K222112 · Medical Precision BV · QRN · Mar 30, 2023 · Radiology
Device Facts
Record ID
K222112
Device Name
Comfort Marker 2.0
Applicant
Medical Precision BV
Product Code
QRN · Radiology
Decision Date
Mar 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5785
Device Class
Class 2
Indications for Use
The device is indicated for use for applying ink to the skin to identify the margins for radiation therapy. The device is intended to be used in clinical settings by Radiotherapy professionals.
Device Story
Comfort Marker 2.0 is an application system for placing reference tattoo markers on patients undergoing radiotherapy. The device comprises a control unit, a pen module, and a safety needle. The pen module uses a driving rod to actuate the safety needle; a measuring coil monitors needle movement to control depth settings. The device is used in clinical radiotherapy departments by radiotherapy professionals. The control unit is rechargeable via a docking station. The device provides accurate, durable reference points, reducing procedural pain and improving cosmesis compared to manual lancets. It enhances clinical workflow by providing a standardized, easier-to-use method for marking patients, with integrated safety features to prevent needle stick injuries.
Clinical Evidence
Clinical trial (n=100) compared Comfort Marker 2.0 to lancets. Primary endpoints: patient comfort (pain) and effectiveness (visibility). Results: 44% of Comfort Marker patients reported painless procedure vs 16% for lancets (p=0.008). Effectiveness: 100% of Comfort Marker markings remained good/excellent throughout treatment vs 67% for lancets (p=0.003). Secondary endpoints: professional satisfaction (98% vs 78% ease of use, p=0.008) and cosmesis (median score 4.5 vs 3.5, p<0.001). No sharps injuries reported.
Technological Characteristics
Battery-powered (3.7V, 2.4Ah Li-ion) handheld device with control unit and pen. Materials: Zylar 960 (housing), AISI 304H stainless steel (needles), UGI S4310-6 (spring), Dymax 1405-M-UR-SC (glue), ABS/PC+ABS plastics. Sensing: Measuring coil for depth control. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993 (biocompatibility), IEC 60601-1/1-2/1-6 (electrical safety/EMC), ISO 11135 (sterilization), ISO 14971 (risk management).
Indications for Use
Indicated for applying ink to skin to identify radiation therapy margins in patients with undamaged skin undergoing repeated radiotherapy. Contraindicated for use on damaged or dermatitis skin.
Regulatory Classification
Identification
A radiation therapy marking device is a powered device that transdermally delivers a permanent or temporary colorant to the skin for the purpose of placing marks to guide radiation therapy. The Comfort Marker 2.0 is an application system for placing reference marks on or in patient's skin to facilitate set-up for radiation therapy treatment, consisting of a battery-powered control unit with pen, a main-powered docking station, and a sterile, single-use safety needle.
Special Controls
In combination with the general controls of the FD&C Act, the radiation therapy marking device is subject to the following special controls:
- (1) Design verification and validation must include:
- Documentation of performance data from studies that demonstrate: (i)
- (A) The indicated colorant is compatible with the device and its method of delivery:
- (B) The device can reproducibly deliver the indicated colorant with the specifications described; and
- (C) The length of time that compatible colorants remain visible on the skin following device application.
- (ii) Documentation of performance data from studies that demonstrate:
- Accuracy and reproducibility of needle penetration depth; (A)
- Device protection from cross-contamination, including fluid ingress (B) protection:
- Adequacy of the cleaning and disinfection instructions to ensure that the (C) reusable components of the device can be cleaned and disinfected; and
- (D) The sterility of all patient-contacting components (e.g., safety needle).
- Documentation of performance data from studies that demonstrate electrical (iii) safety and electromagnetic compatibility (EMC) of all electrical components of the device.
- Documentation of performance data from studies that demonstrate continued (iv) sterility, package integrity, and device functionality over the intended shelf life.
- (v) Documentation of software verification, validation, and hazard analysis.
- The labeling required under 21 CFR 801.109(c) must include: (2)
- An explanation of the device and the mechanism of operation; (i)
- Validated methods and instructions for reprocessing of any reusable components; (ii)
- (iii) Disposal instructions: and
- A shelf life for all sterile components. (iv)
{0}------------------------------------------------
March 30, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medical Precision BV % Cherita Jones Regulatory Consultant M Squared Associates, Inc 127 West 30th St. Floor 9 NEW YORK NY 10001
Re: K222112
Trade/Device Name: Comfort Marker 2.0 Regulation Number: 21 CFR 892.5785 Regulation Name: Radiation therapy marking device Regulatory Class: Class II Product Code: QRN Dated: February 27, 2023 Received: February 27, 2023
Dear Cherita Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is for Lora D. Weidner -S. The date of the signature is 2023.03.30, and the time is 20:36:13 -04'00'.
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222112
Device Name
Comfort Marker 2.0
Indications for Use (Describe)
The device is indicated for use for applying ink to the skin to identify the margins for radiation therapy. The device is intended to be used in clinical settings by Radiotherapy professionals.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <div style="display:inline-block;"> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
The following information is provided as required by 21 CFR § 807.87 for Traditional 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
| Contact Details | |
|-----------------------------------|----------------------------------|
| Applicant Name | Medical Precision b.v. |
| Applicant Address | Telfordstraat 9 - 30 |
| | NL 8013 RL, ZWOLLE |
| | The Netherlands |
| | |
| Applicant Contact | Blerta Kukaj |
| Applicant Contact E-mail | kukaj@medicalprecision.nl |
| Correspondent Name | M Squared Associates, Inc |
| Correspondent Address | 901 King Street, Suite 200 |
| | Alexandria, VA 22314 |
| | |
| Correspondent Telephone Number | 347-954-0624 |
| Correspondent Contact | Cherita James |
| Correspondent E-mail | cjames@msquaredassociated.com |
| Submission Date | 27 February 2023 |
| Device Name | |
| Device Trade Name | Comfort Marker 2.0 |
| Common Name | Radiation Therapy Marking Device |
| Classification | II |
| Regulation | 21 CRF 892.5785 |
| Product Code | QRN |
| Legally Marketed Predicate Device | |
| Predicate Trade Name | Comfort Marker 2.0 (DEN200041) |
| Product Code | QRN |
| | |
# Device Description Summary
Comfort Marker 2.0 is an application system for placing reference marks on patients eligible for Radiotherapy treatments.
The Comfort Marker 2.0 allows the user to place accurate tattoo markers (reference points) on patients enabling radio therapy. The device consists of a Control and Pen module which drives a Safety Needle.
Different depth settings can be chosen to accommodate for different skin types. When the tattoo marking has been placed the Control Unit can be stored and charged in the Docking station.
## Intended Use/Indications for Use
The device is indicated for use for applying ink to the skin to identify the margins for radiation therapy. The device is intended to be used in clinical settings by Radiotherapy professionals.
## Indications for Use Comparison
No change in intended use
#### Technological Comparison
Both the subject (Comfort Marker 2.0 in this 510k) and the predicate device (Comfort Marker 2.0, De Novo DEN200041) are the same in this case. No technological differences exist. Comfort Marker 2.0 received De novo clearance on 10th of December 2021. This 510(K) is prepared to include clinical
{4}------------------------------------------------
data on pain experience by patients undergoing radiotherapy treatment. Therefore, the predicate and the legally marketed device are substantially equivalent.
| Item | Subject Device | Predicate Device<br>DEN200041 | Similarities/Differences |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Product Code | QRN | QRN | No difference in<br>FDA product code |
| Trade Name | Comfort Marker 2.0 | Comfort Marker 2.0 | No difference |
| 510 (K) number | Not assigned | De Novo: DEN200041 | N/A |
| Intended use | The device is intended to be<br>used in clinical setting by<br>Radiotherapy professionals | The device is intended to be<br>used in clinical setting by<br>Radiotherapy professionals | No difference |
| Indications for<br>use | The device is indicated for<br>use for applying ink to the<br>skin to identify the margins<br>of radiation therapy. | The device is indicated for<br>use for applying ink to the<br>skin to identify the margins<br>of radiation therapy. | No difference |
| Target<br>population | Patients with undamaged<br>skin undergoing repeated<br>radiotherapy. | Patients with undamaged<br>skin undergoing repeated<br>radiotherapy. | No difference: |
| Contra<br>indications | Do not use the device on<br>damaged or dermatitis skin. | Do not use the device on<br>damaged or dermatitis skin. | No difference |
| Where used | In a Radiotherapy<br>department of a hospital | In a Radiotherapy<br>department of a hospital | No difference |
| Energy used /<br>delivered | Battery: Rechargeable<br>battery, 3.7 V, 2.4 Ah,<br>Lithium Ion<br><br>AC/DC power supply:<br>18W, 12V, 1.5A | Battery: Rechargeable<br>battery, 3.7 V, 2.4 Ah,<br>Lithium Ion<br><br>AC/DC power supply:<br>18W, 12V, 1.5A | No difference |
| Design /<br>operating<br>principle | Longitudinal movement of<br>the driving rod in the pen<br>actuates the safety needle.<br>The movement of the safety<br>needle is measured by a<br>measuring coil and to<br>control desired depth<br>setting. | Longitudinal movement of<br>the driving rod in the pen<br>actuates the safety needle.<br>The movement of the safety<br>needle is measured by a<br>measuring coil and to<br>control desired depth<br>setting. | No difference |
| Clinical<br>Performance | Clinical trial was performed<br>on 100 patients. Pain and<br>user experience was<br>recorded. | N/A | Clinical data on pain<br>experience, user<br>experience and<br>clinical effectiveness<br>was recorded in this<br>study. |
| Bench<br>Performance | • Needle depth<br>• Needle sharpness test<br>• Drop test<br>• Ink migration test<br>• Needle pattern test<br>• Glue strength test<br>• Needle protrusion test | • Needle depth<br>• Needle sharpness test<br>• Drop test<br>• Ink migration test<br>• Needle pattern test<br>• Glue strength test<br>• Needle protrusion test | No difference |
| | Durability of reference<br>•<br>points on human<br>volunteers | Durability of reference<br>•<br>points on human<br>volunteers | |
| Standards met | Biological evaluation:<br>• ISO 10993-1, 2018<br>• ISO 10993-10,<br>2009,<br>• ISO 10993-5, 2009<br>Sterilization:<br>• ISO<br>11135,2014/Amd<br>1:2018<br>Packaging:<br>• ISO 11607-1, 2019<br>• ISO 11607-2, 2019<br>Risk Management:<br>• ISO 14971:2012<br>Electrical safety:<br>• IEC 60601-1,<br>2005/A1:2012<br>• IEC 60601-1-2,<br>2014<br>• IEC 60601-1-6,<br>2010<br>Clean rooms:<br>• ISO 14644-1, 2015<br>• ISO 14644-2, 2015<br> | Biological evaluation:<br>• ISO 10993-1, 2018<br>• ISO 10993-10, 2009,<br>• ISO 10993-5, 2009<br>Sterilization:<br>• ISO 11135,2014/Amd<br>1:2018<br>Packaging:<br>• ISO 11607-1, 2019<br>• ISO 11607-2, 2019<br>Risk Management:<br>• ISO 14971:2012<br>Electrical safety:<br>• IEC 60601-1,<br>2005/A1:2012<br>• IEC 60601-1-2, 2014<br>• IEC 60601-1-6, 2010<br>Clean rooms:<br>• ISO 14644-1, 2015<br>• ISO 14644-2, 2015<br> | No difference |
| Device<br>materials | Safety Needle:<br>• Needle housing:<br>Zylar 960 (Methyl<br>Methacrylate<br>Butadiene Styrene)<br>• Needles: Stainless<br>Steel AISI 304H<br>• Spring: Stainless<br>steel UGI S4310-6<br>• Glue: Dymax<br>1405-M-UR-SC<br>Pen :<br>• Pen shield : Ferro<br>Magnetic steel<br>1018<br>• Pen ring: Ferro<br>Magnetic iron-<br>ETG100<br>• Pen actuator: ABS<br>PA<br>757_CHIMET_Pol<br>ylacR<br>• Pen coilformer:<br>ABS PA<br>757_CHIMET_Pol<br>ylacR | Safety Needle:<br>• Needle housing:<br>Zylar 960 (Methyl<br>Methacrylate<br>Butadiene Styrene)<br>• Needles: Stainless<br>Steel AISI 304H<br>• Spring: Stainless<br>steel UGI S4310-6<br>• Glue: Dymax<br>1405-M-UR-SC<br>Pen :<br>• Pen shield : Ferro<br>Magnetic steel<br>1018<br>• Pen ring: Ferro<br>Magnetic iron-<br>ETG100<br>• Pen actuator: ABS<br>PA<br>757_CHIMET_Pol<br>ylacR<br>• Pen coilformer:<br>ABS PA<br>757_CHIMET_Pol<br>ylacR | No difference |
| | | | |
| | Control Unit:<br>PC+ABS PC510 CHIMEI WONDERLOY Silicone rubber Docking station:<br>PC+ABS PC510 CHIMEI WONDERLOY | Control Unit:<br>PC+ABS PC510 CHIMEI WONDERLOY Silicone rubber Docking station:<br>PC+ABS PC510 CHIMEI WONDERLOY | |
| Biocompatibility | Complies with ISO 10993 series of standards for Cytotoxicity Skin Irritation Skin Sensitization Material Mediated pyrogenicity Acute Systemic Toxicity | Complies with ISO 10993 series of standards for Cytotoxicity Skin Irritation Skin Sensitization Material Mediated Pyrogenicity Acute Systemic Toxicity | No difference |
| Electrical safety<br>and EMC | Complies with IEC 60601 series of standards: IEC 60601-1 ,2005/A1:2012 IEC 60601-1-2, 2014 IEC 60601-1-6, 2010 | Complies with IEC 60601 series of standards: IEC 60601-1 ,2005/A1:2012 IEC 60601-1-2, 2014 IEC 60601-1-6, 2010 | No difference |
| Sterility of<br>safety needle | SAL: 10-6<br>Ethylene oxide sterilization<br>Single use component of the device system | SAL: 10-6<br>Ethylene oxide sterilization<br>Single use component of the device system | No difference |
| Mechanical<br>safety | The safety Needle extends max up to 1 mm during operation (mechanical stop) The safety needle is covered by housing to prevent needle stick injuries while not in operation | The safety Needle extends max up to 1 mm during operation (mechanical stop) The safety needle is covered by housing to prevent needle stick injuries while not in operation | No difference |
| Software and<br>Cybersecurity | The software does not impact any risk related solutions and no injury or damage to health is possible due to software. The software is therefore classified as class A Complies with IEC 62304:2006-05 and IEC62304:2015-06 | The software does not impact any risk related solutions and no injury or damage to health is possible due to software. The software is therefore classified as class A Complies with IEC 62304:2006-05 and IEC62304:2015-06 | No difference |
| MRI safety | MRI unsafe | MRI unsafe | No difference |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## Non-Clinical and/or Clinical test Summary and Conclusions
Non-Clinical and Clinical Summary
Product functionality: The Control unit, the pen and the safety needle all meet the requirements presented in bench testing.
Biocompatibility: The Safety Needle meets cytotoxicity requirements of ISO 10093-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity.
Biocompatibility: The Safety Needle meets sensitization requirements of ISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitization.
Biocompatibility: The safety Needle meets the Material Mediated pyrogenicity requirements of ISO 10993-11:2017 Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity.
Biocompatibility: The safety Needle meets the Acute Systemic Toxicity requirements of ISO 10993-11:2017 Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity.
Sterilization: A Sterility Assurance Level (SAL) of 10-6 has been validated in accordance with the requirements of ISO 11135:2014 for Ethylene Oxide.
Electrical Safety and EMC: The device meets safety and EMC standards requirements of IEC 60601-1 ,2005/A1:2012, IEC 60601-1-2, 2014 and IEC 60601-1-6, 2010
Software and Cybersecurity: The software is compliant with Complies with IEC 62304:2006-05 and IEC62304:2015-06 Medical device software life cycle processes.
Clinical simulated use testing: The performance of Comfort Marker 2.0 and durability of reference points (at 3 different depth settings) were tested on 6 healthy volunteers. The reference points were tracked for 12 weeks for visibility, migration and clearance. The healthy volunteer study showed that Comfort Marker 2.0 can place well defined reference points at all three depth settings using three inks. The reference points did not show any migration on the skin. All reference points from three inks were clearly visible during week 2 until 8, a period in which radiotherapy treatment is provided.
Clinical testing: The aim of the randomized, multi-arm, double-blind study with concurrent ("active") control was to establish whether the use of Comfort Marker 2.0 translates into a benefit in terms of comfort, satisfaction, effectiveness, and cosmesis compared to the use of lancets (standard, control). The clinical study was performed in Portugal between October 2021 and January 2022. Target one hundred patients (18 years or older) were enrolled (50 assigned to lancet arm and 50 assigned to Comfort Marker 2.0 arm.
Patient demographic and clinical characteristics: The median age of all patients included in the trial was 61 (25-85 years). The majority of patients within the clinical trial were women (73%) with men making up 27% of the patient population. The number of set-up markings was also well balanced: 64.0% of the patients in the lancets group and 68.0% of those in the CM group had received >4 set-up markings, with a median of 9 in both groups. Most of the patients included were referred to irradiate breast or chest wall (61.0%), followed by pelvis (22.0%) and thorax (11.0%).
Patient accountability for the four endpoints defined for the trial is presented in the table below:
{8}------------------------------------------------
| Stage | Investigational<br>device arm<br>(Comfort<br>Marker 2.0) | Control<br>arm<br>(lancets) | Total |
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------|-------|
| Enrollment | 50 | 50 | 100 |
| Treatment (reference points tattooing) | 50 | 50 | 100 |
| Primary outcome endpoint analysis: Patient's comfort<br>(experienced pain during tattooing process) | 50 | 50 | 100 |
| Primary outcome endpoint analysis: effectiveness<br>(reference points visibility for radiotherapy professionals) | 50 | 48 | 98 |
| Secondary outcome endpoint analysis: Radiotherapy<br>professional satisfaction | 50 | 50 | 100 |
| Secondary outcome endpoint analysis: Cosmesis<br>(aesthetic appearance of reference points) | 50 | 48 | 98 |
The Comforttatto trial met all endpoints (outcome measures) set at the start of the trial. No adverse events were reported during the trial.
- 1. Patients comfort endpoint: The percentage of patients that graded the tattooing process as painless was significantly higher for patients receiving Comfort Marker 2.0 compared to lancets (44.0% vs. 16.0%, respectively; p = 0.008). (Table 2 and Figure S1)
- 2. Effectiveness endpoint: No fading of the reference points for both lancet and Comfort Marker 2.0 arms was recorded. During tattoo quality assessment by radiotherapy professionals on 4point scale (bad, reasonable, good and excellent), patient's receiving Comfort Marker 2.0 had a significantly higher proportion of radiotherapy professionals reported good and excellent quality markings compared to those receiving lancets. The median score of set-up markings graded as good/excellent on the last evaluation compared to the first evaluation was significantly worse in patients receiving lancets (67% v 89% respectively; p =0.003). The same effect wasn't found on patients receiving CM (100.0% for both evaluations; p=0.173).
- Satisfaction endpoint: Compared to those receiving lancets, patients receiving Comfort Marker 3. 2.0 had significantly higher radiotherapy professionals reported tattooing processes evaluated as easy (78.0% vs. 98.0%, respectively; p=0.008).
- 4. Cosmesis endpoint: The photographic assessment (reference points photos were taken on one of the last three days of treatment and grading radiotherapy professionals were blind to patient identify and trial arm) of the reference points in both trial arms was performed by 20 observers (both physicians and radiotherapy professionals) on 4-point scale (bad, reasonable, good and excellent). Patients receiving Comfort Marker 2.0 had a significantly higher score on the photographic assessment (Table 5), with a median score of 3.5 and 4.5 for the lancets and the Comfort Marker 2.0 group, respectively (p < 0.001).While 84.0% of the patients in the Comfort Marker 2.0 group had a mean score of at least 4, in the lancets group that number is five times lower (16.7%; p < 0.001).
Additionally, data on sharps injuries was collected. Sharp injuries were considered any incident which causes the needle of the lancet or the CM to inadvertently penetrate the skin of the radiotherapy professionals performing the tattooing during any of the tattooing process (material assembly, tattooing, or sharps disposal). Outcome: no sharp injuries were registered during the Comforttattoo trial.
Conclusion: Based on the two device comparison characteristics presented in this 510K, the subject device is substantially equivalent to the legally marketed predicate device DEN200041. Clinical data
{9}------------------------------------------------
provided supports the Comfort Marker 2.0 claims as: a) Patients marked with Comfort Marker 2.0 reported less procedural pain than patients marked with lancets, b) Physicians and radiation therapist doing photographic assessment of cosmesis rated Comfort Marker 2.0 better than lancets. - Limitation/ additional claim information: study performed in Europe where majority of the population is Caucasian race, specific to Portugal where the study was performed the majority is from a Mediterranean ethnicity,
c) Radiation therapists rated placing tattoos with Comfort Marker 2.0 easier compared to the use of lancets.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.