K222072 · Patterson Dental Supply, Inc. · LLZ · Aug 8, 2022 · Radiology
Device Facts
Record ID
K222072
Device Name
Dolphin Imaging 12.0
Applicant
Patterson Dental Supply, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 8, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Dolphin Imaging 12.0 software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
Device Story
Dolphin Imaging 12.0 is a dental imaging software suite for specialized dental practices; used by orthodontists and oral surgeons. Inputs include 2D/3D patient radiographs and images. Software modules perform cephalometric tracing, VTO treatment simulation, Arnett/Gunson FAB analysis, McLaughlin VTO, 2D implant planning, and 3D orthodontic/surgical case planning. Operates on local PC workstations (64-bit Windows). Provides diagnostic and treatment planning tools; output is visualized by the clinician to assist in case diagnosis and treatment planning. Benefits include efficient management of patient records, accurate simulation of surgical/orthodontic outcomes, and improved communication for multidisciplinary cases. Interpretation of results remains the responsibility of the licensed practitioner.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including manual, integration, system, and regression testing, alongside adherence to recognized standards (IEC 62366, IEC 62304, ISO 14971, NEMA PS 3.1-3.20).
Technological Characteristics
Software-based medical image management and processing system. Operates on 64-bit Microsoft Windows PC workstations. Complies with DICOM (NEMA PS 3.1-3.20) for interoperability. Software life cycle follows IEC 62304; risk management follows ISO 14971. Usability engineering follows IEC 62366. Moderate level of concern.
Indications for Use
Indicated for specialized dental practices, including orthodontic and oral maxillofacial surgery, for capturing, storing, and presenting patient images to assist in treatment planning and case diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
August 8, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Patterson Dental Supply, Inc. % Ms. Erin Daly Sr. Regulatory Manager Dolphin Imaging & Management Solutions 9200 Oakdale Avenue, Ste 500 CHATSWORTH CA 91311
Re: K222072
Trade/Device Name: Dolphin Imaging 12.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: July 13, 2022 Received: July 14, 2022
Dear Ms. Daly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222072
Device Name Dolphin Imaging 12.0
# Indications for Use (Describe)
Dolphin Imaging 12.0 software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) SUMMARY
#### Dolphin Imaging 12.0 K22072
#### Administrative Information
| Date Prepared: | 13 July 2022 |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Patterson Companies<br>1031 Mendota Heights Road<br>Saint Paul, MN 55120<br>Contact: Erin Daly<br>Telephone: (608) 345-4888 |
| Trade or Proprietary Name: | Dolphin Imaging 12.0 |
| Common or Usual Name: | Dental imaging software |
| Classification Name: | Medical image management and processing system |
| CFR Reference: | 21 CFR 892.2050 |
| Product Code: | LLZ |
| Regulatory Class: | Class II |
| Predicate Device: | K110430 - Dolphin Imaging |
| Reference Device(s): | Not applicable |
| Predicate Device Trade or Proprietary<br>Name: | Dolphin Imaging 11.5<br>K110430 |
| Common or Usual Name: | Dental imaging software |
| Classification Name: | Medical image management and processing system |
| CFR Reference: | 21 CFR 892.2050 |
| Product Code: | LLZ |
#### Product Description
Regulatory Class:
Dolphin Imaging 12.0 software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
Class II
Dolphin Imaging 12.0 software provides imaging, diagnostics, and case presentation capabilities for dental specialty professionals. The Dolphin Imaging 12.0 suite of software products is a collection of modules that together provide a comprehensive toolset for the dental specialty practitioner. Users can easily manage 2D/3D images and x-rays; accurately diagnose and treatment plan, quickly communicate and present cases to patients and can work efficiently with colleagues on multidisciplinary cases. The below functionalities make up the medical device modules:
{5}------------------------------------------------
| Module | Overview |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cephalometric Tracing | Digitize landmarks on a patient's radiograph, trace cephalometric<br>structures, view cephalometric measurements, superimpose images for<br>analysis and perform custom analysis. |
| Treatment Simulation (VTO) | Simulate orthodontic and surgical treatment results using Visual Treatment<br>Objective (VTO) and growth features. |
| Arnett/Gunson FAB Analyses | Perform face, airway, bite (FAB) analysis and simulate treatment for<br>orthodontic and surgical cases based on the methodologies of Dr. William<br>Arnett. |
| McLaughlin Dental VTO | Analyze and evaluate orthodontic and surgical visual treatment objective<br>(VTO) based on the theories of Dr. Richard McLaughlin. |
| Implanner™ | Plan dental implant procedures in 2D. |
| Dolphin 3D | Plan, diagnose and present orthodontic and surgical cases, airway analysis,<br>study models, implant planning and surgery treatment simulation in 3D. |
Dolphin Imaging 12.0 is a software of Moderate level of concern. Dolphin Imaging maintains all the same key medical device functionality as the predicate Dolphin Imaging 11.5 (K110430). Usability enhancements to the Cephalometric Tracing and Analysis, Treatment Simulation (VTO), and Dolphin 3D modules are included in Dolphin Imaging 12.0, as well as system updates to allow use with current operating software and hardware capabilities
# Indications for Use
Dolphin Imaging 12.0 software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
# Comparison to Predicate
| Parameter | Subject Device | Predicate |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K222072 Dolphin Imaging 12.0 | K110430 – 11.5 Dolphin Imaging |
| Intended Use | Dolphin Imaging 12.0 software is<br>designed for use by specialized dental<br>practices for capturing, storing and<br>presenting patient images and assisting<br>in treatment planning and case<br>diagnosis. Results produced by the<br>software's diagnostic and treatment<br>planning tools are dependent on the<br>interpretation of trained and licensed<br>practitioners. | Dolphin Imaging software is designed<br>for use by specialized dental practices<br>for capturing, storing and presenting<br>patient images and assisting in<br>treatment planning and case diagnosis.<br>Results produced by the software's<br>diagnostic and treatment planning tools<br>are dependent on the interpretation of<br>trained and licensed practitioners. |
| Classification | Class II | Class II |
| Classification Code | LLZ | LLZ |
| Type of Use (Rx / OTC) | Rx | Rx |
| User | Orthodontic, oral maxillofacial surgery,<br>and dental-specialty practices. | Orthodontic, oral maxillofacial surgery,<br>and dental-specialty practices. |
| User Interface/OS | Local PC workstation with 64-bit<br>Microsoft Windows OS | Local PC workstation with Microsoft<br>Windows OS |
| Medical Device Features | | |
{6}------------------------------------------------
| 2D Image Digitize and<br>Measurements | Yes | Yes |
|-----------------------------------------|-----|-----|
| 2D Image Superimposition | Yes | Yes |
| 2D Growth Forecast | Yes | Yes |
| 2D Treatment Simulation | Yes | Yes |
| 2D Custom Analysis Editor | Yes | Yes |
| 3D Image Digitize and<br>Measurements | Yes | Yes |
| 3D Airway Analysis | Yes | Yes |
| 3D Image (Volume)<br>Superimposition | Yes | Yes |
| 3D Implant Simulation | Yes | Yes |
| 2D/3D Orthognathic Surgical<br>Planning | Yes | Yes |
# Non-Clinical Performance Testing
Dolphin Imaging 12.0 complies with applicable FDA recognized standards:
| Usability (IEC) | Medical devices- Part 1: Application of usability engineering to |
|---------------------------------------------------|------------------------------------------------------------------|
| IEC 62366 | medical devices |
| Software (IEC)ANSI/AAMI/IEC 62304: 2006 & A1:2016 | Medical device software- Software life cycle process |
| DICOM (NEMA)NEMA PS 3.1-3.20 (2016) | Digital Imaging and Communications in Medicine (DICOM) |
| Risk Management | Medical devices- Application of risk management to medical |
| ISO 14971:2019 | devices |
Dolphin Imaging 12.0 design and performance verification involved the following tests:
- Performance Testing ●
- Manual Testing/ Integration Testing ●
- . System and Regression testing
#### Clinical Performance Testing
No clinical testing was required to support substantial equivalence.
#### Conclusions
Dolphin Imaging 12.0 is substantially equivalent to the predicate device, Dolphin Imaging, and differences do not raise any concerns about safety or efficacy of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.