K222018 · Peep Medical, LLC Dba Go2 Devices · BWF · Jun 2, 2023 · Anesthesiology
Device Facts
Record ID
K222018
Device Name
Breathe+
Applicant
Peep Medical, LLC Dba Go2 Devices
Product Code
BWF · Anesthesiology
Decision Date
Jun 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Breathe+ is intended for use as a Positive Expiratory Pressure Device to help prevent or reverse atelectasis in adult patients needing PEP therapy. Intended for single-patient, multi-use in a hospital or home care setting.
Device Story
Breathe+ is an oral Positive Expiratory Pressure (PEP) device; used in hospital or home settings. Input: patient exhalation. Operation: inhalation opens a flap valve with limited resistance; exhalation closes the flap, forcing air through a resistance mechanism (set by bands) to generate ~8 cm H2O pressure. Output: positive expiratory airway pressure. Benefits: enhances expiratory muscle strength; prevents/reverses atelectasis. Hands-free operation. Healthcare providers use the device to manage lung pressure and improve oxygenation in patients with lung disease.
Clinical Evidence
Bench testing only. Testing included resistance to flow, PEP value repeatability, and accuracy. Performance was evaluated on finished samples pre- and post-aging and post-repeated cleaning. No clinical data provided.
Technological Characteristics
Oral PEP device; resistor-based pressure generation; hands-free design. Materials: surface-contact, mucosal, externally communicating tissue (prolonged duration >24h, <30 days). Connectivity: none. Sterilization: not specified (single-patient, multi-use, cleanable with soapy water).
Indications for Use
Indicated for adult patients requiring PEP therapy to prevent or reverse atelectasis.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 2, 2023
PEEP Medical, LLC % Paul Dryden Consultant 131 Bay Point Drive St. Petersburg, Florida 33704
Re: K222018
Trade/Device Name: Breathe+ Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: April 16, 2023 Received: April 17, 2023
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222018
Device Name Breathe+
Indications for Use (Describe)
The Breathe+ is intended for use as a Positive Expiratory Pressure Device to help prevent or reverse atelectasis in adult patients needing PEP therapy. Intended for single-patient, multi-use in a hospital or home care setting.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Date Prepared: | 02-Jun-2023 |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | PEEP Medical, LLC<br>9230 Katy Freeway, Suite 600<br>Houston, TX 77055 |
| Official Contact: | Sean Boutros, MD<br>Founder |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name:<br>Common/Usual Name:<br>Classification Name: | Breathe+<br>Incentive Spirometer<br>Product Code – BWF – Spirometer, Therapeutic (Incentive) |
| Predicate Device:<br>Common/Usual Name:<br>Classification Name: | DHD Diemolding Healthcare Division – Therapep – K962749<br>Incentive Spirometer<br>Product Code – BWF - Spirometer, Therapeutic (Incentive) |
| Reference Device:<br>Common/Usual Name:<br>Classification Name: | D R Burton - iPEP System and vPEP – K160636<br>Incentive Spirometer<br>Product Code – BWF – Spirometer, Therapeutic (Incentive) |
| Device Description: | The subject device provides PEP only. |
| | The applications of PEEP (PEP) in patients with lung disease is a well-known entity; oxygenation<br>depends on the FiO2 and PEEP, and manipulation of any of these parameters should raise oxygenation in<br>intubated patients. Patients with emphysema instinctively practice pursed lip breathing in order to<br>increase the lung pressure, decrease alveolar collapse and dead space, and improve oxygenation. |
| | The subject device is an oral device that provides positive expiratory airway pressures to enhance<br>expiratory muscle strength while preventing and reversing atelectasis. Furthermore, the device is hands-<br>free. |
| Principle of Operation: | Upon inhalation, the "flap" valve opens and one can inhale with limited resistance. During exhalation, the<br>"flap" valve closes and the front plate is pushed open under resistance, ~ 8 cm H2O, that is set by the<br>bands. The exhaled breath exits through the sides. |
| Indications for Use: | The Breathe+ is intended for use as a Positive Expiratory Pressure Device to help prevent or reverse<br>atelectasis in adult patients needing PEP therapy. Intended for single-patient, multi-use in a hospital or<br>home care setting. |
{4}------------------------------------------------
| Description | Subject Device | Predicate Device | Reference Device | Differences |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Sponsor | PEEP Medical | DHD Diemolding Healthcare<br>Division | D R Burton | |
| 510(k) Number | TBD | K962749 | K160636 | |
| Model Name | Breath+ | TheraPEP | iPEP System and vPEP | |
| Classification | Class II Device<br>BWF<br>21 CFR 858.5690 | Class II Device<br>BWF<br>21 CFR 858.5690 | Class II Device<br>BWF<br>21 CFR 858.5690 | Identical |
| Indications for use | The Breathe+ is intended for use as<br>a Positive Expiratory Pressure<br>Device to help prevent or reverse<br>atelectasis in adult patients needing<br>PEP therapy.<br><br>Intended for single-patient, multi-<br>use in a hospital or home care<br>setting. | The DHD TheraPEP is intended for<br>use as a Positive Expiratory<br>Pressure Device for patients<br>suffering from Cystic Fibrosis, lung<br>diseases with secretory problems,<br>and to prevent or reverse atelectasis. | The D R Burton iPEP Therapy<br>System is intended for use as a<br>Positive Expiratory Pressure<br>(PEP) by patients capable of<br>generating an exhalation flow of<br>10 lpm for 3-4 seconds and an<br>Incentive Spirometer as an<br>inspiratory, deep breathing<br>positive exerciser.<br><br>Intended for single-patient, multi<br>use.<br><br>iPEP System for hospital and<br>clinical settings. | The subject device is<br>intended to prevent or<br>reverse atelectasis. |
| | | | vPEP for hospital, clinical, and home care setting. | |
| Environment of Use | Hospital and home under HCP direction | Clinical settings | Hospital, clinical, and home settings | Similar |
| Patient Population | Patients requiring inspiratory exercise and / or PEP therapy | Patients requiring inspiratory exercise and / or PEP therapy | Patients requiring inspiratory exercise and / or PEP therapy and capable of generating an exhalation flow of 10 lpm for 3-4 seconds | Similar |
| Principle of Operation (PEP) | Resistor<br>Mouthpiece | Resistor<br>Pressure range indicator<br>Pressure adapter<br>Mouthpiece | Flap valve, which generates oscillation during exhalation<br>One-way valve<br>Mouthpiece | Similar |
| Use with a nebulizer | No | Yes | No | Use without a nebulizer does not raise new concerns of safety or effectiveness. |
| Duration of Use | Prolonged < 30 days | Prolonged < 30 days | Prolonged < 30 days | Similar |
| Biocompatibility ISO 10993-1 | Surface Contact<br>Mucosal membrane<br><br>Externally Communicating Tissue<br><br>Duration of Use - prolonged (>24 hours, <30 days) | Surface Contact<br>Mucosal membrane<br><br>Externally Communicating Tissue<br><br>Duration of Use - prolonged (>24 hours, <30 days) | Surface Contact<br>Mucosal membrane<br><br>Externally Communicating Tissue<br><br>Duration of Use – limited (<24 hours) | Similar |
| Performance Testing | | | | |
| Exhalation Resistance (cmH2O) | 4.4 cmH2O @ 20 lpm<br>8.5 cmH2O @ 30 lpm<br>9.7 cmH2O @ 40 lpm | Unknown | Min - 0 cmH2O @ 5 lpm<br>Min - 2.5 cmH2O @ 25 lpm<br>Max - 5.5 cmH2O @ 5 lpm<br>Max - 25.1 cmH2O @ 25 lpm | The exhalation resistance does not raise new concerns of safety or effectiveness. |
| Single patient, multi-use | Yes | Yes | Yes | Similar |
| Clean with soapy water | Yes | N/A | N/A | Cleaning the device does not raise concerns of safety or effectiveness. Test results did not show a degradation |
{5}------------------------------------------------
{6}------------------------------------------------
#### Indications for Use -
The indications for use are similar for the proposed device when compared to the predicate -Discussion - Each device is indicated to provide PEP during patient exhalation, though the subject device makes no claims regarding Cystic Fibrosis.
#### Technology and construction -
The technology of using a resistor to create PEP during exhalation is common to both devices. Discussion - There are no technological differences which would raise different questions of safety or effectiveness between the proposed device and the predicate.
#### Environment of Use -
The environments of use are similar - home and clinical settings. Discussion - The environments of use are similar. Home use is supported by the reference device.
#### Patient Population -
The patient population of the proposed device and predicate are the same. Discussion - The patient population is equivalent to the predicate.
# Non-Clinical Testing Summary -
#### Bench testing -
We performed the following tests: Resistance to flow, and PEP values for repeatability and accuracy. These tests were performed on final, finished samples both pre- and post-aging as well as post-repeated cleaning. Discussion - The device performance was tested and does not raise any new concerns of safety or effectiveness.
# Biocompatibility -
Contact Type and Duration: Surface Contact, Mucosa, Externally Communicating, Tissue; prolonged (>24 hours, <30 days)
Discussion - The proposed device materials, contact type, and duration are identical to the device materials that were cleared under K062104. A Material Certification was used to support the Biocompatibility.
# Discussion of Differences –
The design, performance and function of the device does not raise any significant differences that would raise concerns of safety or effectiveness.
# Substantial Equivalence Conclusion
The sponsor has demonstrated through performance testing, design and non-clinical testing that the subject device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.