Saneso Lens Wash Bottle Assembly (SAN-LWB-A)

K222173 · Saneso, Inc. · OCX · Feb 27, 2023 · Gastroenterology, Urology

Device Facts

Record IDK222173
Device NameSaneso Lens Wash Bottle Assembly (SAN-LWB-A)
ApplicantSaneso, Inc.
Product CodeOCX · Gastroenterology, Urology
Decision DateFeb 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Saneso Lens Wash Bottle Assembly (SAN-LWB-A) is intended to be used with FDA-cleared Saneso endoscopes. It facilitates the following: 1. Sterile water for lens wash function 2. Air / CO2 for luminal insufflation function

Device Story

Saneso Lens Wash Bottle Assembly (SAN-LWB-A) is a reusable accessory for Saneso endoscopes. It consists of a cylindrical bottle, cap, tubing set, and air/water connector. The device functions as the proximal, non-patient-contacting portion of the lens wash and insufflation system. It connects to the endoscope's air/water port to deliver sterile water for lens cleaning and air or CO2 for luminal insufflation. The system is operated by clinicians during endoscopic procedures. The device was modified from single-use to multi-use, requiring manual cleaning and steam sterilization validation to ensure safety and efficacy. It does not contact the patient directly; the distal portion of the system (valves, channels, nozzles) remains part of the previously cleared Saneso endoscope.

Clinical Evidence

No clinical or animal studies were conducted. Substantial equivalence is supported by non-clinical performance testing, including manual cleaning validation and steam sterilization validation, which met all acceptance criteria.

Technological Characteristics

Reusable lens wash bottle assembly consisting of a cylindrical bottle, cap, tubing set, and air/water connector. Operates via manual application for fluid/gas delivery. Compatible with Saneso-A series endoscopes and Processor-A. Sterilization method: Steam sterilization. No software or electronic components.

Indications for Use

Indicated for use with FDA-cleared Saneso endoscopes to provide sterile water for lens washing and air/CO2 for luminal insufflation during endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 27, 2023 Saneso, Inc. % Parul Chansoria Founder and CEO Elexes Medical Consulting, LLC 30 N Gould St Ste R Sheridan, WY 82801 K222173 Re: > Trade/Device Name: Saneso Lens Wash Bottle Assembly (SAN-LWB-A) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: January 24, 2023 Received: January 24, 2023 Dear Parul Chansoria: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any {2}------------------------------------------------ Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. # Sivakami Venkatachalam -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K222173 Device Name Saneso Lens Wash Bottle Assembly (SAN-LWB-A) Indications for Use (Describe) The Saneso Lens Wash Bottle Assembly (SAN-LWB-A) is intended to be used with FDA-cleared Saneso endoscopes. It facilitates the following: 1. Sterile water for lens wash function 2. Air / CO2 for luminal insufflation function | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------|--| | <div><span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | | | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Saneso. The word "Saneso" is written in large, red, bold letters. Below the word "Saneso" is the phrase "Endoscopy Reinvented" in smaller, gray letters. This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92. #### 5.1. Submitter Saneso, Inc. One Oxford Center, 301 Grant Street Suite 4300 Pittsburgh, PA 15219 ### Contact Person: Parul Chansoria, MS, RAC, CQA CEO & Founder, Elexes Medical Consulting Telephone: +408-475-8091 Email: parul@elexes.com Summary prepared: 07/15/2022 #### 5.2. Device Information Common/Usual Name: Endoscope, Water Bottle with cap and tubing Trade Name: Saneso Lens Wash Bottle Assembly (SAN-LWB-A) Regulation Name: Endoscope and accessories Regulatory Class: Class II Classification Panel: Gastroenterology/Urology Product Code: OCX Regulation Number: 21 CFR Part 876.1500 #### 5.3. Predicate Device Information Saneso Lens Wash Bottle Assembly is substantially equivalent to the following cleared device: | Company | Predicate Device<br>Priority | Product | 510(k) Number | |-------------|------------------------------|-------------------|---------------| | Saneso Inc. | Primary | Saneso - A series | K210052 | #### 5.4. Device Description The Saneso Lens Wash Bottle Assembly is intended to be used with Saneso endoscopes. It facilitates the following: - Sterile water for lens wash function 1. - 2. Air / CO2 for luminal insufflation function Saneso lens wash & insufflation system is divided into proximal and distal sections: Proximal Section: It comprises the Lens Wash Bottle Assembly and terminates at its {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Saneso. The word "Saneso" is written in red, with a trademark symbol in the upper right corner. Below the word "Saneso" is the phrase "Endoscopy Reinvented" in a smaller, gray font, also with a trademark symbol. connection with the air/water port on the endoscope. It consists of the lens wash bottle, lens wash bottle cap, airline, water line, CO2 line, and air/water connector. This section is the non-patient contacting portion of the system and is covered in this submission. Distal Section: It starts at the air/water port and terminates at the air/water nozzles on the distal end of the endoscope. It consists of the air/water valve, air/water valve, air/water channel, and the lens wash nozzles, which are all located in the endoscope. This section is the patient contacting portion of the system and is part of the Saneso endoscopes, which is FDA 510k approved (K210052). #### ર.ડ. Intended Use ### 5.5.1. Subject Device The Saneso Lens Wash Bottle Assembly (SAN-LWB-A) is intended to be used with FDA-cleared Saneso endoscopes. It facilitates the following: - Sterile water for lens wash function ● - Air / CO2 for luminal insufflation function ● ### 5.5.2. Primary Predicate Device #### Saneso Colonoscope 360-A The Saneso Colonoscope 360-A is intended for diagnostic visualization of the lower gastrointestinal tract (including the rectum, colon, and Saneso Colonoscope 360-A is not indicated for ileoscopy procedures. The system also provides access to therapeutic interventions using standard endoscopy tools. The Saneso system consists of Saneso Colonoscope 360-A, Processor- A, and other ancillary equipment. ### Saneso Gastroscope 360-A The Saneso Gastroscope 360-A is intended for diagnostic visualization of the upper gastrointestinal tract (including the esophagus, stomach and duodenum). The system also provides access for therapeutic interventions using standard endoscopy tools. The Saneso system consists of Saneso Gastroscope 360-A, Processor-A and other ancillary equipment. ### Saneso Single Camera Colonoscope-A The Saneso Single Camera Colonoscope is intended for diagnostic visualization of the lower gastrointestinal tract (including the rectum, colon and ileocecal valve). The system also provides access for therapeutic interventions using standard endoscopy tools. The Saneso system consists of Saneso Single Camera Colonoscope, Processor-A and other ancillary equipment. {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Saneso Endoscopy Reinvented. The word "Saneso" is in large, bold, red letters. Below it, in smaller, gray letters, is the phrase "Endoscopy Reinvented." ### Saneso Gastroscope 360- A The Saneso Single Camera Gastroscope is intended for diagnostic visualization of the upper gastrointestinal tract (including the esophagus, stomach, and duodenum). The system also provides access to therapeutic interventions using standard endoscopy tools. The Saneso system consists of Saneso Single Camera Gastroscope, Processor-A, and other ancillary equipment. Note: The scope of the submission is not the endoscope (Colonoscope and the gastroscope) but the Lens wash bottle alone. Lens wash bottle is intended to be used only as an accessory to the endoscope system (SAN-A series) which was originally cleared by the FDA with K number K210052. There is no modification in the system, but there is a modification in the accessory where it is changing from single-use to multi-use, and it is supported by the reprocessing validation results. #### 5.6. Technological Characteristics | Table 1: Substantial Equivalence Table for Saneso Lens Wash Bottle Assembly | | | | |-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Parameter | Subject Device | Primary Predicate<br>Device | Equivalence | | 510(k) Number | - | K210052 | NA | | Product code | OCX | FDS, FDF | Equivalent | | Regulation No. | 21 CFR Part 876.1500 | 21 CFR Part 876.1500 | Equivalent | | Classification: | Class II | Class II | Equivalent | | Regulation<br>Name | Endoscope and<br>Accessories | Endoscope and<br>Accessories | Equivalent | | Prescription<br>Use/OTC | Prescription Use | Prescription Use | Equivalent | | Intended Use | The Saneso Lens Wash<br>Bottle Assembly is<br>intended to be used<br>with the FDA cleared<br>Saneso endoscopes. It<br>facilitates the<br>following:<br>a. Sterile water for<br>lens wash function<br>b. Air / CO2 for | Saneso<br>Colonoscope<br>360-A<br>The Saneso<br>Colonoscope 360-A is<br>intended for diagnostic<br>visualization of the<br>lower gastrointestinal<br>tract (including the<br>rectum, colon, and<br>cecum) and Saneso | Equivalent<br>Note: The scope of<br>the submission is not<br>the endoscope<br>(Colonoscope and the<br>gastroscope) but the<br>Lens wash bottle<br>alone. | | | | | | | | luminal insufflation<br>function | Colonoscope 360-A is<br>not indicated for<br>ileoscopy procedures.<br>The system also<br>provides access for<br>therapeutic interventions<br>using standard<br>endoscopy tools. The<br>Saneso system consists<br>of Saneso Colonoscope<br>360-A, Processor- A and<br>other ancillary<br>equipment.<br><br><b>Saneso Gastroscope<br/>360-A</b><br><br>The Saneso Gastroscope<br>360-A is intended for<br>diagnostic visualization<br>of the upper<br>gastrointestinal tract<br>(including the<br>esophagus, stomach and<br>duodenum). The system<br>also provides access for<br>therapeutic interventions<br>using standard<br>endoscopy tools. The<br>Saneso system consists<br>of Saneso Gastroscope<br>360-A, Processor-A and<br>other ancillary<br>equipment. | Lens wash bottle is<br>intended to be used<br>only as an accessory<br>to the endoscope<br>system (SAN-A<br>series) which was<br>originally cleared by<br>the FDA with K<br>number K210052.<br>There is no<br>modification in the<br>system, but there is a<br>modification in the<br>accessory where it is<br>changing from<br>single-use to<br>multi-use, and it is<br>supported by the<br>reprocessing<br>validation results. | | | | <b>Saneso Single Camera<br/>Colonoscope-A</b><br><br>The Saneso Single<br>Camera Colonoscope is<br>intended for diagnostic<br>visualization of the<br>lower gastrointestinal<br>tract (including the<br>rectum, colon and<br>ileocecal valve). The<br>system also provides | | | | | access for therapeutic<br>interventions using<br>standard endoscopy<br>tools. The Saneso<br>system consists of<br>Saneso Single Camera<br>Colonoscope,<br>Processor-A and other<br>ancillary equipment. | | | | | Saneso Gastroscope<br><b>360-A</b> | | | | | The Saneso Single<br>Camera Gastroscope is<br>intended for diagnostic<br>visualization of the<br>upper gastrointestinal<br>tract (including the<br>esophagus, stomach and<br>duodenum). The system<br>also provides access for<br>therapeutic interventions<br>using standard<br>endoscopy tools. The<br>Saneso system consists<br>of Saneso Single<br>Camera Gastroscope,<br>Processor-A and other<br>ancillary equipment. | | | Construction/Co<br>mponents | The Saneso Lens Wash<br>Bottle Assembly<br>consists of a reusable<br>lens wash water bottle<br>(cylindrical bottle), cap,<br>tubing set and air/water<br>connector. | The Saneso Lens Wash<br>Bottle consists of a<br>single-use lens wash<br>water bottle (cylindrical<br>bottle), cap, tubing set<br>and air/water connector. | Equivalent | | Sterile/Non<br>Sterile | Sold Non - Sterile | Sold Non - Sterile | Equivalent | | Accessories | Cap CO2 outlet Water line Cap water outlet | Cap CO2 outlet Water line Cap water outlet CO2 line | Equivalent | | | CO2 line Air Line CO2 clamp Cap Air inlet | Air Line CO2 clamp Cap Air inlet | | | Usage | Reusable | Single Use | Different | | Method of<br>Application | Manual | Manual | Equivalent | | Compatible<br>Endoscopes/proce<br>ssors | Saneso - A series<br>(K210052), Saneso<br>Processor - A | Saneso - A series<br>(K210052), Saneso<br>Processor - A | Equivalent | {7}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for Saneso. The word "Saneso" is in red, and the words "Endoscopy Reinvented" are in gray below it. The logo is simple and modern. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Saneso. The word "Saneso" is in large, red, bold letters. Below that, in smaller gray letters, is the phrase "Endoscopy Reinvented" with a trademark symbol. {9}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Saneso. The word "Saneso" is written in large, red, bold letters. Below the word "Saneso" is the phrase "Endoscopy Reinvented" written in smaller, gray letters. The logo is simple and modern. ### 5.6.1. Similarities The Saneso Lens Wash Bottle Assembly SAN-LWB-A is similar to the Lens Wash Bottle which is cleared under K210052 in terms of intended use, construction/components, method of application. ### 5.6.2. Differences The difference between the Saneso Lens Wash Bottle Assembly SAN-LWB-A and the Lens Wash Bottle which is cleared under K210052 is its usage, ie; from single use to multi-use. However, this difference does not raise new questions of safety or efficacy. Differences were justified using Non-Clinical performance testing - Manual Cleaning Validation and Steam Sterilization Validation and demonstrates that the Lens wash bottle Assembly is safe and effective. #### 5.7. Summary of Non-Clinical Tests Performance testing was performed as per the relevant standards and test procedures listed in the Declaration of Conformity. - 1. Manual Cleaning Validation Cleaning validation was performed using Saneso's recommended cleaning procedure. Saneso Lens Wash Bottle Assembly met all the acceptance criteria and demonstrated that the cleaning process does not impact the functionality of the device. - 2. Steam Sterilization Validation Sterilization was performed and Saneso Lens Wash Bottle Assembly met all the acceptance criteria. The device is found to be safe and effective. #### Summary of Clinical Tests 5.8. {10}------------------------------------------------ Image /page/9/Picture/0 description: The image contains the logo for Saneso Endoscopy Reinvented. The word "Saneso" is in red, and the words "Endoscopy Reinvented" are in gray. The logo is simple and modern, and it is likely used to promote a medical device or service. ### 5.8.1. Animal Study No animal studies have been conducted for the Subject Device. ### 5.8.2. Clinical Study No clinical studies have been conducted for the Subject Device. #### 5.9. Conclusion Saneso Lens Wash Bottle Assembly SAN-LWB-A is substantially equivalent to the Lens Wash Bottle which is cleared under K210052 in terms of intended use and technological characteristics. Differences were justified using Non-Clinical performance testing. Thus, Saneso Inc. has concluded that the device does not introduce any significant questions of safety and efficacy and is substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...