Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Nov 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The Sterile Hypodermic Syringe with/without needle is intended for use in the aspiration and injection of fluids for medical purposes. The Sterile Hypodermic Needle is intended to be used with a syringe for fluids aspiration and injection for medical purposes.
Device Story
Disposable manual syringe and hypodermic needle; used in professional healthcare facilities for fluid aspiration and injection. Device components include needle cap, needle tube, needle hub (6% luer connector), barrel with volume scale, plunger, and plunger stopper. Operation is entirely manual; clinician-operated. Device is MR-unsafe due to potential for projectile injury, thermal burns, and image artifacts. Benefits include reliable fluid delivery/withdrawal for medical procedures.
Clinical Evidence
No clinical data was submitted. Substantial equivalence is supported by bench testing, including performance testing (ISO 7886-1, ISO 9626, ISO 7864, ISO 80369-7/20, ISO 6009), biocompatibility (ISO 10993 series), and sterilization/packaging validation (ISO 11135, ISO 11737, ASTM F1980, etc.).
Technological Characteristics
Materials: Polypropylene (PP), stainless steel, rubber piston. Sensing/Actuation: Manual. Energy: None. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectors: 6% luer per ISO 80369-7/20. Needle hub color-coded per ISO 6009. Form factor: Disposable syringe (1-100 ml) and needle (15-30G).
Indications for Use
Indicated for aspiration and injection of fluids for medical purposes in patients including adults, transitional adolescents, adolescents, and children.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K212857 — Sterile Syringe for Single Use (with Needle), Sterile Hypodermic Needle for Single Use · Jiangsu Zhiyu Medical Instrument Co, Ltd. · May 9, 2023
K211728 — Sterile hypodermic syringes for single use · Jiangsu Kangyou Medical Instrument Co., Ltd. · Oct 5, 2021
K211482 — Disposable Syringe with Needle · Shandong Qinkai Medical Industry Co., Ltd. · Jan 24, 2022
K163161 — Sterile Single-use Syringe with Needle · Jiangxi Hongda Medical Equipment Group , Ltd. · Mar 20, 2017
K212846 — Sterile Hypodermic Syringe for Single Use, with/without Needle, Sterile Hypodermic Needle for Single Use · Jiangsu Jichun Medical Devices Co., Ltd. · Aug 30, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
November 4, 2022
Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd % Charles Mack Principal Engineer Irc 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K222124
Trade/Device Name: Sterile Hypodermic Syringe with/without Needle, Sterile Hypodermic Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: August 3, 2022 Received: August 8, 2022
Dear Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -S3
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K222124
Device Name Sterile Hypodermic Syringe with/without needle Sterile Hypodermic Needle
Indications for Use (Describe)
Sterile Hypodermic Syringe with/without needle
The Sterile Hypodermic Syringe with/without needle is intended for use in the aspiration and injection of fluids for medical purposes.
Sterile Hypodermic Needle
The Sterile Hypodermic Needle is intended to be used with a syringe for fluids aspiration and injection for medical purposes.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo and text. The logo is on the left side of the image and consists of a red plus sign and a blue stylized letter "m". To the right of the logo is the text "安徽宏宇五洲医疗器械股份有限公司 ANHUI HONGYU WUZHOU MEDICAL MANUFACTURER CO., LTD". The text is in both Chinese and English.
# K222124 510(k) SUMMARY
| Preparation Date: | November 4, 2022 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer's Name and Address: | Anhui Hongyu Wuzhou Medical<br>Manufacturer Co., Ltd.<br>No. 2 Guanyin Road, Economic Development<br>Zone, Taihu, Anqing City, Anhui Province, China<br>246400 |
| Corresponding Official: | Charles Mack |
| Telephone Number: | 931-625-4938 |
| Email Address: | charliemack@irc-us.com |
| Trade Name: | Sterile Hypodermic Syringe<br>with/without Needle<br>Sterile Hypodermic Needle |
| Common Name(s): | syringe, piston |
| Regulation Name(s): | Piston Syringe |
| Regulation Number(s): | 21CFR 880.5860 |
| Product Code: | FMF, FMI |
| Device Class: | Class II |
| Predicate Device: | K060211<br>Anhui Hongyu Wuzhou Medical<br>Manufacturer Co., Ltd. |
{4}------------------------------------------------
# Device Description:
The Sterile Hypodermic Syringe with/without needle is intended for use in the aspiration and injection of fluids for medical purposes.
The Sterile Hypodermic Needle is intended to be used with a syringe for fluids aspiration and injection for medical purposes.
It is intended for the patient population, including adults, transitional adolescents, adolescents, and children.
The device is intended for use in a professional healthcare facility.
Warning: MR-unsafe!
Do not expose the device to a magnetic resonance (MR) environment.
- -The device may present a risk of projectile injury due to the presence of ferromagnetic materials that can be attracted by the MR magnet core.
- -Thermal injury and burns may occur due to the metal components of the device that can heat during MR scanning.
- -The device may generate artifacts in the MR image.
- -The device may not function properly due to the strong magnetic and radio frequency fields generated by the MR scanner.
The device is a disposable syringe made of the following components:
Needle cap: Covers the needle tube until the syringe is used.
Needle tube: The needle tube penetrates the patient's skin to inject/withdraw fluid from the body.
Needle hub: The needle hub is connected to the syringe by a 6% luer connector, which is colored to distinguish the gauge of the needle.
Barrel: The barrel has a scale showing the capacity of the syringe. It is connected to the Hypodermic Needle by a 6% luer connector.
Plunger: Assembled with the plunger stopper to inject/withdraw fluid from the body. Plunger stopper: Sealing when injecting/withdrawing fluid from the body.
{5}------------------------------------------------
### Indications for Use
### Sterile Hypodermic Syringe with/without needle
The Sterile Hypodermic Syringe with/without needle is intended for use in the aspiration and injection of fluids for medical purpose.
### Sterile Hypodermic Needle
The Sterile Hypodermic Needle is intended to be used with a syringe for fluids aspiration and injection for medical purposes.
{6}------------------------------------------------
| Item | Subject Device | Predicate Device (own) | Discussion |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) | Pending | K060211 | - |
| Manufacturer | Anhui Hongyu Wuzhou Medical<br>Manufacturer Co., Ltd. | Anhui Hongyu Wuzhou<br>Medical Manufacturer<br>Co., Ltd. | - |
| Name | Sterile Hypodermic Syringe<br>with/without needle<br>Sterile Hypodermic Needle | Wuzhou Syringe<br>with/without Needle | - |
| Product code | FMF, FMI | FMF, FMI | - |
| Indication for Use | <b>Sterile Hypodermic Syringe<br/>with/without needle</b><br>The Sterile Hypodermic Syringe<br>with/without needle is intended for<br>use in the aspiration and injection of<br>fluids for medical purposes.<br><br><b>Sterile Hypodermic Needle</b><br>The Sterile Hypodermic Needle is<br>intended to be used with a syringe for<br>fluids aspiration and injection for<br>medical purposes. | The Wuzhou Syringe,<br>with/without a needle is<br>intended to be used for<br>medical purposes to<br>inject fluids into or<br>withdraw fluids from the<br>body. | Note 0 |
| Item | Subject Device | Predicate Device (own) | Discussion |
| Operation Mode | Manual use only | Manual use only | Identical |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Identical |
| Material | PP, Stainless Steel, Rubber Piston | PP, Stainless Steel, Rubber Piston | Identical |
| Needle Length | 1/6"-2" | 1/6" ~ 1 1/2" | Note 1 |
| Syringe Volume | 1,2,2.5,3,5,10,20,30,50,60,100 ml | 1,2,<br>5,10,20,30,50,60,100 ml | Note 2 |
| Needle Gauge | 15,16,17,18,19,20,21,22,23,24,25,26,<br>27,28,29,30G | 16,18,20,21,22,23,24,25,26G | Note 3 |
| Needle Hub | Color-coded per ISO 6009 | Color-coded per ISO 6009 | Note 4 |
| Performance | Complied with the current applicable<br>performance standards:<br>- ISO 9626<br>- ISO 7864<br>- ISO 80369-7<br>- ISO 80369-20<br>- ISO 7886-1 | Complied with the current applicable<br>performance standards:<br>- ISO 9626<br>- ISO 7864<br>- ISO 594-1<br>- ISO 594-2<br>- ISO 7886-1 | Note 5 |
| Single Patient Use | Yes | Yes | Identical |
| Biocompatibility | Conforms to ISO 10993-1 | Conforms to ISO 10993-1 | Identical |
| How Supplied | Sterile<br>Single Patient Use | Sterile<br>Single Patient Use | Identical |
| Sterilization | EO | EO | Identical |
| SAL | 10-6 | 10-6 | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
#### Note 0: Indication for Use
The Indications for Use for the subject device is slightly different from the predicate device. The intended use is the same for both devices.
#### Note 1: Needle Length
The needle length of the subject devices is available in more sizes than the predicate device. The difference does not raise new questions on the safety and effectiveness of the proposed device.
### Note 2: Syringe Volume
The subject devices' syringe volume is different than the predicate device but is within the same range. The difference does not raise new questions about the safety and effectiveness of the proposed device.
#### Note 3: Needle Gauge
The needle gauge of the subject devices is a more different from the predicate device. The difference does not raise new questions about the safety and effectiveness of the proposed device.
#### Note 4: Needle Hub
The needle hub's material and colorant of the subject device is identical to the predicate device and all conform to ISO 6009 and ISO 10993.
### Note 5: Performance
The subject device conforms to the current FDA recognize standards ISO80369-7 and ISO80369-20, which will replace ISO594-1 and ISO594-2.
{9}------------------------------------------------
## Performance Testing
To establish substantial equivalence to the identified predicate device, we performed the tests noted below on the subject devices. The testing results proved that the device complies with the applicable standards requirement and is substantially equivalent to the predicate devices.
## Non-Clinical Performance Testing
Testing was performed to evaluate the functional performance and safety of the subject device with the following standards:
## Performance:
- ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - - Part 1: Syringes for manual use
- ISO 9626 Second edition 2016-08-01 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864 Fourth edition 2016-08-01 Sterile hypodermic needles for single use --Requirements and test methods
- -ISO 80369-7 First edition 2016-10-15 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- ISO 80369-20 First edition 2015-05-15, small-bore connectors for liquids and gases in healthcare applications part 20: standard test methods. (General I (QS/RM))
- ISO 6009 Fourth edition 2016-08-01 Hypodermic needles for single use Colour coding for identification
{10}------------------------------------------------
# Biocompatibilitv
The subject device is classified as an external communication device → Circulating Blood -> Limited Contact Duration (≤24h). ISO 10993-5: 2009 In Vitro Cytotoxicity ISO 10993-10: 2010 Sensitization ISO 10993-10: 2010 Intracutaneous Reactivity ISO 10993-11: 2017 Acute Systemic Toxicity ISO 10993-4:2017 Coagulation ISO 10993-4: 2017 Complement activity
ISO 10993-4: 2017 & ASTM F756-17: Hemolytic Properties
# Sterilization Validation:
ISO11135-1:Sterilization of health care products - ethylene oxide - part 1: requirements for the development, validation, and routine control of a sterilization process for medical devices.
ISO11737-1:Sterilization of medical devices-Microbiological methods-Part 1:
Determination of the population of microorganisms on product.
ISO11737-2: Sterilization of medical devices -- Microbiological methods -- Part 2: Tests of sterility performed in the validation of a sterilization process.
ISO 10993-7: Biological evaluation of medical devices - Part 7: Test of Ethylene Oxide Residuals.
AAMI / ANSI ST72: Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing.
# Package and Shelf Life:
# Applicable Standards:
AAMI/ANSI/ISO 11137-1:2006/(R) 2010 Sterilization of health care products -Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices
AAMI/ANSI/ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation, and maintenance of a sterilization process
AAMI/ANSI/ISO 11607-1:2006/(R) 2010 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems, 3ed.
ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
{11}------------------------------------------------
ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
ASTM D3078-02 (2021), Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission. (Sterility)
DIN58953-6: 2016 Sterilization - Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized.
ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
# The package integrity test noted below was conducted for the sterilization package of the subject devices by accelerated aging testing followed by simulated shipping distribution testing:
- Accelerated Aging Test ।
- Simulated shipping distribution testing -
- Visual inspection -
- Performance Inspection (Chemical performance and Physical performance) -
- Sterile Test -
- Vacuum Leak Test -
- Dye penetration test -
- Agar Contact-Attack Test -
- Tensile Seal Strength Test -
# Clinical Performance Data
No clinical data was submitted in this submission.
# Conclusions:
Based on the verification test results, the subject devices conform to the same applicable standards requirements such as performance and biocompatibility as the predicate device. The subject device uses the same fundamental scientific technology, same indications for use, sterilization methods, and the same shelf life and packaging. The differences do not raise new questions about the safety and effectiveness of the proposed device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.