br>Configuration | Same | Sherlock 3CG® Procedure Kit<br>• Sterile Components:<br>Uncoated Tyvek® double<br>pouch.<br>• Non-sterile Components: Foil<br>Pouch | | |
| Indications for<br>use | Same | Catheter stylets provide internal<br>reinforcement to aid in catheter<br>placement. When used with the<br>Sherlock 3CG® Tip Confirmation<br>System (TCS), the Sherlock 3CG®<br>TPS Stylet also provides the placer<br>rapid feedback on catheter tip<br>location and orientation through the<br>use of passive magnets and cardiac<br>electrical signal detection. | | |
| Duration of<br>use | Same | Limited (<24 hours) | | |
| | | | | |
| General<br>Device<br>Description | Same | Bard Access Systems, Inc.'s<br>Sherlock 3CG® Tip Positioning<br>System (TPS) Stylet is a sterile,<br>single use device 0.49 mm (0.019 in)<br>outer diameter x 78.5 cm, made of<br>specially-formulated materials<br>designed to aid in the placement of<br>specific Bard catheters, as well as<br>any open ended, non-valved,<br>polyurethane, peripherally inserted<br>central catheters (PICCs) that meet<br>the dimensional specifications of the<br>stylet. The Sherlock 3CG® TPS<br>Stylet is designed to work with<br>catheters containing a minimum<br>lumen diameter of 0.51mm (0.020<br>in). The stylet provides internal<br>reinforcement to aid in catheter<br>placement. The Sherlock 3CG® TPS<br>Stylet may be used with the Sherlock<br>3CG® Tip Confirmation System<br>(TCS) to provide catheter tip<br>placement information during the<br>procedure. | | |
| Means of<br>insertion | Same | Percutaneous | | |
| Insertion site | Same | Peripheral | | |
| Stylet<br>Materials | Same | Stainless steel core wire Polyimide casing Hydrophilic coating Magnets (Neodymium-Iron-Boron) Conductive epoxy UV adhesive | | |
| T-Lock<br>Connector<br>Extension | DESIGN | DESIGN | | |
| | Female Luer Lock Connector with wings | Female Luer Lock Connector without wings | Functional, Dimensional<br>testing per ISO 594-2. | |
| Set<br>Assembly | Septum captured in plastic cap Cap removed from rotating collar | Septum placed in shrink wrap ring Cap provided on rotating collar | Particulate Matter testing<br>per USP <788>. | |
| | MATERIAL | MATERIAL | | |
| | Female Luer Lock Connector: PVC with colorants | Female Luer Lock Connector: Rigid PVC with colorant | Biocompatibility testing<br>per ISO 10993-4, ISO<br>10993-5, ISO 10993-10,<br>and ISO 10993-11. | |
| | Slide Clamp – Polypropylene, white | Slide Clamp – ABS, white | | |
| | T-Fitting (T-Connector) – MABS, clear | T-Fitting (T-Connector) – PVC, clear | | |
| | Spin Lock (Rotating Collar) – MABS, clear | Spin Lock (Rotating Collar) – Polycarbonate, clear | | |
| | Same | Tubing: PVC | | |
| | Same | Injection Stopple (Septum) –<br>Polyisoprene | | |
| Stylet Pull<br>Tab Material | Same | Polypropylene | | |
| Stylet Tether<br>Material and<br>Length | Same | Medical grade Santoprene TPV wire insulation Copper tinsel wire Tin coated brass crimp 72 cm length…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.