K222035 · Avatar Medical · LLZ · May 24, 2023 · Radiology
Device Facts
Record ID
K222035
Device Name
Avatar Medical Software V1
Applicant
Avatar Medical
Product Code
LLZ · Radiology
Decision Date
May 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.
Device Story
AMS V1 is a software-only medical imaging system; inputs DICOM-compatible CT and MR images from local storage or PACS. Transforms data into 3D and VR formats for visualization; allows manual measurements, annotations, and image filtering. Operates on off-the-shelf (OTS) workstations; VR interface requires compatible OTS headsets. Used by clinicians, surgeons, and imaging technicians in hospitals/clinics. Intraoperative use restricted to non-sterile review rooms or display on existing monitors; not for use in sterile fields or direct patient contact. Provides 3D/VR visualization to assist preoperative planning and intraoperative review; clinician remains responsible for final patient management decisions.
Clinical Evidence
Bench testing only. Software verification and validation performed per ANSI AAMI IEC 62304. Measurement accuracy validated against reference digital phantoms and Osirix MD (K101342) using t-test statistics; no statistical difference found. Display quality evaluated against AAPM guidance for luminance/contrast. VR platform optical testing performed per IEC 63145-20-20. Filter technology and VR fluidity (FPS) verified against thresholds.
Technological Characteristics
Software-only device; runs on OTS workstations. Supports DICOM standard. Features 3D rendering and VR interface. Measurement tools include linear and diameter. Connectivity via LAN for remote streaming. Cybersecurity controls implemented. Complies with ANSI AAMI IEC 62304 for software lifecycle and IEC 63145-20-20 for VR optical performance.
Indications for Use
Indicated for use by healthcare professionals for processing, reviewing, and analyzing multi-dimensional CT or MR images for preoperative surgical planning and intraoperative display. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Avatar Medical % Rory Carrillo Quality and Regulatory Consultant Cosm 45 Bartlett St. SAN FRANCISCO CA 94110
May 24, 2023
Re: K222035
Trade/Device Name: Avatar Medical Software V1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: April 21, 2023 Received: April 24, 2023
Dear Rory Carrillo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222035
#### Device Name AVATAR MEDICAL Software V1
#### Indications for Use (Describe)
AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5. 510(k) Summary
#### 1. General Information
#### K222035
| 510(k) Sponsor | Avatar Medical |
|-----------------------|---------------------------------------------------------------|
| Address | 11 rue de Lourmel<br>75015 Paris France |
| Correspondence Person | Rory A. Carrillo<br>Quality and Regulatory Consultant<br>Cosm |
| Contact Information | Email: rory@cosmhq.com<br>Phone: 562-533-7010 |
| Date Prepared | November 22, 2022 |
## 2. Subject Device
| Proprietary Name | Avatar Medical Software V1 |
|---------------------|------------------------------------------------|
| Common Name | AMS V1 |
| Classification Name | System, Image Processing, Radiological |
| Regulation Number | 21 CFR 892.2050 |
| Regulation Name | Medical Image Management and Processing System |
| Product Code | LLZ |
| Regulatory Class | II |
#### Predicate Device 3.
| Proprietary Name | Ceevra Reveal 2.0 |
|------------------------|------------------------------------------------|
| Premarket Notification | K173274 |
| Classification Name | System, Image Processing, Radiological |
| Regulation Number | 21 CFR 892.2050 |
| Regulation Name | Medical Image Management and Processing System |
| Product Code | LLZ |
| Regulatory Class | II |
## 4. Device Description
The Avatar Medical Software V1 (AMS V1) is a software-only device that allows trained medical professionals to review CT and MR image data in three-dimensional (3D) format and/or in virtual reality (VR) interface. The 3D and VR images are accessible through the software desktop application and, if desired, through compatible VR headsets which are used by users for preoperative surgical planning and for display during intervention/surgery.
{4}------------------------------------------------
The AMS V1 product is to be used to assist in medical image review. Intended users are trained medical professionals, including imaging technicians, clinicians and surgeons.
AMS V1 includes two main software-based user interface components, the Desktop Interface and the VR Interface. The Desktop Interface runs on a compatible off-the-shelf (OTS) workstation provided by the hospital and only accessed by authorized personnel. The Desktop Interface contains a graphical user interface where a user can retrieve DICOM-compatible medical images locally or on a Picture Archiving Communication System (PACS). Retrieved CT and MR images can be viewed in 2D and 3D formats. Users are able to make measurements, annotations, and apply fixed and manual image filters.
The VR Interface is accessible via a compatible OTS headset to allow users to review the medical images in a VR format. VR formats can be viewed only when the user connects a compatible VR headset directly to the workstation being used to view the Desktop Interface. Additionally. AMS V1 enables the intended users to remotely stream the Desktop Interface to another workstation on the same local area network (LAN).
The 3D images generated using AMS V1 are intended to be used in relation to surgical procedures in which CT or MR images are used for preoperative planning and/or during intervention/surgery.
The intraoperative use of the AMS V1 solely corresponds to the two following cases:
- Display of the AMS V1 Desktop Interface on existing monitors/screens in the operating ● room
- . Use in a non-sterile imaqe review room accessible from the operating room during the procedure (AMS V1 operates on VR headsets which are not approved to be used in the sterile environment of the operating room)
#### Usage During Interventional and Surgical Procedures
The AVATAR MEDICAL Software V1 is not meant to be used in a sterile environment.
The AVATAR MEDICAL Software V1 is not meant to register 3D images to patients.
Two usage scenarios of the AVATAR MEDICAL Software V1 software are possible during an interventional or surqical procedure:
#### 1. Inside the operating room
The Desktop Interface (no VR) can be displayed on a computer monitor (previously installed respecting the management of sterility) inside the operating room. It is not expected that users will analyze the image, optimize filters, and realize measurements
{5}------------------------------------------------
in this intraoperative context. This usage is for visualizing results of a previous surgery planning session. The user will follow their scrub-in and scrub-out facility protocols when exiting and entering the sterile field of the operating room.
#### 2. Outside the operating room
The full software experience, including VR, can be accessed in an adjacent non-sterile image review room. In this usage, the user will need to scrub-out to access the review room and, after review, scrub-in following their facility protocols as required. The user will not be in contact nor in the presence of the patient while wearing the VR headset
The software is not meant to be used with the user's hands on the patient during an interventional or surqical procedure.
### 5. Intended Use and Indications for Use
AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.
### 6. Comparison of Technological Characteristics
| Feature/<br>Function | Subject Device:<br>AMS V1 | Predicate Device:<br>Ceevra Reveal 2.0<br>(K173274) |
|-------------------------|--------------------------------------------------------|--------------------------------------------------------|
| Intended Users | Health care Professionals | Health care Professionals |
| Intended<br>Environment | Healthcare facilities such as<br>hospitals and clinics | Healthcare facilities such as<br>hospitals and clinics |
| Device Class | Class II | Class II |
{6}------------------------------------------------
| Feature/<br>Function | Subject Device:<br>AMS V1 | Predicate Device:<br>Ceevra Reveal 2.0<br>(K173274) |
|----------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------|
| Preoperative<br>Use | Yes | Yes |
| Intraoperative<br>Display | Yes | Yes |
| Image Analysis<br>Features | Interactive<br>manipulation,<br>filtering, create annotations and<br>measurements | Interactive<br>manipulation and<br>segment |
| Measurements | Yes - linear and diameter<br>measurements only | No |
| Remote<br>Streaming | Yes | No |
| Pan image | Pan image in any direction | Pan image in any direction |
| PIN Code | Yes, for remote streaming only | Optional 4-digit security code |
# 7. Predicate Device Comparison
The subject device and predicate device have similar indications for use in that they are both intended for viewing CT and MR data in multi-dimensional views as well as for preoperative and intraoperative surgical planning using similar technologies. Additionally, both devices have similar technological characteristics with only minor differences that do not raise questions of safety or effectiveness.
# 8. Performance Data
Safety and performance of the AMS V1 has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with ANSI AAMI IEC 62304:2006/41:2016 - Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices." All functional and
AMS V1 Traditional 510(k)
{7}------------------------------------------------
performance testing occurred on the minimum hardware configuration and the specified operating systems and VR platforms. Measurement performance testing was conducted by leveraging reference digital phantoms and the comparison with a cleared device (Osirix MD K101342). The table below summarizes the acceptance criteria for the measurement tools. The quality of the display was successfully evaluated against the AAPM guidance recommendation for visual evaluation of luminance and contrast . Additional optical testing was conducted on compatible VR platforms as per IEC63145-20-20 and passed as expected. The homogeneity of luminance, resolution, and contrast was evaluated as acceptable per these standards in the center of the displays for the specified VR platforms. The functioning of the filter technology was assessed by visual inspection. Using a reference DICOM, the opacity and color of specific voxels in the image was demonstrated to be controllable as intended by the filtering principle, which is similar to the cleared device Osirix MD (K101342) . The fluidity of the VR experience was assessed by evaluating the average Frame Per Second rate. The averaged FPS was superior to the specific threshold for the minimal hardware configuration.
| Feature/<br>Function | Tool in Subject<br>Device:<br>AMS V1 | Tool in Reference<br>Device:<br>Osirix MD<br>(K101342) | Acceptance criteria |
|---------------------------------------|--------------------------------------|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Linear<br>Measurements<br>(polylines) | Curve | Close Polygon | No statistical difference<br>between distributions of<br>measurements obtained for<br>AMS V1 or the reference<br>device Osirix MD, as evaluated<br>per t-test statistics.<br>Tests performed for a series of<br>objects in reference MR and<br>CT phantom images. |
| Diameter<br>Measurements | Ruler | Length | |
### 9. Conclusion
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the AMS V1 raises no new questions of safety and effectiveness and demonstrates substantial equivalence to the predicate device in terms of safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.