← Product Code [IZF](/productcode/IZF) · K221949

# Ortho Device, ADAPTIX 3D Orthopedic Imaging System (K221949)

_Pausch Medical GmbH · IZF · Jan 26, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221949

## Device Facts

- **Applicant:** Pausch Medical GmbH
- **Product Code:** [IZF](/productcode/IZF.md)
- **Decision Date:** Jan 26, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1740
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Ortho Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of the hand, elbow and foot in patients of all ages. The imaging will provide the physician visualized information about anatomical structures to facilitate assessment in orthopedic cases such as: · Fractures of bones in finger, metacarpus or wrist - · Fractures of foot, ankle or elbow joint • Arthritis

## Device Story

Portable 3D tomographic X-ray system; generates cross-sectional images of hand, elbow, foot. Input: X-ray projections from flat panel source; output: 3D tomosynthesis images and synthetic 2D images. Used in orthopedic/radiological practices and hospital emergency departments; operated by medical professionals. C-Arm manually rotates up to 90 degrees; system mounted on stand or trolley. Reconstruction software processes individual images into 3D volumes. Radiologists interpret output to support clinical decision-making for fractures and arthritis. Benefits: cost-effective, portable, detailed multi-slice 3D visualization of orthopedic anatomy.

## Clinical Evidence

Bench testing only. Includes image quality assessment (spatial/contrast resolution, homogeneity, linearity) and radiologist evaluation of sample clinical images to confirm diagnostic capability for intended anatomy (fingers, wrist, elbow, foot, ankle) and conditions (fractures, arthritis).

## Technological Characteristics

Portable 3D tomographic X-ray system; C-Arm configuration. Flat panel detector (15x11 cm, 99 µm pixel pitch, 14-bit ADC). Fixed 60 kV X-ray source. Wired (Ethernet) connectivity. DICOM 3.0 output. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, and IEC 61910-1. Software: Moderate Level of Concern.

## Regulatory Identification

A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- DR 800 with Tomosynthesis ([K183275](/device/K183275.md))
- OrthoScan Mobile DI Mini C-Arm ([K113708](/device/K113708.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 26, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" in a smaller font below.

Pausch Medical GmbH % Oliver Eikenberg Lead Consultant Quality and Regulatory Affairs Emergo Global Consulting LLC 2500 Bee Cave Road Building 1. Suite 300 AUSTIN TX 78746

Re: K221949

Trade/Device Name: ADAPTIX 3D Orthopedic Imaging System ("Ortho Device") Regulation Number: 21 CFR 892.1740 Regulation Name: Tomographic x-ray system Regulatory Class: Class II Product Code: IZF , MQB Dated: December 22, 2022 Received: December 27, 2022

Dear Oliver Eikenberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang
2023.01.26
17:03:19
-05'00'

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221949

#### Device Name

ADAPTIX 3D Orthopedic Imaging System ("Ortho Device")

Indications for Use (Describe)

The Ortho Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of the hand, elbow and foot in patients of all ages.

The imaging will provide the physician visualized information about anatomical structures to facilitate assessment in orthopedic cases such as:

· Fractures of bones in finger, metacarpus or wrist

- · Fractures of foot, ankle or elbow joint
• Arthritis

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## ADAPTIX 3D Orthopedic Imaging System ("Ortho Device")

### 1. Submission Sponsor

Pausch Medical GmbH Graf - Zeppelin - Str. 1 91056 Erlangen, Bavaria, Germany Contact: Christian Stoian (CEO) Email: Christian.Stoian@mavig.com_ Phone: +49 (0)9131 / 99 92 - 0

## 2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746, US Email: LST.AUS.ProjectManagement@ul.com Office Phone: (512) 327-9997 Contact: Oliver Eikenberg, PhD Title: Senior Consultant, Quality & Regulatory Affairs, Emergo Group

### 3. Date Prepared

01/26/2023

### 4. Device Identification

| Trade/Proprietary Name: | ADAPTIX 3D Orthopedic Imaging System (“Ortho Device”)              |
|-------------------------|--------------------------------------------------------------------|
| Common/Usual Name:      | Tomographic x-ray system                                           |
| Regulation Number:      | §892.1740 Tomographic x-ray system, §892.1680 Stationary x-ray sys |
| Product Code:           | IZF Tomographic x-ray system, MQB solid state x-ray imager         |
| Class:                  | Class II                                                           |
| Classification Panel:   | Radiology                                                          |

## 5. Legally Marketed Predicate Device(s)

| Primary predicate device |                                                                                              |
|--------------------------|----------------------------------------------------------------------------------------------|
| Device name:             | DR 800 with Tomosynthesis                                                                    |
| Regulation Number:       | §892.1740 Tomographic x-ray system<br>§892.1650 fluoroscopic, image-intensified x-ray system |
| Product Code:            | IZF Tomographic x-ray system                                                                 |
| Subsequent Product Code: | JAA fluoroscopic, image-intensified x-ray system                                             |
| 510(k) number:           | K183275                                                                                      |
| Manufacturer:            | Agfa N.V.                                                                                    |

#### Secondary predicate device

| Device name:             | OrthoScan Mobile DI Mini C-Arm                         |
|--------------------------|--------------------------------------------------------|
| Regulation Number:       | §892.1650 fluoroscopic, image-intensified x-ray system |
| Product Code:            | OXA mobile image-intensified fluoroscopic x-ray system |
| Subsequent Product Code: | JAA fluoroscopic, image-intensified x-ray system       |
| 510(k) number:           | K113708                                                |
| Manufacturer:            | ORTHOSCAN, INC.                                        |

{4}------------------------------------------------

#### 6. Indications for Use Statement

The Ortho Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of the hand, elbow and foot in patients of all ages.

The imaging will provide the physician visualized information about anatomical structures to facilitate assessment in orthopedic cases such as:

- Fractures of bones in finger, metacarpus or wrist
- Fractures of foot, ankle or elbow joint
- Arthritis

### 7. Device Description

The Ortho Device is a 3D tomographic X-ray device intended to be used to produce radiological images of a specific cross-sectional plane of the body. The device is comprised of a Flat Panel X-ray source combined with a digital detector within a mounting frame, a control unit and a workstation. It is intended to offer 3D imaging of orthopedic structures by using a panel of X-ray sources that construct a 3D tomosynthesis image with the associated reconstruction software from individual images; it is also possible to create synthetic 2D images of the desired anatomy.

The Ortho Device is a portable system that can be mounted on a stand for tabletop applications or on a trolley cart for added mobility with motorized vertical positioning. The C-Arm and Control Unit components are both designed to be carriable by a single person. To allow for the ideal positioning of the anatomy (hand and weight-bearing foot images) in the beam path and to achieve the desired plane of view, the Ortho Device C-Arm can be manually rotated by up to 90°. The central beam is aligned perpendicularly to the image receptor.

The "Ortho Device" was created to fill a diagnostic niche in orthopedic medicine for cost effective and portable imaging for patients and is used, amongst other applications, for 3D-radiographic diagnostic imaging of hand, elbow and foot in orthopedic and radiological practices as well as in emergency departments of hospitals. The Ortho Device results are detailed multi-slice 3D images of patients that allow radiologist interpretation of clinical image data and by this support medical professionals decisionmaking on human anatomy.

The Ortho Device system is designed to meet the requirements in accordance with relevant sections of 21CFR 1020.30-1020.31.

#### 8. Substantial Equivalence Discussion

The following table compares the Ortho Device to the predicate devices with respect to indications for use, principles of operation, technological characteristics, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

{5}------------------------------------------------

| Attribute                                                                   | Primary Predicate Device<br>DR 800 with Tomosynthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Secondary Predicate Device<br>OrthoScan<br>Mobile DI Mini C-Arm                                                                                                                                                                                                                                                                                   | Subject Device<br>Ortho Device                                                                                                                                                                                                                                                                                                                                                                                                                                           | DEVICE<br>COMPARISON                                                             |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Manufacturer                                                                | Agfa N.V.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ORTHOSCAN, INC.                                                                                                                                                                                                                                                                                                                                   | Pausch Medical GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                      | n/a                                                                              |
| US Premarket<br>Notification                                                | K183275                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K113708                                                                                                                                                                                                                                                                                                                                           | K221949                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | n/a                                                                              |
| FDA Product<br>Code                                                         | IZF Tomographic X-ray system,<br>JAA image-intensified<br>fluoroscopic X-ray system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | OXO Mobile Image-intensified<br>fluoroscopic X-ray system,<br>JAA image-intensified<br>fluoroscopic X-ray system                                                                                                                                                                                                                                  | IZF Tomographic X-ray<br>system,<br>MQB solid state X-ray imager                                                                                                                                                                                                                                                                                                                                                                                                         | Similar                                                                          |
| Intended Use<br>/ Indications<br>for Use                                    | The DR 800 system is indicated for<br>performing dynamic imaging<br>examinations (fluoroscopy and/or<br>rapid sequence) of the following<br>anatomies/procedures:<br>•Positioning fluoroscopy<br>procedures<br>•Gastro-intestinal examinations<br>•Urogenital tract examinations<br>•Angiography<br>It is intended to replace<br>fluoroscopic images obtained<br>through image intensifier<br>technology. In addition, the system<br>is intended for project radiography<br>of all body parts.<br>In addition, the system provides the<br>Agfa Tomosynthesis option, which<br>is intended to acquire tomographic<br>slices of human anatomy and to be<br>used with Agfa DR X-Ray systems.<br>Tomosynthesis is used to<br>synthesize tomographic slices from<br>a single tomographic sweep.<br>The DR 800 is not intended for<br>mammography applications. | The Orthoscan Mobile DI Mini<br>C-Arm is designed to provide<br>the physician with general<br>fluoroscopic visualization of<br>the patient including, but not<br>limited to, surgical orthopedic<br>procedures and critical and<br>emergency care procedures in<br>hospital, emergency care,<br>critical care or physician office<br>environment. | The Ortho Device is intended<br>to generate tomosynthesis<br>images of human anatomy for<br>diagnostic purposes of the<br>hand, elbow and foot in<br>patients of all ages.<br>The imaging will provide the<br>physician visualized<br>information about anatomical<br>structures to facilitate<br>assessment in orthopedic<br>cases such as:<br>• Fractures of bones in<br>finger, metacarpus or<br>wrist<br>• Fractures of foot, ankle<br>or elbow joint<br>• Arthritis | Similar                                                                          |
| Device                                                                      | Flat Panel detector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Flat Panel detector                                                                                                                                                                                                                                                                                                                               | Flat Panel detector (C-Arm)                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar design                                                                   |
| Components                                                                  | X-ray Generator<br>X-ray table with tube, housing<br>and Collimator<br>MUSICA Acquisition Workstation<br>with Control Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | X-ray Generator<br>X-ray tube and housing                                                                                                                                                                                                                                                                                                         | Board generating High<br>Voltage (C-Arm)<br>Flat Panel Source (C-Arm)<br>Emission Control Board (C-Arm)<br>Acquisition Board (Control Unit)<br>Workstation                                                                                                                                                                                                                                                                                                               | and intended use<br>to OrthoScan and<br>different to DR<br>800                   |
| Attribute                                                                   | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Secondary Predicate Device                                                                                                                                                                                                                                                                                                                        | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                           | DEVICE<br>COMPARISON                                                             |
|                                                                             | DR 800 with Tomosynthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | OrthoScan                                                                                                                                                                                                                                                                                                                                         | Ortho Device                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                  |
|                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Mobile DI Mini C-Arm                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                  |
| Flat<br>Panel<br>Detector                                                   | Thales Pixium RF4343 FL<br>Gadolinium Oxysulfide (GOS) or<br>Cesium lodide (Csl) Scintillator<br>Field Sizes: 43 x 43 cm, 30 x 30<br>cm, 20 x 20 cm, 15 x 15 cm<br>Resolution: 2840 x 2874<br>Dynamic Range 16 bit<br>Pixel size 148 µm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | CMOS detector<br>Detector size 15 x 12 cm<br>Field of View (full) 13.8 x 10.9cm<br>Resolution 2000 x 1500 Pixel<br>pitch 75 µm                                                                                                                                                                                                                    | Flat Panel Detector<br>Detector dimensions: 15x11 cm<br>Resolution: 1488 x 1148<br>ADC Conversion 14 bit<br>Pixel Pitch: 99 µm                                                                                                                                                                                                                                                                                                                                           | Similar design<br>and intended use<br>to OrthoScan and<br>different to DR<br>800 |
| X-ray<br>Generator(s)                                                       | kV range: 40 to 110 kV<br>mA range: 1.5 to 50.8 mA<br>Choice of three models:<br>50, 65KW, 80 KW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | kV range: 40 to 78 kV<br>mA range: 0.040 to 0.160 mA                                                                                                                                                                                                                                                                                              | kV range: 60 kV (fixed)<br>mA range: 0.01 to 0.05 mA                                                                                                                                                                                                                                                                                                                                                                                                                     | Different                                                                        |
| Collimator                                                                  | Ralco R 302 MLP/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Fixed Aperture at fixed SID                                                                                                                                                                                                                                                                                                                       | Fixed Aperture at fixed SID                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar to<br>OrthoScan but<br>different to DR<br>800                            |
| Software                                                                    | MUSICA Dynamic<br>MUSICA2<br>MUSICA3/3+<br>MUSICA DTS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthotouch                                                                                                                                                                                                                                                                                                                                        | Adaptix Software Package<br>(Acquisition &<br>Reconstruction)                                                                                                                                                                                                                                                                                                                                                                                                            | Different                                                                        |
| Software<br>Image<br>Manipulation<br>Functions                              | Image acquisition control and display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                             |
| Software<br>Image file<br>format                                            | DICOM 3.0 File Output                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                             |
| Type of<br>Software<br>Application                                          | Windows-based software application                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                             |
| Panel<br>Interface                                                          | Wired or wireless (optional)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                   | Wired (Ethernet)                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar                                                                          |
| Power Source                                                                | 230 V AC, 50/60 Hz (X-ray table)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | AC: 100-230 V , 50-60 Hz                                                                                                                                                                                                                                                                                                                          | AC: 110 - 240 V, 50Hz/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar                                                                          |
| Accessories                                                                 | Ankle Rest<br>Foot Rest<br>Head Rest                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Cart                                                                                                                                                                                                                                                                                                                                              | Trolley (Cart)<br>Foot Rest<br>Patient supporting arm rest                                                                                                                                                                                                                                                                                                                                                                                                               | Similar                                                                          |
| Protection<br>type and level<br>against<br>electric shock                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Class 1                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                             |
| Electrical<br>Safety,<br>Electro-<br>magnetic<br>Compatibility<br>standards |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-28<br>IEC 60601-2-54                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                             |

## Table 5A – Comparison of Characteristics between Subject Device and Predicate Devices

{6}------------------------------------------------

{7}------------------------------------------------

#### 9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of Ortho Device and to show substantial equivalence to the predicate devices, Pausch Medical GmbH completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The Ortho Device passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate devices.

- . In vitro Cytotoxicity testing per ISO 10993-5 – Passed
- . Irritation and skin Sensitization testing per ISO 10993-10 – Passed
- Systemic toxicity testing per ISO 10993-11 – Passed
- . Electrical safety testing per IEC 60601-1 – Passed
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 – Passed
- Radiation protection testing in diagnostic X-ray equipment per IEC 60601-1-3 – Passed
- . Medical Electrical Equipment Usability testing per IEC 60601-1-6 – Passed
- . Safety and essential performance testing of X-ray tube assemblies for medical diagnosis and X-ray equipment for radiography and radioscopy per IEC 60601-2-28 and IEC 60601-2-54 -Passed
- . Particular electrical testing performance requirements for Radiation dose documentation for X-ray equipment per IEC 61910-1 and Digital Imaging and Communications in Medicine (DICOM) per NEMA PS 3.1 - Passed
- Transportation Testing per ASTM D4169 demonstrates packaging integrity maintained
- . Testing for image quality - spatial and contrast resolution, homogeneity, and linearity etc. – Passed
- . Evaluation of sample clinical images by radiologists to demonstrate that the device is able to image all intended body parts (fingers, metacarpus/wrist, elbow, foot, ankle)
- . Evaluation of sample clinical images by radiologists to demonstrate that the device is able to provide imaging data to help clinician for the assessment of bone fracture and arthritis
- . Software verification and validation testing has been completed on a functional level for a Moderate Level of Concern software including system compatibility testing, risk analysis per IEC 62304/FDA Guidance
- . Risk Management per EN ISO 14971, all requirements were met and risks reduced as far as possible.

#### 10. Statement of Substantial Equivalence

The Ortho Device has the same indications for use as the predicate devices OrthoScan Mobile DI Mini C-Arm and DR 800 with Tomosynthesis. Any minor differences in the technological characteristics of the subject device when compared to the predicate devices have been successfully evaluated through appropriate safety and performance testing which demonstrates that the subject device, when compared to the predicate devices, does not raise any new questions of safety and effectiveness. Therefore, the Ortho Device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K221949](https://fda.innolitics.com/device/K221949)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
