← Product Code [QTZ](/productcode/QTZ) · K222222

# Artemis, Artemis TPO, Artemis MX (K222222)

_Eigen · QTZ · Oct 12, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K222222

## Device Facts

- **Applicant:** Eigen
- **Product Code:** [QTZ](/productcode/QTZ.md)
- **Decision Date:** Oct 12, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

Artemis along with the Needle Guide Attachment is used for image-guided interventional and diagnostic procedures of the prostate gland. It provides 2D and 3D visualization of Ultrasound (US) images and the ability to fuse and register these images with those from other imaging modalities such as Ultrasound, Magnetic Resonance, Computed Tomography, etc. It also provides the ability to display a simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organ and the current and the projected future path of the interventional instrument taking into account patient movement. The software also provides a virtual grid on the live ultrasound for performing systematic sampling of the target organ. Other software features include patient data management, multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, reporting, and pathology management. Artemis is intended for treatment planning and guidance for clinical, interventional and/or diagnostic procedures. The device is intended to be used in interventional and diagnostic procedures in a clinical setting. Example procedures include, but are not limited to image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, soft tissue ablations and placement of fiducial markers. Artemis is also intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes. Artemis Cryo Treatment Planning module is an add on to the existing Artemis software that allows physicians to prepare for cryo treatment planning based on positive pathology cores obtained during Artemis guided biopsies and registration results with other imaging modalities such as MRI, CT. The module allows accurate placement of cryo probes on targets, 3D tracking, real-time feedback on extend of cryo ice formation. The technology provided by Artemis generates ice models based on the specifications provided by the cryo device manufacturers and displays the models on the live ultrasound to provide guidance to the users during the procedure. The module also allows outlining or segmenting other organs that surround the prostate. Organs include bladder, and urethra.

## Device Story

Artemis is a medical image processing workstation for prostate interventions. It inputs live 2D ultrasound video from standard ultrasound machines and tracks probe position via a mechanical semi-robotic arm with encoders. The system reconstructs 3D ultrasound models and fuses/registers them with imported DICOM images (MRI, CT, US). It displays a virtual grid, simulated needle/probe trajectories, and real-time ice-ball growth models for cryoablation. Used in clinical settings by physicians, the device provides 3D visualization and navigation to assist in biopsy and ablation targeting. It aids clinical decision-making by enabling precise instrument placement and monitoring of ice formation relative to target anatomy and surrounding organs (bladder, urethra). Benefits include improved accuracy for diagnostic procedures and longitudinal tracking for active surveillance patients.

## Clinical Evidence

Bench testing only. Verification and validation testing evaluated all input/output functions and operational modes. Measurement validation was performed using phantoms and clinical CT/MRI images to confirm performance against predicate devices. No clinical trial data presented.

## Technological Characteristics

PC-based workstation with video digitizing card; mechanical assembly with encoders for probe/needle tracking. Supports DICOM connectivity. Software features include 3D reconstruction, multi-planar reformation, segmentation, and image registration (rigid/elastic). Complies with IEC 60601-1, IEC 60601-1-2 (EMC), and ISO 14971 (Risk Management).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Artemis ([K162474](/device/K162474.md))

## Reference Devices

- ICEfx Cryoablation System ([K181153](/device/K181153.md))
- 3-D Imaging Workstation ([K081093](/device/K081093.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Eigen % William Mandel Director of Operations, Regulatory Affairs, and Quality Assurance 13300 Grass Valley Avenue, Suite A GRASS VALLEY CA 95945

Re: K222222

October 12, 2023

Trade/Device Name: Artemis Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QTZ, GEH Dated: September 28, 2023 Received: October 3, 2023

Dear William Mandel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K222222

Device Name Artemis

#### Indications for Use (Describe)

Artemis along with the Needle Guide Attachment is used for image-guided interventional and diagnostic procedures of the prostate gland. It provides 2D and 3D visualization of Ultrasound (US) images and the ability to fuse and register these images with those from other imaging modalities such as Ultrasound, Magnetic Resonance, Computed Tomography, etc. It also provides the ability to display a simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organ and the projected future path of the interventional instrument taking into account patient movement. The software also provides a virtual grid on the live ultrasound for performing systematic sampling of the target organ. Other software features include patient data management, multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, reporting, and pathology management.

Artemis is intended for treatment planning and guidance for clinical, interventional and/or diagnostic procedures. The device is intended to be used in interventional and diagnostic procedures in a clinical setting. Example procedures include, but are not limited to image fusion for diagnostic clinical examinations and procedures, soft tissue ablations and placement of fiducial markers. Artemis is also intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes.

Artemis Cryo Treatment Planning module is an add on to the existing Artemis software that allows physicians to prepare for cryo treatment planning based on positive pathology cores obtained during Artemis guided biopsies and registration results with other imaging modalities such as MRI, CT. The module allows accurate placement of cryo probes on targets, 3D tracking, real-time feedback on extend of cryo ice formation. The technology provided by Artemis generates ice models based on the specifications provided by the cryo device manufacturers and displays the models on the live ultrasound to provide guidance to the users during the procedure.

The module also allows outlining or segmenting other organs that surround the prostate. Organs include bladder and urethra. □

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

leading innovation
in medical imaging

Image /page/3/Picture/1 description: The image shows the word "eigen" in a light blue font against a teal background. The font appears to be sans-serif and slightly rounded. The lighting on the text suggests a subtle glow or highlight effect.

## 510(k) Summary

### K222222

| Submitter's Name:          | Eigen                                                                                                                                                                                                    |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address:       | 13366 Grass Valley Avenue, Grass Valley, CA 95945                                                                                                                                                        |
| Submitter's Telephone:     | 530-274-1240                                                                                                                                                                                             |
| Submitter's Fax:           | 530-274-3536                                                                                                                                                                                             |
| Contact Name:              | William Mandel                                                                                                                                                                                           |
| Date Summary was Prepared: | Feb 11, 2022                                                                                                                                                                                             |
| Trade or Proprietary Name: | Artemis                                                                                                                                                                                                  |
| Common or Usual Name:      | Medical Image Processing Workstation                                                                                                                                                                     |
| Classification Name:       | Primary - Radiological Image Processing for Ablation Therapy<br>Planning and Evaluation, QTZ<br>Secondary - Unit, Cryosurgical, Accessories, GEH<br>Picture Archiving and Communications. 21CFR 892.2050 |

Predicate Devices:

| Device Name               | 510(k) Number | Type      |
|---------------------------|---------------|-----------|
| Artemis                   | K162474       | Primary   |
| ICEfx Cryoablation System | K181153       | Reference |

The design, function, and specifications of Artemis are similar to the identified legally marketed predicate devices. Similar to the devices from Eigen, LLC (K162474), ICEfx Cryoablation System (K181153), Artemis provides image-guided interventional planning and navigation for prostate procedures, the ability to view and capture live 2D ultrasound data to create reconstructed 3D ultrasound images/models and the ability to fuse and register these images with those acquired and imported from other modalities like Magnetic Resonance Imaging, X-ray Computed Tomography, and Ultrasound. Artemis also performs other viewing and image-processing functions such as image registration, multi-planar reformats and includes tools to segment, measure and annotate images. Similar to predicate devices, the subject device can import data from other DICOM based imaging devices and also output selected image views, processed data and user-defined reports. Similar to ICEfx Cryoablation System (K181153) listed predicate device, Artemis provides Navigation of cryo-needles and visual growth of ice-balls.

The ICEfx Cryoablation System (K181153) utilizes its own cryo-needles while Artemis utilizes a

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the text "leading innovation in medical imaging". The text is in a light gray color. The background is a gradient of blue and green.

Image /page/4/Picture/1 description: The image shows the word "eigen" in a light blue font against a teal background. The font appears to be sans-serif and slightly rounded. The letters are close together, creating a compact and unified visual element.

mechanical semi-robotic arm with encoders to navigate third party cryo-needles.

This difference in navigation and tracking technique does not significantly affect the use of the device, nor does it raise new or additional safety risks. This difference between Artemis and ICEfx Cryoablation System (K181153) devices does not impact device safety or effectiveness. Artemis and 3D-Imaging Workstation (K162474) utilize the identical navigation and tracking mechanism.

The current Artemis for this 510(k) and the previous Artemis (K162474) also share common software source code for basic system functionality such as multi-modality image viewing, segmentation, registration, navigation, annotation, reporting and DICOM functionality.

#### Description of the Device and Summary of the Technological Characteristics

Artemis is designed to display the 2-D live video received from commercially available ultrasound machines and use this 2-D video to reconstruct a 3-D ultrasound image. The system has been designed to work with the clinicians' existing ultrasound machine, probe, commercially available biopsy needle guide, needle gun combination, and cryoablation systems. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement, reporting and 3-D image registration.

Artemis is comprised of a mechanical assembly that holds the ultrasound probe and tracks probe position. The mechanical tracker is connected to a PC-based workstation containing a video digitizing card and running the image processing software. Control of the ultrasound probe and ultrasound system is done manually by the physician, just as it would be in the absence of Artemis. However, by tracking the position and orientation of the ultrasound probe while capturing the video image, the workstation is able to reconstruct and display a 3-D image and 3-D rendered surface model of the prostate.

The reconstructed 3-D image can be further processed to perform various

{5}------------------------------------------------

leading innovation
in medical imaging

Image /page/5/Picture/2 description: This image contains text describing a medical procedure. The text mentions measurements including volume estimation, which can be examined for abnormalities by a physician. Patient information, notes, and images may be stored for future retrieval, and locations for biopsies may be selected by the physician. The text also mentions 3-D models that may be recalled, aligned, or registered to the current 3-D model of the prostate, which is especially useful for patients under active surveillance.

The physician may attach a commercially available biopsy needle guide compatible to the ultrasound probe and use the probe and needle to perform tissue biopsy and or cryoablation. Whenever the ultrasound machine is turned on by the physician, the live 2-D ultrasound image is displayed on the screen of Artemis during the procedure. As the ultrasound probe with attached needle guide is maneuvered by the physician, the position and orientation of the probe with respect to the organ is tracked. Artemis is able to add, display and edit loaded plans for the procedure as well as provide the probe position and needle trajectory relative to the 3-D image and 3-D rendered surface model of the prostate.

In addition to standard transrectal needle guidance procedures, Artemis also supports transperineal needle guidance by mounting a Needle Guide Attachment (NGA). A commercially available needle guide compatible with the NGA is used. This NGA will be used for both biopsy and cryo needles. The NGA provides additional data to track the needle direction angle. When using transperineal mode, the procedure planning, segmentation, registration and navigation are performed in the same way as the standard transrectal procedure. The only difference lies in how the needle guide needs to be moved to target the different planned locations. For the transrectal procedure, the needle guide is always attached to the probe. Therefore moving the probe moves the needle guide. In transperineal needle guidance procedures the needle is not attached to the probe. Therefore the NGA needs to be moved to move the needle guide. Artemis highlights the closed target to the current needle guide position.

Artemis offers the physician additional 3-D information for assessing prostate abnormalities, planning and implementing biopsy procedures. The additional image processing features are generated with minimal changes to previous Ultrasound probe based procedures,

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the text "Leading innovation in medical imaging." The text is in a sans-serif font and is white. The background is a gradient of blue and white.

Image /page/6/Picture/1 description: The image shows the word "eigen" in a light blue font against a teal background. The font appears to be sans-serif and has a slightly blurred or glowing effect. The word is presented in lowercase letters and is centered within the frame.

and the physician always has access to the live 2-D ultrasound image during prostate assessment or biopsy procedure. The device also provides automated reports with information and pictures from the procedure.

### Intended Use

Artemis along with the Needle Guide Attachment is used for image-guided interventional and diagnostic procedures of the prostate gland. It provides 2D and 3D visualization of Ultrasound (US) images and the ability to fuse and register these images with those from other imaging modalities such as Ultrasound, Magnetic Resonance, Computed Tomography, etc. It also provides the ability to display a simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organ and the current and the projected future path of the interventional instrument taking into account patient movement. The software also provides a virtual grid on the live ultrasound for performing systematic sampling of the target organ. Other software features include patient data management, multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, reporting, and pathology management.

Artemis is intended for treatment planning and guidance for clinical, interventional and/or diagnostic procedures. The device is intended to be used in interventional and diagnostic procedures in a clinical setting. Example procedures include, but are not limited to image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, soft tissue ablations and placement of fiducial markers. Artemis is also intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes.

Artemis Cryo Treatment Planning module is an add on to the existing Artemis software that allows physicians to prepare for cryo treatment planning based on positive pathology cores obtained during Artemis guided biopsies and registration results with other imaging modalities such as MRI, CT. The module allows accurate placement of cryo probes on targets, 3D tracking, real-time feedback on extend of cryo ice formation. The technology

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the text "leading innovation in medical imaging" in a light gray font. To the right of this text is the word "eiger" in a teal color. The word "eiger" is in a bold, sans-serif font. The background of the image is a light teal color.

provided by Artemis generates ice models based on the specifications provided by the cryo device manufacturers and displays the models on the live ultrasound to provide guidance to the users during the procedure.

The module also allows outlining or segmenting other organs that surround the prostate. Organs include bladder, and urethra.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the text "leading innovation in medical imaging". The text is stacked on top of each other, with "leading innovation" on the first line and "in medical imaging" on the second line. The text is written in a sans-serif font and is light gray in color.

Image /page/8/Picture/1 description: The image shows the word "eigen" in a light blue color. The background is a darker shade of blue. The text is clear and easy to read.

# Substantial Equivalence

Following is a comparison of technological characteristics between Artemis and predicate devices:

| Product                          | Submitted Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate Device: Eigen, LLC<br>3-D Imaging Workstation K081093                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Predicate Device: ICEfx<br>Cryoablation System (K181153)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | Eigen<br>Artemis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended Use                     | Artemis along with the Needle Guide<br>Attachment is used for image-guided<br>interventional and diagnostic<br>procedures of the prostate gland. It<br>provides 2D and 3D visualization of<br>Ultrasound (US) images and the<br>ability to fuse and register these<br>images with those from other<br>imaging modalities such as<br>Ultrasound, Magnetic Resonance,<br>Computed Tomography, etc. It also<br>provides the ability to display a<br>simulated image of a tracked<br>insertion tool such as a biopsy<br>needle, guidewire or probe on a<br>computer monitor screen that shows<br>images of the target organ and the<br>current and the projected future path<br>of the interventional instrument<br>taking into account patient<br>movement. The software also<br>provides a virtual grid on the live<br>ultrasound for performing systematic<br>sampling of the target organ. Other<br>software features include patient<br>data management, multi-planar<br>reconstruction, segmentation, image<br>measurements, 2D/3D image<br>registration, reporting, and<br>pathology management.<br><br>Artemis is intended for treatment<br>planning and guidance for clinical,<br>interventional and/or diagnostic | Artemis along with the Needle Guide<br>Attachment is used for image-guided<br>interventional and diagnostic<br>procedures of the prostate gland. It<br>provides 2D and 3D visualization of<br>Ultrasound (US) images and the<br>ability to fuse and register these<br>images with those from other<br>imaging modalities such as<br>Ultrasound, Magnetic Resonance,<br>Computed Tomography, etc. It also<br>provides the ability to display a<br>simulated image of a tracked<br>insertion tool such as a biopsy<br>needle, guidewire or probe on a<br>computer monitor screen that shows<br>images of the target organ and the<br>current and the projected future path<br>of the interventional instrument<br>taking into account patient<br>movement. The software also<br>provides a virtual grid on the live<br>ultrasound for performing systematic<br>sampling of the target organ. Other<br>software features include patient<br>data management, multi-planar<br>reconstruction, segmentation, image<br>measurements, 2D/3D image<br>registration, reporting, and<br>pathology management.<br><br>Artemis is intended for treatment<br>planning and guidance for clinical,<br>interventional and/or diagnostic | The ICEfx Cryoablation System is<br>indicated for use as a cryosurgical<br>tool in the fields of<br>general surgery, dermatology,<br>neurology (including<br>cryoanalgesia), thoracic surgery<br>(with the<br>exception of cardiac tissue), ENT,<br>gynecology, oncology, proctology<br>and urology. This<br>System is designed to destroy<br>tissue (including prostate and<br>kidney tissue, liver metastases,<br>tumors, skin lesions) by the<br>application of extremely cold<br>temperatures.<br><br>The ICEfx Cryoablation System<br>has the following specific<br>indications:<br>• Urology Ablation of prostate<br>tissue in cases of prostate cancer<br>and Benign Prostate<br>Hyperplasia (BPH)<br>• Oncology Ablation of cancerous<br>or malignant tissue and benign<br>tumors, and palliative<br>intervention<br>• Dermatology Ablation or<br>freezing of skin cancers and other<br>cutaneous disorders;<br>Destruction of warts or lesions,<br>angiomas, sebaceous hyperplasia,<br>basal cell tumors of<br>the eyelid or canthus area,<br>ulcerated basal cell tumors,<br>dermatofibromas, small<br>hemangiomas, mucocele cysts,<br>multiple warts, plantar warts,<br>actinic and seborrheic<br>keratosis, cavernous<br>hemangiomas, peri-anal<br>condylomata, and palliation of<br>tumors of<br>the skin |
|                                  | procedures. The device is intended to<br>be used in interventional and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | procedures. The device is intended<br>to be used in interventional and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | diagnostic procedures in a clinical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | diagnostic procedures in a clinical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | setting. Example procedures include,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | setting. Example procedures include,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | but are not limited to image fusion for<br>diagnostic clinical examinations and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | but are not limited to image fusion<br>for diagnostic clinical examinations<br>and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | procedures, soft tissue biopsies, soft                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | procedures, soft tissue biopsies,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | tissue ablations and placement of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | soft tissue ablations and placement<br>of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | fiducial markers. Artemis is also                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | fiducial markers. Artemis is also                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | intended to be used for patients in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | intended to be used for patients in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                  | active surveillance to keep track of<br>previous procedures information and<br>outcomes.<br>Artemis Cryo Treatment Planning<br>module is an add on to the existing<br>Artemis software that allows<br>physicians to prepare for cryo<br>treatment planning based on positive<br>pathology cores obtained during<br>Artemis guided biopsies and<br>registration results with other<br>imaging modalities such as MRI, CT.<br>The module allows accurate<br>placement of cryo probes on targets,<br>3D tracking, real-time feedback on<br>extend of cryo ice formation. The<br>technology provided by Artemis<br>generates ice models based on the<br>specifications provided by the cryo<br>device manufacturers and displays<br>the models on the live ultrasound to<br>provide guidance to the users during<br>the procedure.<br>The module also allows outlining or<br>segmenting other organs that<br>surround the prostate. Organs include<br>bladder, and urethra.                                                                                                                                                                                                                                                 | active surveillance to keep track of<br>previous procedures information and<br>outcomes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | • Gynecology Ablation of<br>malignant neoplasia or benign<br>dysplasia of the female genitalia<br>• General surgery Palliation of<br>tumors of the rectum, anal<br>fissures, pilonidal cysts, and<br>recurrent cancerous lesions,<br>ablation of breast fibroadenomas<br>• ENT Palliation of tumors of the<br>oral cavity and ablation of<br>leukoplakia of the mouth<br>• Thoracic surgery (with the<br>exception of cardiac tissue)<br>• Proctology Ablation of benign or<br>malignant growths of the anus or<br>rectum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product                          | Submitted Device<br>Eigen<br>Artemis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device: Eigen, LLC<br>3-D Imaging Workstation K081093                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Predicate Device: ICEfx<br>Cryoablation System (K181153)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                     | QTZ, GEH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | LLZ…

---

**Source:** [https://fda.innolitics.com/device/K222222](https://fda.innolitics.com/device/K222222)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
