K222148 · Pixcell Medical Technologies, Ltd. · GKZ · Aug 16, 2023 · Hematology
Device Facts
Record ID
K222148
Device Name
HemoScreen Hematology Analyzer
Applicant
Pixcell Medical Technologies, Ltd.
Product Code
GKZ · Hematology
Decision Date
Aug 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Attributes
AI/ML, Pediatric
Intended Use
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
Device Story
HemoScreen is a tabletop POC hematology analyzer; provides CBC and 5-part leukocyte differential. System includes reader, disposable cartridge with preloaded reagents, and sampler. User collects capillary or venous blood via sampler; inserts into cartridge; inserts cartridge into reader. Reader uses viscoelastic focusing to align cells in microfluidic chamber; captures high-resolution microscopic images; analyzes images via machine vision algorithms to differentiate and count cells. WBCs stained for subtype differentiation; HGB calculated from optical density of individual cells. Results displayed on touchscreen within six minutes. Enables rapid hematological assessment in clinical settings; supports clinical decision-making by providing quantitative blood parameters.
Clinical Evidence
Prospective clinical study (N=42) compared direct capillary sampling to indirect capillary sampling (predicate). Data evaluated via Passing-Bablok regression and Pearson's correlation for 20 parameters. Results showed high correlation (e.g., WBC r=0.978, RBC r=0.97, MCV r=0.994). Precision study (N=23) evaluated operator-to-operator variability using two fingersticks per subject, demonstrating adequate precision. No changes to nonclinical data (Limit of Blank, Linearity, Interference, etc.) from K180020.
Technological Characteristics
Tabletop analyzer using viscoelastic focusing for single-cell alignment in microfluidic chambers. Employs high-resolution microscopic imaging and machine vision algorithms. Sample volume: 40μL. Connectivity: USB export. Factory calibrated. Disposable reagent Cartridge and Sampler system. Software-driven interface.
Indications for Use
Indicated for adults and children at least 2 years of age requiring enumeration and classification of CBC parameters (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, and 5-part leukocyte differential) using capillary or venous K2EDTA anticoagulated whole blood in clinical or POC settings.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
K240636 — HemoScreen Hematology Analyzer · Pixcell Medical Technologies · May 2, 2024
K180020 — HemoScreen Hematology Analyzer · Pixcell Medical Technologies, Ltd. · Oct 29, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 16, 2023
PixCell Medical Technologies, Ltd. % Erika Ammirati President Ammirati Regulatory Consulting 575 Shirlynn Court Los Altos, California 94022
Re: K222148
Trade/Device Name: HemoScreen Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: March 29, 2023 Received: March 30, 2023
Dear Erika Ammirati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Min Wu-S
Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222148
Device Name HemoScreen Hematology Analyzer
#### Indications for Use (Describe)
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K222148.
| 807.92 (a)(1): Name: | PixCell Medical Technologies, Ltd. |
|----------------------|------------------------------------|
| Address: | 6 Hayezira St. |
| | Yoknaem Ilit, Israel 2069202 |
| Phone: | +972-4-9593516 |
| Email: | yaara@pixcell-medical.com |
| Contact: | Yaara Ben-Yosef, PhD |
### 807.92 (a)(2): Device name- trade name and common name, and classification
Trade name: HemoScreen Hematology Analyzer
Common Name: Automated differential cell counter
Classification: 21 CFR 864.5220
### 807.92 (a)(3): Identification of the legally marketed predicate devices
HemoScreen Hematology Analyzer (PixCell Medical Technologies, Ltd), cleared under K180020.
#### 807.92 (a)(4): Device Description
HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent Cartridge. In addition to the Cartridge, the system includes a disposable Sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the Cartridge.
Once the Cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (Sampler) into the reagent compartments (Cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the Cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.
The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The Cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the Cartridge.
{4}------------------------------------------------
The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.
Quality Control: Commercial 3-level liquid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.
Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.
## 807.92 (a)(5): Intended Use
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
## 807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between HemoScreen and the predicate.
| Comparison | ON-TEST<br>HemoScreen with direct fingerstick<br>sampling | PREDICATE<br>HemoScreen with indirect fingerstick<br>sampling (K180020) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Intended Use | Automated hematology analyzer | Same |
| Parameters<br>Measured | • Red Blood Cells (RBC),<br>• White Blood Cells (WBC),<br>• Platelets (PLT),<br>• Hemoglobin (HGB),<br>• Hematocrit (HCT),<br>• Mean Corpuscular (erythrocyte)<br>Volume (MCV),<br>• Mean Cell (erythrocyte) Hemoglobin<br>(MCH),<br>• Mean Cell (erythrocyte) Hemoglobin<br>Concentration (MCHC),<br>• Red Blood Cell Distribution Width<br>(RDW)-CV<br>• Mean Platelets Volume (MPV),<br>• Neutrophils (NEUT; #/%), | Same |
| Comparison | ON-TEST<br>HemoScreen with direct fingerstick<br>sampling | PREDICATE<br>HemoScreen with indirect fingerstick<br>sampling (K180020) |
| | Lymphocytes (LYMP; #/%), Eosinophils (EO; #/%) and Basophiles (BASO; #/%) | |
| Class | Class II | Same |
| Regulation<br>Number | 21 CFR 864.5220 | Same |
| Product Code | GKZ | Same |
| FDA Branch | Hematology | Same |
| Throughput | 10 samples/hour | Same |
| Test Principle | The HemoScreen uses a novel focusing<br>method called viscoelastic focusing which<br>causes the cells to perfectly align into a<br>plane. High resolution microscopic images<br>are taken of the flowing cells. Each image<br>is analyzed using machine vision<br>algorithms and the different cell types are<br>differentiated and counted. WBCs are<br>stained prior to analysis so as to enable<br>differentiation between their subtypes and<br>abnormal cells.<br>HGB is calculated based on the optical<br>density measured on intact individual cells. | Same |
| Calibration | Factory calibrated | Same |
| Sample Type-<br>venous | K2EDTA anticoagulated whole blood | Same |
| Sample Type-<br>fingerstick | Direct and indirect<br><br><i>Direct:</i> Capillary blood from fingertip<br>drawn directly into Sampler.<br><br><i>Indirect:</i> Capillary blood from fingertip<br>delivered to microtube, then transferred<br>into Sampler | Indirect, only |
| Sample Volume | 40μL | Same |
{5}------------------------------------------------
{6}------------------------------------------------
## 807.92 (b)(1): Brief Description of Nonclinical Data- no changes from K180020
Limit of Blank
Please refer to the 510(k) Summary for K180020.
Limits of Detection and Quantitation
Please refer to the 510(k) Summary for K180020.
Linearity
Please refer to the 510(k) Summary for K180020.
Repeatability of Venous Samples (internal and external operators)
Please refer to the 510(k) Summary for K180020.
Reproducibility of OC Materials Please refer to the 510(k) Summary for K180020.
Interference
Please refer to the 510(k) Summary for K180020.
Reference Intervals- Adult Males and Females (CLSI EP28-A3C)
Please refer to the 510(k) Summary for K180020.
### 807.92 (b)(2): Brief Description of Clinical Data-
No changes from K180020 for venous sampling and indirect capillary sampling into a K2EDTA microtube. The data in this section reflect the comparisons between direct capillary sampling (ontest), and indirect capillary sampling (predicate).
Comparisons between Direct Capillary Sampling and Indirect Capillary Sampling.
A comparison HemoScreen to HemoScreen study was performed where results from direct sampling of fingerstick blood into the Sampler were compared to results from indirect sampling, where fingerstick blood was first collected into a K2EDTA microtube, and then transferred to the Sampler. This was a prospective clinical study that included 42 subjects and matched fingerstick collections. The data were evaluated by Passing-Bablok regression and Pearson's correlation for all 20 parameters, and the summarized data are provided below.
{7}------------------------------------------------
| Parameter | N | Result Range<br>(Indirect) | Intercept [95% CI] | Slope [95% CI] | Pearson<br>Correlation |
|---------------|----|----------------------------|-------------------------|----------------------|------------------------|
| WBC (10³/μL) | 42 | 4.54-12.52 | -0.216 [-0.893, 0.412] | 1.028 [0.955, 1.108] | 0.978 |
| RBC (10⁶/μL) | 42 | 4.14-5.78 | 0 [-0.401, 0.498] | 0.994 [0.881, 1.078] | 0.97 |
| HGB (g/dL) | 42 | 11.84-17.12 | -0.749 [-2.348, 1.097] | 1.048 [0.913, 1.156] | 0.969 |
| HCT (%) | 42 | 35.76-50.11 | -1.77 [-7.048, 3.527] | 1.032 [0.901, 1.157] | 0.967 |
| MCV (fL) | 42 | 76.6-93.84 | 1.013 [-2.325, 3.887] | 0.988 [0.954, 1.025] | 0.994 |
| MCH (pg) | 42 | 24.67-32.98 | -0.191 [-0.907, 0.605] | 1.007 [0.981, 1.032] | 0.998 |
| MCHC (g/dL) | 42 | 32.2-35.83 | -1.519 [-4.767, 1.859] | 1.047 [0.949, 1.143] | 0.961 |
| RDW (%) | 42 | 11.64-14.16 | -0.25 [-0.876, 0.266] | 1.019 [0.977, 1.068] | 0.991 |
| PLT (10³/μL) | 42 | 142.4-399.6 | 3.646 [-16.602, 21.01] | 0.963 [0.894, 1.051] | 0.979 |
| MPV (fL) | 42 | 9.18-13.46 | 0.572 [-1.036, 1.837] | 0.949 [0.832, 1.113] | 0.92 |
| NEUT (10³/μL) | 42 | 2.19-7.95 | -0.221 [-0.531, 0.062] | 1.049 [0.995, 1.109] | 0.984 |
| LYMP (10³/μL) | 42 | 1.51-4.17 | -0.153 [-0.562, 0.251] | 1.075 [0.925, 1.273] | 0.953 |
| MONO (10³/μL) | 42 | 0.27-0.86 | -0.087 [-0.256, 0.023] | 0.996 [0.796, 1.358] | 0.825 |
| EO (10³/μL) | 42 | 0.04-0.53 | -0.021 [-0.032, -0.005] | 1.017 [0.945, 1.081] | 0.988 |
| BASO (10³/μL) | 42 | 0.01-0.07 | 0.003 [-0.01, 0.012] | 1.225 [0.842, 1.659] | 0.598 |
| NEUT (%) | 42 | 43.75-68.8 | -2.615 [-11.774, 4.007] | 1.05 [0.941, 1.194] | 0.957 |
| LYMP (%) | 42 | 19.95-45.05 | -1.418 [-4.219, 1.404] | 1.073 [0.983, 1.162] | 0.969 |
| MONO (%) | 42 | 3.95-11.25 | -2.108 [-3.977, -0.95] | 1.147 [0.985, 1.459] | 0.823 |
| EO (%) | 42 | 0.7-7.3 | -0.27 [-0.429, -0.028] | 1.029 [0.968, 1.097] | 0.987 |
| BASO (%) | 42 | 0.1-0.7 | -0.075 [-0.3, 0.15] | 1.5 [1, 2] | 0.614 |
Passing-Bablok Regression Data and Pearson Correlation
CI = confidence interval
# Precision of Direct Capillary Sampling
Twenty-three (23) subjects were enrolled for the direct capillary precision study. Two capillary punctures were performed on each subject, one puncture by each of two operators. The fingersticks were done on the same hand, with one operator collecting blood using the HemoScreen Sampler from one finger and completing the test, and the other operator collecting blood using the HemoScreen Sampler from another finger and completing the test. The operator order was randomized to counteract order effects, and the two punctures were done within 10 minutes of each other. One HemoScreen analyzer and multiple Cartridge lots were used for testing. Neither operator was aware of the test results from the other operator. Precision in capillary blood was shown to be adequate.
## 807.92 (b)(3): Conclusions from Nonclinical and Clinical Data
The conclusions drawn from the updated clinical data demonstrate that the device is safe and effective for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.