← Product Code [OVD](/productcode/OVD) · K221936

# Standalone ALIF Interbody Fusion System (K221936)

_Eminent Spine · OVD · Oct 17, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221936

## Device Facts

- **Applicant:** Eminent Spine
- **Product Code:** [OVD](/productcode/OVD.md)
- **Decision Date:** Oct 17, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Eminent Spine Standalone ALIF Interbody Fusion is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The Eminent Spine Standalone ALIF Interbody Fusion System is a stand-alone interbody fusion device intended to be used with screws which accompany the spacers for fixation. Hyperlordotic implants (≥25 lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I Spondvlolisthesis. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The device is designed for use with autograft and/or allogenic bone graft composed ofcancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

## Device Story

Standalone ALIF Interbody Fusion System; intervertebral body fusion device for lumbar spine. Implanted via anterior lumbar approach; packed with autogenous/allogenic bone graft to facilitate fusion. Serrated superior/inferior surfaces grip vertebral endplates; integrated fixation screws inserted through anterior face for stability. Provides mechanical support until biologic fusion occurs. Used by surgeons in clinical settings. Benefits patients by stabilizing spinal segments and promoting fusion in DDD cases.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing included static and dynamic compression, compression-shear, torsion, subsidence, and expulsion testing.

## Technological Characteristics

Materials: PEEK (ASTM F2026) with tantalum pins (ASTM F560), or Ti-6Al-4V ELI (ASTM F136), or additively manufactured Ti-6Al-4V (ASTM F3001). Integrated fixation screws/plates: Ti-6Al-4V ELI (ASTM F136). Form factor: Interbody spacer with integrated screw fixation. Non-software device.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Globus HEDRON Lumbar Spacers ([K191391](/device/K191391.md))

## Reference Devices

- Eminent Cervical Stand-Alone System ([K212853](/device/K212853.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 17, 2022

Eminent Spine % Daniel Johnson Engineer Jalex Medical 27865 Clemens Rd Suite 3 Westlake, Ohio 44145

Re: K221936

Trade/Device Name: Standalone ALIF Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: September 8, 2022 Received: September 9, 2022

Dear Daniel Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K221936

### Device Name

Standalone ALIF Interbody Fusion System

#### Indications for Use (Describe)

The Eminent Spine Standalone ALIF Interbody Fusion is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The Eminent Spine Standalone ALIF Interbody Fusion System is a stand-alone interbody fusion device intended to be used with screws which accompany the spacers for fixation. Hyperlordotic implants (≥25 lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I Spondvlolisthesis. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The device is designed for use with autograft and/or allogenic bone graft composed ofcancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

| Type of Use (Select one or both, as applicable) |   |
|-------------------------------------------------|---|
| ☑                                               | ☐ |

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Submitted By:               | Eminent Spine<br>2004 Ventura Dr. Suite #100<br>Plano, TX 75093                                                       |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Date:                       | 06/29/2022                                                                                                            |
| Contact Person:             | Daniel Johnson, Engineer                                                                                              |
| Contact Telephone:          | (216) 333-2127                                                                                                        |
| Contact Fax:                | (440) 933-7839                                                                                                        |
| Device Trade Name:          | Standalone ALIF Interbody Fusion System                                                                               |
| Common Name:                | Intervertebral Body Fusion Device                                                                                     |
| Device Classification Name: | Intervertebral Body Fusion Device with Integrated Fixation, Lumbar                                                    |
| Device Classification:      | Class II                                                                                                              |
| Reviewing Panel:            | Orthopedic                                                                                                            |
| Product Code:               | OVD                                                                                                                   |
| Primary Predicate Device:   | Globus HEDRON Lumbar Spacers (K191391)<br>The primary predicate device has never been subject to a recall.            |
| Reference Predicate Device: | Eminent Cervical Stand-Alone System (K212853)<br>The reference predicate devices have never been subject to a recall. |

### Device Description:

The Standalone ALIF devices are inserted through an anterior lumbar approach and packed with autogenous bone graft to facilitate fusion. Serrations on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebral bodies to aid in expulsion resistance, while screws are inserted through the anterior face of the implant for bone fixation. The device is intended to provide mechanical support to the implanted level until biologic fusion is achieved.

The cages are made from medical grade polyetheretherketone (PEEK) per ASTM F2026 with tantalum per ASTM F560 pins, from titanium alloy Ti-6Al-4V ELI per ASTM F136, or from additively manufactured Ti-6Al-4V per ASTM F3001. The integrated fixation screws and screw anti-backout plate are manufactured from Ti-6A1-4V ELI per ASTM F136.

### Indications for Use:

The Eminent Spine Standalone ALIF Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The Eminent Spine Standalone ALIF Interbody Fusion System is a standalone interbody fusion device intended to be used with screws which accompany the spacers for fixation. Hyperlordotic implants (>=25° lordosis) are intended for use with supplemental fixation ( e.g. facet screws or posterior fixation). The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I Spondylolisthesis or retrolisthesis. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The device is designed for use with autograft and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

{4}------------------------------------------------

## Summary of Technological Characteristics:

The Eminent Spine Standalone ALIF Interbody Fusion System and the predicates have the same intended use and fundamental scientific technology. All devices compare similarly in:

- Design features
- Intended use
- Materials ●
- Dimensions ●
- Function ●

## Mechanical Testing:

Substantial equivalence is supported by the results of mechanical testing in the following test modes:

- Static compression ●
- Static compression-shear ●
- Static torsion ●
- Subsidence
- Expulsion
- Dynamic compression ●
- Dynamic compression-shear
- Dynamic torsion

Results support that the subject device performs as well as or better than the chosen acceptance criteria.

## Conclusion:

Based on the indications for use, technological characteristics, and comparison with the predicate device, the subject device has demonstrated substantial equivalence.

---

**Source:** [https://fda.innolitics.com/device/K221936](https://fda.innolitics.com/device/K221936)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
